K201116 · Abiomed, Inc. · DYB · Jun 15, 2020 · Cardiovascular
Device Facts
Record ID
K201116
Device Name
Abiomed 23 Fr Sheath
Applicant
Abiomed, Inc.
Product Code
DYB · Cardiovascular
Decision Date
Jun 15, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1340
Device Class
Class 2
Indications for Use
The Abiomed 23 Fr Sheath is intended for introduction of pacing leads and catheters into the body.
Device Story
Abiomed 23 Fr Sheath is a sterile introducer kit; includes sheath with sideport tubing, stopcock, hemostasis valve assembly, dilators, and guidewire. Used by clinicians to facilitate percutaneous introduction of pacing leads or catheters into the vasculature. Device is identical to predicate; no complex electronics or software involved.
Clinical Evidence
No clinical data; bench testing only (not applicable as device is identical to predicate).
Technological Characteristics
Catheter introducer kit; includes sheath, sideport tubing, stopcock, hemostasis valve, dilators, and guidewire. Sterile. No software or electronic components.
Indications for Use
Indicated for the introduction of pacing leads or catheters into the body. Prescription use only.
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
K192769 — Impella CP Introducer · Abiomed, Inc. · Nov 18, 2019
K252766 — 14Fr Low Profile Introducer Kit · Abiomed, Inc. · Oct 8, 2025
K222184 — Introducer Sheath Kits · Beijing Demax Medical Technology Co.,Ltd · Feb 21, 2023
K022513 — MEDCOMP VASCU-SHEATH INTRODUCER SET · Medcomp · Oct 25, 2002
K092575 — ENGAGE/ENGAGE TR · St Jude Medical · Oct 8, 2009
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 15, 2020
Abiomed Inc. Sandy Fowler Regulatory Affairs Specialist 22 Cherry Hill Drive Danvers, Massachusetts 01923
Re: K201116
Trade/Device Name: Abiomed 23 Fr Sheath Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II Product Code: DYB Dated: April 24, 2020 Received: April 27, 2020
Dear Sandy Fowler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Hetal Odobasic Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201116
Device Name Abiomed 23 Fr Sheath
Indications for Use (Describe) The Abiomed 23 Fr Sheath is intended for introduction of pacing leads or catheters.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### Abiomed 23 Fr Sheath 510(k) Summary
This summary of 510(k) information is being submitted in accordance with the Safe Medical Devices Act (SMDA) of 1990 and Title 21 of the Code of Federal Regulations, Part 807.92.
### A. Application Information:
| Date Prepared: | April 24th, 2020 |
|-----------------------------|------------------------------------------------------------------------------------------------------------------|
| Submitter's Name & Address: | ABIOMED, Inc.<br>22 Cherry Hill Drive<br>Danvers, MA 01923 |
| Contact Person: | J. Kenneth Ryder<br>Senior Director, Global Regulatory Affairs<br>Ph: 978-646-1707<br>E-mail: kryder@abiomed.com |
### B. Device Information:
| Trade or Proprietary Name: | Abiomed 23 Fr Sheath |
|----------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| Common or Usual Name: | Introducer, Catheter |
| FDA Classification: | Class II, DYB, 21 CFR- 870.1340 |
| Regulation Description: | Catheter Introducer |
| Performance Standard: | Performance standards do not currently exist for these devices.<br>(i.e. none established under section 514 of the F D & C Act.) |
#### ﻥ Predicate Device:
Oscor Adelante-S Introducer Series, K122084
### D. Device Description:
The Abiomed 23 Fr Sheath is packaged as a sterile kit, which includes an introducer sheath (consisting of sheath with sideport tubing, stopcock and hemostasis valve assembly), and dilators, and a guidewire. It is identical to the commercially available predicate device, the Oscor 23 Fr Adelante-S2 Introducers, K122084.
### E. Intended Use:
The Abiomed 23 Fr Sheath is intended for introduction of pacing leads and catheters into the body.
{4}------------------------------------------------
## F. Comparison of Required Technological Characteristics:
All technological characteristics of the Abiomed 23 Fr Sheath are identical to its commercially available predicate, including its design, packaging, manufacturing, sterilization, and labeling,
## G. Comparison to Predicate Device:
The Abiomed 23 Fr Sheath is identical to its commercially available predicate device.
### H. Qualification Testing:
Since the Abiomed 23 Fr Sheath is identical to its commercially available predicate device, no additional qualification testing was required for the substantial equivalence determination.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.