ARMADA 14XXT PTA CATHETER

K121352 · Abbott Vascular · DQY · Aug 15, 2012 · Cardiovascular

Device Facts

Record IDK121352
Device NameARMADA 14XXT PTA CATHETER
ApplicantAbbott Vascular
Product CodeDQY · Cardiovascular
Decision DateAug 15, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1250
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The device is indicated to dilate stenoses in femoral, popliteal, infra popliteal and renal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. The 2.0 to 5.0 mm balloon diameters are also indicated for postdilatation of balloon-expandable stents.

Device Story

Armada 14 XT PTA Catheter is an over-the-wire co-axial balloon catheter used for percutaneous transluminal angioplasty. Device features distal balloon for dilation, radiopaque markers for positioning, and proximal brachial/femoral markers for depth gauging. Operated by physicians in clinical settings; catheter is advanced over a guide wire to the target stenosis; inflation lumen allows contrast fluid delivery to expand balloon to specific diameter/length. Device facilitates vessel patency in peripheral arteries and dialysis fistulae; also used for stent post-dilatation. Benefits include restoration of blood flow in stenotic vessels.

Clinical Evidence

Bench testing only. Testing included balloon crossing/refolded profiles, burst strength, compliance, inflation/deflation times, fatigue, tensile strength, flexibility, kink resistance, torque strength, and radiopacity. Biocompatibility testing included cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, hemolysis, pyrogen, and complement activation. All tests met acceptance criteria.

Technological Characteristics

Over-the-wire co-axial catheter design. Materials, design, and sterilization processes are identical to predicate TREK/MINI-TREK catheters. Features include radiopaque markers, inflation/deflation lumen, and proximal markers. No electronic or software components.

Indications for Use

Indicated for patients requiring dilation of stenoses in femoral, popliteal, infra-popliteal, and renal arteries, and treatment of obstructive lesions in native or synthetic arteriovenous dialysis fistulae. 2.0-5.0 mm balloon sizes also indicated for post-dilatation of balloon-expandable stents.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## 510(k) Summary . The 510(k) Summary is submitted in accordance with 21 CFR 8807,92 and the requirements of the Safe Medical Device Act (SMDA) of 1990. AUG 1 5 2012 Submitter's Name Submitter's Address Telephone Fax Contact Person Date Prepared Device Trade Name Device Common Name Device Classification Name Predicate Device Names Abbott Vascular 3200 Lakeside Drive, Santa Clara, CA 95054 (408) 845-0682 (408) 845-3743 Ivalee Cohen May 2, 2012 Armada 14 XT PTA Catheter PTA Catheter Catheter, angioplasty, peripheral, transluminal, LIT • Abbott Vascular TREK OTW & MINI TREK OTW Coronary Dilatation Catheter (K103011, cleared 1/10/11 and K110617, cleared 6/2/11) · Abbott Vascular Armada 14 PTA Catheter (K102705, cleared 12/7/10) #### Device Description The Armada 14 XT PTA Catheter is an over-the-wire co-axial design with a balloon at the distal tip. The balloon segment expands to a known diameter and length at a specific inflation pressure and has radiopaque marker(s) under the balloon to aid in positioning the balloon in a stenosis. The co-axial shaft consists of a tubular inner and outer member. The inner lumen permits the use of a guide wire to facilitate advancement of the catheter to and through the stenosis to be dilated. The outer lumen provides for inflation and deflation of the balloon with contrast fluid. Along the proximal portion of the shaft are brachial and femoral markers to aid in gauging the catheter's position relative to the guiding catheter tip when introducing the catheter through the guiding catheter. An adaption arm is located at the proximal end to provide access to the inflation lumen and guide wire lumen and allows connection with an inflation device. #### Indication for Use The device is indicated to dilate stenoses in femoral, popliteal, infra popliteal and renal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. The 2.0 to 5.0 mm balloon diameters are also indicated for postdilatation of balloon-expandable stents. {1}------------------------------------------------ # Technological Characteristics The subject Armada 14 XT PTA Catheter has the same fundamental scientific technology (operating principle) and indication for use as the Armada 14 predicate device and shares the same materials, design, labeling, packaging materials, shelf life and sterilization process as the TREK and MINI-TREK predicate devices. ### Performance Data Performance testing was successfully completed on the Armada 14 XT PTA Catheter. The following tests were conducted: - . Balloon Crossing Profile - Refolded Balloon Profile . - Guide wire Lumen ID . - . Total Catheter Length - Minimum Balloon Burst Strength (Rated Burst Pressure, RBP) . - Balloon Compliance (Diameter vs. Pressure) . - . Balloon Inflation/ Deflation Time - Balloon Fatigue (Repeat Balloon Inflations) ● - . Tensile strength - Proximal Adaptation - . Tensile strength - Proximal Balloon Seal - Tensile strength Distal Outer Member to Proximal Shaft . - Tensile strength Soft Tip to Inner Member Seal ● - Catheter Flexibility & Kink Test - Torque Strength . - . Radiopacity - Minimum Balloon Burst Strength (RBP) Within Stent � - Balloon Fatigue Within Stent (Repeat Balloon Inflations in Stent) ● - Shelf Life (Accelerated Aging) Biocompatibility testing included cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, hemolysis, pyrogen, and complement activation. ### Conclusion The in vitro bench and biocompatibility tests demonstrated that the Armada 14 XT PTA Catheters met all acceptance criteria and performed similarly to the predicate devices. No new safety or effectiveness issues were raised during the testing program and, therefore, these devices may be considered substantially equivalent to the predicate devices. {2}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle with its wings spread. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 AUG 15 2012 Abbott Vascular c/o Ivalee Cohen 3200 Lakeside Drive Santa Clara, CA 95054 Re: K121352 Trade/Device Name: Armada 14 XT PTA Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: LIT, DQY Dated: May 02, 2012 Received: May 04, 2012 Dear Ms. Cohen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a provisions of the Federal rood, Drily, You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate informations against missianuly and warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations offecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ # Page 2 - Ms. Ivalee Cohen Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of modical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electionic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm1118001,htmfor for for for for for for for for for for for for for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, Jease note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours. .s.kiel Go Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use K121352 510(k) Number (if known): Device Names: 、 Armada 14 XT PTA Catheter Indications for Use: The device is indicated to dilate stenoses in femoral, popliteal, infra popliteal and renal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. The 2.0 to 5.0 mm balloon diameters are also indicated for post-dilatation of balloon-expandable stents. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Cuañillele (Division Sign-Off) Division of Cardiovascular Devices Page 1 of 化1252 510(k) Number_
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...