K151409 · Biotronik, Inc. · DQY · Aug 7, 2015 · Cardiovascular
Device Facts
Record ID
K151409
Device Name
Selectra Catheters and Selectra Accessory Kit
Applicant
Biotronik, Inc.
Product Code
DQY · Cardiovascular
Decision Date
Aug 7, 2015
Decision
SESK
Submission Type
Special
Regulation
21 CFR 870.1250
Device Class
Class 2
Indications for Use
The Selectra CS lead introducer system is used to facilitate lead implantation in the left side of the heart via the coronary sinus.
Device Story
Selectra CS lead introducer system facilitates access to coronary venous system for lead/catheter implantation. System comprises guiding catheters (5F inner, 7F outer) and accessory kit (slitter tool, guide wire, TVI tools, syringe, torque tool, valves, stopcocks, caps). Telescoping catheter design allows probing of coronary sinus ostium, coronary sinus access, contrast media delivery, and left ventricular lead positioning. Used by clinicians in cardiac procedures. Output is physical access to coronary venous anatomy; assists physician in lead placement. Benefits include improved procedural access for cardiac rhythm management device implantation.
Clinical Evidence
Bench testing only. Biocompatibility evaluation per ISO 10993 (cytotoxicity, hemocompatibility). Performance testing included particulate matter, endotoxin, bioburden, visual inspection, simulated use, leak testing, kink testing, tensile strength, friction characteristics, X-ray visibility, and package integrity. Accelerated aging studies supported extended shelf life.
Technological Characteristics
Percutaneous guiding catheters and accessory kit. Materials include polycarbonate (stopcock). Telescoping 5F/7F catheter design. External communicating device with direct blood path contact (<24h). Sterilized. No software or electronic components.
Indications for Use
Indicated for patients requiring lead implantation in the left side of the heart via the coronary sinus.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized representation of three human profiles facing right, stacked on top of each other. The profiles are rendered in a dark color, creating a sense of depth and unity. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the profile graphic.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 7, 2015
Biotronik, Inc. c/o Mr. Jon Brumbaugh Vice President, Regulatory Affairs And Compliance 6024 Jean Road Lake Oswego, OR 97035
Re: K151409
Trade/Device Name: Selectra CS Lead Introducer System Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: II Product Code: DQY Dated: July 7, 2015 Received: July 8, 2015
Dear Mr. Brumbaugh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{1}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
# Bram D. Zuckerman -S
Bram D. Zuckerman, M.D. Director Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known): K151409
Selectra Cathteters and Selectra Accessory Kit Device Name:
Indications For Use:
The Selectra CS lead introducer system is used to facilitate lead implantation in the left side of the heart via the coronary sinus.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use ____________ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
{3}------------------------------------------------
#### 510(k) Summary
BIOTRONIK Selectra CS Lead Introducer System Selectra Catheters and Selectra Accessory Kit
## 1. Submitter
BIOTRONIK 6024 SW Jean Road Lake Oswego, OR 97035 Phone: (888) 345-0374 Fax: 503-451-8519
Contact Person: Jon Brumbaugh
Date Prepared: May 14, 2015
## 2. Device
| Name of Device | Selectra CS Lead Introducer System, including<br>Selectra Catheters and Selectra Accessory Kit |
|----------------------|------------------------------------------------------------------------------------------------|
| Common or Usual Name | Lead Introducer System |
| Classification Name | Percutaneous Catheter (21 CFR 870.1250) |
| Regulatory Class | II |
| Product Code | DQY |
### 3. Predicate Devices
- BIOTRONIK Selectra 7F Catheters (K123324, cleared January 11, 2013) ●
- BIOTRONIK Selectra 5F Catheters (K11154, cleared May 23, 2011)
- . BIOTRONIK Selectra Accessory Kit (K131978, cleared August 23, 2013)
No reference devices were used in this submission.
# 4. Device Description
BIOTRONIK's Selectra CS lead introducer system is a combination of guiding catheters and implantation accessories used to facilitate access to the coronary venous system for suitable leads and catheters. The Selectra CS lead introducer system consists of several individually available guiding catheters with various different curve shapes and the Selectra accessory kit.
The Selectra Accessory Kit includes the following components in a single sterile package:
- . 1 Selectra Slitter Tool
- 1 quide wire .
- 4 transvalvular insertion (TVI) tools ●
- 1 syringe ●
- 1 torque tool
- . 2 check valves
- 2 stopcocks ●
- 2 sealing caps ●
{4}------------------------------------------------
#### 510(k) Summary
BIOTRONIK Selectra CS Lead Introducer System Selectra Catheters and Selectra Accessory Kit
The catheters are available as inner (5F) and outer (7F) catheters which jointly form a telescope system, and allow probing of the ostium of the coronary sinus, access to the coronary sinus, introduction of contrast media for angiography, and positioning of left ventricular leads, also called coronary sinus (CS) leads.
The Selectra catheters are compatible with one another as well as the Selectra Accessory Kit.
# 5. Indications for Use
The Selectra CS lead introducer system is used to facilitate lead implantation in the left side of the heart via the coronary sinus.
The Indications for Use statement for the Selectra system is identical to the predicate device.
## 6. Comparison of Technological Characteristics with the Predicate Device
The technological principles of the subject and predicate devices are the same. The differences represent minor modifications to the currently marketed catheters as follows:
- . Additional models with a shorter overall working length and the same diameter
- A modified blister package ●
- A longer shelf life ●
- Manufacturing changes with no effect on performance criteria ●
- . Stopcock produced from a different polycarbonate material
In addition to the longer shelf life, labeling was updated to facilitate internal inventory management, to comply with UDI requirements (CFR 801.20 and 801.40) and to comply with international standards.
# 7. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
A biocompatibility evaluation for the Selectra Catheters and Selectra Accessory Kit was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing," May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The Selectra CS lead introducer system is classified as an external communicating device with direct contact to blood path and limited contact duration (< 24 h). The testing included:
- Cytotoxicity ●
- Hemocompatibility
In addition, the subject devices were tested for the following:
- . Particulate matter
- Endotoxin ●
- Bioburden ●
{5}------------------------------------------------
#### 510(k) Summary
BIOTRONIK Selectra CS Lead Introducer System Selectra Catheters and Selectra Accessory Kit
To demonstrate that the modified catheters meet the same performance criteria, the following tests were conducted using the same test methods and acceptance criteria as for the predicate devices.
- Visual inspection .
- Simulated use testing (handling, ) ●
- . Leak testing
- Kink testing
- Tensile strength .
- Friction characteristics ●
- X-ray visibility
- Package integrity ●
- . Process validation for manufacturing changes.
When appropriate, test samples were subjected to environmental preconditioning and accelerated aging prior to testing.
# 8. Conclusions
The subject devices result from minor modifications to the predicate devices. The performance testing demonstrates that the subject devices meet the same functional acceptance criteria for the same intended use. In addition, testing conducted after accelerated aging support the longer labeled shelf life of the catheters.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.