Wrist Blood Pressure Monitor, Blood Pressure Monitor

K210768 · Andon Health Co, Ltd. · DXN · Jul 30, 2021 · Cardiovascular

Device Facts

Record IDK210768
Device NameWrist Blood Pressure Monitor, Blood Pressure Monitor
ApplicantAndon Health Co, Ltd.
Product CodeDXN · Cardiovascular
Decision DateJul 30, 2021
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 870.1130
Device ClassClass 2

Indications for Use

Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist. The cuff circumference is limited to 14cm-25cm.

Device Story

The Fully Automatic Electronic Blood Pressure Monitor (models KD-743V, KD-743B, KD-752) is a non-invasive device for measuring systolic/diastolic blood pressure and pulse rate. It utilizes an inflatable wrist cuff and an oscillometric pressure sensor to detect arterial pressure oscillations. The device processes these signals to calculate blood pressure and pulse, displaying results on an LCD screen. It includes features for irregular heartbeat detection and blood pressure classification. Used in clinical or home settings by patients or medical professionals, the device provides immediate feedback to assist in monitoring cardiovascular health. Data is stored in internal memory, with some models offering averaging functions (e.g., last 3 readings, 7-day averages). The device operates on DC power (batteries).

Clinical Evidence

No new clinical trials were conducted. Substantial equivalence is supported by clinical test reports from reference devices KD-7961 (K090769) and KD-972 (K121470), which share the same algorithm and design principles. Accuracy was verified against ISO 81060-2 criteria.

Technological Characteristics

Oscillometric measurement via silicon integrated pressure sensor. Powered by 2x1.5V AAA batteries. Wrist cuff circumference 14-25cm. LCD display (backlight on KD-752). Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, and IEC 80601-2-30. Algorithm: Amplitude-based.

Indications for Use

Indicated for adult individuals requiring non-invasive measurement of systolic and diastolic blood pressure and pulse rate via a wrist-worn inflatable cuff (14cm-25cm circumference). Suitable for use by medical professionals or for home use.

Regulatory Classification

Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. July 30, 2021 Andon Health Co., Ltd. Liu Yi President No. 3 Jin Ping Street, Ya An Road, Nankai District Tianjin, Tianjin 300190 China Re: K210768 Trade/Device Name: Wrist Blood Pressure Monitor, Wireless Blood Pressure Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: June 24, 2021 Received: July 1, 2021 Dear Liu Yi: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K210768 Device Name Fully Automatic Electronic Blood Pressure Monitor Indications for Use (Describe) Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist. The cuff circumference is limited to 14cm-25cm. Type of Use (Select one or both, as applicable)| | Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ K210768 # 510(K) SUMMARY (In accordance with 21 CFR 807.92) ## 1.0 Submitter's Information | Name: | Andon Health Co., Ltd. | |----------------------|-----------------------------------------------------------------------------------| | Address: | No 3, Jinping Street, Ya An Road, Nankai District, Tianjin,<br>300190, P.R. China | | Phone Number: | 86-22-87611660 | | Fax Number: | 86-22-87612379 | | Contact: | Mr. Liu Yi | | Date of Preparation: | March 1, 2021 | ## 2.0 Device Information | Device Name: | Fully Automatic Electronic Blood Pressure Monitor | |----------------------|---------------------------------------------------| | Common Name: | Non-Invasive Blood Pressure Measurement System | | Classification Name: | Non-Invasive Blood Pressure Measurement System | ## 3.0 Classification | Product Code: | DXN | |--------------------|--------------------| | Regulation Number: | 21 CFR 870.1130 | | Classification: | II | | Review Panel: | 870 Cardiovascular | ### 4.0 Predicate Device Information | Manufacturer: | Andon Health Co., Ltd. | Andon Health Co., Ltd. | |-------------------|------------------------------------------------------|------------------------------------------| | Device: | Fully Automatic Electronic Blood<br>Pressure Monitor | Wireless Blood Pressure Wrist<br>Monitor | | 510(k)<br>Number: | K183535 | K163276 | | Classification | II | II | | Product Code | DXN | DXN | ## 5.0 Indications for Use Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist. The cuff circumference is limited to 14cm-25cm. {4}------------------------------------------------ ## 6.0 Device Descrintion Fully Automatic Electronic Blood Pressure Monitor (KD-743V, KD-743B, KD-752) is designed and manufactured according to IEC 80601-2-30. The operational principle is based on oscillometric and silicon integrates pressure sensor technology. It can calculate the systolic and diastolic blood pressure, and display the result on the LCD. The measurements results can also be classified by the function of blood pressure classification indicator. If any irregular heartbeat is detected, it can be shown to the user. ## 7.0 Comparison of Technological Characteristics with Predicate Device The following table is the summary of the technological characteristics of the proposed subject device and predicate device. | Item | Subject Device<br>(KD-743V,<br>KD-743B,<br>KD-752) | Predicate<br>Device<br>(KD-753<br>K183535) | Predicate<br>Device<br>(KD-721)<br>(K163276) | Comparison Result | |----------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------| | Name and<br>mode | Fully Automatic<br>Electronic Blood<br>Pressure Monitor | Fully<br>Automatic<br>Electronic<br>Blood Pressure<br>Monitor | Fully Automatic<br>Electronic Blood<br>Pressure Monitor | Same | | Model | KD-743V, KD-743B<br>KD-752 | KD-753 | KD-721 | -- | | Rx or OTC | OTC | OTC | OTC | Same | | Population | Adult | Adult | Adult | Same | | Cuff Location | Wrist | Wrist | Wrist | Same | | <b>Physical Attributes</b> | | | | | | Weight<br>(exclude<br>batteries and<br>cuff) | KD-743V: 80g<br>KD-743B: 78g<br>KD-752: 80g | 71g | About 67g | Changed | | Dimensions | KD-743V:<br>87mm×66.3mm×33.3mm<br>KD-743B:<br>87mm×66.3mm×33.3mm<br>KD-752:<br>100mm×70mm×23.5mm | 83mm×74mm<br>×26mm | 78mm×60mm<br>×28mm | Changed | | Item | Subject Device<br>(KD-743V,<br>KD-743B,<br>KD-752) | Predicate<br>Device<br>(KD-753<br>K183535) | Predicate<br>Device<br>(KD-721)<br>(K163276) | Comparison Result | | Memory | KD-743V& KD-743B:<br>1×120 times<br>1×60 times<br>2×60 times<br>4×30 times<br><br>KD-752:<br>2×120 times<br>4×60 times<br>2×60 times<br>4×30 times | 1×120 times<br>1×60 times<br>2×60 times<br>4×30 times | 2×60 times | KD-743V&<br>KD-743B are same<br>with predicate<br>device KD-753<br>KD-752 add more<br>memory function on<br>meter | | Displayed<br>Calculated<br>Parameters | SYS<br>DIA<br>Pulse<br>IHB | SYS<br>DIA<br>Pulse<br>IHB | SYS<br>DIA<br>Pulse<br>IHB | Same | | Display<br>component | KD-743V&<br>KD-743B:<br>LCD<br><br>KD-752:<br>LCD with<br>backlight | LCD | LCD | KD-743V&<br>KD-743B are same<br>with predicate<br>device<br>KD-752 add<br>additional backlight | | Average<br>function | KD-743V&<br>KD-743B:<br>1. Average value of all<br>results in the current user<br>memory zone.<br>2. Average the value of<br>the latest 3 times<br><br>KD-752:<br>1. Average value of all<br>results in the current user<br>memory zone.<br>2. Average the value of<br>the latest 3 times<br>3. Average value of all<br>the results in last 7 days<br>AM<br>4. Average value of all | 1.Average<br>value of all<br>results in the<br>current user<br>memory zone.<br>2. Average the<br>value of the<br>latest 3 times | average value of<br>all the results<br>measured from 5<br>o'clock to 9<br>o'clock in last 7<br>days | KD-743V&<br>KD-743B are same<br>with predicate<br>device KD-753<br>KD-752 add more<br>functions | | Item | Subject Device<br>(KD-743V,<br>KD-743B,<br>KD-752)<br>the results in last 7 days PM | Predicate Device<br>(KD-753<br>K183535) | Predicate Device<br>(KD-721)<br>(K163276) | Comparison Result | | Other<br>Displayed<br>Information | Date<br>Time<br>Memory<br>Battery usage<br>Blood pressure<br>classification | Date<br>Time<br>Memory<br>Battery usage<br>Blood pressure<br>classification | Date<br>Time<br>Memory<br>Battery usage<br>Bluetooth<br>symbol<br>Blood pressure<br>classification<br>(Displayed on<br>device) | Similar | | Electrical Power | | | | | | DC Mains | 3V | 3V | 3V | Same | | Battery | 2 ×1.5V SIZE AAA | 2 ×1.5V SIZE AAA | 2 ×1.5V SIZE AAA | Same | | Environmental Operation | | | | | | Temperature | 10~40°C | 10~40°C | 10~40°C | Same | | Humidity | ≤85% | ≤85% | ≤85% | Same | | Environmental Storage | | | | | | Temperature | -20~50°C | -20~50°C | -20~50°C | Same | | Humidity | ≤85% | ≤85% | ≤85% | Same | | Performance NIBP | | | | | | Pulse Rate<br>Range | 40 -180times/min | 40 -180times/min | 40 -180times/min | Same | | Pulse Rate<br>Accuracy | Within ±5% | Within ±5% | Within ±5% | Same | | Technique/<br>Method | Oscillometric | Oscillometric | Oscillometric | Same | | Measure<br>process | KD-743V&<br>KD-743B:<br>Measure during deflating<br><br>KD-752:<br>Measure during inflating | Measure<br>during<br>deflating | Measure<br>during<br>inflation | KD-743V&<br>KD-743B are same<br>with predicate<br>device KD-753<br>KD-752 are same<br>with predicate<br>device KD-721 | | Pressure<br>Accuracy | Within ±3mmHg | Within ±3mmHg | Within ±3mmHg | Same | | Item | Subject Device<br>(KD-743V,<br>KD-743B,<br>KD-752) | Predicate Device<br>(KD-753<br>K183535) | Predicate Device<br>(KD-721)<br>(K163276) | Comparison Result | | Cuff Pressure<br>Range | 0-300mmHg | 0-300mmHg | 0-300mmHg | Same | | Overpressure<br>Limit | 300mmHg | 300mmHg | 300mmHg | Same | | Algorithm | Amplitude | Amplitude | Amplitude | Same | {5}------------------------------------------------ {6}------------------------------------------------ {7}------------------------------------------------ There are no significant differences between the two products and they are identical in terms of intended use, materials, design, manufacturing methods. ## 8.0 Discussion of Non-Clinical Testing Non-clinical tests conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards: - IEC 60601-1:2005/(R)2012 And A1:2012, C1:2009/(R)2012 And . A2:2010/(R)2012 (Consolidated Text) Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance (IEC 60601-1:2005, MOD) - IEC 60601-1-2:2014,Medical Electrical Equipment -- Part 1-2: General . Requirements For Basic Safety And Essential Performance -- Collateral Standard: Electromagnetic Disturbances -- Requirements And Tests - IEC 60601-1-11 Edition 2.0 2015-01 Medical electrical equipment Part 1-11: ● General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment - IEC 80601-2-30:2009 & A1:2013,Medical Electrical Equipment - Part 2-30: Particular Requirements For The Basic Safety And Essential Performance Of Automated Noninvasive Sphygmomanometers None of the tests demonstrate that the new Blood Pressure Monitors raise new questions of safety and effectiveness as compared to the predicate. ## 9.0 Clinical Test Comparison and evaluation are carried out between the subject device and cleared devices, and it is showed that: The proposed devices KD-743V and KD-743B has the same algorithm and design principle with cleared device KD-7961 (K090769), so the clinical test report of KD-7961 can be used as a reference when considering the clinical effect of KD-743V and KD-743B. {8}------------------------------------------------ The proposed devices KD-752 has the same algorithm and design principle with cleared device KD-972 (K121470), and so the clinical test report of KD-972 can be used as a reference when considering the clinical effect of KD-752. Accuracy of the blood pressure monitors for the clinical test report was verified by meeting criteria 1 and criteria 2 of ISO 81060-2. ## 10. Comparison to the Predicate Device and Conclusion The conclusion drawn from the nonclinical tests and clinical test demonstrate that the subject device, Fully Automatic Electronic Blood Pressure Monitor (KD-743V, KD-743B, KD-752), is substantially equivalent to the Fully Automatic Electronic Blood Pressure Monitor KD-753 whose 510(k) number is K183535 and KD-721 whose 510(k) number is K163276. The subject devices are very similar with its predicate device in the intended use, design principle, materials, performance and applicable standards. Their appearance, the memory capacity, the average function is different, and MCU used is different. However, the tests in this submission demonstrates that these differences do not raise any new questions of safety and effectiveness and the subject device is as safe, as effective and performs as well as or better than the legally marketed predicate K183535 and K163276.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...