K022170 · Rex Medical · DYB · Jan 21, 2003 · Cardiovascular
Device Facts
Record ID
K022170
Device Name
INNER-LOCK LOCKING INTRODUCER SHEATHS
Applicant
Rex Medical
Product Code
DYB · Cardiovascular
Decision Date
Jan 21, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1340
Device Class
Class 2
Indications for Use
The Inner-Lock Locking Introducer Sheaths are used to facilitate placing a catheter through the skin into a graft. The Inner-Lock dilator is an accessory device, which is used by placing it into the sheath to create an occlusion and further provide support to the wall of the indwelling sheath.
Device Story
Vascular access device consisting of central lumen, angled side arm extension, and hemostasis valve. Includes accessory dilator for occlusion and sheath wall support. Features distal tip retention mechanism operated via proximal hub control arm. Used by clinicians to facilitate catheter placement into grafts. Retention mechanism serves as alternative to suturing hub to skin; angled side port designed to reduce shear force during fluid movement. Device provides stable access for catheterization procedures.
Clinical Evidence
No clinical data; bench testing only. Comparative functional testing performed per ISO 10555-1, ISO 11070, and FDA guidance for intravascular catheters.
Technological Characteristics
6F and 7F vascular access sheath. Components: central lumen, angled side arm, hemostasis valve, distal retention mechanism. Materials: biocompatible (ISO 10993). Standards: ISO 10555-1, ISO 11070. Connectivity: None. Energy source: None (mechanical).
Indications for Use
Indicated for patients requiring catheter placement into a graft via percutaneous access.
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
Predicate Devices
PINNACLE™ R/O or RADIFOCUS® Introducer R/O device (K954234, K984260)
Related Devices
K051468 — INNER-LOCK INTRODUCER SHEATH · Rex Medical · Aug 4, 2005
K032569 — SHORT INTRODUCER SHEATH · Rex Medical · Sep 16, 2003
K042399 — MODIFICATION TO SHORT INTRODUCER SHEATH · Rex Medical, L.P. · Dec 17, 2004
K971608 — CORDIS BRITE TIP CATHETER SHEATH INTRODUCER SYSTEM · Cordis Corp. · Jul 21, 1997
Submission Summary (Full Text)
{0}------------------------------------------------
# JAN 21 2003
### EXHIBIT #1
### 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 510(k) number is: KO22170
### Submitter's Identification: 1.
Rex Medical 585 County Line Road Radnor, PA 19087
Contact: Mr. Paul Tashjian
Date Summary Prepared:
July 1, 2002
#### 2. Name of the Device:
Inner-Lock Locking Introducer Sheaths
### 3. Predicate Device Information:
K#954234, K#984260: PINNACLE™ R/O or RADIFOCUS® Introducer R/O device, Terumo Medical Corporation, Elkton, MD
### 4. Device Description:
The Rex Medical, Inner-Lock Locking Introducer Sheaths, 6F and 7F is a vascular access device consisting of a central lumen, an angled side arm extension and a hemostasis valve. The distal tip contains a retention mechanism that is operated via a control arm on the proximal hub. The device is used in exactly the same manner as the predicate device and other substantially equivalent 510(k) cleared devices.
#### 5. Intended Use:
The Inner-Lock Locking Introducer Sheaths are used to facilitate placing a catheter through the skin into a graft. The Inner-Lock dilator is an accessory device, which is used by placing it into the sheath to create an occlusion and further provide support to the wall of the indwelling sheath.
{1}------------------------------------------------
### Comparison to Predicate Devices: 6.
### Discussion of Similarities
The Inner-Lock Locking Introducer Sheaths are similar to the Terumo RADIFOCUS ® Introducer R/O device, in that both devices make use of a tapered distal tip with a central lumen to pass over a guidewire and dilator combination; both devices have an elastomer valve, side port and a radio opaque indicator and proximal hubs and extensions are similar in function and material characteristics. Both devices can be used for dialysis graft access.
### Discussion of Differences
The Inner-Lock Locking Introducer Sheaths are different from the Terumo RADIFOCUS ® Introducer R/O device, by means of an angled side port and a device retention feature. The angled sidearm may reduce shear force during fluid movement. The retention feature may reduce the risk of the device accidentally being dislodged from the patient. The retention mechanism provides an alternative to suturing the device hub to the skin.
### Discussion of Non-Clinical Tests Performed for Determination of 7. Substantial Equivalence are as follows:
Comparative functional testing (to the predicate) was performed based on ISO 10555-1, ISO 11070 and FDA's Reviewer Guidance for Short-Term and Long-Term Intravascular Catheters. Material testing also included ISO 10993 Biocompatibility Testing. Testing results revealed the subject device to be substantially equivalent to the predicate device.
### Discussion of Clinical Tests Performed: 8.
Not applicable, as there are no new indications for use which must be supported by clinical data.
### 9. Conclusions:
The subject device, Inner-Lock Locking Introducer Sheaths, have the same intended use and characteristics as the predicate device, the Terumo RADIFOCUS® Introducer R/O device. Moreover, bench testing contained in our submission and non-clinical testing supplied demonstrates that there are no differences in their technological characteristics, thereby not raising any new questions of safety or effectiveness. Thus, the Inner-Lock Locking Introducer
{2}------------------------------------------------
Sheaths are substantially equivalent to the predicate device, the Terumo RADIFOCUS® Introducer R/O device.
-
:
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle or bird with three curved lines representing its wings or feathers. The bird is positioned to the right of a circular text element that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA". The text is arranged around the circumference of the circle.
### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 2 1 2003
Ms. Susan D. Goldstein-Falk Rex Medical c/o MDI Consultants, Inc. 55 Northern Blvd. Suite 200 Great Neck, NY 11021
Re: K022170
Rex Medical Inner-Lock Locking Introducer Sheath Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II Product Code: DYB Dated: October 24, 2002 Received: October 28, 2002
Dear Ms. Goldstein-Falk:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Susan D. Goldstein-Falk
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
Also, please If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address
http://www.fda.qov/cdrh/dsma/dsmamain.html
Sincerely yours.
N. Harrington
am D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Exhibit B
Page _ 1 _ _ of _ _ 1 _
# 510(k) Number (if known): _ KO2Q170 __________________________________________________________________________________________________________________________________________
Device Name: Inner-Lock Locking Introducer Sheaths
Indications For Use:
The Inner-Lock Locking Introducer Sheaths are used to facilitate placing a catheter through the skin into a graft. The Inner-Lock dilator is an accessory device, which is used by placing it into the sheath to create an occlusion and further provide support to the wall of the indwelling sheath.
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | | |
|--------------------------------------------------------|---------|-----------------------------------------------|
| (Division Sign-Off) | | |
| Division of Cardiovascular Devices | | |
| 510(k) Number | K022170 | |
| Prescription Use (Per 21 CFR 801.109) | OR | Over-The-Counter Use (Optional Format 1-2-96) |
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.