Super Sheath Introducer Sheath S-3N5/1005 S-3N7/1006 S-3W5/1007 S-3W7/1008
Applicant
Togo Medikit Co., Ltd.
Product Code
DYB · Cardiovascular
Decision Date
Sep 17, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1340
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Super Sheath Introducer Sheath is indicated for use in the introduction of diagnostic and interventional devices inserted into the human vasculature of adult and pediatric patients of all ages.
Device Story
The Super Sheath Introducer Sheath facilitates the introduction of diagnostic or interventional devices into blood vessels. The device consists of a sheath (with hub, shaft, and three-way stopcock valve), a dilator, and an optional guidewire. It is used in clinical settings by healthcare professionals. The sheath hub includes a hemostatic valve and suture wing. The dilator features a tapered tube for guidewire insertion. The device is provided sterile for single use. By providing a stable access point, it allows clinicians to safely insert catheters or other interventional tools into the vasculature, potentially benefiting patients by enabling minimally invasive diagnostic or therapeutic procedures.
Clinical Evidence
Bench testing only. Testing included tensile, kink, connection strength, pressure, lubricity, corrosion resistance, and radiopacity tests. Biocompatibility testing included cytotoxicity, intracutaneous reactivity, sensitization, acute systemic toxicity, hemocompatibility, pyrogen, and LAL tests.
Technological Characteristics
3.3F sheath available in 5cm and 7cm lengths. Components: sheath hub, shaft, three-way stopcock (ISO 80369-7:2021 compliant), dilator, and 0.018" guidewire. Materials include plastic components with minor changes from predicate. Sterilization via Ethylene Oxide. Shelf-life: 3 years. Single-use.
Indications for Use
Indicated for introduction of diagnostic and interventional devices into human vasculature for adult and pediatric patients of all ages.
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
Predicate Devices
Super Sheath Introducer Sheaths and Super Sheath Introducer Sheath Set (K121504)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 17, 2024
Togo Medikit Co., Ltd. Fumiaki Kanai President MIC International Corp. 4-32-16 Ryogoku Sumida-ku Tokyo, 130-0026 Japan
Re: K241230
Trade/Device Name: Super Sheath Introducer Sheath Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II Product Code: DYB, DRE Dated: May 1, 2024 Received: May 2, 2024
Dear Fumiaki Kanai:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
# Sincerely, Sevan R. Sevan R. _____________________________________________________________________________________________________________________________________________________________________
For
Misti Malone, PhD Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K241230
Device Name Super Sheath Introducer Sheath
Indications for Use (Describe)
The Super Sheath Introducer Sheath is indicated for use in the introduction of diagnostic and interventional devices inserted into the human vasculature of adult and pediatric patients of all ages.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary
#### a. Owner/Company name, address
Togo Medikit Co., Ltd. 17148-6 Aza Kamekawa, Oaza Hichiya Hyuga City, Miyazaki Prefecture, 883-0062, Japan Daisuke Nagamizu, Manager, Regulatory Affairs Section, Quality Assurance Department Phone: 011-81-982-53-8000 Email: qc@togomedikit.co.jp
#### b. Official Correspondent
Fumiaki Kanai, President MIC International 4-32-16 Ryogoku, Sumida-ku Tokyo, 130-0026, Japan Phone: 011-81-3-6659-5482 Email: kanaif@mici.co.jp
c. Date prepared September 10, 2024
#### d. Name of device
Trade Name: Super Sheath Introducer Sheath Common Name: Introducer, Catheter Classification Name: Catheter Introducer Classification Regulation: 21 CFR 870.1340 Product Code: DYB Classification Panel: Cardiovascular
#### e. Predicate device
510(k): K121504 Trade name: Super Sheath Introducer Sheaths and Super Sheath Introducer Sheath Set Regulation Number: 21 CFR 870.1310 Product code: DYB
{5}------------------------------------------------
## f. Description of the device
The Super Sheath Introducer Sheath (hereinafter referred to as the "Subject Device") is used to introduce diagnostic or interventional devices into a blood vessel. The Subject Device is available in two products: one consisting of a Sheath and Dilator (Super Sheath Introducer Sheath without Guidewire), and the other consisting of a Sheath, Dilator, and Guidewire (Super Sheath Introducer Sheath with Guidewire), similar to the Predicate Device K121504. These devices are provided sterile and intended for single use, and the shelf-life is three years.
The main component of the Sheath consists of a Sheath hub, a Sheath shaft connected to the Sheath hub, and a three-way stopcock valve connected to the hub with a side port tube. The Sheath shaft is 3.3 French size and available in 5cm and 7cm lengths. The Sheath hub contains a hemostatic valve and has a suture wing. The shape and dimensions of the three-way stopcock luer connector are modified to comply with the new applicable standard ISO 80369-7:2021.
The Dilator has an open, tapered plastic tube for Guidewire insertion. The Dilator tube is press-fit into the inner hub with a bushing. The length of the Dilator tube corresponds to the length of the Sheath shaft, which is longer and available in 10.5 cm and 12.5 cm.
The 0.018" Guidewire is the smallest size Guidewire and a 40 cm long, thin, flexible wire with a inserter attached to the tip. This Guidewire is a legally marketed product which is cleared by the Food and Drug Administration (K021990).
The upper cap and suture wing of the Sheath hub, and the Dilator hub are colored purple. The Dilator tube and the inserter are colored blue. And the unvented cap of the three-way stopcock is pink.
#### g. Indications for Use
The Super Sheath Introducer Sheath is indicated for use in the introduction of diagnostic and interventional devices inserted into the human vasculature of adult and pediatric patients of all ages.
#### h. Technological Characteristics
The following table compares the Super Sheath Introducer Sheath to the predicate device with respect to intended use and technological characteristics, providing more detailed information regarding the basis for the determination of substantial equivalence.
| Feature | Subject Device -Super<br>Sheath Introducer<br>Sheath - K241230 | Predicate Device - Super<br>Sheath Introducer<br>Sheaths and Super<br>Sheath Introducer<br>Sheath Sets - K121504 | Comparison |
|--------------------------------------------|---------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------|
| French Sizes<br>(available) | 3.3F | 3.3 F | Same |
| Indications for Use | Super Sheath Introducer<br>Sheath is indicated for use<br>in the introduction of<br>diagnostic and | The Super Sheath<br>Introducer Sheaths and<br>Super Sheath Introducer | Same |
| | | | |
| | interventional devices<br>inserted into the human<br>vasculature of adult and<br>pediatric patients of all<br>ages. | Sheath Sets are indicated<br>for<br>use in the introduction of<br>diagnostic and<br>interventional devices<br>inserted into the human<br>vasculature of adult and<br>pediatric patients of all<br>ages. | |
| Regulation Number | 870.1340 | 870.1340 | Same |
| FDA Product Code | DYB | DYB | Same |
| Prescription/<br>OTC Use | Prescription | Prescription | Same |
| Components | Sheath, Dilator and<br>Guidewire | Sheath, Dilator and<br>Guidewire | Same |
| Single-Use/<br>Reusable | Single-Use | Single-Use | Same |
| Dimensional Comparison | | | |
| Effective Length | $50 mm - 70 mm$ | $50 mm – 70 mm$ | Same |
| Sheath Shaft Inner<br>Diameter (ID) | 3.3Fr | 3.3Fr | Same |
| Three-way stopcock:<br>luer lock connecter | Meets ISO 80369-7-21 | Meets ISO 80369-7-21.<br>This change has already<br>been implemented and<br>documented in the LTF. | Same |
| Dilator Tube Effective<br>Length | $105 mm + 5/-5 mm$<br>$125 mm + 5/-5 mm$ | $105 mm + 5/-5 mm$<br>$125 mm + 5/-5 mm$ | Same |
| Dilator Distal Tip | 0.018"<br>(Only 0.018" size is<br>available) | 0.018"<br>0.021"<br>0.025" | Different<br>Removal of the 0.021"<br>and 0.025" sizes |
| Guidewire | | | |
| Tip Shape | Straight type | Straight type | Same |
| Outer Diameter | Max OD 0.018" | Max OD 0.018" | Same |
| Length - Nominal | $400 mm \pm 5 mm$ | $400 mm \pm 5 mm$ | Same |
| Sterilization and Shelf Life | | | |
| Sterilization Method | Ethylene Oxide | Ethylene Oxide | Same |
| Sterile Package | Pouch | Pouch | Same |
| Shelf-Life (Use By<br>Date) | Three (3) years | Three (3) years | Same |
{6}------------------------------------------------
# i. Substantial Equivalence
Many of the materials of the Predicate Device are identical to that of the subject device excepting minor material changes for the Dilator tube, inserter and case tube components. In addition, there are changes in solvents in the manufacturing process. The changes were validated via performance
{7}------------------------------------------------
testing, sterilization validation, shelf-life testing, and biocompatibility testing including a biocompatibility risk assessment. The results of the validation testing and the risk assessment conducted demonstrated substantial equivalence when compared to the Predicate Device (K121504).
## j. Bench Testing
The following bench tests were performed to demonstrate substantial equivalence to the Predicate Device (K121504) and verify conformity to the international standards and in-house requirements.
- 1. Sheath Shaft Tensile Test
- 2. Sheath Shaft Kink Test
- Sheath Hub to Sheath Shaft Connection Strength Test 3.
- 4. Sheath Hub to Side Tube Connection Strength Test
- న. Three-Way Stop Cock to Side Tube Connection Strength Test
- 6. Sheath Hemostatic Valve Pressure Test
- 7. Sheath Pressure Test
- Sheath Lubricity Test 8.
- 9. Dilator Corrosion Resistance Test
- 10. Dilator Tube Tensile Test
- 11. Dilator Tube to Hub Connection Strength Test
- 12. Sheath Radiopacity Test
- Guidewire Radiopacity Test 13.
#### k.Biocompatibility Testing
We performed the following biocompatibility tests after risk assessment of biocompatibility.
- 1. Cytotoxicity
- 2. Intracutaneous reactivity
- 3. Sensitization
- 4. Acute Systemic Toxicity
- 5. Hemocompatibility
- 6. Pyrogen test
- 7. LAL test
#### l. Conclusion
Based on the above discussion and the enclosed sections regarding substantial equivalence to the predicate device, Togo Medikit Co., Ltd. concludes that the Subject Device is substantially equivalent to the Predicate Device (K121504), and the Subject Device does not raise any new questions regarding safety or effectiveness.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.