FlexSeal Introducer Sheath with Hydrophilic Coating
K190628 · Freudenberg Medical Mis, Inc. · DYB · May 30, 2019 · Cardiovascular
Device Facts
Record ID
K190628
Device Name
FlexSeal Introducer Sheath with Hydrophilic Coating
Applicant
Freudenberg Medical Mis, Inc.
Product Code
DYB · Cardiovascular
Decision Date
May 30, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1340
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
The FlexSeal® Introducer Sheath with Hydrophilic Coating is intended to be placed in the peripheral vasculature to provide a conduit for the introduction of diagnostic or interventional devices, and to minimize blood loss associated with such insertions.
Device Story
The FlexSeal Introducer Sheath is a sterile, disposable device consisting of an introducer sheath with a hydrophilic coating, an active spring-loaded hemostatic valve, and a radiopaque dilator. It is used in the peripheral vasculature to provide a conduit for diagnostic or interventional devices. The device is operated by a physician. The active hemostatic valve, controlled by a proximal Ring Lock, allows the user to maintain the valve in a partially open position to facilitate device introduction while minimizing blood loss. The device is inserted into the vasculature to enable clinical procedures; the output is the creation of a secure access port. The Ring Lock feature assists the physician by eliminating the need to manually hold the valve open during device passage. The device benefits the patient by providing a controlled access point that reduces blood loss during interventional procedures.
Clinical Evidence
Bench testing only. Performance testing included visual inspection, dimensions, guidewire compatibility, leakage, kink resistance, bond tensile strength, hydrophilic coating integrity, luer compliance, cracking pressure, USP particulate testing, packaging, distribution, simulated use, and shelf life. Biocompatibility testing was performed per ISO 10993-1 for short-duration contact (<24 hours) with circulating blood, including cytotoxicity, sensitization, irritation, systemic toxicity, genotoxicity, and hemocompatibility. Radiopacity, dilator hub dimensions, and corrosion testing were leveraged from the predicate device.
Technological Characteristics
Device consists of a 21Fr, 56 cm sheath with a hydrophilic coating and a radiopaque dilator. Features an active spring-loaded hemostatic valve and a proximal Ring Lock. Dilator lumen is compatible with 0.035" guidewires. Sterilized via ethylene oxide to SAL 10^-6. Materials and design are consistent with the predicate device.
Indications for Use
Indicated for patients requiring peripheral vascular access for the introduction of diagnostic or interventional devices while minimizing blood loss.
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
Predicate Devices
FlexSeal Introducer Sheath with Hydrophilic Coating (K161659)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 30, 2019
Freudenberg Medical MIS, Inc % Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114
Re: K190628
Trade/Device Name: FlexSeal Introducer Sheath with Hydrophilic Coating Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter introducer Regulatory Class: Class II Product Code: DYB Dated: May 21, 2019 Received: May 22, 2019
Dear Mark Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Kenneth Cavanaugh Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190628
Device Name
FlexSealTM Introducer Sheath with Hydrophilic Coating
Indications for Use (Describe)
The FlexSeal® Introducer Sheath with Hydrophilic Coating is intended to be placed in the peripheral vasculature to provide a conduit for the introduction of diagnostic or interventional devices, and to minimize blood loss associated with such insertions.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|------------------------------------------------------------|-----------------------------------------------------------|
|------------------------------------------------------------|-----------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 1 510(K) SUMMARY
#### ADMINISTRATIVE INFORMATION 1.1
May 29th, 2019 Date of Summary Preparation:
#### 1.2 CONTACT INFORMATION
| Primary Submission Contact | Ming Cheng Chew<br>Regulatory Consultant,<br>Libra Medical Inc.<br>Tel: 763-232-3701<br>Email: mcchew@libramed.com |
|------------------------------|--------------------------------------------------------------------------------------------------------------------|
| Secondary Submission Contact | Sew-Wah Tay, PhD<br>Regulatory Consultant,<br>Libra Medical Inc.<br>Tel: 612-801-6782 |
Sponsor Larry Bender Freudenberg Medical MIS, Inc 2301 Centennial Boulevard Jeffersonville, Indiana 47130 Tel: 812-280-2354 Email: Larry.Bender@freudenbergmedical.com
Email: swtay@libramed.com
#### 1.3 DEVICE INFORMATION
| Trade Name | FlexSeal™ Introducer Sheath with Hydrophilic<br>Coating |
|---------------------------|---------------------------------------------------------|
| Common Name | Introducer Sheath |
| Classification Name | Catheter Introducer |
| Classification Regulation | 870.1340 |
| Class | II |
| Panel | Cardiovascular |
| Product Code | DYB |
| Previous Submissions | None |
#### 1.4 510(K) TYPE AND REASON FOR SUBMISSION
This 510(k) is a special 510(k) and is submitted to obtain marketing clearance for a line extension of the FlexSeal device family.
{4}------------------------------------------------
### 1.5 PREDICATE DEVICE
The FlexSeal Introducer Sheath with Hyrdophilic Coating is substantially equivalent to the FlexSeal Introducer Sheath with Hydrophilic Coating devices (K161659).
### 1.6 DEVICE DESCRIPTION
The FlexSeal Introducer set contains the following disposable components:
- One sterile introducer sheath with an active spring loaded hemostatic valve ●
- One sterile radiopaque dilator
The FlexSeal Introducer set consists of an introducer sheath with hydrophilic coating and a dilator. The device size is 21Fr with a 56 cm shaft. An active hemostatic valve is integrated into the sheath. A Ring Lock at the proximal end of the device assists with the control of the hemostatic valve.
### 1.7 INDICATIONS FOR USE
The FlexSeal™ Introducer Sheath with Hydrophilic Coating is intended to be placed in the peripheral vasculature to provide a conduit for the introduction of diagnostic or interventional devices, and to minimize blood loss associated with such insertions.
#### 1.8 TECHNOLOGICAL CHARACTERISTICS
The sheath shaft's external surface has a hydrophilic coating to facilitate introduction into the cardiovascular system. The introducer sheath has a tapered leading tip and a marker band within the tip for visualization. Proximal to the shaft is a housing that contains an extension tube with a 3-way stopcock. The hub is an active valve which can be manually open for device introduction to provide hemostasis and minimize blood loss. These features are the same as the predicate.
The dilator has a tapered distal tip. Its lumen is compatible with a 0.035" guidewire. This is the same as the predicate.
The Ring Lock at the proximal end of the hub allows the user to keep the valve in partially open positions without the need to manually hold the valve open. This feature is not present on the predicate, but does not change the function of the device. The Ring Lock is an extra feature to assist the physician when using the device, but if it is not engaged, the device still functions the same as the predicate device.
## 1.9 PERFORMANCE DATA
The FlexSeal has been tested to meet the device intended use and to ensure conformance to the product specifications.
The FlexSeal has been tested and meets all its physical and performance specifications on the bench including:
- Visual surface inspection .
- Dimensions .
{5}------------------------------------------------
- Guidewire compatibility ●
- . Leakage
- Kink resistance ●
- Bond tensile strength ●
- Hydrophilic coating integrity ●
- Luer compliance ●
- Cracking pressure ●
- USP particulate test ●
- Packaging tests
- Distribution tests
- Simulated Use
- Shelf Life Testing
The following tests were leveraged from the predicate device:
- Radiopacity ●
- Dilator hub dimensions ●
- Corrosion ●
In addition, the FlexSeal was tested for biocompatibility per ISO 10993-1 for short duration contact with circulating blood (<24 hours). Only cytotoxicity was repeated for the subject device.
| Family of Biocompatibility Tests<br>Performed | Results |
|-----------------------------------------------|--------------------|
| Cytotoxicity | Meets Requirements |
| Sensitization | Same as Predicate |
| Irritation/Intracutaneous Reactivity | Same as Predicate |
| Systemic (Acute) Toxicity | Same as Predicate |
| Genotoxicity | Same as Predicate |
| Hemocompatibility | Same as Predicate |
The device is sterilized by ethylene oxide to an SAL 106 level. These performances are similar to that described by the predicate device.
The conclusions drawn from the nonclinical tests that demonstrate that the subject device's performance was substantially equivalent to the predicate device.
#### SUBSTANTIAL EQUIVALENCE 1.10
The FlexSeal is substantially equivalent to the FlexSeal devices (K161659). It has the same intended use for facilitating the introduction of devices into the cardiovascular system and to
{6}------------------------------------------------
minimize blood loss. The conclusions drawn from the nonclinical tests that demonstrate that the subject device's performance was substantially equivalent to the predicate device.
### 1.11 CONCLUSION
Based on the test data and the same intended use, the FlexSeal is found to be substantially equivalent to its predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.