K150257 · Merit Medical Systems, Inc. · DYB · Jun 19, 2015 · Cardiovascular
Device Facts
Record ID
K150257
Device Name
PreludeEASE Hydrophilic Sheath Introducer
Applicant
Merit Medical Systems, Inc.
Product Code
DYB · Cardiovascular
Decision Date
Jun 19, 2015
Decision
SESK
Submission Type
Special
Regulation
21 CFR 870.1340
Device Class
Class 2
Indications for Use
The Merit PreludeEASE Hydrophilic Sheath Introducer is intended to provide access and facilitate the percutaneous introduction of various devices into veins and/or arteries, including but not limited to the radial artery, while maintaining hemostasis for a variety of diagnostic and therapeutic procedures. The access needle with inner metal needle and outer plastic cannula is used to gain access to the vein or artery for placement of guide wires.
Device Story
PreludeEASE Hydrophilic Sheath Introducer facilitates percutaneous vascular access; provides conduit for diagnostic/therapeutic devices into veins/arteries (e.g., radial artery). Device consists of sheath with integral hemostasis valve, sidearm with 3-way stopcock, rotating suture ring, and vessel dilator. Hydrophilic coating on sheath tubing reduces friction. Used in clinical settings by physicians; provides access for guide wires and catheters. Hemostasis valve prevents blood loss during procedure. 7F size expansion of existing product line; utilizes orange colorant for size differentiation. Benefits include reliable vascular access and maintenance of hemostasis during interventional procedures.
Clinical Evidence
Bench testing only. No clinical data. Testing included ISO 11070 compliance, mechanical performance (kink resistance, pull force, leak, lubricity, stiffness), and comprehensive biocompatibility (cytotoxicity, sensitization, systemic toxicity, pyrogenicity, genotoxicity, hemolysis, thrombogenicity, complement activation).
Technological Characteristics
7F sheath introducer with vessel dilator. Features: hydrophilic coating, hemostasis valve, sidearm with 3-way stopcock, rotating suture ring. Materials: standard medical-grade polymers; orange colorant added for 7F identification. Standards: ISO 11070, ISO 11135, ISO 10993 series, ASTM D4169, ASTM F756, USP <151>. Sterilization: Ethylene oxide.
Indications for Use
Indicated for patients requiring percutaneous vascular access (veins/arteries, including radial) for diagnostic or therapeutic procedures; facilitates device introduction while maintaining hemostasis.
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is an abstract symbol that resembles a stylized caduceus or a series of interconnected human profiles. The symbol is rendered in a dark color, contrasting with the lighter background.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 19, 2015
Merit Medical Systems, Inc. Michaela Rivkowich Principal Regulatory Affairs Specialist 1600 West Merit Parkway South Jordan, Utah 84095
Re: K150257
Trade/Device Name: PreludeEASE Hydrophilic Sheath Introducer Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II Product Code: DYB Dated: May 21, 2015 Received: May 22, 2015
Dear Ms. Rivkowich:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit/tray. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
forBram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
Device Name PreludeEASE Hydrophilic Sheath Introducer
Indications for Use (Describe)
The Merit PreludeEASE Hydrophilic Sheath Introducer is intended to provide access and facilitate the percutaneous introduction of various devices into veins and/or arteries, including but not limited to the radial artery, while maintaining hemostasis for a variety of diagnostic and therapeutic procedures.
The access needle with inner metal needle and outer plastic cannula is used to gain access to the vein or artery for placement of guide wires.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and comblete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 5.0 510(k) Summary
| General<br>Provisions | Submitter Name:<br>Address:<br>Telephone Number:<br>Fax Number:<br>Contact Person:<br>Date of Preparation:<br>Registration Number: | Merit Medical Systems, Inc.<br>1600 West Merit Parkway<br>South Jordan, UT 84095<br>(801) 208-4196<br>(801) 253-6932<br>Michaela Rivkowich<br>February 2, 2015<br>1721504 |
|-----------------------|------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Subject Device | Trade Name:<br>Common/Usual Name:<br>Classification Name:<br>Regulatory Class:<br>Product Code:<br>21 CFR §:<br>Review Panel: | PreludeEASE™ Hydrophilic Sheath Introducer<br>Sheath Introducer<br>Catheter Introducer<br>II<br>DYB<br>870.1340<br>Cardiovascular |
| Predicate<br>Device | Trade Name:<br>Classification Name:<br>Premarket Notification:<br>Manufacturer: | PreludeEASE™ Hydrophilic Sheath Introducer<br>Sheath Introducer<br>K140543<br>Merit Medical Systems, Inc.<br>This predicate has not been subject to a design-related recall. |
| Reference<br>Device | Trade Name:<br>Classification Name:<br>Premarket Notification:<br>Manufacturer: | Glidesheath Slender™<br>Sheath Introducer<br>K142183<br>Terumo Medical Corporation |
{4}------------------------------------------------
| Device<br>Description | The 7F PreludeEASE™ Hydrophilic Sheath Introducer consists of a<br>sheath introducer with compatible vessel dilator that snaps securely<br>into the sheath introducer hub. The sheath hub contains an integral<br>hemostasis valve. A rotating suture ring is affixed to the sheath hub.<br>The sheath tubing is coated with a hydrophilic coating. A sidearm is<br>affixed to the sheath hub and has a 3-way stopcock at its proximal end.<br>The 7F PreludeEASE Hydrophilic Sheath Introducer is available in<br>11cm, 16cm and 23cm lengths and is designed to accept 0.018", 0.021"<br>and 0.035" diameter guide wires.<br>The 7F PreludeEASE™ Hydrophilic Sheath Introducer is marketed with<br>any of the following components, depending on the product<br>configuration: guide wire, metal access needle, access needle with<br>inner metal needle and outer plastic cannula, syringe, scalpel and<br>BowTie™ guide wire insertion device. |
|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | There is no change in the indications for use statement from the<br>predicate to the subject device.<br>The Merit PreludeEASE Hydrophilic Sheath Introducer is intended to<br>provide access and facilitate the percutaneous introduction of various<br>devices into veins and/or arteries, including but not limited to the radial<br>artery, while maintaining hemostasis for a variety of diagnostic and<br>therapeutic procedures.<br>The access needle with inner metal needle and outer plastic cannula is<br>used to gain access to the vein or artery for placement of guide wires. |
| Comparison to<br>Predicate<br>Device | The technological characteristics of the subject 7F PreludeEASE<br>Hydrophilic Sheath Introducer are substantially equivalent to those of<br>the predicate 4F through 6F PreludeEASE Hydrophilic Sheath<br>Introducer. The subject device has the same basic design as the<br>predicate device in that it consists of sheath tubing, hub, sidearm and<br>stopcock and is provided with a vessel dilator. The difference between<br>the subject and the predicate devices is in the French size to expand<br>the product line to include the 7F sheath introducer. The subject device<br>utilizes the same materials as the predicate device with the exception of<br>the orange colorant which is used to differentiate the 7F size from the<br>other sizes. |
{5}------------------------------------------------
No performance standards have been established under Section 514 of the Food, Drug and Cosmetic Act for these devices. Performance testing of the subject 7F PreludeEASE Hydrophilic Sheath Introducer was conducted based on the risk analysis and based on the requirements of the following international standard:
- . ISO 11070:1998, Sterile, single-use intravascular catheter introducers
- . ISO 11135:2014, Sterilization of health care products - Ethylene oxide - Requirements for development, validation and routine control of a sterilization process for medical devices
- . ASTM D4169-09, Standard Practice for performance Testing of Shipping Containers and Systems
- . ISO 2233:2000. Packaging - Complete. filled transport packages and unit loads - Conditioning for testing
- . ISO 10993-1:2009, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a risk management process, and FDA guidance Required Biocompatibility Training and Toxicology Profiles for Evaluation of Medical Devices, May 1, 1995
- . ANSI/AAMI/ISO 10993-3:2003, Biological Evaluation of Medical Devices - Part 3: Tests for Genotoxicity, Carcinogenicity and Reproductive Toxicity
- ISO 10993-4:2002 (Amd.1:2006), Biological evaluation of medical . devices – Part 4: Selection of tests for interactions with blood
- . ISO 10993-5:2009, Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
- ISO 10993-7:2008, Biological evaluation of medical devices Part . 7: Ethylene oxide sterilization residuals
- ISO 10993-10:2010, Biological evaluation of medical devices -Part 10: Tests for irritation and skin sensitization
- . ISO 10993-11:2006, Biological evaluation of medical devices -Part 11: Tests for systemic toxicity
- . ASTM F756-08, Standard practice for assessment of hemolytic properties of materials
- . United States Pharmacopeia 35, National Formulary 30, 2012 <151> Pyrogen Test
The following tests were performed to demonstrate there were no unacceptable risks associated with the changes made to the device:
Performance Testing - Bench
- . Size Designation
- Kink Resistance ●
- Sheath / Dilator Tip Peel Back ●
- . Surface
- Hydrophilic Coating Lubricity
- .
Performance
Data
{6}------------------------------------------------
- Hydrophilic Coating Coverage and Adherence .
- Hydrophilic Coating Length .
- Sheath Stiffness and Flex Modulus
- Sheath Introducer Hub to Tubing Pull (Force at Break) ●
- Sheath Introducer Tubing Pull (Force at Break) ●
- Dilator Hub to Tubing Pull (Force at Break) ●
- Dilator Stiffness and Flex Modulus ●
- . Sheath Introducer Assembly Leak
- Radiodetectabillity ●
- Simulated Use .
## Performance
## Data
- Cytotoxicity
Biocompatibility
- Sensitization ●
- Intracutaneous Injection ●
- Acute Systemic Toxicity ●
- Pyrogenicity ●
- Genotoxicity ●
- Hemolysis ●
- . Thrombogenicity
- Complement Activation ●
- Chemical Characterization ●
The results of the testing demonstrated that the subject 7F PreludeEASE Hydrophilic Sheath Introducer met the predetermined acceptance criteria applicable to the safety and efficacy of the device.
Based on the indications for use, design, safety and performance testing, the subject 7F PreludeEASE Hydrophilic Sheath Introducer Summary of meets the requirements that are considered essential for its intended Substantial use and is substantially equivalent to the predicate device, the Equivalence PreludeEASE Hydrophilic Sheath Introducer, K140543 manufactured by Merit Medical Systems, Inc..