Micropuncture Pedal Access Set

K172980 · Cook Incorporated · DYB · May 3, 2018 · Cardiovascular

Device Facts

Record IDK172980
Device NameMicropuncture Pedal Access Set
ApplicantCook Incorporated
Product CodeDYB · Cardiovascular
Decision DateMay 3, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1340
Device ClassClass 2
AttributesTherapeutic

Indications for Use

This device is intended to introduce up to an 0.038" wire guide into the peripheral vasculature. It is also used to introduce diagnostic and/or interventional devices able to pass through the ID of the introducer.

Device Story

The Micropuncture Pedal Access Set is a sterile, single-use kit used by physicians to gain peripheral vascular access via the Seldinger technique. The set includes a 21-gauge percutaneous entry needle, a 0.018-inch mandril wire guide with a distal coil tip, a coaxial introducer/dilator pair (4.0 or 5.0 French outer introducer; 3.0 French inner dilator), and a Check-Flo Hemostasis Valve. The physician uses the needle and wire guide to access the vessel, then uses the coaxial pair to introduce the wire guide (up to 0.038") and subsequent diagnostic or interventional devices into the peripheral vasculature. The hemostasis valve minimizes blood loss during the procedure. By facilitating pedal access, the device enables revascularization in patients where traditional antegrade transfemoral approaches have failed or are unsuitable, potentially improving outcomes for complex tibial vessel disease.

Clinical Evidence

Clinical evidence consists of three prospective/observational studies involving 220 patients. These studies demonstrate the successful use of the device for pedal access in patients with infrainguinal artery occlusions and critical limb ischemia, supporting the safety and effectiveness of the device for its intended use.

Technological Characteristics

Coaxial introducer/dilator system (4.0/5.0 French outer, 3.0 French inner). Components: 21G needle, 0.018" nitinol/platinum mandril wire guide with distal coil tip, Check-Flo Hemostasis Valve. Materials tested per ANSI AAMI ISO 10993-1:2009(R)2013. Mechanical testing per BS EN ISO 11070:2014 (tensile, leakage, fracture, flex, radiopacity). Sterile, single-use.

Indications for Use

Indicated for patients requiring peripheral vascular access for diagnostic or interventional procedures, including those with infrainguinal artery occlusions or critical limb ischemia requiring below-the-knee revascularization.

Regulatory Classification

Identification

A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. May 3, 2018 Cook Incorporated Mr. Daniel Corbin Regulatory Affairs Specialist 750 Daniels Way Bloomington, Indiana 47404 Re: K172980 Trade/Device Name: Micropuncture Pedal Access Set Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II Product Code: DYB Dated: April 4, 2018 Received: April 5, 2018 Dear Mr. Corbin: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); {1}------------------------------------------------ and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Kenneth J. Cavanaugh -S for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K172980 Device Name Micropuncture® Pedal Access Set Indications for Use (Describe) Indications for Use for the Micropuncture® Pedal Access Set: This device is intended to introduce up to an 0.038" wire guide into the peripheral vasculature. It is also used to introduce diagnostic and/or interventional devices able to pass through the ID of the introducer. Type of Use (Select one or both, as applicable) | <span>☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> | |-------------------------------------------------------------| | <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image is the logo for Cook Medical. The logo is set on a red background. The word "COOK" is in white, sans-serif, block letters. Below the word "COOK" is a smaller rectangle, also in red, with the word "MEDICAL" in white, sans-serif, block letters. O DANIELS WAY P.O. ### 510(k) SUMMARY # Micropuncture® Pedal Access Set Traditional 510(k) Summary 21 CFR §807.92 ### Submitter Information | Applicant: | Cook Incorporated | |-----------------------|---------------------------------| | Address: | 750 Daniels Way | | | Bloomington, IN 47404 | | Contact: | Daniel J. Corbin | | Email: | RegSubmissions@cookmedical.com | | Contact Phone Number: | 812-335-3575 ext. 104018 | | Contact Fax Number: | 812-332-0281 | | Date Prepared: | 03 May 2018 | | Device Information | | | Trade Name: | Micropuncture® Pedal Access Set | | Common Name: | Introducer, Catheter | | Classification Name: | Catheter Introducer | ## Predicate Device The predicate device of the Micropuncture® Pedal Access Set is the Micropuncture® Introducer Set, cleared under K133114. DYB (21 CFR §870.1340) ### Reference Device The Performer® Introducer, cleared under K171999, is a reference device for the subject device, the Micropuncture® Pedal Access Set. {4}------------------------------------------------ ### Comparison to Predicate It has been demonstrated that the Micropuncture® Pedal Access Set and the predicate device are substantially equivalent in terms of intended use, duration of use, principles of operation, fundamental technological characteristics, and insertion method. The design, dimensions, indications for use, duration of use, and materials of the subject device are similar to the attributes of the predicate device. The predicate coaxial introducer/dilator pair are available in 4.0 or 5.0 French diameters and 10 centimeter length; the subject coaxial introducer/dilator pair is available in 4.0 or 5.0 French diameters and 7 centimeter length. The wire guide component in the predicate device is available in length of 40 centimeters and subject set is available as 40 centimeters in length. The predicate wire guide is manufactured from nitinol and palladium; the wire guide in the subject device set is made from nitinol and platinum. The needle components are identical in both the predicate and subject devices. The subject device comes with a Check-Flo® Hemostasis Valve, as where the predicate device does not contain a hemostasis valve. The differences between the subject device and the predicate device, including components, device dimensions, materials, and indications for use, do not raise new questions of safety and effectiveness as demonstrated by performance testing. #### Device Description The device is utilized to gain access to the peripheral vasculature using the Seldinger technique. These introducer sets are comprised of either 4.0 French or 5.0 French outer introducer and 3.0 French inner dilator coaxial pair, a 21 gauge percutaneous entry needle, an 0.018 inch mandril wire guide with a distal coil tip, and a Check-Flo® Hemostasis Valve. These sets are supplied sterile and are intended for one-time use. #### Intended Use This device is intended to introduce up to an 0.038" wire guide into the peripheral vasculature. It is also used to introduce diagnostic and/or interventional devices able to pass through the ID of the introducer. {5}------------------------------------------------ ### Clinical Literature The following articles provide evidence for the subject device, the Micropuncture® Pedal Access Set, for its indications for use: - Walker CM, Mustapha J, Zeller T, et al. Tibiopedal Access for Crossing of Infrainguinal Artery Occlusions: A Prospective Multicenter Observational Study. Journal of Endovascular Therapy. 2016; 23(6): 839-846. - Goltz JP, Planert M, Horn M, et al. Retrograde Transpedal Access for Revascularization of Below-the-Knee Arteries in Patients with Critical Limb Ischemia after an Unsuccessful Antegrade Transfemoral Approach. Fortschr Röntgenstr 2016: 188: 940-948. - El-Sayed H, Bennett M, Loh T, Davies M. Retrograde Pedal Access and Endovascular Revascularization: A Safe and Effective Technique for High-Risk Patients with Complex Tibial Vessel Disease. Annals of Vascular Surgery. 2016: February; 31: 91-98. These articles describe successful use of the subject device for pedal access in a total of 220 patients, and support the use of the subject device for its indications for use. ### Test Data The Micropuncture® Pedal Access Set. subject of this submission, was subjected to applicable testing to assure reliable design and performance under the testing parameters. The following tests were conducted to ensure reliable design and performance: - Biocompatibility testing Testing (i.e., cytotoxicity, sensitization, intracutaneous . reactivity, systemic toxicity, pyrogenicity, hemocompatibility, complement activation, and partial thromboplastin time) demonstrated that the device is biocompatible. In conformance with the applicable sections of ANSI AAMI ISO 10993-1:2009(R)2013, the predetermined acceptance criteria were met. - Tensile Testing of the Hub to Shaft Bond Testing verified that the hub of the . outer introducer and inner dilator would not loosen or separate from the shaft when tested in accordance with BS EN ISO 11070:2014. The predetermined acceptance criteria were met. {6}------------------------------------------------ - . Tensile Testing of the Connection Bonds of the Check-Flo® Valve - Testing verified that under proper clinical use of the hemostasis valve, the connection bonds are expected to withstand peak load values. The predetermined acceptance criterion was met. - . Liquid Leakage Testing - Testing in accordance with BS EN ISO 11070:2014 -Sections 7.3 and 7.4 verified that under proper clinical use of the hemostasis valve, there shall be no liquid leakage. The predetermined acceptance criterion was met. - . Dimensional Verification Testing - Testing verified that the components of the Micropuncture® Pedal Access Set would meet the specified dimensional requirements. The predetermined acceptance criteria were met. - . Wire Guide Fracture Test - Testing in accordance with BS EN ISO 11070:2014 -Section 8.4 and Annex H verified that the wire guide would show no signs of fracture when subjected to the resistance to fracture testing. The predetermined acceptance criterion was met. - . Wire Guide Tensile Test - Testing in accordance with BS EN ISO 11070:2014 -Section 8.6 and Annex H verified that the distal tip of the wire guide would not separate under clinically relevant conditions. The predetermined acceptance criterion was met. - Radiopacity Testing Testing verified that the wire guide would be detectable . under fluoroscopy. The predetermined acceptance criterion was met. - Resistance to Damage by Flex Testing Testing in accordance with BS EN ISO . 11070:2014 - Section 8.5 and Annex G verified that the wire guide would not show any damage or defects when subjected to repeated flexing. The predetermined acceptance criterion was met. In conclusion, the results of these tests support a determination of substantial equivalence.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...