K162769 · C.R. Bard, Inc. · DYB · Nov 4, 2016 · Cardiovascular
Device Facts
Record ID
K162769
Device Name
Pinpoint GT Introducer Needle
Applicant
C.R. Bard, Inc.
Product Code
DYB · Cardiovascular
Decision Date
Nov 4, 2016
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1340
Device Class
Class 2
Indications for Use
The Pinpoint™ GT Introducer Needle is intended for patients requiring percutaneous access to place a guidewire for subsequent placement of catheters or other medical procedures requiring introducer needle access. The Pinpoint™ GT Introducer Needle may be used in any appropriate patient population.
Device Story
Pinpoint™ GT Introducer Needle is an 18G needle for percutaneous vascular access and guidewire placement. Device contains an integral passive magnet in the hub; magnet emits a passive magnetic field detectable by ultrasound systems equipped with Pinpoint™ GT Technology. When used with the Pinpoint™ GT System, the device creates a virtual image of the needle on the ultrasound display, providing clinicians with a visual representation of the needle during insertion. Used in clinical settings by healthcare providers. Output allows clinicians to visualize needle position in real-time, potentially improving accuracy of vascular access and reducing complications. Device is provided sterile.
Clinical Evidence
Bench testing only. Tests included needle tip geometry, dimensions, cannula surface finish, tensile force, stiffness, echogenicity, fluid path, guidewire interface, magnetic axis orientation, and sterilization validation. All tests met predetermined acceptance criteria.
Technological Characteristics
18G stainless steel needle; integral passive magnet in hub; non-patient contacting adhesive; HDPE protective cover. Sensing principle: passive magnetic field detection via external ultrasound system. Sterilization: Ethylene Oxide (SAL 10⁻⁶). Standards: ISO 594-1/2 (luer fittings), ISO 9626 (tubing), ISO 7864 (needles), ISO 10993 (biocompatibility), ISO 11135 (sterilization).
Indications for Use
Indicated for patients requiring percutaneous access to place a guidewire for subsequent placement of catheters or other medical procedures. May be used in any appropriate patient population.
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol is a stylized representation of a human figure, with three profiles overlapping to create a sense of depth and connection.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 4, 2016
C.R. Bard, Inc. Jamie Howell Associate Regulatory Affairs Specialist 605 North 5600 West Salt Lake City, Utah 84116
Re: K162769
Trade/Device Name: Pinpoint™ GT Introducer Needle Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: II Product Code: DYB Dated: September 29, 2016 Received: October 3, 2016
Dear Jamie Howell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Image /page/1/Picture/8 description: The image contains a signature and the name "Tina Kiang". The signature is a complex, looping design to the left of the name. The name is written in a simple, sans-serif font, with a dash following the last name. There is also a small letter "s" below the name.
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K162769
Device Name
Pinpoint™ GT Introducer Needle
#### Indications for Use (Describe)
The Pinpoint™ GT Introducer Needle is intended for patients requiring percutaneous access to place a guidewire for subsequent placement of catheters or other medical procedures requiring introducer needle access.
The Pinpoint™ GT Introducer Needle may be used in any appropriate patient population.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| <div> <span> <b> ☑ </b> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for BARD ACCESS SYSTEMS. The word "BARD" is in large, bold, sans-serif font, with a stylized "A" that resembles an upside-down "V". Below the word "BARD" is the phrase "ACCESS SYSTEMS" in a smaller, sans-serif font. The logo is black and white.
# 510(k) Summary
21 CFR 807.92(a)
| | Submitter Name:<br>Address: | Bard Access Systems, Inc.<br>605 North 5600 West<br>Salt Lake City, UT 84116 |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| General<br>Provisions | Contact Person: | Ms. Jamie Howell<br>Regulatory Affairs Specialist |
| | Telephone Number:<br>Fax Number:<br>Date of Preparation: | 801-522-5465<br>801-522-5425<br>29 September 2016 |
| Subject<br>Device | Trade Name:<br>Common Name:<br>Regulation Number:<br>Regulation Name:<br>Regulatory Class:<br>Product Code: | PinpointTM GT Introducer Needle<br>Introducer Needle<br>21 CFR §870.1340<br>Catheter Introducer<br>II<br>DYB |
| Predicate<br>Devices | Premarket Notification:<br>Trade Name:<br>Common Name:<br>Regulation Number<br>Regulation Name:<br>Regulatory Class:<br>Product Code: | K142445, date of clearance April 13, 2015<br>PinpointTM GT Safety Introducer Needle<br>Introducer Needle<br>21 CFR §870.1340<br>Catheter Introducer<br>II<br>DYB |
| Device<br>Description | Bard Access Systems, Inc.'s, PinpointTM GT Introducer Needle is an 18G x<br>2.75 inch needle designed for percutaneous access to introduce a guidewire.<br>The PinpointTM GT Introducer Needle contains a magnet which emits a<br>passive magnetic field that can be detected by ultrasound systems equipped<br>with PinpointTM GT Technology. The PinpointTM GT Introducer Needle, when<br>used with the PinpointTM GT System creates a virtual image of the needle on<br>the ultrasound display, providing clinicians with a visual representation of the<br>needle throughout the insertion process. | |
| Intended Use | The PinpointTM GT Introducer Needle is designed for percutaneous vascular<br>access or procedures requiring the placement of a guidewire. | |
| Indications For<br>Use | The PinpointTM GT Introducer Needle is intended for patients requiring<br>percutaneous access to place a guidewire for subsequent placement of<br>catheters or other medical procedures requiring introducer needle<br>access. The PinpointTM GT Introducer Needle may be used in any<br>appropriate patient population. | |
{4}------------------------------------------------
Technological characteristics of the subject 18G Pinpoint™ GT Introducer Needle is substantially equivalent with respect to the basic design and function to those of the cited predicate device. Modifications made to the subject device when compared to the predicate device are as follows: . Dimensional specification modification from a 21G needle to an 18G needle; . Tip geometry, primary grind angle modification to the needle from 12° to 19° (required to accommodate the larger diameter of 18G needle shaft); . Removal of the active safety mechanism on the subject 18G Pinpoint™ GT Introducer Needle; . Modification to the hub design owing to the removal of the safety mechanism: . Use of a different, non-patient contacting, adhesive that is used to bond the needle to the magnet, in the subject 18G Pinpoint™ GT Introducer Needle; and Change to protective cover (packaging retainer) . . Change in the type of polyethylene used in the needle protective cover; and Technological ● Protective cover changed from clam shell to tube Characteristics over cannula. The following table provides a comparison between the technological characteristics of the subject and predicate device.
| Subject and Predicate Device Comparison Table | | | |
|-----------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Attribute | Subject:<br>Pinpoint™ GT Introducer<br>Needle | Predicate:<br>Pinpoint™ GT Safety<br>Introducer Needle | |
| Owner | Bard Access Systems, Inc. | Bard Access Systems, Inc. | |
| Classification | Same as predicate device. | DYB – 21 CFR §870.1340<br>Catheter Introducer | |
| 510k Status | Subject of this<br>Premarket Notification | K142445 - date of clearance<br>April 13, 2015 | |
| Commercial<br>Name | Pinpoint™ GT Introducer<br>Needle | Pinpoint™ GT Safety<br>Introducer Needle | |
| Indications for<br>Use | The Pinpoint™ GT<br>Introducer Needle is<br>intended for patients<br>requiring percutaneous<br>access to place a<br>guidewire for subsequent<br>placement of catheters or<br>other medical procedures | The Pinpoint™ GT Safety<br>Introducer Needle is<br>intended for patients<br>requiring percutaneous<br>access to place a guidewire<br>for subsequent placement of<br>catheters or other medical<br>procedures requiring | |
| | | requiring introducer needle<br>access. The Pinpoint™ GT<br>GT Introducer Needle may<br>be used in any appropriate<br>patient population. | introducer needle<br>access. The Pinpoint™ GT<br>Safety Introducer Needle<br>may be used in any<br>appropriate patient<br>population. |
| | | The primary intent of the<br>subject device, Pinpoint™<br>GT Introducer Needle, is<br>to assist with<br>percutaneous vasculature<br>access or procedures<br>requiring the placement of<br>a guidewire. | The primary intent of the<br>predicate device, Pinpoint™<br>GT Safety Introducer<br>Needle, is to assist with<br>percutaneous vasculature<br>access or procedures<br>requiring the placement of a<br>guidewire. |
| Technological<br>Characteristics<br>(cont.) | Scientific<br>Technology<br>Description | Additionally, the Pinpoint™<br>GT Introducer Needle<br>contains, integral, within<br>the needle hub a passive<br>magnet. The needle's<br>incorporated passive<br>magnet can be detected<br>by an ultrasound system<br>equipped with Pinpoint™<br>GT Technology. The<br>Pinpoint™ GT Technology<br>(not the subject of this<br>submission) consists of a<br>magnet sensing probe and<br>software loaded on<br>ultrasound equipment,<br>creating the Pinpoint™ GT<br>System; this system is<br>capable of displaying a<br>visual representation of<br>the needle on an<br>ultrasound image. The<br>detection of the needle's<br>passive magnet by the<br>Pinpoint™ GT technology<br>is an optional feature and<br>tool/device offered to<br>clinicians for visual<br>representation of a needle<br>throughout the insertion<br>process; the presence of<br>the passive magnet does<br>not impact the ability of the<br>device to perform as an<br>introducer needle. | Additionally, the Pinpoint™<br>GT Safety Introducer Needle<br>contains, integral, within the<br>needle hub an active safety<br>mechanism and a passive<br>magnet. The needle's<br>incorporated passive<br>magnet can be detected by<br>an ultrasound system<br>equipped with Pinpoint™ GT<br>Technology. The Pinpoint™<br>GT Technology (not the<br>subject of this submission)<br>consists of a magnet<br>sensing probe and software<br>loaded on ultrasound<br>equipment, creating the<br>Pinpoint™ GT System; this<br>system is capable of<br>displaying a visual<br>representation of the needle<br>on an ultrasound image.<br>The detection of the<br>needle's passive magnet by<br>the Pinpoint™ GT<br>technology is an optional<br>feature and tool/device<br>offered to clinicians for<br>visual representation of a<br>needle throughout the<br>insertion process; the<br>presence of the passive<br>magnet does not impact the<br>ability of the device to<br>perform as an introducer<br>needle. |
| | Needle<br>Components | Needle Shaft:<br>Same as predicate device.<br>Needle tip:<br>Same as predicate device.<br>A Bevel 19°<br>Hub:<br>Same as predicate device. | Needle Shaft:<br>Silicone coated<br>Needle tip:<br>Echogenic<br>A Bevel 12°<br>Hub<br>Open ended luer locking hub<br>Bevel Indicator<br>Passive Magnet |
| Technological<br>Characteristics<br>(cont.) | Needle<br>Materials | Luer Hub:<br>Same as predicate device<br>Needle Shaft:<br>Same as predicate device.<br>Magnet:<br>Same as predicate device.<br>Needle Protective Cover<br>(packaging retainer):<br>High Density Polyethylene<br>(HDPE)<br>No colorant<br>Safety Mechanism:<br>None | Luer Hub:<br>Clear Polycarbonate<br>Needle shaft:<br>Silicone Coated Alloy<br>Magnet:<br>Neodymium Iron Boron<br>(NdFeB)<br>Nickel (Ni) Coated<br>Needle Protective Cover<br>(packaging retainer):<br>Clear Amorphous<br>Polyethylene Terephtalate<br>(APET)<br>Safety Mechanism:<br>Translucent Green<br>Polycarbonate<br>Silicone O-ring<br>301 Stainless Steel<br>Cap polycarbonate |
| | Needle<br>Dimensions | Diameter:<br>18 gauge<br>Length:<br>Same as predicate | Diameter:<br>21 gauge<br>Length:<br>2.75" or 7cm |
| | Needle<br>Labeling | Gauge Size:<br>No gauge size coloring<br>due to removal of the<br>safety mechanism<br>Bevel up Indicator:<br>V in ribs of the hub | Gauge Size:<br>Safety mechanism a<br>translucent green color<br>Bevel up Indicator:<br>Raised embossed arrow on<br>the hub |
| Sterility | Same as predicate device. | Provided Sterile<br>SAL 10⁻⁶<br>Ethylene Oxide | |
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
The differences between the subject and predicate device are not critical to the intended use of the device and do not raise any new or different questions regarding safety or effectiveness.
Verification and validation tests were designed and performed in accordance with Design Controls as per 21 CFR §820.30. The following tests were conducted per guidance documents and standards in conjunction with inhouse protocols to determine appropriate methods for evaluating the performance of the device:
Performance Testing - Bench:
- Needle Tip ●
- . Needle Dimensions
- Cannula Surface Finish
- Needle-Hub Tensile Force ●
- Needle Stiffness
- Effective Needle Lenath
- Needle Echogenicity
- Needle Bevel Up Indicator
- Needle Hub ●
- Needle Fluid Path ●
- Introducer Needle Interface with Guidewire
- Needle Protective Cover ●
- . Magnetic Axis Orientation
- Needle Tip to Magnet Length ●
#### Safety & Performance Tests
The following guidance documents and standards in conjunction with in-house protocols were used to determine the appropriate methods for evaluating the performance of the device:
- · ISO 594-1: 1986, Conical fittings with 6% luer taper for syringes, needles and certain other medical equipment – Part 1: General Requirements
- · ISO 594-2: 1998, Conical fittings with 6% luer taper for syringes, needles and certain other medical equipment - Part 2: Lock Fittings
- · ISO 9626: 2001. Stainless steel needle tubing for the manufacturer of medical devices
- · ISO 7864: 1993, Sterile hypodermic needles for single use
- Guidance Document; Medical Devices with Sharps Injury Prevention Features, August 9, 2005
- Guidance on the Content of Premarket Notification [510(k)] Submissions for Hypodermic Single Lumen Needles, April 1993
- · Guidance Document; Applying Human Factors and Usability Engineering to Optimize Medical Device Design, June 22, 2011
- ISO 10993-1:2009, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing, and the FDA Modified ISO 10993 Test Profile
{8}------------------------------------------------
| | • ISO 10993-7:2008, Biological Evaluation of Medical Devices Part 7:<br>Ethylene Oxide Sterilization Residuals<br>• ISO 11135:2014 Annex B Sterilization Of Health Care Products-<br>Ethylene Oxide—Requirements For Development, Validation And<br>Routine Control Of A Sterilization Process For Medical Devices |
|------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The subject device met all predetermined acceptance criteria derived from the<br>above listed references and demonstrated substantially equivalent<br>performance as compared to the cited predicate device. |
| Summary of<br>Substantial<br>Equivalence | Based on the intended use, technological characteristics, and performance<br>testing, the subject 18G Pinpoint™ GT Introducer Needle meets the<br>requirements that are considered sufficient for its intended use and<br>demonstrates substantial equivalence to the cited predicate device. |
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.