Automatic Upper Arm Blood Pressure Monitor is intended for use by medical professionals or at home to monitor and display diastolic, systolic blood pressure and pulse rate on adult each time, with an air cuff buckled around one's arm according to the instruction in the user's guide manual.
Device Story
Device is an automatic upper arm blood pressure monitor using oscillometric method. Input: cuff pressure changes detected by electronic capacitive pressure sensor. Processing: sensor converts pressure alterations to electrical signals; algorithm analyzes signals to determine systolic/diastolic pressure and pulse rate. Output: digital display of blood pressure, pulse rate, date, time, WHO classification bar, and status icons (low battery, heart). Used in clinical or home settings by patients or medical professionals. Benefits: non-invasive, convenient blood pressure monitoring for hypertension management.
Clinical Evidence
No new clinical data presented. Substantial equivalence is based on identical hardware and software to predicate devices, which were previously demonstrated to conform to AAMI/ANSI/ISO 81060-2 for clinical validation of automated measurement.
Technological Characteristics
Materials: ABS plastic, Nylon, Latex. Sensing: Electronic capacitive pressure sensor. Energy: DC 6V (4 x AA alkaline batteries). Connectivity: Standalone. Software: V01. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 80601-2-30, ISO 10993-5/10, ISO 81060-2.
Indications for Use
Indicated for adult patients to monitor and display systolic/diastolic blood pressure and pulse rate using an upper arm cuff. Intended for use by medical professionals or at home.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
AGE Automatic Upper Arm Blood Pressure Monitor (K153552)
AGE Automatic Upper Arm Blood Pressure Monitor (K172895)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 12, 2020
Dongguan E-Test Technology Co., LTD % Cassie Lee, Manager Dongguan E-Test Technology Co., Ltd Room 201, 301. Building 1, Changping Section No.1, Dongshen Road Dongguan City, Guangdong, China 523588
Re: K193627
Trade/Device Name: Automatic Upper Arm Blood Pressure Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive blood pressure measurement system Regulatory Class: Class II Product Code: DXN Dated: December 24, 2019 Received: December 26, 2019
Dear Cassie Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
LT Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K193627
Device Name
Automatic Upper Arm Blood Pressure Monitor
Indications for Use (Describe)
Automatic Upper Arm Blood Pressure Monitor is intended for use by medical professionals or at home to monitor and display diastolic, systolic blood pressure and pulse rate on adult each time, with an air cuff buckled around one's arm according to the instruction in the user's guide manual.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.
### 1. Submitter's Information
510(k) Owner's Name: DONGGUAN E-TEST TECHNOLOGY CO., LTD
Establishment Registration Number: Applying
Address: Room 201,301. Building 1, Changping Section No.1,Dongshen Road, Changping
Town, Dongguan City, Guangdong, China.
Tel: +86-0769-81158038
Contact Person (including title): Victor Wan (Vice-president)
E-mail: victor@agelh.com
# Application Correspondent:
Contact Person: Ms. Cassie Lee
DONGGUAN E-TEST TECHNOLOGY CO., LTD
Address: Room 201,301, Building 1, Changping Section No.1, Dongshen Road, Changping
Town, Dongguan City, Guangdong, China.
Tel: +86-0769-81158038
{5}------------------------------------------------
K193627
Email: requlatory @glomed-info.com
# 2. Subject Device Information
Type of 510(k): Traditional
Common Name: Noninvasive blood pressure measurement systems
Classification Name: System, Measurement, Blood-Pressure, Non-Invasive
Trade Name: Automatic Upper Arm Blood Pressure Monitor
Model Name: BA-801X, BA-802X, BA-803X, BA-805X, BA-806X, BA-811X, BA-812X, BA-813X,
BA-821X, BA-822X, BA-823X, BA-826X, BA-818, BA-819
Review Panel: Cardiovascular
Product Code: DXN
Regulation Number: 870.1130
Regulatory Class: 2
# 3. Predicate Device Information
| Sponsor | Dongguan Ageless Health Industrial Co., Ltd<br>Dongguan Ageless Health Industrial Co., Ltd |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name<br>and Model | AGE Automatic Upper Arm Blood Pressure Monitor<br>Models: BA-801X, BA-802X, BA-803X, BA-805X, BA-806X, BA-811X, BA-812X, BA-813X, BA-821X, BA-822X, BA-823X, BA-826X<br>AGE Automatic Upper Arm Blood Pressure Monitor<br>Models: BA-815, BA-816, BA-818, BA-819 |
{6}------------------------------------------------
| 510(k) Number | K153552 | K172895 |
|----------------------|----------|----------|
| Product Code | DXN | DXN |
| Regulation<br>Number | 870.1130 | 870.1130 |
| Regulation<br>Class | II | II |
### 4. Device Description
Automatic Upper Arm Blood Pressure Monitor is designed to measure the systolic, diastolic and pulse rate of an individual by using a non-invasive technique which an inflatable cuff is wrapped around upper arm. Our method to define systolic and diastolic pressures is similar to the auscultatory method but using an electronic capacitive pressure sensor rather than stethoscope and mercury manometer. The sensor converts tiny alteration in cuff pressure to electrical signals; by analyzing those signals to define the systolic, diastolic and calculating pulse rate is a well-known technique in the market so called "oscillometric method". The device also has low voltage indication, which will be triggered when the battery is low.
#### 5. Intended Use / Indications for Use
Automatic Upper Arm Blood Pressure Monitor is intended for use by medical professionals or at home to monitor and display diastolic, systolic blood pressure and pulse rate on adult each time, with an air cuff buckled around one's arm according to the instruction in the user's guide manual.
{7}------------------------------------------------
### 6. Test Summary
The whole product and manufacturing used for the Automatic Upper Arm Blood Pressure Monitor are identical to those of the predicate device, which were demonstrated to conform with the following standards:
- ◆ IEC 60601-1, Medical Electrical Equipment - Part 1: General Requirements for Safety, 2005 + A1:2012
- � IEC 60601-1-2, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic Disturbances -Requirements and tests, 2014
- IEC 60601-1-11, Medical Electrical Equipment Part 1-11: General Requirements for Basic � Safety And Essential Performance - Collateral Standard: Requirements For Medical Electrical Equipment And Medical Electrical Systems Used In The Home Healthcare Environment [Including: Technical Corrigendum 1 (2011)]
- � IEC 80601-2-30, Medical Electrical Equipment - Part 2-30: Particular Requirements For The Basic Safety And Essential Performance Of Automated Noninvasive Sphygmomanometers
- � ISO 10993-5, Biological evaluation of medical devices -- Part 5: Tests for In Vitro cytotoxicity, 2009
- � ISO 10993-10, Biological evaluation of medical devices -- Part 10: Tests for irritation and skin sensitization, 2010
- AAMI / ANSI / ISO 81060-2 Second Edition, Non-Invasive Sphygmomanometers Part 2: � Clinical Validation of Automated Measurement Type. (Cardiovascular)
# 7. Comparison to predicate device and conclusion
{8}------------------------------------------------
The technological characteristics, features, specifications, materials, mode of operation, and intended use of Automatic Upper Arm Blood Pressure Monitor is substantially equivalent to the predicate devices quoted above.
The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness.
| Elements of<br>Comparison | Subject Device | Predicate Device1 | Predicate Device2 | Verdict | |
|--------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|----|
| Company | DONGGUAN E-TEST<br>TECHNOLOGY CO.,<br>LTD | Dongguan Ageless Health<br>Industrial Co., Ltd | Dongguan Ageless Health<br>Industrial Co., Ltd | -- | |
| Product Name | Automatic Upper Arm<br>Blood Pressure Monitor | AGE Automatic Upper<br>Arm Blood Pressure<br>Monitor | AGE Automatic Upper<br>Arm Blood Pressure<br>Monitor | -- | |
| Model Name | BA-801X, BA-802X, BA-<br>803X, BA-805X, BA-<br>806X, BA-811X, BA-<br>812X, BA-813X, BA-<br>821X, BA-822X, BA-<br>823X, BA-826X, BA-818,<br>BA-819 | BA-801X, BA-802X, BA-<br>803X, BA-805X, BA-<br>806X, BA-811X, BA-<br>812X, BA-813X, BA-<br>821X, BA-822X, BA-<br>823X, BA-826X | BA-815, BA-816, BA-818,<br>BA-819 | -- | |
| Classification<br>Name | System, Measurement,<br>Blood-Pressure, Non-<br>Invasive | System, Measurement,<br>Blood-Pressure, Non-<br>Invasive | System, Measurement,<br>Blood-Pressure, Non-<br>Invasive | SE | |
| 510(k) Number | K193627 | K153552 | K172895 | -- | |
| Product Code | DXN | DXN | DXN | SE | |
| Elements of<br>Comparison | Subject Device | Predicate Device1 | Predicate Device2 | Verdict | |
| Intended Use /<br>Indications for<br>Use | Automatic Upper Arm<br>Blood Pressure Monitor<br>is intended for use by<br>medical professionals or<br>at home to monitor and<br>display diastolic, systolic<br>blood pressure and pulse<br>rate on adult each time,<br>with an air cuff buckled<br>around one's arm<br>according to the<br>instruction in the user's<br>guide manual. | AGE Automatic Upper<br>Arm Blood Pressure<br>Monitor is intended for<br>use by medical<br>professionals or at home<br>to monitor and display<br>diastolic, systolic blood<br>pressure and pulse rate<br>on adult each time, with<br>an air cuff buckled around<br>one's arm according to<br>the instruction in the<br>user's guide manual. | AGE Automatic Upper<br>Arm Blood Pressure<br>Monitor is intended for<br>use by medical<br>professionals or at home<br>to monitor and display<br>diastolic, systolic blood<br>pressure and pulse rate<br>on adult each time, with<br>an air cuff buckled around<br>one's arm according to<br>the instruction in the<br>user's guide manual. | SE | |
| Energy source | DC 6V (4 X AA<br>1.5V alkaline batteries) | DC 6V (4 X AA<br>1.5V alkaline batteries) | DC 6V (4 X AA<br>1.5V alkaline batteries) | SE | |
| Design | Software | Same version software | | | SE |
| | hardware | Same version hardware | | | |
| Material | ABS plastic, Nylon, Latex | ABS plastic, Nylon, Latex | ABS plastic, Nylon, Latex | SE | |
| Measurement<br>Site | Upper Arm | Upper Arm | Upper Arm | SE | |
| Measuring<br>range | Pressure: 0~280 mmHg<br>Pulse: 40~199<br>beats/minute | Pressure: 0~280 mmHg<br>Pulse: 40~199<br>beats/minute | Pressure: 0~280 mmHg<br>Pulse: 40~199<br>beats/minute | SE | |
| Pressure<br>resolution | 1 mmHg or 0.1kPa | 1 mmHg or 0.1kPa | 1 mmHg or 0.1kPa | SE | |
| Measuring<br>accuracy | Pressure: ± 3mmHg | Pressure: ± 3mmHg | Pressure: ± 3mmHg | SE | |
| Elements of<br>Comparison | Subject Device | Predicate Device1 | Predicate Device2 | Verdict | |
| | Pulse: ±5% | Pulse: ±5% | Pulse: ±5% | | |
| Cuff<br>Circumference | There are 6 size:<br>size A: 17cm--22cm<br>(SMALL ADULT CUFF)<br>size B: 22cm--30cm<br>(ADULT CUFF-1)<br>size C: 24cm--34cm<br>(ADULT CUFF-2)<br>size D: 22cm--42cm (L-<br>LARGE ADULT CUFF)<br>size E: 30cm--42cm<br>(LARGE ADULT CUFF)<br>size F: 42cm--50cm<br>(EXTRA LARGE ADULT<br>CUFF) | size A: 17cm--22cm<br>(SMALL ADULT CUFF)<br>size B: 22cm--30cm<br>(ADULT CUFF-1)<br>size C: 24cm--34cm<br>(ADULT CUFF-2)<br>size D: 22cm--42cm (L-<br>LARGE ADULT CUFF)<br>size E: 30cm--42cm<br>(LARGE ADULT CUFF)<br>size F: 42cm--50cm<br>(EXTRA LARGE ADULT<br>CUFF) | For BA-815: 22-34 cm;<br>For BA-816: 28-42 cm;<br>For BA-818 and BA-819,<br>there are 6 size:<br>size A: 17cm--22cm<br>(SMALL ADULT CUFF)<br>size B: 22cm--30cm<br>(ADULT CUFF-1)<br>size C: 24cm--34cm<br>(ADULT CUFF-2)<br>size D: 22cm--42cm (L-<br>LARGE ADULT CUFF)<br>size E: 30cm--42cm<br>(LARGE ADULT CUFF)<br>size F: 42cm--50cm<br>(EXTRA LARGE ADULT<br>CUFF) | SE | |
| Size and<br>Weight of<br>model BA-<br>801X | About 460g (not include<br>battery);<br>(L)115.0mm X<br>(D)103.0mm X (H)<br>67.5mm | About 460g (not include<br>battery);<br>(L)115.0mm X<br>(D)103.0mm X (H)<br>67.5mm | / | SE | |
| Size and<br>Weight of<br>model BA-<br>802X | About 460g (not include<br>battery);<br>(L)115.0mm X<br>(D)103.0mm X (H)<br>67.5mm | About 460g (not include<br>battery);<br>(L)115.0mm X<br>(D)103.0mm X (H)<br>67.5mm | / | SE | |
| Elements of | Subject Device | Predicate Device1 | Predicate Device2 | Verdict | |
| Comparison | | | | | |
| Size and<br>Weight of<br>model BA-<br>803X | About 460g (not include<br>battery)<br>(L)119.0mmX<br>(D)100.0mm X<br>(H)63.2mm | About 460g (not include<br>battery)<br>(L)119.0mmX<br>(D)100.0mm X<br>(H)63.2mm | / | SE | |
| Size and<br>Weight of<br>model BA-<br>805X | About 460g (not include<br>battery)<br>(L) 118.3mm X<br>(D)100.0mm X<br>(H)62.9mm | About 460g (not include<br>battery)<br>(L) 118.3mm X<br>(D)100.0mm X<br>(H)62.9mm | / | SE | |
| Size and<br>Weight of<br>model BA-<br>806X | About 460g (not include<br>battery)<br>(L) 146.0mm X<br>(D)113.0mm X<br>(H)66.0mm | About 460g (not include<br>battery)<br>(L) 146.0mm X<br>(D)113.0mm X<br>(H)66.0mm | / | SE | |
| Size and<br>Weight of<br>model BA-<br>811X | About 460g (not include<br>battery)<br>(L) 144.0mm X<br>(D)113.0mm X<br>(H)67.0mm | About 460g (not include<br>battery)<br>(L) 144.0mm X<br>(D)113.0mm X<br>(H)67.0mm | / | SE | |
| Size and<br>Weight of<br>model BA-<br>812X | About 460g (not include<br>battery)<br>(L) 113.5mm X<br>(D)92.0mm X (H)48.0mm | About 460g (not include<br>battery)<br>(L) 113.5mm X<br>(D)92.0mm X (H)48.0mm | / | SE | |
| Size and<br>Weight of<br>model BA-<br>813X | About 460g (not include<br>battery)<br>(L) 115.3mm X<br>(D)92.0mm X (H)48.0mm | About 460g (not include<br>battery)<br>(L) 115.3mm X<br>(D)92.0mm X (H)48.0mm | / | SE | |
| Elements of<br>Comparison | Subject Device | Predicate Device1 | Predicate Device2 | Verdict | |
| | | | | | |
| Size and<br>Weight of<br>model BA-<br>821X | About 610g (not include<br>battery)<br>(L) 150.0mm X<br>(D)100.0mm X<br>(H)60.0mm | About 610g (not include<br>battery)<br>(L) 150.0mm X<br>(D)100.0mm X<br>(H)60.0mm | / | SE | |
| Size and<br>Weight of<br>model BA-<br>822X | About 380g (not include<br>battery)<br>(L) 135.0mm X<br>(D)90.0mm X<br>(H)47.65mm | About 380g (not include<br>battery)<br>(L) 135.0mm X<br>(D)90.0mm X<br>(H)47.65mm | / | SE | |
| Size and<br>Weight of<br>model BA-<br>823X | About 460g (not include<br>battery)<br>138.0mm X(D)95.0mm<br>X (H)67.0mm | About 460g (not include<br>battery)<br>138.0mm X (D)95.0mm X<br>(H)67.0mm | / | SE | |
| Size and<br>Weight of<br>model BA-<br>826X | About 420g (not include<br>battery)<br>143.0mm X(D)115.0mm<br>X (H)57.0mm | About 420g (not include<br>battery)<br>143.0mm X (D)115.0mm<br>X (H)57.0mm | / | SE | |
| Size and<br>Weight of<br>model BA-818 | About 465g (not include<br>battery)<br>(L) 103.0mm X<br>(D)129.0mm X<br>(H)161.0mm | / | About 465g (not include<br>battery)<br>(L) 103.0mm X<br>(D)129.0mm X<br>(H)161.0mm | SE | |
| Size and<br>Weight of<br>model BA-819 | About 438g (not include<br>battery)<br>(L) 160.0mm X<br>(D)98.0mm X (H)55.0mm | / | About 438g (not include<br>battery)<br>(L) 160.0mm X<br>(D)98.0mm X (H)55.0mm | SE | |
| Elements of | Subject Device | Predicate Device1 | Predicate Device2 | Verdict | |
| Comparison | | | | | |
| Inflation and<br>Deflation | Automatic | Automatic | Automatic | SE | |
| Measuring<br>Method | Oscillometry | Oscillometry | Oscillometry | SE | |
| Patient<br>Population | Adult | Adult | Adult | SE | |
| Display | Blood Pressure (Systolic<br>and Diastolic), Pulse,<br>Date, Time, WHO BP<br>Classification Indicating<br>Bar, Low Battery Icon,<br>Heart Icon, Memory<br>Record Number | Blood Pressure (Systolic<br>and Diastolic), Pulse,<br>Date, Time, WHO BP<br>Classification Indicating<br>Bar, Low Battery Icon,<br>Heart Icon, Memory<br>Record Number | Blood Pressure (Systolic<br>and Diastolic), Pulse,<br>Date, Time, WHO BP<br>Classification Indicating<br>Bar, Low Battery Icon,<br>Heart Icon, Memory<br>Record Number | SE | |
| Software<br>Version | V01 | V01 | V01 | SE | |
| Operation<br>condition | Temperature: 5°C ~<br>40°C<br>Humidity: 15~90%RH<br>Atmospheric Pressure:86<br>kPa~106 kPa | Temperature: 5°C ~ 40°C<br>Humidity: 15~90%RH<br>Atmospheric Pressure:86<br>kPa~106 kPa | Temperature: 5°C ~ 40°C<br>Humidity: 15~90%RH<br>Atmospheric Pressure:86<br>kPa~106 kPa | SE | |
| Transport/stora<br>ge environment | Temperature: -20°C ~<br>+65°C<br>Humidity: 10~95%RH<br>Atmospheric Pressure:86<br>kPa~106 kPa | Temperature: -20°C ~<br>+65°C<br>Humidity: 10~95%RH<br>Atmospheric Pressure:86<br>kPa~106 kPa | Temperature: -20°C ~<br>+65°C<br>Humidity: 10~95%RH<br>Atmospheric Pressure:86<br>kPa~106 kPa | SE | |
| Safety | IEC 60601-1<br>IEC 60601-1-11 | IEC 60601-1<br>IEC 60601-1-11 | IEC 60601-1<br>IEC 60601-1-11 | SE | |
| Elements of | Subject Device | Predicate Device1 | Predicate Device2 | Verdict | |
| Comparison | | | | | |
| | IEC 80601-2-30 | IEC 80601-2-30 | IEC 80601-2-30 | | |
| EMC | IEC 60601-1-2 | IEC 60601-1-2 | IEC 60601-1-2 | SE | |
| Biocompatibility | ISO 10993 series | ISO 10993 series | ISO 10993 series | SE | |
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
{12}------------------------------------------------
{13}------------------------------------------------
{14}------------------------------------------------
# Comparison in Detail(s):
# Final Conclusion:
The subject device noninvasive sphygmomanometer (Automatic Upper Arm Blood Pressure Monitor) (Model: BA-801X, BA-802X, BA-803X, BA-805X, BA-806X, BA-811X, BA-812X, BA-813X, BA-821X, BA-822X, BA-823X, BA-826X, BA-818, BA-819) has all features of the predicate device. The few differences do not affect the safety and effectiveness of the subject device. Thus, the subject device is substantially equivalent to the predicate device.
All hardware and software of the subject device are based on that of the predicate device K 153552 and K 172895 since no new testing is presented in the submission.
8. Date of the summary prepared: July 8, 2020
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.