PRESSUREWIRE SENSOR, MODEL #12001

K983506 · Radi Medical Systems AB · DQX · Feb 3, 1999 · Cardiovascular

Device Facts

Record IDK983506
Device NamePRESSUREWIRE SENSOR, MODEL #12001
ApplicantRadi Medical Systems AB
Product CodeDQX · Cardiovascular
Decision DateFeb 3, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1330
Device ClassClass 2

Indications for Use

PressureWire™ Sensor is intended to provide pressure signals from the coronary and peripheral vessels and to guide the positioning of interventional devices.

Device Story

PressureWire™ Sensor is a steerable guidewire (0.014" diameter, 175 cm length) with an integrated pressure sensor mounted proximal to a 3 cm shapeable radiopaque tip. Used in coronary and peripheral vasculature during percutaneous interventional procedures. Operates by connecting to a RADI Medical Systems PressureWire™ Interface via a detachable cable containing an identification memory chip with sensor-specific calibration parameters. Provides real-time invasive pressure signals. Clinicians use these signals, often in conjunction with guiding catheter pressure readings, to calculate indices such as Fractional Flow Reserve (FFR) or pressure gradients. Output assists in clinical decision-making regarding vessel status and interventional device placement. Benefits include real-time hemodynamic assessment during minimally invasive procedures.

Clinical Evidence

Bench testing only. Testing confirmed that the presence of the PressureWire Sensor within a guiding catheter (6 Fr. or greater) does not clinically significantly impact the pressure signal recorded by the catheter.

Technological Characteristics

Steerable guidewire, 0.014" diameter, 175 cm length. Integrated pressure sensor proximal to 3 cm shapeable radiopaque tip. Detachable cable with identification memory chip for calibration parameters. Connects to proprietary PressureWire™ Interface. Analog-to-digital signal processing via interface.

Indications for Use

Indicated for use in coronary and peripheral blood vessels to measure blood pressure during percutaneous procedures and to facilitate the placement of interventional devices in patients undergoing such procedures.

Regulatory Classification

Identification

A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.

Special Controls

*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ #### 510(K) SUMMARY 9 This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of Safe Medical Device Act 1990 and CFR 21 §807.92. ## Submitter's Information: | Name: | RADI Medical Systems AB | |-------------------------|-------------------------------------------------------------------------------| | Address: | Palmbladsgatan 10, SE-754 50 Uppsala, Swede | | Phone/Fax: | +46-18-161000 / +46-18-161099 | | Contact Person: | Mats Granlund | | Date of Preparation: | October 5, 1998 | | Device Name: | | | Trade Names: | PressureWire™ Sensor | | Common Name: | Pressure Guidewire | | Classification Name: | Catheter Guide Wire (870.1330)<br>Catheter Tip Pressure Transducer (870.2870) | | Predicate Device Names: | PressureWire™ Sensor (K972793) | ## Device Description: The PressureWire™ Sensor consists of a pressure sensor mounted steerable guidewire to perform real-time invasive pressure measurement in the vasculature and a detachable cable for connection to RADI Medical Systems PressureWire™ Interface only. The guidewire has an outer diameter of .014" and is 175 cm in length. The pressure sensor is mounted just proximal of the 3 cm shapeable radiopaque tip. PressureWire™ Sensor is connected to the PressureWire™ Interface via the Interface Cable which has an identification memory chip containing the individual calibration parameters for the sensor. The readings from the PressureWire™ Sensor and the one measured through a guiding catheter can be used as input for calculation of various indices which require one or two input values e.g. gradient or Fractional Flow Reserve (FFR) etc. ## Intended Use: PressureWire™ Sensor is intended to provide pressure signals from the coronary and peripheral vessels and to guide the positioning of interventional devices. ## Technical Characteristics Summary: The technical characteristics of PressureWire™ Sensor is almost identical with the predicate device. ## Performance Data: Due to the extreme similarity in design and materials between subject and predicate devices further product performance testing has not been considered necessary. A bench test have been conducted to show that the presence of the PressureWire Sensor within a guiding catheter (6 Fr. or greater) does not clinically significant impact the pressure signal recorded by the catheter ## Conclusions The device is almost identical with predicate device and found to be suitable for its intended use. {1}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized emblem featuring a symbol that resembles a bird or an abstract human figure, composed of three curved lines. ﺰ Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 3 1099 FEB Mr. Mats Granlund RADI Medical Systems AB Palmbladsqatan 10, SE-754 50 Uppsala, Sweden Re : K983506 Pressure Wire™ Sensor Trade Name: Requlatory Class: II Product Code: DQX Dated: December 21, 1998 December 23, 1998 Received: Dear Mr. Granlund: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to leqally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, qood manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to A substantially equivalent determination assumes compliance with 895. the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Druq Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to {2}------------------------------------------------ Page 2 - Dr. Susan Silavin your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in_vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4586. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Thomas J. Callahan Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory And Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Statement of Indications for Use K983506 510(k) Number: PressureWire™ Sensor Device Name: Indications for Use: The PressureWire Sensor is intended for use in coronary and peripheral blood vessels to measure blood pressure during percutaneous procedures and to facilitate the placement of interventional devices. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Obito Uchiha for TJC (Division Site Off) (Division Sien-Off) Division of Carciovascular, Respiratory, and Neurological Devices K983506 510(k) Number CONCURRENCE OF CDRH, OFFICE OF DEVICE EVALUATION (ODE) V Prescription Use OR Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (Per 21 CFR 801.109) (Optional Format 1/2/96)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...