K161102 · St. Jude Medical, Inc. · DYB · May 19, 2016 · Cardiovascular
Device Facts
Record ID
K161102
Device Name
Nanostim Introducer Kit
Applicant
St. Jude Medical, Inc.
Product Code
DYB · Cardiovascular
Decision Date
May 19, 2016
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1340
Device Class
Class 2
Indications for Use
The Nanostim™ Introducer Kit is intended to provide a conduit into the venous system for insertion of diagnostic and other interventional devices.
Device Story
Nanostim Introducer Kit functions as guiding sheath for venous access; facilitates insertion of diagnostic/interventional devices. Comprised of introducer sheath and dilator; features hemostasis valve to minimize air entry, sideport with three-way stopcock, and radiopaque tip marker for visualization. Compatible with 0.035" and 0.038" guidewires. Available in 18F size; lengths 30cm and 50cm. Used by clinicians in clinical settings to provide conduit into venous system. Output is physical access path for secondary devices; aids clinical decision-making by enabling minimally invasive procedures. Benefits patient by providing stable, controlled venous entry.
Clinical Evidence
No clinical data. Bench testing only, including visual, dimensional, torsional strength, and simulated use/kink testing, confirmed performance meets acceptance criteria.
Technological Characteristics
Catheter introducer system consisting of sheath and dilator. Features hemostasis valve, sideport with three-way stopcock, and radiopaque tip marker. Compatible with 0.035" and 0.038" guidewires. 18F diameter; 30cm and 50cm lengths. Mechanical device; no software or energy source.
Indications for Use
Indicated for patients requiring venous access for the insertion of diagnostic or interventional devices.
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
K140793 — NANOSTIM INTRODUCER KIT · St. Jude Medical, Inc. · Dec 17, 2014
K160716 — The Nanostim Introducer Kit · St. Jude Medical, Inc. · Apr 14, 2016
K222184 — Introducer Sheath Kits · Beijing Demax Medical Technology Co.,Ltd · Feb 21, 2023
K092575 — ENGAGE/ENGAGE TR · St Jude Medical · Oct 8, 2009
K123445 — NMI COAXIAL MICROINTRODUCER SET · Navilyst Medical, Inc. · Apr 1, 2013
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized emblem. The emblem consists of three abstract human profiles facing right, stacked on top of each other. The profiles are rendered in a dark color, contrasting with the white background.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 19, 2016
St Jude Medical, Inc. Mr. Hassan Labay Sr. Manager, Regulatory Affairs 645 Almanor Avenue Sunnyvale, CA 94085
Re: K161102
Trade/Device Name: Nanostim Introducer Kit Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II Product Code: DYB Dated: April 18, 2016 Received: April 19, 2016
Dear Mr. Labay:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kenneth J. Cavanaugh -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K161102
Device Name The Nanostim™ Introducer Kit
Indications for Use (Describe)
The Nanostim™ Introducer Kit is intended to provide a conduit into the venous system for insertion of diagnostic and other interventional devices.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
This summary of 510(k) substantial equivalence information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
# 510(k) Number:
# Applicant Information:
| Owner Name: | St. Jude Medical, Inc. |
|-----------------|--------------------------------------------------------------------|
| Address: | 645 Almanor Avenue<br>Sunnyvale, CA. 94085<br>Office: 408 522 6622 |
| Contact Person: | Hassan Labay |
| Phone Number: | 408 522 6622 |
| Date Prepared: | 4/18/2016 |
### Device Information:
| Classification: | Class II |
|----------------------|-------------------------------------------|
| Trade Name: | St. Jude Medical Nanostim™ Introducer Kit |
| Common name: | Introducer, Catheter |
| Classification name: | Catheter Introducer (21 CFR 870.1340/DYB) |
### Predicate Devices:
The modified Nanostim Introducer Kit is substantially equivalent in intended use and method of operation to the predicate Nanostim Introducer Kit (K160716).
### Device Description:
The St Jude Medical Nanostim Introducer Kit is designed to perform as a guiding sheath for introduction of diagnostic and interventional devices. The Nanostim™M Introducer Kit is comprised of an introducer sheath and a dilator. The Introducer Kit is compatible with 0.035" and 0.038" guidewires and is available in one size (18F) and two lengths (30cm and 50cm).
The introducer is fitted with a hemostasis valve to minimize air introduction during insertion and/or exchange, and a sideport with a three-way stopcock. The introducer features a radiopaque tip marker incorporated within the sheath material to identify the location of the distal tip of the sheath. Design modifications were made to the length and packaging of the current device. The changes made to the Nanostim Introducer Kit do not affect the intended use of the device or alter the fundamental scientific technology of the device.
{4}------------------------------------------------
#### Indications For Use:
The Nanostim™ Introducer Kit is intended to provide a conduit into the venous system for insertion of diagnostic and other interventional devices.
#### Summary of Technological Characteristics in Comparison to Predicate Device:
The modified device and the predicate device (K160716) have the same indication for use, technological characteristics and performance. Both the subject device and predicate device are intended to provide a conduit into the venous system for insertion of diagnostic and other interventional devices. The modified device is identical to the existing Nanostim Introducer Kit, with the exception of the device length. The predicate device is 30cm while the modified device is 50cm. The modification described herein do not affect the intended use of the device or alter the fundamental scientific technology associated with the device, and, therefore, the subject device is substantially equivalent to the predicate device with respect to indications, design, and function.
### Summary of Non-Clinical Testing:
Design verification testing was performed to verify the modified device remains substantially equivalent to the predicate device. Testing performed on the modified device included the following:
- . Visual test;
- Dimensional measurement test; ●
- Torsional Strength; ●
- Simulated use/kink test
All of the pre-determined acceptance criteria were met.
#### Summary of Clinical Testing:
Clinical evaluation is not required for this device.
#### Substantial equivalence:
The subject device has the following similarities to the predicate device cleared under K160716:
- the same indication for use;
- the same fundamental scientific technology; ●
- the same technological characteristics; ●
- the same materials ●
- the same principles of operation;
- . incorporates the same basic introducer design and has the same sterilization process:
{5}------------------------------------------------
The changes made to the device do not affect the intended use of the device or clinical effect and scientific technology of the device, the subject device is substantially equivalent to currently cleared Nanostim Introducer Kit 30cm (K160716).
#### Summary:
In summary, the modified device that is subject to this submission is substantial equivalent to the predicate device. It has the same indication for use, the same technological characteristics, the same materials, the same sterilization process and the same principles of operation as the predicate device. The differences between the modified device and the predicate device raise no new issues of safety or effectiveness. Performance data demonstrate that the subject device performs as the predicate device and is therefore substantially equivalent to the predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.