INNERDYNE RADIALLY EXPANDING VASCULAR ACCESS SYSTEM (REVAS)
K992668 · Innerdyne, Inc. · DYB · May 26, 2000 · Cardiovascular
Device Facts
Record ID
K992668
Device Name
INNERDYNE RADIALLY EXPANDING VASCULAR ACCESS SYSTEM (REVAS)
Applicant
Innerdyne, Inc.
Product Code
DYB · Cardiovascular
Decision Date
May 26, 2000
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1340
Device Class
Class 2
Indications for Use
The InnerDyne, Inc., REVAS product is intended to provide percutaneous access to the vasculature. The REVAS product is indicated for the following use: Percutaneous vascular access.
Device Story
REVAS is a radially expanding vascular access system used during minimally invasive surgery. Device consists of an expandable sleeve assembly and a dilator/sheath assembly with an integral hemostasis valve. Operation: sleeve is inserted over a guidewire into a vessel; sleeve is radially compressed for insertion and expands to accommodate the dilator/sheath. This process dilates surrounding tissue. Once dilated, the dilator is removed, leaving the sleeve and sheath in place to provide a sealed port for instrument passage. System allows for initial placement of a small-diameter sheath, which can be removed and replaced with a larger-diameter dilator/sheath through the same sleeve to create a larger port. Used by clinicians in a surgical setting to facilitate instrument access to the vasculature. Benefits include temporary, sealed percutaneous access.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on design, material, and functional similarity to legally marketed predicate devices.
Technological Characteristics
System comprises an expandable sleeve, tapered dilator, and sheath with integral hemostasis valve. Materials are medical grade, assembled via adhesives, welds, and mechanical interlocks. Operates via mechanical radial expansion of the sleeve to accommodate dilators. No energy source or software components.
Indications for Use
Indicated for percutaneous vascular access in patients requiring minimally invasive surgery for temporary passage of instruments into the vasculature (arterial or venous). Contraindicated for use in the presence of vascular grafts at the access site.
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for InnerDyne, Inc. The words "InnerDyne" are in a bold, black font, with a registered trademark symbol after the word "Dyne". The word "Inc." is to the right of "InnerDyne" and is outlined in black.
## SUMMARY OF SAFETY & EFFECTIVENESS INFORMATION
Submitter: InnerDyne, Inc. 5060 Amelia Earhart Dr Salt Lake City, Utah (801)-350-3600
Date Prepared: August 6, 1999
Contact: Rick Gaykowski Corporate Vice President, Regulatory Affairs and Quality Assurance
Classification Name: Percutaneous Introducer Common/Usual Name: Percutaneous Introducer With Sheath Trade/Proprietary Name: InnerDyne, Inc., Radially Expanding Vascular Access System (REVAS™)
The REVAS product consists of an expandable sleeve assembly. The tubular member of the expandable is configured so as to be radially compressed to reduce the outside diameter of the device prior to insertion. Upon use, the expandable sleeve assembly is inserted through the skin and surrounding tissue into the selected vessel over a previously placed guidewire according to the techniques of Seldinger. (An alternative placement method would have the needle placed within the lumen of the expandable sleeve, and the combined needle and sleeve inserted into the target vessel. The guidewire would be placed and the needle removed.) The dilator/sheath assembly, which consists of a tapered dilator and sheath with integral hemostasis valve, is inserted over the guidewire through the lumen of the expandable sleeve which expands radially to accommodate the dilator/sheath. This process in turn, radially expands the surrounding tissue. Following dilation the dilator is removed, leaving the expandable sleeve and sheath with integral hemostasis valve in place to provide a sealed port for passage of instruments.
This system configuration allows the user to initially place a small diameter sheath for passage of small instruments. The sheath can then be removed from the lumion the sleeve while leaving the sleeve inserted in the vessel. A larger dirmeter dilator/sheath can then be inserted through the sleeve to create a larger port for passage of the larger instruments.
The device is assembled from medical grade materials under GMP and ISO conditions. Components are molded and machined by qualified suppliers. The components are assembled and secured by adhesives, welds, and mechanical interlocks.
The subject InnerDyne, Inc., REVAS product is substantially equivalent to predicate
{1}------------------------------------------------
InnerDyne, Inc., REVAS products. The subject REVAS device is similar to the referenced predicate in size, function, product dimensions and indications for use.
The REVAS product is intended for use during minimally invasive surgery for temporary percutaneous vascular access for passage of instruments into the vasculature. The device is configured to be used for arterial or venous access.
The basic design principles for the subject InnerDyne, Inc., REVAS product and the predicate REVAS devices are similar, and remain essentially unchanged from information provided to the Agency. The product configuration, composition, and utilized materials are similar in each of the products. The principles of operation for the subject InnerDyne, Inc., REVAS product and the predicate REVAS devices are similar. That is, each of these products employs a similar insertion technique, indications for use, contraindications for use, warnings and precautions. The subject device has an incorporated contraindication against use in the presence of vascular grafts at the access site.
From the foregoing, we conclude that the subject REVAS product is as safe and effective as currently marketed devices for the stated indications.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the text "Public Health Service" in a bold, sans-serif font. The text is arranged on a single line, with each word clearly legible. The background is plain white, providing a stark contrast to the black text.
Image /page/2/Picture/2 description: The image is a black and white circular logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with three lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper half of the circle, following its curvature.
MAY 2 6 2000
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Rick Gaykowski Vice President, RA/QA InnerDyne, Inc. 5060 West Amelia Earhart Drive Salt Lake City, UT 84116
K992668 Re: Trade Name: REVAS™ Radially Expanding Vascular Access System Regulatory Class: II (two) Product Code: 74 DYB Dated: April 25, 2000 Received: April 26, 2000
Dear Mr. Gaykowski:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to A substantially equivalent determination assumes compliance with 895. the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Please note: this response to your premarket Federal Register. notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
Page 2 - Mr. Rick Gaykowski
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4586. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also. please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Buim E. Hansenym
James E. Dillard III Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page 1 of 1
510(k) Number (if known):
Device Name: InnerDyne, Inc., Radially Expanding Vascular Access System (REVAS™)
Indications for Use: The InnerDyne, Inc., REVAS product is intended to provide percutaneous access to the vasculature. The REVAS product is indicated for the following use:
> Percutaneous vascular access. .
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Suinn E. Hanson
Divi and N 510k) N
Prescription Use (Per 21 CFR 801.109)
OR
Over-the-Counter Use
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.