BRIDGEPOINT MEDICAL MANTARAY(TM) CATHETER

K111963 · Bridgepoint Medical · DQY · Aug 10, 2011 · Cardiovascular

Device Facts

Record IDK111963
Device NameBRIDGEPOINT MEDICAL MANTARAY(TM) CATHETER
ApplicantBridgepoint Medical
Product CodeDQY · Cardiovascular
Decision DateAug 10, 2011
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1250
Device ClassClass 2

Indications for Use

The BridgePoint Medical Mantaray™ Catheter is indicated for directing, steering, controlling, and supporting a guidewire in order to access discrete regions of the peripheral vasculature.

Device Story

Mantaray™ Catheter is a single-use, over-the-wire, dual-lumen percutaneous catheter. It facilitates guidewire placement, support, and steering in peripheral vasculature. Device features a distal flexible shaft with a non-compliant balloon for fluoroscopic orientation and two sideports with radiopaque markers. Sideports connect to the central guidewire lumen, allowing the guidewire to exit at an angle for steering. Used by physicians in clinical settings. Output is mechanical guidewire positioning; assists clinicians in navigating complex vascular anatomy. Benefits include improved guidewire control and access to discrete vascular regions.

Clinical Evidence

In vivo testing conducted in a porcine model (n=6 animals, 4 arteries each). Evaluation included angiographic assessment, histology, pathology, hematology, and serum chemistry. No complications, vessel injury, or downstream embolism reported. Bench testing confirmed performance characteristics (tensile, burst, fatigue, inflation/deflation, trackability, torque, etc.) against predicate. Biocompatibility data leveraged from predicate device.

Technological Characteristics

Dual-lumen percutaneous catheter; non-compliant balloon for fluoroscopic orientation; radiopaque markers. Materials identical to predicate. Mechanical steering via sideports. Sterilized device. No software or electronic components.

Indications for Use

Indicated for directing, steering, controlling, and supporting a guidewire to access discrete regions of the peripheral vasculature.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ BridgePoint Medical Mantaray™ Catheter Special 510(k) # 9. 510(K) SUMMARY AUG 1 0 2011 This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. 510(k) Number: # Applicant Information: Date Prepared: July 8, 2011 Name: Address: BridgePoint Medical 2800 Campus Drive, #50 Plymouth, MN 55441 Phone: 763-225-8500 Fax: 763-225-8718 | Contact Person: | Jill Munsinger | |-----------------|-------------------------------------------| | Phone Number: | office: 763-225-8510 / cell: 651-270-0572 | | E-mail: | jmunsinger@bridgepointmedical.com | # Device Information: | Classification: | Class II Percutaneous Catheter | |----------------------|--------------------------------| | Trade Name: | Mantaray™ Catheter | | Common Name: | Percutaneous Catheter | | Classification Name: | Percutaneous Catheter | #### Predicate Devices: The BridgePoint Medical Mantaray™ Catheter is substantially equivalent in intended use, method of operation and technical aspects to the following predicate device: K080987 and K101591 - Stingray™ Catheter # Device Description: The Mantaray™ Catheter is a single use, over-the-wire, disposable, dual lumen percutaneous catheter that facilitates the placement, support and steering of a guidewire into discrete regions of the peripheral vasculature through the central guidewire lumen or through one of two sideports (identified by radiopaque markers). The sideports connect with the central guidewire lumen and facilitate guidewire steering (at an angle to the central lumen) by allowing the guidewire to exit the catheter. The catheter contains a small non-compliant balloon segment used for fluoroscopic orientation on the distal tip of the flexible shaft. {1}------------------------------------------------ #### Intended Use: The BridgePoint Medical Mantaray™ Catheter is indicated for directing, steering, controlling, and supporting a guidewire in order to access discrete regions of the peripheral vasculature. ## Comparison to Predicate Device(s): The Mantaray™ Catheters are substantially equivalent to the Stingray™ Catheters, K080987 and K101591 in that they are both designed to direct, steer, control and support a guidewire in accessing discrete regions of the peripheral vasculature. The Mantaray™ Catheter is constructed of the same materials as the Stingray Catheter. Modifications were made to the balloon size and proximal shaft stiffness. The Stingray™ Catheters and Mantaray™ Catheters are manufactured using the same processes and components and have similar physical attributes (balloon performance, trackability, tensile, radiopacity, and torque, etc.). Both devices include radiopaque markers located within the balloon segment to indicate the location of the guidewire lumen ports. #### Performance Data: The Mantaray™ Catheters have been evaluated using the following in vitro bench testing to confirm the performance characteristics as compared to the predicate device: - Tensile . - Burst . - Fatigue . - Inflation & Deflation Time . - Dimensional . - . Guidewire Insert & Withdrawal - Flexibility . - Trackability . - Guidewire Re-Direction ● - Markerband Movement & Removal . - Markerband & Guidewire Interaction . - Kink Resistance ● - Coating . - Torque . - Surface Defects ● - Corrosion Resistance . - Luer and Hub Tests . - Balloon Protector Removal, and ● - Radiopacity . In vivo testing was also completed in accordance with 21 CFR Part 58, "Good Laboratory Practices for Nonclinical Laboratory Studies." The functional performance and safety of the Mantaray™ Catheters were evaluated in a porcine animal model. MantarayTM {2}------------------------------------------------ Catheters were inserted into four arteries in each of six animals used for the evaluation. The vessels were evaluated angiographically followed by histology and pathology. Hematology and serum chemistry along with gross necropsy were also used for evaluations. There were no reported complications during each treatment. All sixanimals survived the in-life period with no angiographic evidence of vessel injury or downstream embolism and no abnormal pathologic findings. Biocompatibility testing on the Mantaray™ Catheters was not completed as all materials are included in the predicate device, Stingray™ Catheter. Biocompatibility testing was successfully completed previously and demonstrated the materials used in the design and manufacture of the device are non-toxic and non-sensitizing to biological tissues consistent with the intended use. The following Biocompatibility tests were previously completed: - Cytotoxicity L929 MEM ISO . - Kligman Sensitization (Maximization) ISO � - Irritation- Intracutaneous Injection ISO . - Acute Systemic Cytotoxicity ISO . - . Pyrogen - ISO - Hemocompatibility (Direct and Indirect) Hemolysis ASTM . - In Vitro Hemocompatibility Assay ISO . - Complement Activation (Direct) Assay ISO ● - In Vivo Thrombogenicity Assay ISO, and . - Unactivated Partial Thromboplastin Time ISO . All test results demonstrated the materials, manufacturing processes, and design of the Mantaray™ Catheters met the established performance criteria and will perform as intended. #### Summary: Based upon the intended use and descriptive information provided in this pre-market notification, the BridgePoint Mantaray™ Catheters have been shown to be substantially equivalent to the currently marketed predicate device. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three bars representing its wings and tail feathers. The eagle is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA", which is arranged in a circular fashion around the left side of the logo. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 BridgePoint Medical, Inc. C/O Jill Munsinger 2800 Campus Drive, Ste. #50 Plymouth, MN 55441 Re: K111963 AUG 1 0 2011 Trade/Device Name: Mantaray™ Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DOY Dated: July 08, 2011 Received: July 11, 2011 Dear Ms. Munsinger: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act {4}------------------------------------------------ Page 2 – Ms. Jill Munsinger or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also. please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. ML Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # 8. INDICATIONS FOR USE STATEMENT 510(k) Number: (TBA) BridgePoint Medical Mantaray™ Catheter Device Name: Indications For Use: The BridgePoint Medical Mantaray™ Catheter is indicated for directing, steering, controlling, and supporting a guidewire in order to access discrete regions of the peripheral vasculature. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) l Page 1 of 1 (Division Sign-Off) Division of Cardiovascular Devices 510(k) Number_ Page 19 of 80
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...