The Braun Percutaneous Introducer Kits are designed to facilitate percutaneous introduction of devices into the vasculature.
Device Story
Braun Percutaneous Introducer Kits consist of introducer sheath, dilator, guidewire, and introducer needle. Used by clinicians to facilitate percutaneous vascular access for subsequent device placement. Sheath/dilator assembly inserted into vasculature; provides conduit for catheter or lead delivery. Benefits include simplified vascular access and reduced trauma during device introduction.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Materials: Teflon FEP or HDPE sheath, HDPE dilator. Dimensions: 3-9 Fr, 7-24 cm length. Mechanical device; no energy source or software. Non-active.
Indications for Use
Indicated for percutaneous introduction of medical devices into the vasculature in patients requiring vascular access.
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
K991647 — TFX MEDICAL INTRODUCER KIT · Tfx Medical Group · Jan 4, 2000
K140768 — TYPE I BRILLIANT INTRODUCER KIT, TYPE II BRILLIANT INTRODUCER KIT, TYPE III BRILLIANT INTRODUCER KIT, TYPE IV BRILLIANT · Lepu Medical Technology (Bejing) Co., Ltd. · Nov 25, 2014
K123974 — PERCUTANEOUS INTRODUCER KIT · Teleflex, Inc. · Feb 21, 2013
K162097 — InTRAkit · Medtronic Vascular · Nov 17, 2016
K172331 — Type I, Type II, Type III, Type IV · Lepu Medical Technology (Beijing) Co., Ltd. · Mar 2, 2018
Submission Summary (Full Text)
{0}
K961745
SEP 17 1996
# II. SUMMARY OF SAFETY AND EFFECTIVENESS
Applicant Name & Address: B. Braun Medical Inc., 824 Twelfth Avenue
Bethlehem, PA 18018 (610)691-5400
Contact: Mark Alsberge, Regulatory Affairs Manager
Date Prepared: 05/01/96
Common or Usual Name: Braun Percutaneous Introducer Set
Device Classification Name: Catheter Introducers & vessel dilators
## Substantial Equivalence:
K810460
K894446
K894343 and/or K884494
K771982 and/or K770561
K780126
K850820
K932733
Burron Percutaneous Introducer Set
Angeion Hemostasis Introducer Set
Venous Introducer
"888" Percutaneous Catheter Introducer
PCI Kit
Arterial/Venous Percutaneous Catheter Introducer
Valveless/Portless Sheath Introducer
B. Braun Medical, Inc.
B. Braun Medical, Inc.
Daig
USCI
Argon
Universal Medical Instrument(UMI)
Cordis
## Device Description:
Braun Percutaneous Introducer Kits consist of an introducer sheath packaged with a corresponding dilator. A guidewire and introducer needle may also be packaged with the sheath and dilator.
## Statement of Intended Use:
The Braun Percutaneous Introducer Kits are designed to facilitate percutaneous introduction of devices into the vasculature.
## Summary/Comparison of Technological Characteristics:
| | Braun PCI | K810460 | K894446 | Daig | USCI | Argon | UMI | Cordis |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Fr sizes | 3 - 9 | 6-8 | 4-15 | 4 - 9 | 4 - 9 | 4 - 9 | 4 - 8 | 6 - 8 |
| Lengths | 7 - 24 cm | not specified in 510(k) | 12 - 24 cm | 8.5 - 12 cm | 8 - 12 cm | unknown | 14 cm | 11 cm |
| Matls | Teflon FEP or HDPE sheath with HDPE dilator | Teflon® FEP sheath & dilator | Teflon® FEP sheath and HDPE dilator | unknown - "proprietary materials" | Teflon | unknown | white teflonsheath with polyethylene dilator | unknown |
| Sideport | no | yes | yes | no | no | no | no | no |
| Gasket | no | no | yes | no | no | no | no | no |
| Use | percutaneous introduction of devices into the vasculature | percutaneous introduction of devices into the vasculature | percutaneous introduction of catheters into the vasculature while minimizing the backflow of blood. | Designed for introduction of electrophysiology and other catheters into the venous system | percutaneous introduction of catheters into the vasculature (either venous or arterial) | percutaneous introduction of balloon catheters, closed end catheters and temporary pacing leads into the vasculature. | percutaneous introduction of catheters into the vasculature (either venous or arterial) | percutaneous introduction of catheters into the vasculature |
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.