PERFORMR SERIES EP DIAGNOSTIC CATHETERS
Device Facts
| Record ID | K964272 |
|---|---|
| Device Name | PERFORMR SERIES EP DIAGNOSTIC CATHETERS |
| Applicant | Medtronic Cardiorhythm |
| Product Code | DRF · Cardiovascular |
| Decision Date | Jan 23, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 870.1220 |
| Device Class | Class 2 |
Indications for Use
A diagnostic electrophysiology catheter is used to record electrical activity from within the heart and to stimulate the heart in accordance with various programmed electrical stimulation protocols incorporated into an EP diagnostic study. All catheters connect physician stimulator/recorder equipment to the desired region of the heart and carry analog signals. Data from the EP study is used to assess arrhythmia patient prognosis, guide therapy selection and evaluate the effectiveness of previously selected therapeutic interventions.
Device Story
Performr Series diagnostic electrophysiology catheters; closed-lumen design; 2-11 platinum alloy electrodes fixed to shaft. Physician-operated; used in clinical EP labs. Manipulator handle allows tip curve angle adjustment; some models feature radius of curvature control or lateral deflection. Catheter connects to external stimulator/recorder equipment; transmits analog cardiac electrical signals for recording and delivers stimulation pulses. Enables assessment of arrhythmia prognosis, therapy guidance, and intervention evaluation.
Clinical Evidence
No clinical data. Non-clinical and in-vivo testing conducted per GLP and FDA guidance to confirm electrical and mechanical performance. Biocompatibility fulfilled via predicate device comparison.
Technological Characteristics
Closed-lumen catheter; 5-7 French diameter; 60-125 cm length. Platinum alloy electrodes. Manipulator handle for tip deflection/curve radius control. Analog signal transmission. Biocompatible materials identical to predicate.
Indications for Use
Indicated for patients undergoing electrophysiology (EP) diagnostic studies to record cardiac electrical activity and provide cardiac stimulation for arrhythmia prognosis, therapy selection, and evaluation of therapeutic interventions.
Regulatory Classification
Identification
An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.
Predicate Devices
- Medtronic CardioRhythm Marinr EP catheter
- Bard EP-XT Steerable catheter
- Elecath Genesis catheter
Related Devices
- K953185 — PLATINUM SERIES EP DIAGNOSTIC CATHETERS · Medtronic Cardiorhythm · Apr 15, 1996
- K043513 — STERILMED REPROCESSED DEFLECTABLE ELECTROPHYSIOLOGY DIAGNOSTIC CATHETER · Sterilmed, Inc. · Mar 1, 2005
- K231312 — PENTARAY® NAV ECO High Density Mapping Catheter, DECANAV® Mapping Catheter, Webster® CS Catheter with Auto ID, Webster® CS Catheter with EZ Steer Technology, Webster® CS Catheter with EZ Steer Technology with Auto ID · Biosense Webster, Inc. · Jul 10, 2023
- K101789 — REPROCESSED ELECTROPHYSIOLOGY DIAGNOSTIC CATHETERS · Sterilmed, Inc. · Aug 19, 2010
- K981642 — STABLEMAPR STEERABLE INTRACARDIAC CATHETERS · Medtronic Cardiorhythm · Aug 5, 1998