GLIDESHEATH

K033681 · Terumo Corp. · DYB · Dec 11, 2003 · Cardiovascular

Device Facts

Record IDK033681
Device NameGLIDESHEATH
ApplicantTerumo Corp.
Product CodeDYB · Cardiovascular
Decision DateDec 11, 2003
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1340
Device ClassClass 2

Indications for Use

The Glidesheath is used to facilitate placing a catheter through the skin into a vein or artery. The Mini Guide Wire is an accessory device which is used for placement of the sheath into the vein or artery. The RADIFOCUS Obturator is also an accessory device which is used by placing it into the sheath to create an occlusion and further provide support to the wall of the indwelling sheath while it remains in place within the vein or artery after removal of a catheter.

Device Story

Glidesheath is an introducer sheath and dilator system used to facilitate catheter placement into veins or arteries. System includes a Mini Guide Wire for placement assistance and a RADIFOCUS Obturator for maintaining occlusion and providing wall support to the indwelling sheath after catheter removal. Sheath, dilator, and obturator contain bismuth for fluoroscopic visibility. Device is operated manually by clinicians in a clinical setting. By dilating the vessel and providing a stable access port, the device enables subsequent catheter insertion and maintains vascular access. Biocompatible materials are used for prolonged contact (24 hours to 30 days).

Clinical Evidence

Bench testing only. Verification tests performed: leakage/clogging, tensile strength of connections, separation force of dilator and sheath, internal/external sliding resistance, and penetration force. Biocompatibility testing conducted per ISO-10993. No clinical data presented.

Technological Characteristics

Introducer sheath and dilator system. Materials include bismuth for fluoroscopic visibility. Biocompatible per ISO-10993. Sterilized via ethylene oxide per ANSI/AAMI/ISO 11135-1994 and EN 550. Manual operation. Sizes: 5 & 6 French; lengths: 10-25 cm.

Indications for Use

Indicated for facilitating catheter placement through the skin into a vein or artery. No specific patient population, age, or gender restrictions provided.

Regulatory Classification

Identification

A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ # SECTION II. SUMMARY AND CERTIFICATION # SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION PERTAINING TO SUBSTANTIAL EQUIVALENCE ### Device Name A. | Proprietary Name | Glidesheath | |---------------------|----------------------| | Classification Name | Introducer, Catheter | | Common Name | Introducer Sheath | ### B. Intended Use The Glidesheath is used to facilitate placing a catheter through the skin into a vein or artery. The Mini Guide Wire is an accessory device which is used for placement of the sheath into the vein or artery. The RADIFOCUS Obturator is also an accessory device which is used by placing it into the sheath to create an occlusion and further provide support to the wall of the indwelling sheath while it remains in place within the vein or artery after removal of a catheter. Note: This is the same intended use as the RADIFOCUS® Introduccr II Kit , K954234. # C. Device Description The Glidesheath is comprised of an introducer sheath and a dilator. The Glidesheath is used to facilitate placement of a catheter through the skin into a vein or artery. A Mini Guide Wire (with Inserter) may be included with the Glidesheath. The Inserter docs not contact blood and is used strictly for guiding the Guide Wire into a cannula or Introducer. The Mini Guide Wire is inserted through a cannula placed in the patient's blood vessel. The Glidesheath is then inserted over the Mini Guide Wire and into the blood vessel. The Mini Guide Wire is then withdrawn from the vessel. The Dilator maintains the integrity of the Sheath and dilates the blood vessel while the Glidesheath is being placed into the vessel. The Dilator can be removed and an appropriate catheter can then be inserted. The RADIFOCUS Obturator is an accessory device which creates an occlusion when inserted into the Sheath. The {2}------------------------------------------------ Obturator also provides support to the indwelling Sheath after the catheter is removed. The Sheath, Dilator and Obturator contain bismuth, making these devices visible under fluoroscopy. # Principle Of Operation / Technology D. and the same of the same of the same of the same of the same of the same of the states of the states of the states of the states of the states of the states of the states of The Glidesheath and its accessories are operated manually or by a manual process. # Design / Materials E. Differences in materials between the Glidesheath and the RADIFOCUS® Introducer II Kit, K954234 raise no new issues of safety and effectiveness. | Part | Glidesheath | RADIFOCUS® Introducer II,<br>K954234 | |-------------------------------------|--------------------------|--------------------------------------| | Introducer Sheath<br>Size<br>Length | 5 & 6 French<br>10-25 cm | 4 – 11 French<br>5 - 110 cm | | Dilator<br>Length | 15.5 - 30.5 cm | 6 - 110 cm | | Guide Wire OD | 0.021" – 0.038" | 0.021" – 0.038" | {3}------------------------------------------------ ### G. Performance ﻧﮯ ﭘﺮ ﺳﻤ The Glidesheath is comprised of an introducer sheath and a dilator. Only the introducer sheath was modified. The dilator was not modified. The following verification tests were performed to demonstrate the substantial equivalence of the modified device (Glidesheath) to the unmodified device (RADIFOCUS® Introducer II). - Leakage/clogging . - Tensile strength of connections . - Separation force of dilator and sheath . - Internal sliding resistance . - External sliding resistance . - Penetration force . None of the data raises any new issues of safety and effectivencss. Additionally, a risk analysis was conducted and there were no new issues of safety and effectiveness. Therefore the performance of the Glidesheath is substantially equivalent to the performance of the RADIFOCUS® Introducer II, cleared under K954234. # H. Additional Safety Information Manufacturing controls include visual, functional, dimensional and sterility tests. Blood contacting materials were tested in accordance with the tests recommended in the FDA General Program Memorandum #G95-1 (5/1/95): Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part-1: Evaluation and Testing". The introducer sheath is classified as Externally Communicating Devices, Circulating Blood, Prolonged Contact (24 hrs to 30 days). Results of the testing demonstrate that the blood contacting materials are biocompatible. Sterilization conditions have been validated in accordance with ANSI / AAMI / ISO 11135-1994, Medical Devices -- Validation and routine control of ethylene oxide sterilization and EN 550. The device is sterilized to a SAL of 106. {4}------------------------------------------------ # I. Substantial Equivalence ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ The Glidesheath is substantially equivalent in intended use, design, technology/principles of operation, materials and performance to the RADIFOCUS® Introducer II, cleared under K954234. Differences between the two devices do not raise any significant issues of safety or effectiveness. # Submitter Information J. | Prepared By: | Mr. Mark Unterreiner<br>Regulatory Affairs Specialist | |----------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Prepared For: | Terumo Medical Corporation<br>125 Blue Ball Road<br>Elkton, MD 21921<br>Phone: (410) 392-7213<br>Fax: (410) 398-6079<br>Email: mark.unterreiner@terumomedical.com | | Date Prepared: | November 21, 2003 | {5}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or other bird. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 DEC 1 1 2003 Mr. Mark Unterreiner Regulatory Affairs Specialist Terumo Medical Corporation 125 Blue Ball Road Elkton, MD 21921 K033681 Re: Trade Name: Glidesheath Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter introducer. Regulatory Class: II Product Code: 74 DYB Dated: November 21, 2003 Received: November 23, 2003 Dear Mr. Unterreiner: We have reviewed your Section 510(k) premarket notification of intent to market the device We have reviewed your Section 3 iQ(x) premaince is substantially equivalent (for the indications referenced above and have determined the device as againers in interstate referenced above and have detellimbed the devices marketed in interstate for use stated in the enclosure) to tegally markets the Medical Device Amendments, or to commerce prior to May 28, 1976, the enactment the of the Federal Food. Drug. commerce prior to May 28, 1976, the enactiven and of the Federal Food, Drug, devices that have been reclassined in accondice was are archarded application (PMA). and Cosmetic Act (Act) that do not require approval of a provisions of the Act. The and Cosmetic Act (Act) that do not require approvate of the general controls provisions of the Act. The Act. The You may, therefore, market the device, subject to the general sommal registration, listing of general controls provisions of the Act include requirements for amount misbrandi general controls provisions of the Act mender requirements on the misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device to such additional controls. Existing major regulations affecting your device can may be subject to such additional controls. Existing major regulation FDA may be subject to such additional controlis. Existing major regalato of the Marketion, FDA may be found in the Code of Federal Regulations, Title 21, Parts 800 to 898 . In a be found in the Code of I cacral regarations, your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA s issuance of a subscribed with other requirements of the Act that FDA has made a determination that your device coderal econoise. You must that FDA has made a delerimilations administered by other Federal agencies. You must a or any Federal statutes and regulations and united to registration and listing (21 comply with all the Act's requirements, including, but not limited to: registration and lis comply with all the Act s requirements, meridants, out how man and one of the many of the many {6}------------------------------------------------ Page 2 - Mr. Mark Unterreiner forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic forul in the quality systems (QD) regalantis (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally premarked notification: "The PDF intrailigstification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and If you desire specific acrison. I for in vitro diagnostic devices), please contact the Office of additionally 21 Of I Creat 586. Additionally, for questions on the promotion and advertising of Compliance at (301) 594-4639. Iffice of Compliance at (301) 594-4639. Also, please note the your de roo, prease ochabranding by reference to premarket notification" (21 CFR Part 807.97). Tegulation on thinked, "Thisorians on your responsibilities under the Act may be obtained from the Other general mionmation on Jose, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Jessa D. Harrison Bram Zuckerman, M.D. Acting Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ Note: This is the same intended use as the predicate device, K954234 | 510(k) Number (if known): | k033681 | |---------------------------|-------------| | Device Name: | Glidesheath | # Indications For Use: The Glidesheath is used to facilitate placing a catheter through the skin into a vein or artery. The Mini Guide Wire is an accessory device which is used for placement of the sheath into the vein or artery. The RADIFOCUS Obturator is also an accessory device which is used by placing it into the sheath to create an occlusion and further provide support to the wall of the indwelling sheath while it remains in place within the vein or artery after removal of a catheter. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use Qlar 12/11/03 vision Sign-off Division of Cardio 510(K) Numbe (Optional Format 1-2-96)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...