K171491 · Terumo Medical Corporation · DYB · Oct 5, 2017 · Cardiovascular
Device Facts
Record ID
K171491
Device Name
R2P Destination Slender Guiding Sheath
Applicant
Terumo Medical Corporation
Product Code
DYB · Cardiovascular
Decision Date
Oct 5, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1340
Device Class
Class 2
Indications for Use
R2P™ Destination Slender™ Guiding Sheath is indicated to be used for the introduction of interventional and diagnostic devices in the lower extremities of the peripheral vasculature through an access site, including but not limited to the radial artery.
Device Story
Low-profile guiding sheath; functions as both guiding catheter and introducer sheath. Input: manual insertion into peripheral vasculature. Operation: manual; coil-reinforced tubing with radiopaque tip and hydrophilic coating; includes dilator and hemostatic valve. Output: provides access channel for interventional/diagnostic devices. Used in clinical settings by physicians. Benefits: facilitates minimally invasive peripheral vascular procedures via radial or other access sites.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including mechanical integrity (tensile, torque, bend), leakage, lubricity, biocompatibility, and simulated use testing.
Indicated for patients requiring introduction of interventional and diagnostic devices into the peripheral vasculature of the lower extremities via an access site (e.g., radial artery).
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
{0}------------------------------------------------
October 5, 2017
Terumo Medical Corporation Liang Lu Senior Regulatory Affairs Specialist 950 Elkton Blvd Elkton, Maryland 21921
Re: K171491
Trade/Device Name: R2PTM Destination Slender™ Guiding Sheath Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II Product Code: DYB Dated: September 6, 2017 Received: September 7, 2017
Dear Liang Lu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Kenneth J. Cavanaugh -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K171491
Device Name
R2PTM Destination Slender™ Guiding Sheath
Indications for Use (Describe)
R2P™ Destination Slender™ Guiding Sheath is indicated to be used for the introduction of interventional and diagnostic devices in the lower extremities of the peripheral vasculature through an access site, including tut not limited to the radial artery.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------|---------------------------------------------|
| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K171491 510(K) SUMMARY
#### A. SUBMITTER INFORMATION (807.92(a)(1))
#### Prepared by:
Liang Lu Senior Regulatory Affairs Specialist Terumo Medical Corporation Tel. (410) 392-7321 Fax (410) 398-6079
Prepared for: Owner/Operator Terumo Medical Corporation 2101 Cottontail Lane Somerset, NJ 08873, USA Registration Number: 2243441
# Manufacturer (510(k) Applicant)
Terumo Medical Corporation 950 Elkton Boulevard, Elkton, MD 21921 Registration Number: 1118880
#### Sterilization Facility
Steris Isomedix Services Inc. 3459 South Clinton Ave South Plainfield, NJ 07080 Registration Number: 2246552
Contact Person: Liang Lu
Senior Regulatory Affairs Specialist Terumo Medical Corporation 950 Elkton Boulevard, Elkton, MD 21921 Tel. (410) 392-7321 Fax (410) 398-6079 E-mail: liang.lu@terumomedical.com
Date prepared: May 11, 2017
{4}------------------------------------------------
#### B. DEVICE NAME (807.92(a)(2))
| Proprietary Name: | R2PTM Destination Slender™ Guiding Sheath |
|-----------------------|-------------------------------------------|
| Common Name: | Guiding Sheath |
| Classification Name: | Catheter Introducer |
| Classification Panel: | Cardiovascular |
| Regulation: | 21 CFR 870.1340 |
| Product Code: | DYB |
| Classification: | Class II |
### C. PREDICATE DEVICE (807.92(a)(3))
The legally marketed device(s) to which substantial equivalence is claimed are:
- . Primary Predicate Device: K052185 - Destination Carotid Guiding Sheath, manufactured by Terumo Medical Corporation, USA
- Reference Device: K161546 R2P SlenGuide, manufactured by Terumo . Corporation
## D. REASON FOR 510(k) SUBMISSION
This premarket notification (510(k)) is being submitted for the R2P™ Destination Slender™ Guiding Sheath, manufactured by Terumo Medical Corporation, for the purposes of establishing substantial equivalence to a legally marketed predicate device.
#### E. DEVICE DESCRIPTION (807.92(a)(4))
The R2PTM Destination Slender™ Guiding Sheath is a low profile guiding sheath designed to perform as a guiding catheter and an introducer sheath. The sheath is coil reinforced, has a radiopaque tip, is hydrophilically coated and is available in 6Fr with a length of 119cm and 149cm. It comes packaged with a dilator and hemostatic valve.
#### F. INDICATIONS FOR USE (807.92(a)(5))
R2P™ Destination Slender™ Guiding Sheath is indicated to be used for the introduction of interventional and diagnostic devices in the lower extremities of the peripheral vasculature through an access site, including but not limited to the radial artery.
# G. SUBSTANTIAL EQUIVALENCE COMPARISON (807.92(a)(6))
The R2PTM Destination Slender™ Guiding Sheath, subject of this 510(k), is substantially equivalent in its intended use/indications for use, technology/principal of operation, materials, and performance to the predicate device, manufactured by Terumo Medical Corporation.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Terumo. The logo consists of a red swoosh above the word "TERUMO" in green, block letters. The swoosh is positioned above and to the left of the word.
A comparison of the technological characteristics is summarized on the table below.
| Device<br>Characteristic | New Device:<br>R2PTM Destination<br>Slender™ Guiding<br>Sheath | Primary Predicate:<br>Destination Carotid<br>Guiding Sheath<br>(K052185) | Reference Device:<br>R2P SlenGuide<br>(K161546) |
|------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Terumo Medical<br>Corporation | Terumo Medical<br>Corporation | Terumo Corporation |
| Intended Use /<br>Indications for<br>Use | R2PTM Destination<br>Slender™ Guiding<br>Sheath is indicated to<br>be used for the<br>introduction of<br>interventional and<br>diagnostic devices in<br>the lower extremities of<br>the peripheral<br>vasculature through an<br>access site, including<br>but not limited to the<br>radial artery. | The Destination®<br>Guiding Sheath is<br>designed to be used for<br>the introduction of<br>interventional and<br>diagnostic devices into<br>the human vasculature<br>including but not<br>limited to the carotid<br>arteries. | The R2P (radial to<br>peripheral)<br>SlenGuide is<br>designed for the<br>introduction of<br>interventional and<br>diagnostic devices<br>into the<br>peripheral<br>vasculature of<br>the lower extremities |
| Operation<br>Principle | Operated manually or<br>by a manual<br>process; | Same | Same |
| Design /<br>Construction | Sheath, Dilator,<br>Hemostatic Valve with<br>side tube and three-way<br>stopcock | Same | Guiding Catheter,<br>Inner Guide |
#### Table 12.1: Summary of Substantial Equivalence
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for Terumo. The logo consists of a red curved line above the word "TERUMO" in green. The font is sans-serif and bolded.
| Materials | Sheath Assembly | Sheath Assembly | Guiding Catheter: |
|-----------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|
| | Tubing: | Tubing: | Outer layer: |
| | Inner Layer: PTFE | Inner Layer: PTFE | Polyamide elastomer<br>(Nylon) |
| | Middle Coil Layer: | Middle Coil Layer: | Middle Braid Layer: |
| | Stainless Steel | Stainless Steel | Stainless steel |
| | Outer Layer: Nylon | Outer Layer: Nylon | Inner layer: |
| | | | Polytetrafluoroethylene |
| | Radiopaque Tip: Nylon<br>with Tungsten | Tip: Nylon<br>Radiopaque Band:<br>Gold | Radiopaque Tip:<br>Polyamide Elastomer<br>(Nylon) with Tungsten |
| | Hydrophilic Coating:<br>Polyvinylpyrrolidone-<br>based coating | Hydrophilic Coating:<br>Polyvinylpyrrolidone-<br>based coating | Hydrophilic Coating:<br>Dimethyl acrylic amide<br>- glycidyl methacrylate |
| | Hub: Nylon | Hub: Nylon | Hub: Polyamide 12 |
| | Anti-kink protector:<br>Nylon | | Anti-kink protector:<br>Polyester Elastomer |
| | Dilator Assembly<br>Tubing: Polypropylene<br>Hub: Polypropylene/<br>/Thermoplastic<br>Elastomer Blend<br>Coating: Silicone<br>Caulking Pin: Stainless<br>steel | Dilator Assembly<br>Same | Inner Guide:<br>Shaft: Polyester<br>Elastomer with<br>Tungsten<br>Hub: Polyamide<br>Lock Adaptor:<br>polycarbonate |
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for Terumo. The logo consists of a red swoosh above the word "TERUMO" in green. The font is sans-serif and bold.
| | Cross Cut Valve | Cross Cut Valve | None |
|---------|------------------------|---------------------|-----------------|
| | Valve Assembly: | Valve Assembly: | |
| | Housing: | Same | |
| | Polypropylene | | |
| | Cap: Polypropylene | | |
| | Luer Lock Collar: | | |
| | Polycarbonate | | |
| | Valve: Silicone Rubber | | |
| | Elastomer | | |
| | Sidetube: | | |
| | polybutadiene | | |
| | Silicone: Non-reactive | | |
| | silicone oil 1000cst | | |
| | Side Tube Assembly: | Side Tube Assembly: | |
| | Body: Polybutadiene | Same | |
| | 3Way (3WSC) | 3Way (3WSC) | |
| | Stopcock Assembly: | Stopcock Assembly: | |
| | Body: Polycarbonate | Same | |
| | Locking | | |
| | Pin: Polyethylene | | |
| | Cap: Polyethylene and | | |
| | Colorant | | |
| | Handle: Polyethylene | | |
| | and Colorant | Tuohy-Borst Valve | |
| | Lock Adaptor: | Similar material as | |
| | Polycarbonate | Cross Cut Valve | |
| Package | Unit Pouch | Unit Pouch | Unit Pouch |
| | Shelf Box | Shelf Box | Shelf Box |
| | Shipping Carton | Shipping Carton | Shipping Carton |
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for Terumo. The logo consists of the word "TERUMO" in green, with a red arc above the letters "TE". The font is bold and sans-serif. The logo is simple and modern.
| Specifications | Sheath Size: 6 Fr. | Sheath Size: 6-7 Fr. | Catheter Size: 7Fr |
|--------------------------|---------------------------------------------------------------------------------------------------------|-------------------------------------------------|-------------------------------------------------------|
| | Sheath ID/OD<br>(nominal): | Sheath ID/OD<br>(nominal): | I.D.: 2.20 mm (0.087") |
| | 6Fr.: 0.087"/0.100"<br>(2.2mm /2.5mm) | 6Fr.: 0.087"/0.111"<br>7Fr.: 0.100"/0.122" | O.D.: 2.37 mm<br>(0.093") |
| | | | |
| | Sheath Length:<br>119cm, 149cm | Sheath Length:<br>80-110cm | Effective lengths:<br>120cm, 150cm |
| | Hydrophilic Coating:<br>full effective length | Hydrophilic Coating:<br>Distal 15cm | |
| | Distal Shape: Straight | Distal Shapes: Straight,<br>Angled (40 degrees) | |
| | Dilator ID/OD<br>(nominal): | Dilator ID/OD<br>(nominal): | Inner Guide<br>Accepts guide wire<br>diameter: 0.035" |
| | 0.039"/0.086" | 6Fr.: 0.045”/0.084"<br>7Fr.: 0.045”/0.097" | |
| | Dilator Extended<br>Length: 5cm | Dilator Extended<br>Length: 5cm | |
| Sterilization | Ethylene Oxide<br>(validated in<br>accordance with ANSI<br>/ AAMI / ISO 11135-1<br>to achieve SAL 10-6) | Same | Same |
| Shelf life | 30 months | Same | 36 months |
| Disposable<br>Single Use | Yes | Same | Same |
# H. NON-CLINICAL TESTS (807.92(b)(1))
# Performance
Performance testing was conducted to ensure that the R2P™ Destination Slender™ Guiding Sheath met the applicable design and performance requirements throughout the shelf life, verify conformity to the applicable external and internal standards, and demonstrate substantial equivalence to the predicate device.
The following table provides a list of the performance tests that were performed on the proposed R2P™ Destination Slender™ Guiding Sheath.
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the logo for Terumo. The logo consists of the word "TERUMO" in green, with a red swoosh above the letters "TE". The font is bold and sans-serif. The logo is simple and modern.
| Table 5.4: Summary of Performance Testing | | |
|--------------------------------------------|--------------------------------------------------------------------------------------------|----------------------------------------------------|
| Test Item | Reference | Component<br>(Sheath,<br>Dilator, CCV<br>Assembly) |
| Three-Point Bend | In-house standard | Sheath |
| Hub Joint Strength | ISO 11070:2014 | Sheath |
| | ISO 10555-1:2009 | |
| | ISO 10555-1:2013 | |
| Tubing Tensile Strength | ISO 11070:2014 | Sheath |
| | ISO 10555-1:2009 | |
| | ISO 10555-1:2013 | |
| Tip Tensile Strength | ISO 11070:2014 | Sheath |
| | ISO 10555-1:2009 | |
| | ISO 10555-1:2013 | |
| Ovalization | In-house standard | Sheath |
| Lubricity and Durability | In-house standard | Sheath |
| Uncoated Length | In-house standard | Sheath |
| Corrosion Resistance | ISO 11070:2014 | Sheath, Dilator |
| | ISO 10555-1:2009 | |
| | ISO 10555-1:2013 | |
| Gauging | ISO 11070:2014 | Sheath |
| | ISO 10555-1:2009 | |
| | ISO 10555-1:2013 | |
| | ISO 594-1:1986 | |
| Liquid Leakage of Luer<br>under Pressure | ISO 11070:2014 | Sheath |
| | ISO 10555-1:2009 | |
| | ISO 10555-1:2013 | |
| | ISO 594-2:1998 | |
| Air Leakage | ISO 11070:2014 | Sheath |
| | ISO 10555-1:2009 | |
| | ISO 10555-1:2013 | |
| | ISO 594-2:1998 | |
| Separation Force | ISO 11070:2014 | Sheath |
| | ISO 10555-1:2009 | |
| | ISO 10555-1:2013 | |
| | ISO 594-2:1998 | |
| Unscrewing Torque | ISO 11070:2014 | Sheath |
| | ISO 10555-1:2009 | |
| | ISO 10555-1:2013 | |
| | ISO 594-2:1998 | |
| Ease of Assembly | ISO 11070:2014 | Sheath |
| | ISO 10555-1:2009 | |
| | ISO 10555-1:2013 | |
| Test Item | Reference | Component<br>(Sheath,<br>Dilator, CCV<br>Assembly) |
| | ISO 594-2:1998 | |
| Resistance to Overriding | ISO 11070:2014<br>ISO 10555-1:2009<br>ISO 10555-1:2013 | Sheath |
| Stress Cracking | ISO 594-2:1998<br>ISO 11070:2014<br>ISO 10555-1:2009<br>ISO 10555-1:2013<br>ISO 594-2:1998 | Sheath |
| Simulated Use and<br>Particulate* | FDA Guidance Doc 1608 | Sheath, Dilator,<br>CCV |
| Radiodetectability | ISO 11070:2014<br>ISO 10555-1:2013 | Sheath, Dilator |
| Visual Inspection of<br>External Surface | ISO 11070:2014<br>ISO 10555-1:2009<br>ISO 10555-1:2013 | Sheath, Dilator |
| Sheath Tip Inner Diameter | In-house standard | Sheath |
| Sheath Hub Inner Diameter | In-house standard | Sheath |
| Dilator Bump Inner<br>Diameter | In-house standard | Dilator |
| Outer Diameter<br>Measurement | In-house standard | Sheath, Dilator |
| Effective Length | In-house standard | Sheath, Dilator |
| Coating Integrity | FDA Guidance Doc 1608 | Sheath |
| Liquid Leakage of the<br>Sheath | ISO 11070:2014 | Sheath |
| Liquid Leakage Through<br>Hemostasis Valve | ISO 11070:2014 | CCV |
| Torque Strength | FDA Guidance Doc 1608 | Sheath |
| Dilator Tip Inner Diameter | In-house standard | Dilator |
| Dilator Tip Penetration | In-house standard | Dilator |
#### Table 5.4: Summary of Performance Testing
{10}------------------------------------------------
The R2P™ Destination Slender™ Guiding Sheath tested met the predetermined acceptance criteria and results support a determination of substantial equivalence. Based on the results of the performance testing, the proposed R2PTM Destination Slender™ Guiding Sheath is substantially equivalent.
{11}------------------------------------------------
Image /page/11/Picture/1 description: The image shows the logo for Terumo. The logo consists of a red curved line above the word "TERUMO" in green. The word "TERUMO" is in a bold, sans-serif font.
#### Biocompatibility
Biocompatibility classification is based on the FDA Guidance Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process."
The R2PTM Destination Slender™ Guiding Sheath is classified as an Externally Communicating Device, Circulating Blood, Limited Contact (<24 hours).
The following tests are recommended by FDA and ISO 10993-1 to be performed per this device classification:
- 1. Cytotoxicity
- 2. Sensitization
- 3. Intracutaneous Reactivity
- 4. Systemic Toxicity (Acute)
- 5. Pyrogenicity
- 6. Hemocompatibility
- 7. Thrombogenicity
- 8. Complement Activation (Immunology)
- 9. Physiochemical Testing
Results of the testing demonstrate biocompatibility of the finished R2PTM Destination Slender™ Guiding Sheath.
#### Sterilization
The sterility of the device is assured using a sterilization method validated in accordance with ISO 11135:2014, Sterilization of health-care products - Ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices.
The sterilization process was validated utilizing the overkill half cycle approach to provide a Sterility Assurance Level (SAL) of 10-6.
R2P™ Destination Slender™ Guiding Sheath is a limited exposure device. After 24 hours of heated aeration, the level of residual ethylene oxide (EO) and ethylene chlorohydrin (ECH) do not exceed an average daily dose of 4 mg and 9 mg respectively per EN ISO 10993-7:2008.
# I. CLINICAL TESTS (807.92(b)(2))
This 510(k) does not include data from clinical tests.
{12}------------------------------------------------
Image /page/12/Picture/1 description: The image shows the logo for Terumo. The logo consists of a red curved line above the word "TERUMO" in green, block letters. The red line appears to be a stylized representation of a blood vessel or a medical device.
## J. CONCLUSION (807.92(b)(3))
In summary, the R2PTM Destination Slender™ Guiding Sheath, subject of this 510(k), is substantially equivalent in its intended use, technology/principal of operation, materials, and performance to the predicate device(s):
- Primary Predicate Device: K052185 Destination Carotid Guiding Sheath, ● manufactured by Terumo Medical Corporation, USA
- Reference Device: K161546 R2P SlenGuide, manufactured by Terumo . Corporation
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.