MODIFICATION TO:FULLY AUTOMATIC ELECTRONIC BLOOD PRESSURE MONITOR, MODEL KD-556
K091500 · Andon Health Co, Ltd. · DXN · Jul 31, 2009 · Cardiovascular
Device Facts
| Record ID | K091500 |
| Device Name | MODIFICATION TO:FULLY AUTOMATIC ELECTRONIC BLOOD PRESSURE MONITOR, MODEL KD-556 |
| Applicant | Andon Health Co, Ltd. |
| Product Code | DXN · Cardiovascular |
| Decision Date | Jul 31, 2009 |
| Decision | SESE |
| Submission Type | Special |
| Regulation | 21 CFR 870.1130 |
| Device Class | Class 2 |
Intended Use
KD-556 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm.
Device Story
Non-invasive blood pressure monitor; uses inflatable upper arm cuff (22-48cm). Operates via oscillometric principle with silicon integrated pressure sensor. Device measures diastolic/systolic blood pressure and pulse rate. Features include LCD display, irregular heartbeat detection, blood pressure classification indicator, and memory for 2x60 measurements; calculates average of last three readings. Used by medical professionals or patients at home. Output displayed on LCD for clinical decision-making regarding hypertension management.
Clinical Evidence
Bench testing only. Device conforms to IEC 60601-1, IEC 60601-1-2, and AAMI/ANSI SP10:2002 (including amendments A1:2003 and A2:2006) for manual, electronic, or automated sphygmomanometers.
Technological Characteristics
Oscillometric measurement principle; silicon integrated pressure sensor; LCD interface; 2x60 memory capacity; irregular heartbeat detection; averaging function. Complies with ANSI/AAMI SP10, IEC 60601-1, and IEC 60601-1-2 standards.
Indications for Use
Indicated for adult individuals to measure diastolic and systolic blood pressure and pulse rate via non-invasive upper arm cuff (22cm-48cm circumference). Suitable for use by medical professionals or at home.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
- KD-556 Fully Automatic Electronic Blood Pressure-Monitor (K090963)
Related Devices
- K102609 — KD-5966 SERIES FULLY AUTOMATIC ELECTRONIC BLOOD PRESSURE MONITOR · Andon Medical Co., Ltd. · Dec 9, 2010
- K151460 — Fully Automatic Electronic Blood Pressure Monitor · Andon Health Co, Ltd. · Sep 8, 2015
- K090771 — FULLY AUTOMATIC ELECTRONC BLOOD PRESSURE MONITOR, MODEL, KD-5962 · Andon Health Co, Ltd. · Jun 4, 2009
- K093387 — FULLY AUTOMATIC ELECTRONIC BLOOD PRESSURE MONITOR MODEL KD-5961, KD-5962, KD-5902 · Andon Health Co, Ltd. · Nov 27, 2009
- K234041 — Fully Automatic Electronic Blood Pressure Monitor (KD-5811VA); Fully Automatic Electronic Blood Pressure Monitor (KD-5811BT); Fully Automatic Electronic Blood Pressure Monitor (KD-5810VA); Fully Automatic Electronic Blood Pressure Monitor (KD-5810BT); Fully Automatic Electronic Blood Pressure Monitor (KD-5920VA); Fully Automatic Electronic Blood Pressure Monitor (KD-5920BT); Fully Automatic Electronic Blood Pressure Monitor (KD-5923TS) · Andon Health Co, Ltd. · Apr 22, 2024
Submission Summary (Full Text)
{0}------------------------------------------------
**JUL 31 2009**
# 510(k) Summary
K041500
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of SMDA and 21 CFR 807.92.
### 1.0 submitter's information
Name: Andon Health Co., Ltd. Address: No 31, Changjiang Road, Nankai District, Tianjin, P.R. China Phone number: 86-22-6052 6161 Fax number: 86-22-6052 6162 Contact: Liu Yi Date of Application: 05/18/2009
### 2.0 Device information
Trade name: Fully Automatic Electronic Blood Pressure Monitor Common name: Noninvasive blood pressure measurement system Classification name: Noninvasive blood pressure measurement system
#### 3.0 Classification
Production code: DXN- Noninvasive blood pressure measurement system. Regulation number: 870.1130 Classification: 11 Panel: Cardiovascular
### 4.0 Predict device information
Manufacturer: · Andon Health Co., Ltd. Device: KD-556 Fully Automatic Electronic Blood Pressure-Monitor 510(k) number: K090963
#### 5.0 Device description
KD-556 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm.
{1}------------------------------------------------
It is designed and manufactured according to ANSI/AAMI SP10--manual, electronic or automated sphygmanometers.
The operational principle is based on oscillometric and silicon integrate pressure sensor technology, the result will be shown on a LCD with an electronic interface module, the result can also be classified and displayed by the function of blood pressure classification indicator, the memory capability is 2×60 times. If any irregular-heartbeat is detected, it can be shown on the LCD. It also has the function of averaging the last three measurements.
#### 6.0 Intended use
KD-556 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm.
The intended use and the indication for use of KD-556, as described in its labeling are the same as the predict device KD-556(K090963).
#### 7.0 Summary comparing technological characteristics with predicate device
| Technological Characteristics | Comparison result |
|-------------------------------|-------------------|
| Design principle | Identical |
| Appearance | Identical |
| Patients contact Materials | Identical |
| Performance | Identical |
| Biocompatibility | Identical |
| Mechanical safety | Identical |
| Energy source | Identical |
| Standards met | Identical |
| Electrical safety | Identical |
| EMC | Identical |
| Function | Identical |
{2}------------------------------------------------
2/3
#### 8.0 Performance summary
KD-556 Fully Automatic Electronic Blood Pressure Monitor conforms to the following standards:
- IEC 60601-1, Medical Electrical Equipment Part 1: General Requirements for Safety, 1988; Amendment 1, 1991-11, Amendment 2, 1995.
- · IEC 60601-1-2, Medical Electrical Equipment Part 1-2: General Requirements for Safety - Collateral standard: Electromagnetic Compatibility - Requirements and Tests (Edition 2:2001 with Amendment 1:2004, Edition 2.1 (Edition 2:2001 consolidated with Amendment 1:2004)).
- · AAMI SP10:2002, Manual, electronic or automated sphygmomanometers.
- · AAMI / ANSI SP10:2002/A1:2003 --, Amendment 1 to ANSI/AAMI SP10:2002 Manual, electronic, or automated sphygmomanometers.
- · AAMI / ANSI SP10:2002/A2:2006 --, Amendment 2 to ANSI/AAMI SP10:2002 Manual, electronic, or automated sphygmomanometers.
## 9.0 Comparison to the predict device and the conclusion
Our device KD-556 Fully Automatic Electronic Blood Pressure Monitor is substantially equivalent to the Fully Automatic Electronic Blood Pressure Monitor KD-556 whose 510(k) number is K090963.
The two devices are identical in the intended use, the design principle, the material, the appearance, the functions, the performance, the energy source and the applicable standards. KD-556 only adds a new cuff (cuff circumstance: 22cm-30cm) when compared with the predicted device( K090963).
However, the test in this submission provides demonstration that these small differences do not raise any new questions of safety and effectiveness.
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The logo is black and white.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 3 1 2009
Andon Health Co., Ltd. c/o Ms. M. Elizabeth Bierman, Partner Morgan, Lewis & Bockius LLP 1111 Pennsylvania Ave., NW Washington, DC 20004-2541
K091500 Re:
> Trade/Device Name: KD-556 Fully Automatic Electronic Blood Pressure Monitor. Regulatory Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II (Two) Product Code: DXN Dated: Undated Received: July 2, 2009
Dear Ms. Bierman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration,
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2 - Ms. M. Elizabeth Bierman
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
M. G. Willemann
عرب Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Statement of Indications for Use
Kog 1500 510(k) Number :
Device name:
KD-556 Fully Automatic Electronic Blood Pressure Monitor
### Indications for use:
KD-556 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systelic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm.
Prescription use Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR YES (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-COUNTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
M. L. Willkenne
ardiovascular Devices
510(k) Number K091500
Page 1 of ﺴﺮ