INNERDYNE RADIALLY EXPANDING VASCULAR ACCESS SYSTEM (HEPARIN COATED)
Applicant
Innerdyne, Inc.
Product Code
DYB · Cardiovascular
Decision Date
Sep 11, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1340
Device Class
Class 2
Indications for Use
The Radially Expanding Vascular Access System (Heparin Coated) is intended for use during minimally invasive surgery for temporary percutaneous vascular access for passage of instruments into the vasculature. The device is configured to be used for arterial or venous access. The InnerDyne, Inc., Radially Expanding Vascular Access System (Heparin Coated) is intended to provide percutaneous access to the vasculature. The Radially Expanding Vascular Access System (Heparin Coated) is indicated for the following usages: Percutaneous vascular access.
Device Story
System consists of expandable sleeve assembly, tapered dilator, and sheath with integral hemostasis valve. Operates via Seldinger technique; expandable sleeve inserted over guidewire to dilate tissue; dilator/sheath assembly inserted through sleeve lumen, further expanding sleeve and surrounding tissue. Dilator removed, leaving sleeve and sheath as sealed port for instrument passage. Allows sequential placement of different diameter sheaths. Used in clinical settings by physicians for minimally invasive procedures. Heparin coating applied to sleeve surface to reduce thrombogenicity. Benefits include controlled, sequential vascular access with reduced trauma via radial expansion.
Clinical Evidence
No clinical data. Bench testing only.
Technological Characteristics
Expandable sleeve assembly, tapered dilator, and sheath with integral hemostasis valve. Materials: medical grade, assembled via adhesives, welds, and mechanical interlocks. Surface treatment: heparin coating. Working diameter range: 6-12F. Sterilization: not specified.
Indications for Use
Indicated for percutaneous vascular access in patients requiring temporary arterial or venous access for instrument passage during minimally invasive surgery.
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
Predicate Devices
InnerDyne, Inc., Radially Expanding Vascular Access System (K972684)
Reference Devices
Research Medical Pediatric Femoral Arterial Cannula (K861641)
Related Devices
K972684 — INNERDYNE RADIALLY EXPANDING VASCULAR ACCESS SYSTEM · Innerdyne, Inc. · Oct 10, 1997
K992668 — INNERDYNE RADIALLY EXPANDING VASCULAR ACCESS SYSTEM (REVAS) · Innerdyne, Inc. · May 26, 2000
K990854 — INNERDYNE RADIALLY EXPANDING DILATION, RED DEVICE · Innerdyne, Inc. · Aug 2, 1999
K160254 — GORE DrySeal Flex Introducer Sheath · W. L. Gore & Associates, Inc. · May 12, 2016
K173750 — Prelude IDeal Hydrophilic Sheath Introducer · Merit Medical Systems, Inc. · Feb 15, 2018
Submission Summary (Full Text)
{0}------------------------------------------------
## DVne. ne-
K980578
## SUMMARY OF SAFETY & EFFECTIVENESS INFORMATION
InnerDyne, Inc. Submitter: 5060 Amelia Earhart Dr Salt Lake City, Utah (801)-350-3600
Date Prepared: February 13, 1998
Rick Gaykowski Contact: Corporate Vice President, Regulatory Affairs and Ouality Assurance
Classification Name: Percutaneous Introducer Common/Usual Name: Percutaneous Introducer With Sheath Trade/Proprietary Name: InnerDyne Radially Expanding Vascular Access System (Heparin Coated)
The Radially Expanding Vascular Access System (Heparin Coated) consists of an expandable sleeve assembly. The tubular member of the expandable is configured so as to be radially compressed to reduce the outside diameter of the device prior to insertion. Upon use, the expandable sleeve assembly is inserted through the skin and surrounding tissue into the selected ver a previously placed guidewire according to the techniques of Seldinger. (An alternative placement method would have the needle placed within the lumen of the expandable sleeve, and the combined needle and sleeve inserted into the target vessel. The guidewire would be placed and the needle removed.) The dilator/sheath assembly, which consists of a tapered dilator and sheath with integral hemostasis valve, is inserted over the guidewire through the lumen of the expandable sleeve which expands radially to accommodate the dilator/sheath. This process in turn, radially expands the surrounding tissue. Following dilation the dilator is removed, leaving the expandable sleeve and sheath with integral hemostasis valve in place to provide a sealed port for passage of instruments.
This system configuration allows the user to initially place a small diameter sheath for passage of small instruments. The sheath can then be removed from the sleeve while leaving the sleeve inserted in the vessel. A larger diameter dilator/sheath can then be inserted through the sleeve to create a larger port for passage of the larger instruments.
The device is assembled from medical grade materials under GMP and ISO conditions. Components are molded and machined by qualified suppliers. The components are assembled and secured by adhesives, welds, and mechanical interlocks.
The subject InnerDyne, Inc., Radially Expanding Vascular Access System (Heparin Coated) device is substantially equivalent to the predicate InnerDyne, Inc., Radially Expanding Vascular Access System device (K972684). The subject Radially Expanding Vascular Access System (Heparin Coated) device is similar to the referenced predicate in size, function, product dimensions and indications for use.
{1}------------------------------------------------
The Radially Expanding Vascular Access System (Heparin Coated) is intended for use during minimally invasive surgery for temporary percutaneous vascular access for passage of instruments into the vasculature. The device is configured to be used for arterial or venous access.
The basic design principles for the subject InnerDyne, Inc., Radially Expanding Vascular Access System (Heparin Coated) device and the predicate Radially Expanding Vascular Access System device (K972684) are similar, and remain essentially unchanged from information previously provided to the Agency. The product configuration, composition, and utilized materials are similar in each of the products. The principles of operation for the InnerDyne, Inc., Radially Expanding Vascular Access System (Heparin Coated) device, and the predicate Radially Expanding Vascular Access System device (K972684) are similar. That is, each of these products employs a similar insertion technique, indications for use, contraindications for use, warnings and precautions. The subject device differs from the referenced Radially Expanding Vascular Access System predicate (K972684) with the addition of a heparin coating to the surface of the expandable sleeve portion of the device, and an expansion of the sheath working diameter range from (6 - 12F). The inclusion of a heparin coating on an offered product is a common industry practice in distributed devices which come in contact with circulating blood. An example of a substantially equivalent heparin coated device is the Research Medical Pediatric Femoral Arterial Cannula, which contains the DuraFlow II Heparin Coating cleared for marketing under K861641.
From the foregoing, we conclude that the Radially Expanding Vascular Access System (Heparin Coated) device is as safe and effective as currently marketed devices for the stated indications.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo features a stylized image of three human profiles facing right, stacked on top of each other. The profiles are rendered in a dark color, creating a silhouette effect. Encircling the image is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 1 1 1998
Mr. Rick Gaykowski Corporate Vice President, RA/QA InnerDyne, Inc. 5060 West Amelia Earhart Dr Salt Lake City, UT 84116
Re: K980578 Trade Name: InnerDyne Radially Expanding Vascular Access System (Heparin Coated) Requlatory Class: II Product Code: DYB Dated: June 12, 1998 Received: June 15, 1998
Dear Mr. Gaykowski:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to leqally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this
{3}------------------------------------------------
Page 2 - Mr. Rick Gaykowski
response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4586. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Judith Lanierison
802 Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page 1 of 1
510(k) Number (if known):
Device Name: InnerDyne, Inc., Radially Expanding Vascular Access System
.
Indications for Use: The InnerDyne, Inc., Radially Expanding Vascular Access System (Heparin Coated) is intended to provide percutaneous access to the vasculature. The Radially Expanding Vascular Access System (Heparin Coated) is indicated for the following usages:
Percutaneous vascular access.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Judith Aanebsox for Boyle Gantt
(Division Sign-Off) Acting Brunch Chief ICOC
Division of Cardiovascular, Respiratory,
and Neurological Devices K980578
510(k) Number
Prescription Use (Per 21 CFR 801.109) OR
Over-the-Counter Use
(Optional Format 1-2-96)
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.