K221440 · Bard Peripheral Vascular, Inc. · DYB · May 31, 2022 · Cardiovascular
Device Facts
Record ID
K221440
Device Name
Liverty TIPS Access Set
Applicant
Bard Peripheral Vascular, Inc.
Product Code
DYB · Cardiovascular
Decision Date
May 31, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1340
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The BD Liverty TIPS Access Set is indicated for percutaneous transjugular liver access during diagnostic and interventional procedures.
Device Story
The Liverty TIPS Access Set is a percutaneous transjugular liver access system used by physicians in clinical settings for diagnostic and interventional procedures, such as TIPS stent placement. The set includes an introducer sheath, a steerable cannula (TipsStar), a hollow needle, a 5F catheter, and dilators. The operator adjusts the cannula curve in-situ via an orange knob. The needle/5F catheter assembly is advanced through the liver parenchyma into the portal vein. The hollow needle allows for immediate confirmation of portal vein access via blood withdrawal and contrast injection. A removable puncture depth adapter on the 5F catheter hub allows for a 1 cm increase in puncture depth. The introducer sheath is advanced over the tract to facilitate the procedure. The device improves procedural flow by reducing the need for component removal and replacement compared to traditional sets.
Clinical Evidence
Bench testing only. No clinical data. Performance evaluated via non-clinical testing including biocompatibility (ISO 10993-1), simulated use, radiopacity, tensile force, torque resistance, kink resistance, radial strength, and sterilization validation (ISO 11135).
Technological Characteristics
Components: introducer sheath, steerable cannula, hollow needle, 5F catheter, 10F/12F dilators. Features: in-situ adjustable cannula curve, hollow needle for contrast/blood aspiration, removable 1cm puncture depth adapter. Sterilization: Ethylene Oxide (ISO 11135, SAL 10^-6). Biocompatibility: ISO 10993-1 compliant. Mechanical testing: tensile, torque, kink, radial strength, and simulated use.
Indications for Use
Indicated for percutaneous transjugular liver access during diagnostic and interventional procedures in patients requiring such access.
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
Predicate Devices
Rösch-Uchida Transjugular Liver Access Set (K171820)
Reference Devices
Ring Transjugular Intrahepatic Access Set (K171820)
K152913 — GORE TIPS Set, GORE TIPS Sheath, GORE TIPS Needle · Creganna Medical Aslo Business AS Cregganna Tactx Medical · Jul 21, 2016
K213002 — SCORPION Portal Vein Access Set · Argon Medical Devices · Oct 14, 2021
K201489 — Traveler38 0.038 Stylet Portal Vein Access Set, Traveler21 21ga Needle Portal Vein Access Set, Traveler16 16ga Needle Portal Vein Access set · Argon Medical Devices, Inc. · Dec 22, 2020
K202141 — Scorpion Stylet Portal Vein Access Set, Scorpion Needle Portal Vein Access Set · Argon Medical Devices, Inc. · Jan 15, 2021
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 31, 2022
Bard Peripheral Vascular, Inc % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114
Re: K221440
Trade/Device Name: Liverty™ TIPS Access Set Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter introducer Regulatory Class: Class II Product Code: DYB Dated: May 6, 2022 Received: May 24, 2022
Dear Prithul Bom:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Misti Malone, PhD Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K221440
Device Name Liverty TIPS Access Set
Indications for Use (Describe)
The BD Liverty TIPS Access Set is indicated for percutaneous transjugular liver access during diagnostic and interventional procedures.
| Type of Use (Select <i>one</i> or <i>both</i> , as applicable) |
|----------------------------------------------------------------|
|----------------------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Liverty™ TIPS Access Set
## 510(k) Summary 21 CFR 807.92
As required by the Safe Medical Devices Act of 1990, coded under Section 513, Part (I)(3)(A) of the Food, Drug and Cosmetic Act, a 510(k) summary upon which substantial equivalence determination is based is as follows:
#### Submitter Information:
| Applicant: | Bard Peripheral Vascular, Inc.<br>1625 West 3rd Street<br>Tempe, Arizona 85281 |
|---------------------|--------------------------------------------------------------------------------|
| Phone: | (602) 830-5517 |
| Fax: | (312) 949-0436 |
| Contact Person: | Scott Johnson, Senior Regulatory Affairs Specialist |
| Date of Submission: | May 6, 2022 |
#### Subject Device:
| Name of Device: | Liverty™ TIPS Access Set |
|-----------------------|------------------------------------------|
| Common or Usual Name: | Introducer, Catheter (Product Code: DYB) |
| Classification Panel: | Cardiovascular |
| Regulatory Class: | II |
| Regulation Number: | 21 CFR 870.1340 |
### Predicate Device:
Rösch-Uchida Transjugular Liver Access Set (K171820, cleared on March 9, 2018)
#### Reference Device:
Ring Transjugular Intrahepatic Access Set (K171820, cleared on March 9, 2018)
### Device Description:
The BD Liverty™ TIPS Access Set consists of an introducer sheath, a steerable cannula, a needle, a 5F catheter, a 10F dilator, and a 12F dilator. This device is intended for percutaneous transjugular liver access during diagnostic and interventional procedures.
The curve at the distal end of the TipsStar™ steerable cannula can be adjusted by the operator in-situ, by turning the orange knob at the end of the cannula to achieve the optimal cannula curve angle. Before introduction of the needle into the cannula, the bright beige 5F catheter must be slid over the needle. This assembly is introduced as one unit through the cannula. The operator
{4}------------------------------------------------
then thrusts the needle / 5F catheter assembly through the liver parenchyma and into the portal vein to create the shunt. The introducer sheath is then advanced until it is positioned across the parenchymal tract. Following completion of diagnostic or interventional procedures, such as placement of a TIPS stent, the introducer sheath is removed.
## Intended Use / Indications for Use:
The BD Liverty™ TIPS Access Set is indicated for percutaneous transjugular liver access during diagnostic and interventional procedures.
## Technological Comparison to Predicate Device:
The Liverty™ TIPS Access Set has the following similarities to the predicate device, the Rösch-Uchida Transjugular Liver Access Set (K171820, cleared on March 9, 2018):
- . Same intended use
- . Same indications for use
- . Same target patient population
- . Same principle of operation
- . Same fundamental scientific technology
- . Same imaging capabilities
- Same sterility level and method of sterilization .
All components of the subject device are of similar design and function as the components of the predicate device. The subject device has three additional features compared to the predicate device. Namely:
- 1. The stiffening cannula on the predicate device comes with a pre-set angle that must be manually adjusted ex vivo if the angulation is not appropriate for obtaining portal vein access. In contrast, the angulation of the TipsStar™ steerable cannula can be adjusted by the operator in situ throughout the procedure.
- 2. The predicate device uses a stylet to puncture through the liver parenchyma. To confirm portal vein access, the stylet must be removed. The needle in the subject device is hollow, which allows confirmation of portal vein access via blood withdrawal and contrast injection to be completed with the needle in place. This feature is leveraged from the design of the reference device, the Ring Transjugular Intrahepatic Access Set (K171820, cleared on March 9, 2018).
{5}------------------------------------------------
- 3. The subject device is provided with a removable puncture depth adapter attached to the 5F catheter hub. Unscrewing and separating the adapter from the 5F catheter hub increases the puncture depth of the needle / 5F catheter assembly by 1 cm. The predicate device does not have a similar depth adapter.
Each of these features are intended to improve procedural flow by minimizing the number of times access set components must be removed / replaced during a TIPS procedure.
## Biocompatibility Testing:
Per ISO 10993-1 and FDA guidance, testing for physical and/or chemical characterization, cytotoxicity, sensitization, irritation/intracutaneous reactivity, material-mediated pyrogenicity, acute systemic toxicity, and hemocompatibilty (direct and indirect hemolysis, complement activation, uPTT, in vivo thrombogenicity) tests were performed on all device components of the subject device or on representative devices. All test results met the acceptance criteria, where applicable, or demonstrated that the device is biocompatible.
## Performance Data:
To demonstrate substantial equivalence of the subject device, the Liverty™ TIPS Access Set, to the predicate device, both technological characteristics and performance criteria were evaluated. Using FDA Guidance Documents on non-clinical testing of medical devices and internal Risk Assessment procedures, tests for the following tests were performed:
- . Introducer Sheath Inner Diameter
- . Needle Assembly Length over Cannula
- Lie Distance .
- Sheath Radiopacity .
- . Needle, 10F Dilator, 12F Dilator, 5F Catheter Radiopacity
- . Needle Tip Ultrasound Visibility
- Device Surface Quality .
- . Needle Point
- Needle Assembly Flexible Portion Stiffness .
- . Needle Assembly Durability
- Cannula and Needle Corrosion Resistance .
- . Contrast Media Resistance
- . Freedom from Leakage from Hemostasis Valve
{6}------------------------------------------------
- . Freedom from Leakage from Sheath
- . 5F Catheter Leakage
- Needle Assembly Air Leak .
- Sheath Peak Tensile Force .
- . Sheath Torque Resistance
- Sheath Tracking Durability .
- . 10F Dilator, 12F Dilator, 5F Catheter Peak Tensile Force
- . Needle Tube and Needle Hub Joint Strength
- . Cannula Curve Angle
- Cannula Curve Durability .
- Luer Performance .
- . Packaging Performance
- Human Factors Evaluation .
- . 10F Dilator Stiffness
- Introducer Sheath Kink Resistance .
- Introducer Sheath Radial Strength .
- Needle Assembly Penetration Force .
- Radiopacity (Introducer Sheath) .
- . Needle Tip Ultrasound Visibility
- . Simulated Use Testing
The results from these tests demonstrate that the technological characteristics and performance criteria of the Liverty™ TIPS Access Set are comparable to the predicate device and that it can perform in a manner equivalent to devices currently on the market for the same intended use.
## Sterilization:
The Liverty™ TIPS Access Set is provided sterile. Sterilizations were performed according to ISO 11135 and demonstrate that the sterilization chambers and cycle deliver a sterility assurance level (SAL) of 10-6 for the sterilization of the Liverty™TIPS Access Set.
## Conclusions:
The subject device, the Liverty™ TIPS Access Set, and the predicate device share the same or similar characteristics: intended use, indications for use, target patient population, principle of operation, fundamental scientific technology, imaging capabilities, and sterility level and method
{7}------------------------------------------------
of sterilization. Additionally, the Liverty™ TIPS Access Set met all predetermined acceptance criteria of design verification and validation as specified by applicable standards, guidance, test protocols and/or customer inputs. Non-clinical testing demonstrated that the subject device, the Liverty™ TIPS Access Set, is as safe and effective as the legally marketed predicate device, the Rösch-Uchida Transjugular Liver Access Set. The available evidence demonstrates that the Liverty™ TIPS Access Set is substantially equivalent to Rösch-Uchida Transjugular Liver Access Set and does not raise new questions of safety or effectiveness.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.