K955820 · B.Braun Medical, Inc. · DYB · Mar 25, 1996 · Cardiovascular
Device Facts
Record ID
K955820
Device Name
BRAUN HEMOSTASIS INTRODUCERS
Applicant
B.Braun Medical, Inc.
Product Code
DYB · Cardiovascular
Decision Date
Mar 25, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1340
Device Class
Class 2
Indications for Use
For percutaneous introduction of catheters into the vasculature while preventing the backflow of blood.
Device Story
Device consists of 16 and 18 French catheter introducers and vessel dilators; includes silicone gasket to prevent blood backflow. Used by clinicians for percutaneous vascular access. Introducer sheath (FEP with BaSO4) and dilator (high-density polyethylene with bismuth subcarbonate) facilitate catheter placement. Device functions as a mechanical conduit for vascular procedures.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Sheath: FEP with BaSO4. Dilator: High-density polyethylene with bismuth subcarbonate. Gasket: Silicone. Sizes: 16 and 18 French. Lengths: 10-40 cm. Packaging: Tyvek/Polymylar pouch. Sterilization: Not specified.
Indications for Use
Indicated for percutaneous introduction of catheters into the vasculature in patients requiring vascular access.
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
K963388 — VESSEL DILATOR · Maxxim Medical · Mar 11, 1997
K983023 — MODIFICATION TO CORDIS BRITE TIP CATHETER SHEATH INTRODUCER SYSTEM · Cordis Corp. · Oct 1, 1998
Submission Summary (Full Text)
{0}
3
90
# II. SUMMARY OF SAFETY AND EFFECTIVENESS
Applicant Name & Address: B. Braun Medical Inc., Cardiovascular Division, 13000 Hwy 55
Minneapolis, MN 55441-3813 (612) 559-3370
Contact: Barbara Atzenhoefer, Regulatory Affairs Manager
Date Prepared: December 19, 1995
Common or Usual Name: Catheter introducer and Vessel dilator for percutaneous catheterization
Device Classification Name:
Introducer: 21 CFR Reference- 870.1340, Class II Panel- 74 DYB
Dilator: 21 CFR Reference- 870.1310, Class II, Panel- 74 DRE
Substantial Equivalence:
| 510(k) # | Product | Manufacturer |
| --- | --- | --- |
| K894446 | Angeion Hemostasis Valve Introducer Kit | B. Braun Cardiovascular - MN |
| K931743 | Burron Cardiovascular Angestat | B. Braun Cardiovascular - MN |
| K871153 | ANGETEAR Tear-away Introducer Kits | B. Braun Cardiovascular - MN |
| K930057 | Fluoro-tip Hemostasis Introducer Set | B. Braun Cardiovascular - MN |
| K945396 | Fluoro-tip Hemostasis Introducer Set | B. Braun Cardiovascular - MN |
Device Description:
16 & 18 french versions of the Braun Hemostasis Introducer with silicone gasket, in lengths of 10 - 40 cm.
Statement of Intended Use:
For percutaneous introduction of catheters into the vasculature while preventing the backflow of blood.
Summary/Comparison of Technological Characteristics:
| Characteristic | Braun Hemostasis Introducers | K894446 Hemostasis Introducer | K931743 Angestat | K871153 ANGETEAR |
| --- | --- | --- | --- | --- |
| A) Labeling/Pkg | | | | |
| Labels & IFUs | See Attachment B | All Braun Hemostasis Introducer Label formats are identical | All Braun Hemostasis Introducer Label formats are identical | same information provided on pouch and box labels as on hemostasis labels |
| Expiration Dating | 2 years | 2 years | 2 years | 4 years |
| Pyrogen Free | Yes | Yes | Yes | Yes |
| Packaging | Tyvek/Polymylar pouch | Tyvek/Polymylar pouch | Tyvek/Polymylar pouch | Tray w/ tyvek lid w/in a tyvek/polymylar pouch |
| B) Intended Use | percutaneous introduction of catheters into the vasculature while minimizing the backflow of blood | percutaneous introduction of catheters into the vasculature while minimizing the backflow of blood | percutaneous introduction of catheters into the vasculature while minimizing the backflow of blood | percutaneous introduction of catheters into the vasculature |
| C) Physical Char. | | | | |
| Available french sizes | 4-18 french | 5-14 french | 4-15 french | 4-16 french |
| Lengths | 10-40 cm | 12-24 cm | 15-32 cm | 16-30 cm |
| gasket material | silicone | Natural rubber | silicone | none |
| sheath material | FEP with BaSO4 | FEP teflon with BaSO4 | FEP with BaSO4 | polyethylene with bismuth subcarbonate |
| dilator material | high density polyethylene with bismuth subcarbonate | polyethylene with bismuth subcarbonate | polyethylene with bismuth subcarbonate | FEP with BaSO4, or polyethylene with bismuth subcarbonate |
| D) Anatomical Sites | vasculature | vasculature | vasculature | vasculature |
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.