Arm Blood Pressure Monitor, Model Name:111, Arm Blood Pressure Monitor, Model Name: B1681, Arm Blood Pressure Monitor, Model Name: B1682, Arm Blood Pressure Monitor, Model Name: B1683
Arm Blood Pressure Monitor, Model Name:111, Arm Blood Pressure Monitor, Model Name: B1681, Arm Blood Pressure Monitor, Model Name: B1682, Arm Blood Pressure Monitor, Model Name: B1683
Applicant
Shenzhen Yuezhongxing Technology Co., Ltd.
Product Code
DXN · Cardiovascular
Decision Date
May 5, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1130
Device Class
Class 2
Indications for Use
Arm Blood Pressure Monitor is intended for use by medical professionals or at home to monitor and display diastolic, systolic blood pressure and pulse rate on adult each time, with an air cuff buckled around one's arm according to the instruction in the user's guide manual.
Device Story
Battery or DC-powered non-invasive blood pressure monitor; uses oscillometric method via arm cuff. Integral pump inflates cuff; electric valve controls deflation. Semiconductor pressure sensor monitors cuff pressure; extracts pulse waveform data. Software analyzes waveform to determine systolic/diastolic blood pressure and pulse rate. Used in hospitals or home by medical professionals or patients. Displays readings on LCD; stores up to 2x99 readings. Includes irregular heartbeat detection (rhythm >25% deviation from average) and WHO blood pressure indicator bar. Provides objective data for clinical decision-making regarding hypertension management.
Clinical Evidence
Clinical study conducted per ISO 81060-2:2018 with 86 adult subjects (16-87 years old). Reference standard: mercury sphygmomanometer. Results: Mean difference (systolic/diastolic) 1.0/0.9 mmHg; SD 3.1/3.4 mmHg. Criterion 2 SD 2.4/2.6 mmHg. All results met ISO 81060-2:2018 acceptance criteria. No adverse effects reported.
Technological Characteristics
Semiconductor pressure sensor; oscillometric measurement; 4xAA batteries or 5V DC adapter. LCD display. Complies with IEC 60601-1, IEC 60601-1-11, IEC 80601-2-30, IEC 60601-1-2. Biocompatibility per ISO 10993-5 and ISO 10993-10.
Indications for Use
Indicated for adult patients requiring blood pressure and pulse rate monitoring. Contraindicated for use in ambulatory environments and in aircraft.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
Automatic Upper Arm Blood Pressure Monitor (K193627)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
May 5, 2023
Shenzhen Yuezhongxing Technology Co., Ltd. Yuchao Chen Sales Manager No.2, Zhenye Road, Liulian Community, Pingshan Avenue Pingshan District Shenzhen, Guangdong 518015 China
Re: K222926
Trade/Device Name: Arm Blood Pressure Monitor, Model Name: 111, Arm Blood Pressure Monitor, Model Name: B1681, Arm Blood Pressure Monitor, Model Name: B1682, Arm Blood Pressure Monitor, Model Name: B1683 Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: March 29, 2023 Received: March 30, 2023
Dear Yuchao Chen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{2}------------------------------------------------
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Robert T. Kazmierski -S
for
LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology. Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K222926
Device Name
Arm Blood Pressure Monitor (Models: 111,B1681,B1682,B1683)
Indications for Use (Describe)
Arm Blood Pressure Monitor is intended for use by medical professionals or at home to monitor and display diastolic, systolic blood pressure and pulse rate on adult each time, with an air cuff buckled around one's arm according to the instruction in the user's guide manual.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(K) Summary
Prepared in accordance with the requirements of 21 CFR Part 807.92
Prepared Date: May 03, 2023
### 1. Submitter's Information
The submitter of this pre-market notification is:
| Name: | Shenzhen Yuezhongxing Technology Co., Ltd. | | |
|-----------------|----------------------------------------------------------------------------------------------------|--|--|
| Address: | No.2, Zhenye Road, Liulian Community, Pingshan Avenue,<br>Pingshan District, Shenzhen,518015,CHINA | | |
| Contact person: | Yuchao Chen | | |
| Title: | Sales manager | | |
| E-mail: | 718145238@qq.com | | |
| Tel: | +86-135 1039 4198 | | |
# 2. Device Identification
| 510(K) number: | K222926 |
|--------------------|---------------------------------------------------|
| Trade/Device Name: | Arm Blood Pressure Monitor |
| Models: | 111, B1681, B1682, B1683 |
| Regulation Number: | 21 CFR 870.1130 |
| Regulation Name: | Noninvasive blood pressure measurement system |
| Common Name: | System, Measurement, Blood-Pressure, Non-Invasive |
| Regulation Class: | Class II |
| Product Code: | DXN |
| Panel: | Cardiovascular |
# 3. Predicate Device
| 510(K) number: | K193627 |
|--------------------|---------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | Automatic Upper Arm Blood Pressure Monitor |
| Models: | BA-801X, BA-802X, BA-803X, BA-805X, BA-806X, BA-811X,<br>BA-812X, BA-813X, BA-821X, BA-822X, BA-823X, BA-826X,<br>BA-818,BA-819 |
| Manufacturer: | DONGGUAN E-TEST TECHNOLOGY CO.,LTD |
| Regulation Number: | 21 CFR 870.1130 |
| Regulation Name: | Noninvasive blood pressure measurement system |
| Common Name: | System, Measurement, Blood-Pressure, Non-Invasive |
| Regulation Class: | Class II |
{5}------------------------------------------------
Traditional 510(k) Submission of Arm Blood Pressure Monitor
| Product Code: | DXN |
|---------------|----------------|
| Panel: | Cardiovascular |
### 4. Device Description
The Arm Blood Pressure Monitor (model: 111, B1681, B1683) are battery-powered or DC powered, automatic, non-invasive blood pressure system and intended to be use in hospital environment or at home. The Arm Blood Pressure Monitor (model: 111, B1682, B1683) are intended for use in adult patient population with arm circumference ranging from 22 cm to 32 cm. These devices are powered by 4 x AA batteries or an AC adaptor (output:5V DC). The device arm cuff inflates using an integral pump and deflates via an electric valve. During deflation, the arm cuff pressure is monitored and pulse waveform data is extracted pulse waveform data is then analyzed by software which determines pulse rate, as well as systolic and diastolic blood pressure. The systolic and diastolic blood pressures are measured using the oscillometric method. The cuff can measure pressure range from 0 to 280 mmHg, and the pulse rate range from 40 to 199
beats/min. When the device detected irregular rhythms, " W " will display on screen. An irregular heartbeat rhythm is defined as a rhythm that is 25% less or 25% more than the average rhythm detected while your monitor is measuring blood pressure. The WHO blood pressure indicator bar can classify by WHO and ISH recommendation. The devices display the latest blood pressure reading, while up to 2x99 readings can be stored in memory.
### 5. Indication for use
Arm Blood Pressure Monitor is intended for use by medical professionals or at home to monitor and display diastolic, systolic blood pressure and pulse rate on adult each time, with an air cuff buckled around one's arm according to the instruction in the user's quide manual.
{6}------------------------------------------------
# Traditional 510(k) Submission of Arm Blood Pressure Monitor
# 6. Comparison to Predicate Device
Compared to the predicate devices, the subject device has the same intended use, similar product design, same performance effectiveness as the predicate device, the summarized comparison information is listed in the following table:
| SE Comparisons | Subject devices | Predicate devices<br>K193627 | Note |
|---------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Company | Shenzhen Yuezhongxing<br>Technology Co., Ltd. | DONGGUAN E-TEST<br>TECHNOLOGY CO.,LTD | / |
| Product Name | Arm Blood Pressure Monitor | Automatic Upper Arm Blood<br>Pressure Monitor | / |
| Model Name | 111, B1681, B1682, B1683 | BA-801X, BA-802X,<br>BA-803X, BA-805X,<br>BA-806X, BA-811X,<br>BA-812X, BA-813X,<br>BA-821X, BA-822X,<br>BA-823X, BA-826X,<br>BA-818,BA-819 | / |
| Product Code | DXN | DXN | Same |
| Regulation name | Noninvasive blood pressure<br>measurement system | Noninvasive blood pressure<br>measurement system | Same |
| Indication for use | Arm Blood Pressure Monitor<br>is intended for use by medical<br>professionals or at home to<br>monitor and display diastolic,<br>systolic blood pressure and<br>pulse rate on adult each time,<br>with an air cuff buckled<br>around one's arm according<br>to the instruction in the user's<br>guide manual. | Automatic Upper Arm Blood<br>Pressure Monitor is intended<br>for use by medical<br>professionals or at home to<br>monitor and display<br>diastolic, systolic blood<br>pressure and pulse rate on<br>adult each time, with an air<br>cuff buckled around one's<br>arm according to the<br>instruction in the user's<br>guide manual. | Same |
| Environment of<br>use | Hospital, Home | Hospital, Home | Same |
| Patient<br>population | Adult | Adult | Same |
| SE Comparisons | Subject devices | Predicate devices<br>K193627 | Note |
| Contraindication | The monitor is contraindicated<br>for use in ambulatory<br>environment.<br>The monitor is contraindicated<br>in aircraft. | The monitor is<br>contraindicated for use in<br>ambulatory environment.<br>The monitor is<br>contraindicated in aircraft. | Same |
| Principle of<br>operation | Cuff oscillometric method | Cuff oscillometric method | Same |
| Measurement<br>site | Upper Arm | Upper Arm | Same |
| Measurement<br>range | Pressure:0 to 280 mmHg<br>Pulse rate: 40 to 199<br>beats/min | Pressure:0 to 280mmHg<br>Pulse rate: 40 to 199<br>beats/min | Same |
| Pressure sensor | Semiconductor pressure<br>sensor | Semiconductor pressure<br>sensor | Same |
| Arm<br>circumference | 22 cm to 32 cm | size A: 17cm-22cm<br>(SMALL ADULT CUFF)<br>size B: 22cm-30cm<br>(ADULT CUFF-1)<br>size C: 24cm-34cm<br>(ADULT CUFF-2)<br>size D: 22cm-42cm<br>(LARGE<br>ADULT CUFF)<br>size E: 30cm-42cm<br>(LARGE ADULT CUFF)<br>size F: 42cm-50cm<br>(EXTRA LARGE ADULT<br>CUFF) | See Note 1 |
| Accuracy of<br>pressure | $ ± 3 mmHg $ | $ ± 3 mmHg $ | Same |
| Accuracy of<br>pulse rate | $ ± 5 % of reading $ | $ ± 5 % of reading $ | Same |
| Inflation method | Automatic inflation with<br>piezoelectric pump | Automatic inflation with<br>piezoelectric pump | Same |
| Deflation method | Automatic rapid deflation<br>valve | Automatic rapid deflation<br>valve | Same |
| display | LCD | LCD | Same |
| Power source | 4 x AAA batteries, or | 4 x AAA batteries | See Note 2 |
| SE Comparisons | Subject devices | Predicate devices<br>K193627 | Note |
| | AC adapter<br>(input:<br>100-240V~,50/60Hz,0.2A<br>output:5V DC, 1.0A,5.0W) | | |
| Operation<br>condition | 5 to 40°C<br>15 to 93% RH<br>700 to 1060hPa | 5 to 40 °C<br>15 to 80% RH<br>800 to 1050 hPa | See Note 3 |
| Storage condition | -25 to 70°C<br>0 to 93%RH | -20 to 65 °C<br>10 to 95% RH<br>86 kPa to 106 kPa | See Note 3 |
| Transportation<br>condition | -25 to 70°C<br>0 to 93%RH | -20 to 65 °C<br>10 to 95% RH<br>86 kPa to 106 kPa | See Note 3 |
| Irregular heart<br>beat feature | Yes | Yes | Same |
| Hypertension<br>indication | Yes | Yes | Same |
{7}------------------------------------------------
Traditional 510(k) Submission of Arm Blood Pressure Monitor
Yuezhongxing
{8}------------------------------------------------
#### Traditional 510(k) Submission of Arm Blood Pressure Monitor
Yuezhongxing
Note 1: Arm cuff size is different. Do not impact safety or performance of device.
Note 2: The subject device can be powered by an AC adaptor (output:5V DC) beside 4 x AAA batteries. IEC 60601-1, IEC 60601-1-11 and IEC 60601-1-2 can demonstrate that the subject device can maintain the safety and performance when is powered by an AC adaptor(output:5V DC). Thus, this difference does not raise different questions of safety and effectiveness.
Note 3: After conducting simulate transportation test, result shows no risk arise.
All the differences don't affect the safety and effectiveness which is concluded after all the required testing, so no safety and effectiveness issues relating to the system come into conclusion.
### 8. Performance Data
#### Clinical test:
We performed a clinical study to verify clinical accuracy of the subject devices in accordance with ISO 81060-2:2018. The study was conducted at Wanxiang Xintian Community Health Serve Center,Xixiang Street, Bao'an District, Shenzhen City China by Shenzhen Cihai Hospital.
We selected auscultatory method (mercury sphygmomanometer) as the reference standard to determine the clinical accuracy by calculating the mean value and standard deviation according to ISO 81060-2:2018.
510(k) Summary
{9}------------------------------------------------
#### Traditional 510(k) Submission of Arm Blood Pressure Monitor
#### Subject
Number: A total of 86 subjects voluntarily participated in this study,
Gender:44 of them were male and 42 were female,
Age: ranging in age from 16 to 87 years old.
Race: Asian
Ethnicity: Not Hispanic or Latino
Arm size distribution:
37 (43%) subjects have an arm circumference which lies within the upper half of the specified range of use of the cuff,
49 (57%) subjects have an arm circumference within the lower half of the specified range of use of the cuff,
22 (26%) subjects I have an arm circumference which lies within the upper quarter of the specified range of use of the cuff,
26 (30%) subjects have an arm circumference within the lower quarter of the specified range of use of the cuff;
15 (17%) subjects have an arm circumference which lies within the upper eighth of the specified range of use of the cuff,
and 10(12%) subjects have an arm circumference within the lower eighth of the specified range of use of the cuff.
#### Procedure:
- a. Either arm may be utilized
- b. Using the reference sphygmomanometer (mercury sphygmomanometer), have the observers determine the subject's blood pressure
- Interchange cuffs and wait at least 60s ﻥ
- d. Use the sphygmomanometer-under-test (subject device) to determine the subject's blood pressure
- e. Clear the sphygmomanometer-under-test memory of the previous determination and then wait at least 60s
- f. Do not use the data points obtained in b) and d) in the calculation of accuracy.
- Using the reference sphygmomanometer (mercury sphygmomanometer), have the g. observers determine the subject's blood pressure
- h. Interchange cuffs and wait at least 60 s.
- i. Use the sphygmomanometer-under-test (subject device) to determine the subject's blood pressure
{10}------------------------------------------------
#### Traditional 510(k) Submission of Arm Blood Pressure Monitor
Interchange cuffs and wait at least 60 s. i.
k. Have the observers use the reference sphygmomanometer (mercury sphygmomanometer) to determine the subject's blood pressure reference reading.
l. Interchange cuffs and wait at least 60 s.
m. Repeat i) to I) until the required number of valid reference readings and determinations have been performed.
#### Result
Blood pressure distribution:
7 (8%) subjects of the reference blood pressure readings have a systolic blood pressure ≤100 mmHg
11 (13%) subjects of the reference blood pressure readings have a systolic blood pressure≥160 mm Hq.
20 (23%) subjects of the reference blood pressure reading have a systolic blood pressure ≥140mmHg.
8 (9%) subjects of the reference blood pressure reading have a diastolic blood pressure ≤ 60 mmHg.
11 (13%) subjects of the reference blood pressure readings have a diastolic blood pressure ≥ 100 mm Hg.
21 (24%)subiects of the reference blood pressure readings have a diastolic blood pressure ≥85 mm Hg.
Adverse effects and complications:
There is no adverse effect and complication was found during study.
Data analysis:
According to the method of criterion 1 in ISO 81060-2:2018,5.2.4.1.2, the mean value of the differences of systolic blood pressure between the sphygmomanometer-under-test and mercury sphygmomanometer was 1.0, and the standard deviation was 3.1. The mean value of the differences of diastolic blood pressure between the sphygmomanometer-under-test and mercury sphygmomanometer was 0.9 and the standard deviation was 3.4.
According to the method of criterion 2 in ISO 81060-2:2018,5.2.4.1.2, the standard deviation of systolic blood pressure of the sphygmomanometer-under-test and mercury sphygmomanometer was 2.4, the standard deviation of diastolic blood pressure of the sphygmomanometer-under-test and mercury sphygmomanometer was 2.6.
#### Conclusion
The statistical analysis results were within the ISO 81060-2-2018 standard acceptance standard,
510(k) Summary
{11}------------------------------------------------
Traditional 510(k) Submission of Arm Blood Pressure Monitor
compared with the results of the mercury sphygmomanometer, determine that the subject device have enough clinical accuracy.
### Non-clinical data
The Arm Blood Pressure Monitor comply with:
Safety and performance:
IEC 60601-1:2005+A1:2012 Medical electrical equipment – Part 1: General requirements for 1. basic safety and essential performance.
2. IEC 60601-1-11:2015 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
3. IEC 80601-2-30:2018 Medical electrical equipment - Part 2-30: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers
Electromagnetic Compatibility:
IEC 60601-1-2:2014 Medical electrical equipment-Part1-2: General requirements for basic 4. safety and essential performance-Collateral Standard: Electromagnetic disturbances-Requirements and tests
Biocompatibility:
ISO 10993-10:2010 Biological evaluation of medical devices - Part 10: Tests for irritation 5. and skin sensitization
ISO 10993-5:2009 Biological evaluation of medical devices - Part 5: Tests for in vitro 6. cytotoxicity
Software Verification and Validation:
FDA software validation quidance "General Principles of Software Validation; Final Guidance for Industry and FDA Staff, Document issued on: January 11, 2002".
Software documentation for moderate level of concern per the FDA Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices
# 9. Conclusion
The conclusions drawn from the nonclinical tests demonstrate that the subject device is as safe and effective as the predicated device(K193627).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.