K141460 · Finapres Medical Systems B.V. · DXN · Mar 20, 2015 · Cardiovascular
Device Facts
Record ID
K141460
Device Name
FINAPRES NOVA NONINVASIVE HEMODYNAMIC MONITOR
Applicant
Finapres Medical Systems B.V.
Product Code
DXN · Cardiovascular
Decision Date
Mar 20, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1130
Device Class
Class 2
Indications for Use
The Finapres NOVA is intended to be used with patients with a need for a noninvasive blood pressure and hemodynamic monitor. The noninvasive blood pressure waveform is measured on the subject's finger. The Finapres NOVA provides a noninvasive characterization of the arterial circulation and its beat-to-beat variability in pressure and flow and in various hemodynamic parameters derived from these pressure and flow signals. The Finapres NOVA has the option to include four additional modules to extend its functionality with ECG and SpO₂ measurements, blood pressure calibration and data transfer to and from the device. The Finapres NOVA is used in a professional medical environment such as hospitals, clinics and research institutions. The Finapres NOVA is a standalone device intended for desktop use. Measurements are to be performed under uninterrupted patient surveillance of the operator. The user of the Finapres NOVA should be a qualified operator. The operator should have knowledge of the system and data interpretation, obtained via medical education, system manuals and/or specific courses. The device does not report any diagnosis but provides numerical values. It is the physician's responsibility to make proper judgments based on these numbers.
Device Story
Finapres NOVA is a desktop noninvasive hemodynamic monitor for professional clinical use. It measures finger arterial blood pressure via an infrared plethysmograph (light source/detector) and finger cuff using the Peñáz arterial volume-clamp method and Wesseling Physiocal criteria. The device computes real-time/beat-to-beat blood pressure and hemodynamic parameters (e.g., cardiac output via Modelflow, total peripheral resistance). Optional modules include ECG, SpO2, and upper-arm blood pressure calibration. Data is displayed for clinician interpretation; the device provides numerical values rather than diagnoses. It benefits patients by enabling continuous hemodynamic monitoring without invasive arterial lines. The system supports 24-hour continuous measurements and includes an optional Height Correction Unit (HCU) for patient movement.
Clinical Evidence
No clinical data was required or presented. Substantial equivalence was established through bench testing and adherence to recognized medical device safety and performance standards, including electrical safety, electromagnetic compatibility, and software life cycle processes.
Technological Characteristics
Desktop standalone device; finger cuff with infrared plethysmograph; arterial volume-clamp sensing principle. Optional modules: ECG, SpO2, upper-arm BP calibration. Connectivity: analog I/O. Software: embedded, NOVAScope/Modelflow algorithm. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-2-27, IEC 80601-2-30, ISO 80601-2-61, ISO 81060-2, IEC 62304, ISO 14971, ISO 10993.
Indications for Use
Indicated for patients >18 years requiring noninvasive blood pressure and hemodynamic monitoring in professional medical environments. Optional modules monitor SpO2, pulse rate, and ECG parameters.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol consisting of three stylized human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 20, 2015
Finapres Medical Systems c/o Ms. Iris van Uitert Quality Manager Hogehilweg 8 NL - 1101 CC Amsterdam The Netherlands
Re: K141460 Trade/Device Name: Finapres Nova Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive blood pressure measurement system Regulatory Class: Class II Product Code: DXN, DSB, DOA, DRT Dated: February 2, 2015 Received: February 5, 2015
Dear Ms. Van Uitert:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and have determined the device is substantially equivalent to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might
{1}------------------------------------------------
Page 2 – Ms. Iris van Uitert
have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. An FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market, but it does not mean that FDA approves your device. Therefore, you may not promote or in any way represent your device or its labeling as being approved by FDA.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
# Kenneth J. Cavanaugh -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K141460
Device Name Finapres NOVA
### Indications for Use (Describe)
The Finapres NOVA is intended to be used with patients who have a need for a noninvasive blood pressure and hemodynamic monitor. The noninvasive blood pressure waveform is measured on the subject's finger. The Finapres NOVA provides a noninvasive characterization of the arterial circulation and its beat variability in pressure and flow and in various hemodynamic parameters derived from these pressure and flow signals. Cardiac output derived from the flow signal requires a calibration with thermal dilution.
The Finapres NOVA has the option to include additional modules to extend its functionality with ECG and SpO2 measurements and blood pressure calibration.
When the SpO2 module is present, the Finapres NOVA can additionally monitor the functional oxygen saturation of arterial hemoglobin (SpO2) and the pulse rate.
When the ECG module is present, the Finapes NOVA can additionally monitor the ECG parameters of a patient and their pulse rate. Alarms concerning the pulse rate will be available from the monitor.
When the blood pressure calibration module is present, the Finapres NOVA can additionally provide an upper arm non-invasive blood pressure measurement to determine the blood pressure value for calibration.
The Finapres NOVA is intended to be used for subjects above 18 years of age.
The Finapres NOVA is intended for use in a professional medical environment.
| Type of Use (Select one or both, as applicable) |
|---------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size:12pt"> <span style="font-family:Symbol"> ☑ </span> </span> Prescription Use (Part 21 CFR 801 Subpart D) |
| <span style="font-size:12pt"> <span style="font-family:Symbol"> ☐ </span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
# Finapres NOVA - K141460
This 510(k) summary is being submitted in accordance with the requirements of SMDA and 21CFR § 807.92.
# 1. Submitter's information
| Submitter: | Finapres Medical Systems B.V.<br>Hogehilweg 8<br>NL-1101 CC Amsterdam<br>The Netherlands<br>Phone number: +31 20 609 0974<br>Fax Number: +31 20 609 0677<br>Operator Number: 9051428<br>Registration number: 3003803088 |
|-----------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact person: | Dr. Ir. Iris van Uitert<br>Quality Manager<br>Phone number: +31 20 609 0974<br>E-mail:iris.van.uitert@finapres.com |
Date of preparation: March 17, 2015
### 2. Device information
| Trade name: | Finapres NOVA Noninvasive Hemodynamic Monitor |
|-----------------------|-----------------------------------------------|
| Common name(s) | Noninvasive Blood Pressure Monitor |
| | Hemodynamic Monitor |
| | Electrocardiograph |
| | Oximeter |
| Classification name | See Table 1 |
| Device classification | Class II |
### Table 1. Classification name
| Classification name | 21 CFR Section | Product Code |
|-------------------------------------------------------|----------------|--------------|
| Noninvasive Blood Pressure Monitor Measurement System | 870.1130 | DXN |
| Plethysmograph, Impedance | 870.2770 | DSB |
| Oximeter | 870.2700 | DQA |
| Cardiac monitor | 870.2300 | DRT |
### 3. Predicate Devices
The Finapres NOVA is substantially equivalent in design (methodology) and indications for use to the devices shown in Table 2 that have previously been cleared:
{5}------------------------------------------------
Table 2. Predicate devices for the Finapres NOVA
| Device name | Manufacturer | 510(k) |
|--------------------------------------------------|--------------------------------------------------------------|---------|
| Finometer® Model-1 | FMS, Finapres Medical Systems,<br>Amsterdam, The Netherlands | K023723 |
| Ohmeda 2350 Finapres ® Blood<br>pressure monitor | Ohmeda Medical, Englewood | K880572 |
| Portapres | FMS, Finapres Medical Systems,<br>Amsterdam, The Netherlands | K023338 |
| NEXFIN_HD™ | BMEYE B.V., Amsterdam, The Netherlands | K072049 |
| Nellcor Oximax N-600X Pulse<br>Oximeter | Nellcor Puriton Bennet, Inc,<br>Pleasanton | K060576 |
| Zoll M Series with real CPR help | Zoll Medical Corporation, 269 Mill<br>Road, Chelmsford | K062537 |
| MOVES | THORNHILL RESEARCH INC, Toronto,<br>Ontario | K093261 |
### 4. Description
The Finapres NOVA is an instrument to noninvasively monitor blood pressure and hemodynamic parameters. The Finapres NOVA provides a characterization of the arterial circulation and its beat variability in pressure and flow and in various hemodynamic parameters derived from these pressure and flow signals. The Finapres NOVA has the option to include four additional modules to extend its functionality with ECG and SpO2 measurements, blood pressure calibration and data transfer to and from the device.
The embedded software in the device provides computation of real-time and beat-tobeat blood pressure as well as hemodynamic parameters from the non-invasely measured blood pressure waveform. Hemodynamic parameters include cardiac output based on the modelflow method and total peripheral resistance.
The measurement of blood pressure in a finger is based on the arterial volume-clamp method of the Czech physiologist J. Peñáz, and the Physiocal - physiological calibration criteria for the proper unloading of the finger arteries of K.H. Wesseling. With this method, finger arterial pressure is measured using a finger cuff and an inflatable bladder in combination with an infrared plethysmograph, which consists of an infrared light source and detector.
The SpO2, upper arm calibration and ECG modules used in the Finapres NOVA are commercially available OEM modules that are used in FDA approved systems. The finger blood pressure measurement module used in the system is similar to other Finapres Medical Systems B.V. devices available on the market. The analog input/output module has been developed by Finapres during the Finapres NOVA development.
# 5. Indications for use
The Finapres NOVA is intended to be used with patients who have a need for a noninvasive blood pressure and hemodynamic monitor. The noninvasive blood pressure waveform is measured on the subject's finger. The Finapres NOVA provides a noninvasive characterization of the arterial circulation and its beat-to-beat variability in
{6}------------------------------------------------
pressure and flow and in various hemodynamic parameters derived from these pressure and flow signals. Cardiac output derived from the flow signal requires a calibration with thermal dilution.
The Finapres NOVA has the option to include additional modules to extend its functionality with ECG and SpO2 measurements and blood pressure calibration.
When the SpO2 module is present, the Finapres NOVA can additionally monitor the functional oxygen saturation of arterial hemoglobin (SpO2) and the pulse rate.
When the ECG module is present, the Finapres NOVA can additionally monitor the ECG parameters of a patient and their pulse rate. Alarms concerning the pulse rate will be available from the monitor.
When the blood pressure calibration module is present, the Finapres NOVA can additionally provide an upper arm non-invasive blood pressure measurement to determine the blood pressure value for calibration.
The Finapres NOVA is intended to be used for subjects above 18 years of age.
The Finapres NOVA is intended for use in a professional medical environment.
# 6. Intended Use
The Finapres NOVA is intended to be used with patients with a need for a noninvasive blood pressure and hemodynamic monitor. The noninvasive blood pressure waveform is measured on the subject's finger. The Finapres NOVA provides a noninvasive characterization of the arterial circulation and its beat-to-beat variability in pressure and flow and in various hemodynamic parameters derived from these pressure and flow signals.
The Finapres NOVA has the option to include four additional modules to extend its functionality with ECG and SpO₂ measurements, blood pressure calibration and data transfer to and from the device.
The Finapres NOVA is used in a professional medical environment such as hospitals, clinics and research institutions. The Finapres NOVA is a standalone device intended for desktop use. Measurements are to be performed under uninterrupted patient surveillance of the operator.
The user of the Finapres NOVA should be a qualified operator. The operator should have knowledge of the system and data interpretation, obtained via medical education, system manuals and/or specific courses. The device does not report any diagnosis but provides numerical values. It is the physician's responsibility to make proper judgments based on these numbers.
# 7. Summary of Technical Comparison with predicate devices
The Finapres NOVA is an updated model of the Ohmeda 2350, the Portapres and the Finometer® Model-1 device. As with predicate devices, the Finapres NOVA is indicated for use as a noninvasive blood pressure monitor. The Finapres NOVA is used to provide real-time and on-line monitoring and trending of hemodynamic parameters just as with the Finometer® Model-1 devices.
The Ohmeda 2350 device and the Finometer® Model-1 device have the same indications for use, the same design, and the same methodology as the Finapres NOVA Noninvasive
{7}------------------------------------------------
Hemodynamic Blood Pressure Monitor. The predicate Ohmeda 2350, the Finometer® Model-1 and the Nexfin HD provide the same noninvasive measurement and display of finger arterial blood pressure and pulse rate monitoring as the Finapres NOVA.
The Finapres NOVA consists of the same device architecture as the Finometer® Model-1 and the Nexfin HD.
The Finapres NOVA uses functionally the same embedded software for controlling the measurements as that used in the Finometer® Model-1 and Portapres device. The Finapres NOVA uses the proprietary NOVAScope/Modelflow software program for the measurement and trending of hemodynamic parameters, which is an updated version of the software that was used for data analysis in the Finometer and for the Portapres.
The Finapres NOVA can be equipped, equivalent to the Finometer® Model-1 and Nexfin_HD, with a blood pressure calibration method and does, therefore, allow for the measurement of accurate absolute blood pressure data complying with the AAMI SP10 :2002 and its successor IEC 81060-2:2009. The Finapres NOVA has, similar to the Nexfin_HD and the Finometer® Model-1, an optional Height correction unit (HCU) that allows the patient free hand movement during measurements with the device. The Finapres NOVA has, similar to the Finometer® Model-1, an optional ECG module that provides an analog ECG which is sampled and stored simultaneously with the blood pressure waveform. The Finapres NOVA has, same as the Ohmeda 2350, an optional SpO2 module that provides an SpO2 signal which is sampled and stored simultaneously with the blood pressure waveform. Similar to the Portapres, the Finapres NOVA enables 24 continuous measurements.
The three optional modules are available in the following predicate devices: MOVES, K093261; Nellcor Oximax N-600X Pulse Oximeter, K060576; Zoll M Series with real CPR help, K062537.
# 8. Non-clinical performance data for substantial equivalence determination
Testing of the Finapres NOVA was performed according to medical device safety standards shown in Table 3.
| Standard number | Standard name |
|------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ANSI/AAMI ES60601-1:2005 /<br>A2:2010 | Medical electrical equipment - Part 1: General requirements for<br>basic safety and essential performance |
| IEC 60601-1:1988 +A1:1991 +<br>A2:1995 (2nd edition) | Medical electrical equipment - Part 1: General requirements for<br>basic safety and essential performance |
| IEC 60601-1-2:2007 | Medical electrical equipment - Part 1-2: General requirements<br>for safety -Electromagnetic compatibility - Requirements and<br>tests |
| IEC 60601-1-4 1996 + Am.1<br>1999 | Medical electrical equipment - Part 1-4 - Collateral Standard:<br>Programmable electrical medical systems |
| IEC 60601-1-8: 2006 | Medical electrical equipment -- Part 1-8: General requirements<br>for basic safety and essential performance -- Collateral<br>standard: General requirements, tests and quidance for alarm<br>systems in medical electrical equipment and medical electrical<br>systems |
| IEC 60601-1-8: 2003 + A1: | Medical electrical equipment -Part 1-8: General requirements |
Table 3.Medical device standards that the Finapres NOVA is tested against
{8}------------------------------------------------
| 2006 | for safety - Collateral standard: General requirements, tests<br>and guidance for alarm systems in medical electrical equipment<br>and medical electrical systems |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-2-27:2005 | Medical electrical equipment - Part 2-27: Particular<br>requirements for the safety, including essential performance, of<br>electrocardiographic monitoring equipment |
| IEC 80601-2-30:2009 | Medical electrical equipment - Part 2-30: Particular<br>requirements for the basic safety and essential performance of<br>automated non-invasive sphygmomanometers |
| ISO 9919:2005 | Medical electrical equipment – Particular requirements for the<br>basic safety and essential performance of pulse oximeter<br>equipment for medical use |
| ISO 80601-2-61:2011 | Medical electrical equipment - Part 2-61: Particular<br>requirements for basic safety and essential performance of<br>pulse oximeter equipment. |
| ISO 81060-2:2009 | Non-invasive sphygmomanometers - Part 2: Clinical validation<br>of automated measurement type |
| ISO 14971:2007 | Medical devices – Risk management - Application of risk-<br>management to medical devices |
| IEC 62304:2006 | Medical device software - software life cycle processes |
| IEC 62366:2009 | Medical devices - Application of usability engineering to medical<br>devices. |
| ISO 10993-1:2009 | Biological evaluation of medical devices - Part 1: Evaluation<br>and testing |
| AAMI / ANSI / ISO 10993-5:2009 | Biological evaluation of medical devices -- Part 5: Tests for In<br>Vitro cytotoxicity |
| ISO 10993-10:2010 | Biological evaluation of medical devices - Part 10: Tests for<br>irritation and skin sensitization |
Software performance was established according to FDA's Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, consistent with moderate level of concern.
The data provided demonstrates that the Finapres NOVA met all applicable requirements and that it is substantially equivalent to its predecessors Finometer® Model-1 and Ohmeda 2350 and to the Nexfin_HD™.
#### 9. Clinical performance data for substantial equivalence determination
Clinical performance data was not required to demonstrate substantial equivalence.
#### 10. Conclusion
On basis of the information above, it is concluded that the device is as safe, as effective, and performs as well as, or better than, the predicate devices.
{9}------------------------------------------------
# 510(k) Summary
# Finapres NOVA - K141460
This 510(k) summary is being submitted in accordance with the requirements of SMDA and 21CFR § 807.92.
# 1. Submitter's information
| Submitter: | Finapres Medical Systems B.V.<br>Hogehilweg 8<br>NL-1101 CC Amsterdam<br>The Netherlands<br>Phone number: +31 20 609 0974<br>Fax Number: +31 20 609 0677<br>Operator Number: 9051428<br>Registration number: 3003803088 |
|-----------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact person: | Dr. Ir. Iris van Uitert<br>Quality Manager<br>Phone number: +31 20 609 0974<br>E-mail:iris.van.uitert@finapres.com |
Date of preparation: March 17, 2015
### 2. Device information
| Trade name: | Finapres NOVA Noninvasive Hemodynamic Monitor |
|-----------------------|-----------------------------------------------|
| Common name(s) | Noninvasive Blood Pressure Monitor |
| | Hemodynamic Monitor |
| | Electrocardiograph |
| | Oximeter |
| Classification name | See Table 1 |
| Device classification | Class II |
### Table 1. Classification name
| Classification name | 21 CFR Section | Product Code |
|-------------------------------------------------------|----------------|--------------|
| Noninvasive Blood Pressure Monitor Measurement System | 870.1130 | DXN |
| Plethysmograph, Impedance | 870.2770 | DSB |
| Oximeter | 870.2700 | DQA |
| Cardiac monitor | 870.2300 | DRT |
### 3. Predicate Devices
The Finapres NOVA is substantially equivalent in design (methodology) and indications for use to the devices shown in Table 2 that have previously been cleared:
{10}------------------------------------------------
Table 2. Predicate devices for the Finapres NOVA
| Device name | Manufacturer | 510(k) |
|--------------------------------------------------|--------------------------------------------------------------|---------|
| Finometer® Model-1 | FMS, Finapres Medical Systems,<br>Amsterdam, The Netherlands | K023723 |
| Ohmeda 2350 Finapres ® Blood<br>pressure monitor | Ohmeda Medical, Englewood | K880572 |
| Portapres | FMS, Finapres Medical Systems,<br>Amsterdam, The Netherlands | K023338 |
| NEXFIN_HD™ | BMEYE B.V., Amsterdam, The Netherlands | K072049 |
| Nellcor Oximax N-600X Pulse<br>Oximeter | Nellcor Puriton Bennet, Inc,<br>Pleasanton | K060576 |
| Zoll M Series with real CPR help | Zoll Medical Corporation, 269 Mill<br>Road, Chelmsford | K062537 |
| MOVES | THORNHILL RESEARCH INC, Toronto,<br>Ontario | K093261 |
### 4. Description
The Finapres NOVA is an instrument to noninvasively monitor blood pressure and hemodynamic parameters. The Finapres NOVA provides a characterization of the arterial circulation and its beat variability in pressure and flow and in various hemodynamic parameters derived from these pressure and flow signals. The Finapres NOVA has the option to include four additional modules to extend its functionality with ECG and SpO2 measurements, blood pressure calibration and data transfer to and from the device.
The embedded software in the device provides computation of real-time and beat-tobeat blood pressure as well as hemodynamic parameters from the non-invasely measured blood pressure waveform. Hemodynamic parameters include cardiac output based on the modelflow method and total peripheral resistance.
The measurement of blood pressure in a finger is based on the arterial volume-clamp method of the Czech physiologist J. Peñáz, and the Physiocal - physiological calibration criteria for the proper unloading of the finger arteries of K.H. Wesseling. With this method, finger arterial pressure is measured using a finger cuff and an inflatable bladder in combination with an infrared plethysmograph, which consists of an infrared light source and detector.
The SpO2, upper arm calibration and ECG modules used in the Finapres NOVA are commercially available OEM modules that are used in FDA approved systems. The finger blood pressure measurement module used in the system is similar to other Finapres Medical Systems B.V. devices available on the market. The analog input/output module has been developed by Finapres during the Finapres NOVA development.
# 5. Indications for use
The Finapres NOVA is intended to be used with patients who have a need for a noninvasive blood pressure and hemodynamic monitor. The noninvasive blood pressure waveform is measured on the subject's finger. The Finapres NOVA provides a noninvasive characterization of the arterial circulation and its beat-to-beat variability in
{11}------------------------------------------------
pressure and flow and in various hemodynamic parameters derived from these pressure and flow signals. Cardiac output derived from the flow signal requires a calibration with thermal dilution.
The Finapres NOVA has the option to include additional modules to extend its functionality with ECG and SpO2 measurements and blood pressure calibration.
When the SpO2 module is present, the Finapres NOVA can additionally monitor the functional oxygen saturation of arterial hemoglobin (SpO2) and the pulse rate.
When the ECG module is present, the Finapres NOVA can additionally monitor the ECG parameters of a patient and their pulse rate. Alarms concerning the pulse rate will be available from the monitor.
When the blood pressure calibration module is present, the Finapres NOVA can additionally provide an upper arm non-invasive blood pressure measurement to determine the blood pressure value for calibration.
The Finapres NOVA is intended to be used for subjects above 18 years of age.
The Finapres NOVA is intended for use in a professional medical environment.
# 6. Intended Use
The Finapres NOVA is intended to be used with patients with a need for a noninvasive blood pressure and hemodynamic monitor. The noninvasive blood pressure waveform is measured on the subject's finger. The Finapres NOVA provides a noninvasive characterization of the arterial circulation and its beat-to-beat variability in pressure and flow and in various hemodynamic parameters derived from these pressure and flow signals.
The Finapres NOVA has the option to include four additional modules to extend its functionality with ECG and SpO₂ measurements, blood pressure calibration and data transfer to and from the device.
The Finapres NOVA is used in a professional medical environment such as hospitals, clinics and research institutions. The Finapres NOVA is a standalone device intended for desktop use. Measurements are to be performed under uninterrupted patient surveillance of the operator.
The user of the Finapres NOVA should be a qualified operator. The operator should have knowledge of the system and data interpretation, obtained via medical education, system manuals and/or specific courses. The device does not report any diagnosis but provides numerical values. It is the physician's responsibility to make proper judgments based on these numbers.
# 7. Summary of Technical Comparison with predicate devices
The Finapres NOVA is an updated model of the Ohmeda 2350, the Portapres and the Finometer® Model-1 device. As with predicate devices, the Finapres NOVA is indicated for use as a noninvasive blood pressure monitor. The Finapres NOVA is used to provide real-time and on-line monitoring and trending of hemodynamic parameters just as with the Finometer® Model-1 devices.
The Ohmeda 2350 device and the Finometer® Model-1 device have the same indications for use, the same design, and the same methodology as the Finapres NOVA Noninvasive
{12}------------------------------------------------
Hemodynamic Blood Pressure Monitor. The predicate Ohmeda 2350, the Finometer® Model-1 and the Nexfin HD provide the same noninvasive measurement and display of finger arterial blood pressure and pulse rate monitoring as the Finapres NOVA.
The Finapres NOVA consists of the same device architecture as the Finometer® Model-1 and the Nexfin HD.
The Finapres NOVA uses functionally the same embedded software for controlling the measurements as that used in the Finometer® Model-1 and Portapres device. The Finapres NOVA uses the proprietary NOVAScope/Modelflow software program for the measurement and trending of hemodynamic parameters, which is an updated version of the software that was used for data analysis in the Finometer and for the Portapres.
The Finapres NOVA can be equipped, equivalent to the Finometer® Model-1 and Nexfin_HD, with a blood pressure calibration method and does, therefore, allow for the measurement of accurate absolute blood pressure data complying with the AAMI SP10 :2002 and its successor IEC 81060-2:2009. The Finapres NOVA has, similar to the Nexfin_HD and the Finometer® Model-1, an optional Height correction unit (HCU) that allows the patient free hand movement during measurements with the device. The Finapres NOVA has, similar to the Finometer® Model-1, an optional ECG module that provides an analog ECG which is sampled and stored simultaneously with the blood pressure waveform. The Finapres NOVA has, same as the Ohmeda 2350, an optional SpO2 module that provides an SpO2 signal which is sampled and stored simultaneously with the blood pressure waveform. Similar to the Portapres, the Finapres NOVA enables 24 continuous measurements.
The three optional modules are available in the following predicate devices: MOVES, K093261; Nellcor Oximax N-600X Pulse Oximeter, K060576; Zoll M Series with real CPR help, K062537.
# 8. Non-clinical performance data for substantial equivalence determination
Testing of the Finapres NOVA was performed according to medical device safety standards shown in Table 3.
| Standard number | Standard name |
|------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ANSI/AAMI ES60601-1:2005 /<br>A2:2010 | Medical electrical equipment - Part 1: General requirements for<br>basic safety and essential performance |
| IEC 60601-1:1988 +A1:1991 +<br>A2:1995 (2nd edition) | Medical electrical equipment - Part 1: General requirements for<br>basic safety and essential performance |
| IEC 60601-1-2:2007 | Medical electrical equipment - Part 1-2: General requirements<br>for safety -Electromagnetic compatibility - Requirements and<br>tests |
| IEC 60601-1-4 1996 + Am.1<br>1999 | Medical electrical equipment - Part 1-4 - Collateral Standard:<br>Programmable electrical medical systems |
| IEC 60601-1-8: 2006 | Medical electrical equipment -- Part 1-8: General requirements<br>for basic safety and essential performance -- Collateral<br>standard: General requirements, tests and guidance for alarm<br>systems in medical electrical equipment and medical electrical<br>systems |
| IEC 60601-1-8: 2003 + A1: | Medical electrical equipment — Part 1-8: General requirements |
Table 3.Medical device standards that the Finapres NOVA is tested against
{13}------------------------------------------------
| 2006 | for safety – Collateral standard: General requirements, tests<br>and guidance for alarm systems in medical electrical equipment<br>and medical electrical systems |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-2-27:2005 | Medical electrical equipment – Part 2-27: Particular<br>requirements for the safety, including essential performance, of<br>electrocardiographic monitoring equipment |
| IEC 80601-2-30:2009 | Medical electrical equipment – Part 2-30: Particular<br>requirements for the basic safety and essential performance of<br>automated non-invasive sphygmomanometers |
| ISO 9919:2005 | Medical electrical equipment – Particular requirements for the<br>basic safety and essential performance of pulse oximeter<br>equipment for medical use |
| ISO 80601-2-61:2011 | Medical electrical equipment – Part 2-61: Particular<br>requirements for basic safety and essential performance of<br>pulse oximeter equipment. |
| ISO 81060-2:2009 | Non-invasive sphygmomanometers – Part 2: Clinical validation<br>of automated measurement type |
| ISO 14971:2007 | Medical devices – Risk management – Application of risk<br>management to medical devices |
| IEC 62304:2006 | Medical device software – software life cycle processes |
| IEC 62366:2009 | Medical devices – Application of usability engineering to medical<br>devices. |
| ISO 10993-1:2009 | Biological evaluation of medical devices – Part 1: Evaluation<br>and testing |
| AAMI / ANSI / ISO 10993-5:2009 | Biological evaluation of medical devices – Part 5: Tests for In<br>Vitro cytotoxicity |
| ISO 10993-10:2010 | Biological evaluation of medical devices – Part 10: Tests for<br>irritation and skin sensitization |
Software performance was established according to FDA's Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, consistent with moderate level of concern.
The data provided demonstrates that the Finapres NOVA met all applicable requirements and that it is substantially equivalent to its predecessors Finometer® Model-1 and Ohmeda 2350 and to the Nexfin_HD™.
#### 9. Clinical performance data for substantial equivalence determination
Clinical performance data was not required to demonstrate substantial equivalence.
#### 10. Conclusion
On basis of the information above, it is concluded that the device is as safe, as effective, and performs as well as, or better than, the predicate devices.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.