FULLY AUTOMATIC ELECTRONIC BLOOD PRESSURE MONITOR MODEL KD-513

K102353 · Andon Health Co, Ltd. · DXN · Sep 17, 2010 · Cardiovascular

Device Facts

Record IDK102353
Device NameFULLY AUTOMATIC ELECTRONIC BLOOD PRESSURE MONITOR MODEL KD-513
ApplicantAndon Health Co, Ltd.
Product CodeDXN · Cardiovascular
Decision DateSep 17, 2010
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1130
Device ClassClass 2

Intended Use

KD-5913 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm.

Device Story

KD-5913 is a non-invasive, fully automatic electronic blood pressure monitor for clinical or home use. Device utilizes an inflatable upper arm cuff (22-48cm) and silicon integrated pressure sensor technology to perform oscillometric measurements. System calculates systolic/diastolic blood pressure and pulse rate; detects irregular heartbeats; and provides an average of the last three measurements. Results are displayed on an integrated LCD. Device is operated by medical professionals or patients for self-monitoring. Output allows healthcare providers to track blood pressure trends and assess cardiovascular status. Benefits include convenient, non-invasive blood pressure monitoring for adult patients.

Clinical Evidence

Bench testing only. Device conforms to IEC 60601-1 (safety), EN 60601-1-2 (EMC), and AAMI/ANSI SP10:2002 (sphygmomanometers) standards.

Technological Characteristics

Oscillometric measurement principle using silicon integrated pressure sensor. Upper arm cuff (22-48cm). LCD display. Powered by internal energy source. Complies with ANSI/AAMI SP10, IEC 60601-1, and EN 60601-1-2 standards.

Indications for Use

Indicated for adult individuals to measure diastolic and systolic blood pressure and pulse rate via non-invasive upper arm cuff (22cm-48cm circumference). Suitable for use by medical professionals or at home.

Regulatory Classification

Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ### K102353 510(k) Summary This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of SMDA and 21 CFR SEP 1 7 2010 807.92. ## 1.0 submitter's information Name: Andon Health Co., Ltd. Address: No 3, Jinping Street Ya An Road, Nankai District, Tianjin, P.R. China Phone number: 86-22-6052 6161 Fax number: 86-22-6052 6162 Contact: Liu Yi Date of Application: 08/16/2010 ### 2.0 Device information | Trade name: | KD-5913 Fully Automatic Electronic Blood Pressure<br>Monitor | |----------------------|--------------------------------------------------------------| | Common name: | Noninvasive blood pressure measurement system | | Classification name: | Noninvasive blood pressure measurement system | ### 3.0 Classification Production code: DXN- Noninvasive blood pressure measurement system. Regulation number: 870.1130 Classification: II Panel: Cardiovascular ## 4.0 Predict device information Manufacturer: Andon Health Co., Ltd. Device: KD-5917 Fully Automatic Electronic Blood Pressure Monitor 510(k) number: K091737 ### 5.0 Device description KD-5913 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper am. The {1}------------------------------------------------ cuff circumference is limited to 22cm-48cm. It is designed and manufactured according to ANSI/AAMI SP10--manual, electronic or automated sphygmanometers. The operational principle is based on oscillometric and silicon integrate pressure sensor technology. It can calculate the systolic and diastolic blood pressure, and display the result on the LCD. If any irregular heartbeat is detected, it can also be shown on the LCD. More over, it also has the function of averaging the last three measurements. ### 6.0 Intended use KD-5913 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systelic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm. The intended use and the indication for use of KD-5913, as described in its labeling are the same as the predict device KD-5917. ### 7.0 Summary comparing technological characteristics with predicate device | Technological Characteristics | Comparison result | |-------------------------------|-------------------| | Design principle | Identical | | Appearance | Similar | | Patients contact Materials | Identical | | Performance | Similar | | Biocompatibility | Identical | | Mechanical safety | Identical | | Energy source | Identical | | Standards met | Identical | | Electrical safety | Identical | | EMC | Identical | | Function | Similar | {2}------------------------------------------------ ## 8.0 Performance summary KD-5913 Fully Automatic Electronic Blood Pressure Monitor conforms to the following standards: - IEC 60601-1, Medical Electrical Equipment Part 1: General Requirements for Safety, 1988; Amendment 1, 1991-11, Amendment 2, 1995. - EN 60601-1-2, Medical Electrical Equipment Part 1-2: General Requirements for Safety - Collateral standard: - Frandard: Electromagnetic Compatibility - Requirements and Tests, 2007 - AAMI SP10:2002, Manual, electronic or automated sphygmomanometers. - AAMI / ANSI SP10:2002/A1:2003 --, Amendment 1 to ANSI/AAMI SP10:2002 Manual, electronic, or automated sphygmomanoment - I To ANSIA AAMI / ANSI - CD10-800040 0000000 0000 - · AAMI / ANSI SP10:2002/A2:2006 --, Amendment 2 to ANSI/AAMI SP10:2002 Manual, electronic, or automated sphygmomanoment . 2 10 ANSIA # 9.0 Comparison to the predict device and the conclusion Our device KD-5913 Fully Automatic Electronic Blood Pressure Monitor is substantially equivalent to the Fully Automatic Electronic Blood Pressure Monitor is Monitor KD-5917 whose 510(k) number is K091737. The two devices are very similar in the intended use, the design principle, the material, the performance and the applicable standards. Only their appearance and some functions are different. Compared to KD-5917, the KD-5913 does not have the display of "Blood pressure classification", and the MCU is changed, either. However, the test in this submission provides demonstration that these small differences do not raise any new questions of safety and effectiveness. ર- 3 {3}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with outstretched wings, symbolizing protection and care. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", indicating the department's name and national affiliation. ### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Andon Health Co., Ltd. c/o Mr. Liu Yi President No. 3 Jin Ping Street, Ya An Road, Nankai District Tianjin China 300190 SEP 1 7 2010 Re: K102353 Trade/Device Name: KD-5913 Fully Automated Electronic Blood Pressure Monitor Regulatory Number: 21 CFR 870.1130 Regulation Name: Non-invasive Blood Pressure Measurement System Regulatory Class: II (two) Product Code: DXN Dated: August 16, 2010 Received: August 19, 2010 Dear Mr. Yi: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ ### Page 2 - Mr. Liu Yi Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## Statement of Indications for Use | 510(k) Number : | K102353 | |-----------------|-----------------------------------------------------------| | Device name: | KD-5913 Fully Automatic Electronic Blood Pressure Monitor | SEP 17 2010### Indications for use: KD-5913 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm. Prescription use Part 21 CFR 801 Subpart D) Over-The-Counter Use YES AND/OR (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-COUNTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Division of Page 1 of 1 (Division Sign-Off) Division of Cardiovascular Devices 510(k) Number KJO235
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%