Arm-Type Fully Automatic Digital Blood Pressure Monitor (BP-1305, BP-1307, BP-1326, BP-1318, BP-1319, BP-1211, BP-1312, BSP-11, BSP-12, BSP-13)

K150545 · Sejoy Electronics & Instruments Co., Ltd. · DXN · Apr 2, 2015 · Cardiovascular

Device Facts

Record IDK150545
Device NameArm-Type Fully Automatic Digital Blood Pressure Monitor (BP-1305, BP-1307, BP-1326, BP-1318, BP-1319, BP-1211, BP-1312, BSP-11, BSP-12, BSP-13)
ApplicantSejoy Electronics & Instruments Co., Ltd.
Product CodeDXN · Cardiovascular
Decision DateApr 2, 2015
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1130
Device ClassClass 2

Indications for Use

The Fully Automatic Blood Pressure Monitors BP series are intended for used by adults with 12 years and older to measure the systolic and diastolic blood pressure and pulse rate.

Device Story

Arm-type digital blood pressure monitor; uses inflatable cuff wrapped around upper arm. Device automatically inflates cuff via internal pump; measures blood pressure using oscillometric method and silicon integrated pressure sensor technology. Mechanical valve controls deflation rate. Device displays systolic, diastolic, and pulse rate on LCD. Includes WHO blood pressure classification indicator; irregular heartbeat detection; and memory storage for measurement results. Used in home or clinical settings by patients or healthcare providers. Provides objective blood pressure data to assist in monitoring cardiovascular health and hypertension management.

Clinical Evidence

Clinical validation performed according to ISO 81060-2:2013 (non-invasive sphygmomanometers - part 2: clinical validation of automated measurement type). Results confirmed compliance with accuracy requirements.

Technological Characteristics

Oscillometric measurement; silicon integrated pressure sensor; inflatable cuff. Powered by 4 AA batteries or AC adapter. Safety features include 300 mmHg maximum pressure limit. Standards: AAMI/ANSI/IEC 80601-2-30:2009, AAMI/ANSI ES60601-1:2005, IEC/EN 60601-1-2:2007. Biocompatibility per ISO 10993-5 and ISO 10993-10.

Indications for Use

Indicated for adults 12 years and older to measure systolic/diastolic blood pressure and pulse rate.

Regulatory Classification

Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an image of three faces in profile, stacked on top of each other. Below the faces are three wavy lines. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 April 2, 2015 Sejoy Electronics & Instruments Co., Ltd. Ren Yunhua General Manager Bldg 2, No. 202, Zhenzhong Rd West Lake Economy & Technology Zone Hangzhou, Zhejiang 310030 CN Re: K150545 Trade/Device Name: Arm-type Fully Automatic Digital Blood Pressure Monitor (BP-1305, BP-1307, BP-1326, BP-1318, BP-1319, BP-1211, BP-1312, BSP-11, BSP-12, BSP-13) Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: December 15, 2014 Received: March 3, 2015 Dear Ren Yunhua: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely vours incerely yours, for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ K150545 世佳电子有限公司 Sejoy Electronics & Instruments Co., Ltd. # Indications for Use 510(k) Number (if known): Blood Pressure Monitors for BP-1211, BP-1312, BP-1305, Device Name: BP-1307, BP-1318, BP-1319, BP-1326, BSP-11, BSP-12, BSP-13 Indications For Use: Measure blood pressure (systolic and diastolic) and pulse rate. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use √ (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for SEJOY. The logo is blue and has a small yellow triangle above the 'J'. The word 'SEJOY' is written in a bold, sans-serif font. Sejoy Electronics & Instruments Co., Ltd. 世佳电子有限公司 ## 510(k) Summary The assigned 510(k) number is:_ #### 1. Date Prepared: 2014.12.15 #### 2. Submitter's Identification: Name: Sejoy Electronics & Instruments Co., Ltd. Add .: Building 2, No.202, Zhenzhong Road, West Lake Economy & Technology Zone, Hangzhou, China 310030 Contact Person: Yunhua Ren Phone: +86-571-81957767 Fax: +86-571-81957750 Email: renyh@sejoy.com ### 3. Name of the Device: Trade Name: Fully Automatic Digital Blood Pressure Monitor Including the following models: - BP-1211, BP-1312, BP-1305, BP-1307, BP-1318, BP-1326, BSP-11, BSP-12, BSP-13 for Arm-type Fully Automatic Digital Blood Pressure Monitor Common Name: Blood Pressure Monitor Classification name: Non-invasive blood pressure measurement System 21 CFR 870-1130, Class II, 74-DXN. #### 4. Classification Information: Regulation Number: 870.1130 Product Code: DXN Device Class: II Panel: 74 Cardiovascular #### 5. Predicate Device Information: The Arm-type Fully Automatic Digital Blood Pressure Monitors are substantially equivalent to the following device: Blood Pressure Monitor (Model BP-1307); FDA 510(k) number: K120554 ; manufactured by SEJOY ELECTRONICS & INSTRUMENTS CO., LTD. {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the word "SEJOY" in blue, with a yellow triangle above the "J". The letters are in a bold, sans-serif font. The triangle is pointing downwards. The background is white. ### 6. Device Description: The arm-type uses an inflatable cuff which is wrapped around the patient's upper arm. The cuff is inflated automatically by an internal pump in the device. The systolic and diastolic blood pressures are determined by oscillometric method and silicon integrate pressure sensor technology. The deflation rate is controlled by a preset mechanical valve at a constant rate. The pressure of the cuff is completely released automatically at the end of the measurement. At the same time, the measurements are displayed on the LCD display for three minute. There is a maximum pressure safety setting at 300 mmHg. The device will not inflate the cuff higher than 300 mmHg. For BP-1305, BP-1307,BP-1211, BP-1312, BP-1318, BP-1319, BP-1326, BSP-11, BSP-12, BSP-13, the blood pressure results are compared with WHO (World Health Organization) Blood Pressure classification, which are severe Hypertension, Moderate Hypertension, Mild Hypertension, High-normal, Normal, and Optimal. The corresponding LCD segment will be turned on along with the systolic, diastolic, and pulse rate information. BP-1305, BP-1307,BP-1312, BP-1318, BP-1319, BP-1326, BSP-11, BSP-12, BSP-13will display an irregular heartbeat symbol " " if an irregular heartbeat was detected during the measurement process. BP-1305,BP-1318, BP-1319, BP-1326, BSP-11, BSP-12, BSP-13 can display average results in three ways: BP-1307 can display the average of all measurements, the average of all AM (5:00 AM-8:59AM) measurements, and the average of all PM (18:00 PM-19:59 PM) measurements. | Features | A | B | C | D | E | F | G | H | I (mm) | J | K(mm) | L | M | N | |---------------------------------------|---|---|-------------------------------|---|---|---|---|---|--------------|---------|------------|---|---|---| | Models | | | | | | | | | | | | | | | | BP-1307<br>(The Predicate<br>Devices) | ● | ● | 120 Memories<br>in two groups | ● | ○ | ● | ● | ● | 166×114×72 | 22-36cm | 84.1×55.1 | ● | ○ | ○ | | BP-1211 | ● | ● | 120 Memories<br>in one groups | ● | ○ | ● | ○ | ○ | 131×102×65 | 22-36cm | 45×30 | ○ | ○ | ○ | | BP-1312 | ● | ● | 120 Memories<br>in two groups | ● | ○ | ● | ● | ● | 131 ×102 ×65 | 22-36cm | 62.4×46.1 | ○ | ○ | ○ | | BP-1318 | ● | ● | 120 Memories<br>in two groups | ● | ○ | ● | ○ | ● | 139× 88× 43 | 22-36cm | 66 ×43 | ○ | ○ | ○ | | BP-1326 | ● | ● | 120 Memories<br>in two groups | ● | ○ | ● | ○ | ● | 139 ×88 ×44 | 22-36cm | 65.9×42.8 | ○ | ○ | ○ | | BP-1305 | ● | ● | 120 Memories<br>in two groups | ● | ○ | ● | ● | ● | 166 ×114× 72 | 22-36cm | 84.1×55.1 | ○ | ○ | ○ | | BP-1307 | ● | ● | 120 Memories<br>in two groups | ● | ● | ● | ○ | ● | 166 ×114× 72 | 22-36cm | 102.1×68.9 | ○ | ○ | ○ | | BP-1319 | ● | ● | 120 Memories<br>in two groups | ● | ○ | ● | ● | ● | 166 ×114× 72 | 22-36cm | 101.6×67.6 | ○ | ○ | ○ | | BSP-11 | ● | ● | 120 Memories<br>in one groups | ● | ○ | ● | ○ | ● | 140 ×98× 48 | 22-36cm | 62.3× 46 | ○ | ○ | ○ | | BSP-12 | ● | ● | 120 Memories<br>in one groups | ○ | ○ | ● | ● | ● | 148× 100× 56 | 22-36cm | 84.1×55.1 | ○ | ○ | ○ | The detail comparisons among devices are listed below: {5}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the logo for SEJOY. The logo is in blue, with a yellow triangle above the "J". The logo is simple and modern. 世佳电子有限公司 | Standard County and Children College of Children | | | Sejoy Electronics & Instruments Co., Ltd. | | | | | | | | | | | |--------------------------------------------------|--|-----|-------------------------------------------|--|--|--|--|----------------------------------|---------|------------|-----------|--|--| | | | | in two groups | | | | | | | | | | | | BSP-13 | | . . | 120 Memories | | | | | ● ○ ● ● ● 164× 111× 60 | 22-36cm | 102.1×68.9 | 0 0 0 | | | | | | | in two groups | | | | | | | | | | | A = Powered by 4 AA size batteries B= Powered by AC adaptor C = Memory Size D = Time & Date E = Results Average in Three way F = WHO (World Health Organization) Classification Indicator G = Last 3 Results Average H = Irregular Heartbeat Detection I = Outside Demission (L x W x H in mm) J = Cuff Size K = LCD Size (Viewing Area in mm) L = LCD Backlight M= Voice N= PC connecter • = Yes o = No O= Optional function depending on clients' needs The devices are all designed and manufactured according to AAMI / ANSI / IEC 80601-2-30:2009, medical electrical equipment - part 2-30: particular requirements for the basic safety and essential performance of automated noninvasive sphygmomanometers. ### 7. Intended Use: The Fully Automatic Blood Pressure Monitors BP series are intended for used by adults with 12 years and older to measure the systolic and diastolic blood pressure and pulse rate. The intended user and the indication for use of the Fully Automatic Blood Pressure Monitors BP series and BSP series as described in the labeling are nearly the same as their Predicate devices, Blood Pressure Monitor (Model: BP-1307). #### Summary comparing technological characteristics with predicate device: 8. | Technological Characteristics | Comparison result | |-------------------------------|-------------------| | Design principle | Identical | | Appearance | Similar | | Patients contact Materials | Identical | | Performance | Identical | | Biocompatibility | Identical | | Mechanical Safety | Identical | {6}------------------------------------------------ Image /page/6/Picture/1 description: The image shows the logo for SEJOY. The logo is in blue and has a yellow triangle above the letter 'J'. There is a period after the letter 'Y'. 世侯甲子有限公 | | Sejoy Electronics & Instruments Co., Ltd | |-------------------|------------------------------------------| | Energy Source | Identical | | Standards Met | Identical | | Electrical Safety | Identical | | EMC | Identical | | Function | Similar | The difference of technological characteristic between the predicate device and the submit arm-type fully automatic blood pressure monitors is the appearance and the function. ## 9. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows: Compliance to applicable voluntary standards including IEC 80601-2-30:2009, medical electrical equipment - part 2-30 as well as AAMI / ANSI ES60601-1:2005/(R)2012 and C1:2009/(R)2012 and, a2:2010/(t)2012 , IEC/EN 60601-1-2:2007, ISO 10993-5: 2009 Biological evaluation of medical devices -Part 5:Tests for in vitro cytotoxicity and ISO 10993-10:2010 Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization. Guidance documents include the "FDA Non-invasive Blood Pressure (NIBP) Monitor Guidance" and "FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and "FDA Bluebook Memorandum G95-1Use of International Standard ISO 10993." ### Non-clinical Tests: Electromagnetic Compatibility Test according to IEC/EN 60601-1-2:2007 General Safety Provisions Test according to AAMI / ANSI ES60601-1:2005(R)2012 and C1:2009/(R)2012 and, a2:2010/(r)2012 Performance Test according to IEC 80601-2-30:2009, medical electrical equipment - part 2-30: particular requirements for the basic safety and essential performance of automated noninvasive sphygmomanometers The test result all meet or exceed the requirement of the standards. Biocompatibility Test according to FDA Bluebook Memorandum G95-1Use of International Standard ISO 10993, ISO 10993-5: 2009 Biological evaluation of medical devices-Part 5: Tests for in vitro cytotoxicity and ISO 10993-10:2010 Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization. ### 10. Discussion of Clinical Tests Performed: Clinical tests were performed and complied the accuracy requirements of ISO 81060-2 Second edition 2013-05-01, non-invasive sphygmomanometers - part 2: clinical validation of automated measurement type. ### 11. Conclusions: {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the word "SEJOY" in blue font. There is a yellow triangle above the letter "J". The word is in a sans-serif font and is slightly bolded. There is a small dot to the right of the word. ## Sejoy Electronics & Instruments Co., Ltd. Our Arm-type Fully Automatic Blood Pressure Monitors BP-1211, BP-1312, BP-1305, BP-1307,BP-1318,BP-1319,BP-1326,BSP-11,BSP-12,BSP-13 have the same intended use and similar technological characteristics as the Blood Pressure Monitor (Model BP-1307); FDA 510(k) number: K120554;manufactured by SEJOY ELECTRONICS & INSTRUMENTS CO., LTD. Moreover, verification and validation tests contained in this submission demonstrate that the difference in the submitted models could maintain the same safety and effectiveness as that of cleared devices. In the other words, those engineering difference do not: (1) affect the intended use or (2) alter the fundamental scientific technology of the device, and the new models as mentioned on this submission are considered substantial equivalent to the predicate devices.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...