FULLY AUTOMATIC ELECTRONIC BLOOD PRESSURE MONITOR

K110330 · Andon Health Co, Ltd. · DXN · Feb 18, 2011 · Cardiovascular

Device Facts

Record IDK110330
Device NameFULLY AUTOMATIC ELECTRONIC BLOOD PRESSURE MONITOR
ApplicantAndon Health Co, Ltd.
Product CodeDXN · Cardiovascular
Decision DateFeb 18, 2011
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1130
Device ClassClass 2

Indications for Use

KD-556J Fully Automatic Electronic Blood Pressure Monitor are for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm.

Device Story

Non-invasive blood pressure monitor; uses inflatable upper arm cuff (22-48cm). Input: oscillometric pressure signals via silicon integrated pressure sensor. Processing: calculates systolic/diastolic pressure and pulse rate; computes average of last three measurements; detects irregular heartbeats. Output: digital display on LCD. Used by medical professionals or patients at home. Provides blood pressure/pulse data to assist in monitoring cardiovascular status; includes JNC hypertension classification.

Clinical Evidence

No new clinical testing performed. Substantial equivalence is supported by the clinical test report of the predicate device (K091500) and compliance with ANSI/AAMI SP10 standards.

Technological Characteristics

Oscillometric measurement principle; silicon integrated pressure sensor. Powered by electrical source. Form factor: upper arm cuff (22-48cm) with LCD display. Standards: ANSI/AAMI SP10, IEC 60601-1, IEC 60601-1-2. Biocompatibility and materials identical to predicate.

Indications for Use

Indicated for adult individuals to measure diastolic and systolic blood pressure and pulse rate via non-invasive upper arm cuff (22cm-48cm circumference). Suitable for use by medical professionals or at home.

Regulatory Classification

Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## 16 1103 # 510(k) Summary This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of SMDA and 21 CFR 807.92. ## 1.0 submitter's information FEB 1 8 2011 | Name: | Andon Health Co., Ltd. | |----------------------|-----------------------------------------------------------------------| | Address: | No 3, Jinping Street Ya An Road, Nankai District, Tianjin, P.R. China | | Phone number: | 86-22-6052 6161 | | Fax number: | 86-22-6052 6162 | | Contact: | Liu Yi | | Date of Application: | 01/27/2011 | ## 2.0 Device information | Trade name: | Fully Automatic Electronic Blood Pressure Monitor | |----------------------|---------------------------------------------------| | Common name: | Noninvasive blood pressure measurement system | | Classification name: | Noninvasive blood pressure measurement system | ## 3.0 Classification Production code: DXN- Noninvasive blood pressure measurement system. Regulation number: 870.1130 Classification: II Panel: Cardiovascular ## 4.0 Predicate device information Manufacturer: Andon Health Co., Ltd. - KD-556 Fully Automatic Electronic Blood Pressure Device: Monitor 510(k) number: K091500 {1}------------------------------------------------ ## 5.0 Device description KD-556J Fully Automatic Electronic Blood Pressure Monitor are for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm. 16110330 It is designed and manufactured according to ANSI/AAMI SP10--manual, electronic or automated sphygmanometers. The operational principle is based on oscillometric and silicon integrates pressure sensor technology. It can calculate the systolic and diastolic blood pressure, and display the result on the LCD. If any irregular heartbeat is detected, it can also be shown on the LCD. More over, it also calculates the average of the last three measurements. ## 6.0 Intended use KD-556J Fully Automatic Electronic Blood Pressure Monitor are for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm. The intended use and the indication for use of KD-556J, as described in the labeling are the same as their predicated device KD-556. | device | | |-------------------------------|-------------------| | Technological Characteristics | Comparison result | | Design principle | Identical | | Appearance | Similar | | Patients contact Materials | Identical | | Performance | Similar | | Biocompatibility | Identical | | Mechanical safety | Identical | | Energy source | Identical | | Standards met | Identical | | Electrical safety | Identical | | EMC | Identical | | Function | Similar | ## 7.0 Summary_comparing_technological_characteristics_with_predicate device {2}------------------------------------------------ ## 8.0 Discussion of non-clinical and clinical test performed ## Non-clinical Tests have been done as follows: - a. Electromagnetic compatibility evaluation according to IEC 60601-1-2; - b. Electrical safety test according test to IEC 60601-1 ; - c. Safety and performance characteristics of the test according to SP10 None of the test demonstrates that KD-556J bring new questions of safety and effectiveness. ## Clinical Test Concerning the Compliance of ANSI/AAMI SP10 From the technical point of view, the subject device KD-556J is identical to its predicate device KD-556. The difference between the subject device and its predicate devices do not affect the clinical accuracy in terms of blood pressure detection. The clinical test report of KD-556(K091500) is applicable to our subject device. {3}------------------------------------------------ ## 9.0 Performance summary KD-556J Fully Automatic Electronic Blood Pressure Monitor conforms to the following standards: - IEC 60601-1, Medical Electrical Equipment Part 1: General Requirements for Safety, 1988; Amendment 1, 1991-11, Amendment 2, 1995. - · EN 60601-1-2, Medical Electrical Equipment Part 1-2: General Requirements for Safety - Collateral standard: Electromagnetic Compatibility - Requirements and Tests, 2007. - · AAMI SP10:2002, Manual, electronic or automated sphygmomanometers. - · AAMI / ANSI SP10:2002/A1:2003 --, Amendment 1 to ANSI/AAMI SP10:2002 Manual, electronic, or automated sphygmomanometers. - AAMI / ANSI SP10:2002/A2:2006 --, Amendment 2 to ANSI/AAMI SP10:2002 Manual, electronic, or automated sphygmomanometers. #### 10.0 Comparison to the predicate device and the conclusion Our device KD-556J Fully Automatic Electronic Blood Pressure Monitor is substantially equivalent to the Fully Automatic Electronic Blood Pressure Monitor KD-556 whose 510(k) number is K091500. KD-556J is very similar with its predicate device in the intended use, the design principle, the material, the performance and the applicable standards. Only the appearance, the time format and the hypertension classification of KD-556J is changed. KD-556J will use the JNC hypertension classification while its predicate device KD-556 uses the WHO hypertension classification. However, the test in this submission provides demonstration that these small differences do not raise any new questions of safety and effectiveness to the new devices. {4}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the bird symbol. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 FEB 1 8 201 Andon Health Co., Ltd. c/o Mr. Liu Yi President No. 3, JinPing Street, Ya An Road, Nankai District Tianjin, P.R. China, 300190 Re: K110330 Trade/Device Name: KD-556J Fully Automatic Electronic Blood Pressure Monitors Regulation Number: 21 CFR 870.1130 Regulation Name: Non-invasive blood pressure measurement systems Regulatory Class: Class II (two) Product Code: DXN Dated: January 31, 2011 Received: February 3, 2011 Dear Mr. Yi: We have reviewed your Section 510(k) premarket notification of intent to market the device w & nave and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for ass stated in the encreat 1976, the enactment date of the Medical Device Amendments, or to econmores that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). Tou may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may, therefer your of the Act include requirements for annual registration, listing of general voltren provincities, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it · If your device is classified (500 above) Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may found in the Code of Features concerning your device in the Federal Register. {5}------------------------------------------------ #### Page 2 – Ms. Liu Yi Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.html for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely vours. W.M.S. Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health #### Enclosure {6}------------------------------------------------ ## Statement of Indications for Use | 510(k) Number : | K110330 | |-----------------|---------| |-----------------|---------| Device name: KD-556J Fully Automatic Electronic Blood Pressure Monitor #### Indications for use: KD-5561 Fully Automatic Electronic Blood Pressure Monitor are for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm. Prescription use Part 21 CFR 801 Subpart D) Over-The-Counter Use AND/OR YES (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-COUNTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) W. Wtast. Page 1 of 1 (Division Sign-Off) Division of Cardiovascular Devices 510(k) Number
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...