K103464 · Medtronic, Inc. · LIT · Dec 22, 2010 · Cardiovascular
Device Facts
Record ID
K103464
Device Name
PACIFIC XTREME PTCA BALLOON DILATATION CATHETER
Applicant
Medtronic, Inc.
Product Code
LIT · Cardiovascular
Decision Date
Dec 22, 2010
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PACIFIC XTREME PTA Balloon Dilatation Catheter in 150mm, 200mm, 250mm and 300mm balloon length is intended to dilate stenoses in the femoral, popliteal, and infrapopliteal arteries.
Device Story
The PACIFIC XTREME PTA Balloon Dilatation Catheter is an over-the-wire percutaneous transluminal angioplasty device used to dilate stenoses in peripheral arteries. The catheter features a proximal hub, dual lumen coaxial shaft, and a distal dilatation balloon. It is compatible with 0.018" guidewires and 4F/5F introducer sheaths. The device is provided with a hydrophilic coating to facilitate navigation. It is used by clinicians in a clinical setting to perform angioplasty procedures. By inflating the balloon at the site of a stenosis, the device mechanically dilates the vessel, potentially improving blood flow and benefiting patients with peripheral arterial disease.
Clinical Evidence
No clinical investigation has been performed for this device. Evidence is based on bench testing and biocompatibility testing conducted in accordance with ISO 10993-1 and FDA guidance.
Technological Characteristics
Over-the-wire PTA catheter; compatible with 0.018" guidewires and 4F/5F introducer sheaths; hydrophilic coating; available in 90cm and 130cm lengths. Testing performed per ISO 10993-1 and FDA guidance for balloon burst strength, compliance, fatigue, torque, and tensile strength.
Indications for Use
Indicated for dilation of stenoses in femoral, popliteal, and infrapopliteal arteries.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
Predicate Devices
Invatec SUBMARINE PLUS PTA Balloon catheter (K042537)
Invatec AMPHIRION DEEP OTW PTA Balloon Dilatation Catheter (K083919)
K100490 — LITEPAC, MODEL 682 · ClearStream Technologies , Ltd. · Mar 16, 2010
Submission Summary (Full Text)
{0}------------------------------------------------
K103464
# 510(k) Summary
DEC 2 2 2010
This 510(k) Summary is submitted in accordance with the requirements of 21 CFR Part 807, Section 807.92(c).
# Applicant:
# Official
| Contact Person: | Heather Morose, Associate Regulatory Specialist<br>Medtronic Inc.<br>37A Cherry Hill Drive<br>Danvers, MA 01923<br>Tel: 978-739-3187<br>Fax: 978-739-7380<br>Email: heather.p.morose@medtronic.com |
|--------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Alternate Contact: | Fred Boucher, Director, RA |
Medtronic, Inc. 37A Cherry Hill Drive Danvers, MA 01923 Tel: 978-739-3116 Fax: 978-739-7380 Email: fred.boucher@medtronic.com
Date Prepared: November 22, 2010
Device Information:
| Trade Name: | Medtronic Invatec PACIFIC XTREME™ PTA Balloon |
|-----------------------|------------------------------------------------|
| Common Name: | Percutaneous Transluminal Angioplasty Catheter |
| Regulation Name: | Percutaneous Catheter |
| Classification: | II |
| Classification Panel: | Peripheral |
| Regulation Number: | 21 CFR 870.1250 |
| Product Code: | LIT, DQY |
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image contains the Medtronic logo. The logo consists of a stylized image on the left and the word "Medtronic" on the right. The stylized image appears to be a representation of a brain. The text is in a bold, sans-serif font.
## Predicate Devices:
Invatec SUBMARINE PLUS PTA Balloon catheter (K042537) Invatec AMPHIRION DEEP OTW PTA Balloon Dilatation Catheter (K083919) Invatec ADMIRAL XTREME PTA Balloon Catheter (K100921)
## Device Description:
The PACIFIC XTREME™ PTA Balloon Dilatation Catheter is an over-the-wire Percutaneous Transluminal Angioplasty (PTA) catheter consisting of a proximal hub, dual lumen coaxial shaft, and a distal dilatation balloon. The PACIFIC XTREME™ PTA Balloon Dilatation Catheter is compatible with guidewires with a maximum diameter of 0.018" and with 4F and 5F introducer sheaths, depending on the diameter and balloon length of the dilatation balloon. The catheter is provided with a hydrophilic coating and is available in useable catheter lengths of 90 and 130cm.
## Indication for Use:
The PACIFIC XTREME PTA Balloon Dilatation Catheter in 150mm, 200mm, 250mm and 300mm balloon length is intended to dilate stenoses in the femoral, popliteal, and infrapopliteal arteries.
### Technological Characteristics:
The PACIFIC XTREME™ PTA Balloon Dilatation Catheter is an over-the-wire percutaneous transluminal angioplasty (PTA) catheter. The catheter is compatible with .018" guidewire. The technological characteristics of the PACIFIC XTREME™ PTA Balloon Dilatation Catheter are substantially equivalent to those of the Submarine Plus PTA Balloon catheter (K042537), Amphirion Deep PTA Balloon Catheter (K083919), and the Admiral Xtreme PTA Balloon Catheter (K100921).
### Summary of Bench Testing:
In vitro bench testing of the PACIFIC XTREME™ PTA Balloon Dilatation Catheter was conducted in accordance with Medtronic Invatec Risk Analysis and all applicable FDA guidance documents and ISO standards, testing included:
- Minimum Balloon Burst Strength (RBP) .
- Balloon Compliance (diameter vs pressure) .
- Balloon Inflation and Deflation time .
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the Medtronic logo. The logo consists of a circular graphic on the left and the word "Medtronic" on the right. The circular graphic appears to be a stylized representation of the earth. The word "Medtronic" is written in a bold, sans-serif font.
- . Balloon Fatigue (repeat balloon inflations)
- Flexibility and Kink Test �
- Torque Strength .
- Balloon Preparation ●
- Guidewire Compatibility .
- . Coating Durability
- . Coating Lubricity
- Catheter Diameter-balloon profile usable length .
- Tensile Strength .
- Catheter Body Burst Pressure 0
- Introducer Sheath Compatibility 0
## Summary of Biocompatibility Testing:
The PACIFIC XTREMETM PTA Balloon Dilatation Catheter is an externally communicating device, which contacts circulating blood for the limited contact duration (<24hours).
Biocompatibility testing was conducted on finished PACIFIC XTREME™ PTA Balloon Catheters in accordance with ISO 10993-1 , "Biological Evaluation of Medical Devices part 1:Evaluation and testing," as specified in the FDA Blue Book Memorandum #G95-1 and FDA guidance: Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters (May 30 2008) All testing documented below was conducted in accordance with the provisions of the FDA GLP regulations 21 CFR 58.
The following Biocompatibility tests were completed and passed:
- Cytotoxicity Study .
- Maximisation Sensitization Study .
- Intracutaneous Study .
- Systemic Toxicity Study .
- Material Mediated Pyrogen Study .
- Hemolysis Study .
- In Vivo Thromboresistance Study .
- Lee White Coagulation Test ●
- Complement Activation .
Medtronic PACIFIC XTREME™ Special 510(k)
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Medtronic logo. The logo consists of a stylized globe-like symbol on the left and the word "Medtronic" in bold, sans-serif font on the right. The globe symbol appears to be made up of interconnected lines or pathways.
#### Assessment of non-clinical performance data for equivalence:
Bench and biocompatibility testing were conducted according to the recommendations from relevant FDA guidance to demonstrate that the PACIFIC XTREME™ PTA Balloon Dilatation Catheter acceptance criteria and performed equally to the predicate devices. No new safety or effectiveness issues were raised during the testing.
#### Summary of Clinical Data:
No clinical investigation has been performed for this device.
#### Conclusion from Data:
Medtronic Invatec has demonstrated that the PACIFIC XTREME™ PTA Balloon Dilatation Catheter is substantially equivalent to the predicate devices based on its indications for use and fundamental scientific technology. Testing demonstrates that the PACIFIC XTREMETM PTA Balloon Dilatation Catheter device is safe, effective and performs as well as the predicate devices
Medtronic PACIFIC XTREME™ Special 510(k)
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" are arranged in a circle around the eagle. The eagle is depicted in a simple, black and white design. The logo is commonly used on official documents and websites related to the department.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Medtronic, Inc. C/O Heather Morose Associate Regulatory Specialist 37A Cherry Hill Drive Danvers, MA 01923
DEC 2 2 2010
Re: K103464
Trade/Device Name: PACIFIC XTREME™ PTA Balloon Dilatation Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous catheter Regulatory Class: Class II · Product Code: LIT, DQY Dated: November 22, 2010 Received: November 24, 2010
Dear Ms. Morose:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Ms. Heather Morose
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.goy/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
#### Enclosure
{6}------------------------------------------------
# Indications for Use
DEC 2 2 2010
510(k) Number (if known): Device Name: PACIFIC XTREMETM PTA Balloon Dilatation Catheter Indications for Use: The PACIFIC XTREME PTA Balloon Dilatation Catheter in 150mm, 200mm, 250mm and 300mm balloon length is intended to dilate stenoses in the femoral, popliteal, and infrapopliteal arteries.
Prescription Use
AND/OR
Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
W.M.D.
510(k) Number K103464
Medtronic PACIFIC XTREME™ Special 510(k)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.