The Selectra accessory kit is used in conjunction with the Selectra CS lead introducer system to facilitate lead implantation in the left side of the heart via the coronary sinus.
Device Story
Selectra accessory kit is a collection of catheter accessories packaged together for use with the Selectra CS lead introducer system. Components include a slitter tool, Seldinger guide wire, transvalvular insertion (TVI) tools, syringe, torquer, stopcocks, check valves, and sealing caps. The kit is used by clinicians during coronary sinus lead placement procedures to facilitate lead implantation. It functions as a procedural support system for guiding catheters. No electronic, software, or algorithmic components are involved.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Kit includes mechanical catheter accessories: slitter tool, guide wire, TVI tools, syringe, torquer, stopcocks, check valves, and sealing caps. No energy source, software, or electronic connectivity. Sterilized components.
Indications for Use
Indicated for use with the Selectra CS lead introducer system to facilitate lead implantation in the left side of the heart via the coronary sinus.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
K183265 — Selectra Catheters, Selectra Accessory Kit · Biotronik, Inc. · Dec 13, 2018
K131978 — SELECTRA ACCESSORY KIT, SELECTRA SLITTER TOOL · Biotronik, Inc. · Aug 23, 2013
K222037 — Selectra Lead Implantation System · Biotronik, Inc. · Jul 19, 2022
K151409 — Selectra Catheters and Selectra Accessory Kit · Biotronik, Inc. · Aug 7, 2015
K020821 — SCOUT (SINUS CORNARIUS UTILITY TOOL), MODEL 335 117 · Biotronik, Inc. · Jun 3, 2002
Submission Summary (Full Text)
{0}------------------------------------------------
# Selectra Accessory Kit Special 510(k) Premarket Notification
SEP 2 6 2011
# 1. 510(K) SUMMARY
Name and Address of Sponsor:
BIOTRONIK, Inc. 6024 Jean Road Lake Oswego, OR 97035
1028232 Establishment Registration Number:
Device Name:
Proprietary Name: Classification:
Selectra Accessory Kit Class II (21 CFR 870.1330; 870.1340; 870.4500; 880.5860) Classification Name: Cardiovascular surgical instruments, catheter guide wire, catheter introducer, piston syringe DWS, DQX, DYB, FMF
#### General Description:
The Selectra accessory kit includes a variety of commonly used catheter accessories which are combined in a single package.
The Selectra accessory kit includes the following components:
Product Code:
- 1 Selectra Slitter Tool .
- . 1 Seldinger Guide Wire
- . 4 Transvalvular Insertion (TVI) Tools
- 1 Syringe .
- . 1 Torquer
- 2 Stopcocks .
- 2 Check valves ●
- . 2 Sealing caps
- . 1 Technical Manual
The Selectra accessory kit may be used with the Selectra CS lead introducer system, BIOTRONIK's family of guiding catheters specifically used for the placement of coronary sinus leads. The Selectra accessory kit is the subject of this Special 510(k).
#### Device Modification:
The changes made to the Selectra accessory kit, as compared to the previously cleared ScoutPro ACS accessory kit, primarily include minor component modifications.
The usage of the Selectra accessory kit remains unchanged and the product characteristics, such as indications for use, contraindications, and function, are the ScoutPro ACS accessory kit cleared on July 23, 2010 (K101776). Therefore, this previously cleared accessory kit will serve as the predicate device for the modified accessory kit included in this Special 510(k).
#### Predicate Device:
BIOTRONIK's ScoutPro ACS Accessory Kit (K101776)
{1}------------------------------------------------
### BIOTRONIK, Selectra Accessory Kit, Special 510(k)
### Indication for Use:
The Selectra accessory kit is used in conjunction with the Selectra CS lead introducer system to facilitate lead implantation in the left side of the heart via the coronary sinus.
・
Phone (888) 345-0374 Fax (503) 635-9936
.
jon.brumbaugh@biotronik.com
:
・
| Name and Address of Manufacturer: | BIOTRONIK SE & Co. KG (reg. no. 9610139) Woermannkehre 1, 12359 Berlin, Germany 011-49-30-689-05-1210 |
|--------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name and Address of Contract Manufacturer: | BIOTRONIK AG (reg. no. 8043892) Ackerstrasse 6 8180 Bülach, Switzerland 011-41-44-864-5169 |
| Name and Address of Contract Sterilizer: | Sterigenics Germany GmbH (reg. no. 3002807090) Kasteler Straße 45 (Rheingaustrasse 190 - 196) D-65203 Wiesbaden, Germany |
| Contact Person(s) and Phone Number: | Jon Brumbaugh VP, Regulatory Affairs and Compliance |
{2}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. In the center of the seal is an abstract symbol that resembles an eagle or other bird-like figure.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -W()66-G609 Silver Spring, MD 20993-0002
Jon Brumbaugh Vice President, Regulatory Affairs and Compliance Biotronik, Inc. 6024 Jean Road Lake Oswego, Oregon 97035
ុទ្ធសាសនា 6 201
. Re: K111839
Trade/Device Name: Selectra Accessory Kit Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide wire Regulatory Class: Class II Product Code: DQY. DQX Dated: June 28, 2011 Received: June 29, 2011
Dear Mr. Brumbaugh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{3}------------------------------------------------
### Page 2 - Mr. Jon Brumbaugh
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting vour device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutfDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Bram D. Zuckerman, M.D.
Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known): TBD
Device Name: Selectra CS Lead Introducer System Indications for Use:
The Selectra accessory kit is used in conjunction with the Selectra CS lead introducer system to facilitate lead implantation in the left side of the heart via the coronary sinus.
V Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
Page 1 of 1
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division St., OR
(Division Sign-Off) Division of Cardiovascular Devices
510(k) Number_K ///839
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.