K131710 · Merit Medical Systems, Inc. · DQX · Jul 29, 2013 · Cardiovascular
Device Facts
Record ID
K131710
Device Name
MERIT HYDROPHILIC GUIDE WIRE
Applicant
Merit Medical Systems, Inc.
Product Code
DQX · Cardiovascular
Decision Date
Jul 29, 2013
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1330
Device Class
Class 2
Indications for Use
The Merit Hydrophilic Guide Wire is intended to be used in the peripheral vascular system to facilitate the placement of devices during diagnostic and interventional procedures.
Device Story
The Merit Hydrophilic Guide Wire is a sterile, single-use device consisting of a jacketed core wire with a hydrophilic coating. It is available in straight and angled configurations in various lengths. The device is used by physicians in clinical settings to navigate the peripheral vascular system and facilitate the placement of diagnostic or interventional devices. The hydrophilic coating reduces friction, aiding in smooth navigation through vasculature. The device is operated manually by the clinician. It does not involve electronic processing, software, or automated decision-making.
Clinical Evidence
No clinical data. Bench testing only. Testing included tensile strength, torque strength, torquability, tip flexibility, coating adherence/integrity (using anatomical model), catheter compatibility, surface, fracture test, flex test, size/dimensions, and radiopacity. Results met predetermined acceptance criteria and were comparable to predicate devices.
Technological Characteristics
Jacketed core wire with hydrophilic coating. Available in straight and angled configurations. Sterilized via ethylene oxide (ISO 11135-1). Biocompatibility evaluated per ISO 10993-1. Mechanical testing performed per FDA guidance for Coronary and Cerebrovascular Guide Wires and ISO 11070.
Indications for Use
Indicated for use in the peripheral vascular system to facilitate device placement during diagnostic and interventional procedures in patients requiring such vascular access.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
| Section 6 | | | | |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|--------------|--|
| 510(k) Summary | | | | |
| General<br>Provisions | Submitter Name:<br>Address: | Merit Medical Systems, Inc.<br>Parkmore Business Park<br>Parkmore, Galway, Ireland | | |
| | Telephone Number:<br>Fax Number:<br>Contact Person:<br>Registration Number: | (+353) 91 703700 (3061)<br>(+353) 91 680 104<br>Mark Mullaney<br>9616662 | JUL 2 9 2013 | |
| | Correspondent Name:<br>Address: | Merit Medical Ireland Ltd.<br>Parkmore Business Park<br>Parkmore, Galway, Ireland | | |
| | Telephone Number:<br>Fax Number:<br>Contact Person:<br>Date of Preparation:<br>Registration Number: | (+353) 91 703700 (3168)<br>(+353) 91 680 104<br>Martha Folan<br>26th June 2013<br>9616662 | | |
| Subject<br>Device | Trade Name:<br>Common/Usual Name:<br>Classification Name: | Merit Hydrophilic Guide Wire<br>Hydrophilic Guide Wire<br>Catheter Guide Wire | | |
| Predicate<br>Device | Trade Name:<br>Classification Name: | Merit Hydrophilic Guide Wire<br>Catheter Guide Wire | | |
| | Premarket Notification Predicate Device # 1:<br>Merit Hydrophilic Guide Wire K120644<br>Manufacturer: Merit Medical Systems, Inc | | | |
| | Premarket Notification Predicate Device # 2:<br>Radiofocus® Glidewire® K863138<br>Manufacturer: Terumo Medical Corp. | | | |
| | Premarket Notification Predicate Device # 3<br>Merit Hydrophilic Guide Wire K130588<br>Manufacturer: Merit Medical Systems, Inc | | | |
{1}------------------------------------------------
| Classification | Class II<br>21 CFR § 870.1330,<br>Product code: DQX<br>Division of Cardiovascular Devices |
|----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The Merit Hydrophilic Guide Wire is intended to be used in the peripheral<br>vascular system to facilitate the placement of devices during diagnostic<br>and interventional procedures. |
| Device<br>Description | The Merit Hydrophilic Guide Wire consists of a jacketed core wire with a<br>hydrophilic coating applied to the jacket. The wire will be offered in<br>straight and angled configurations in various lengths. |
| Comparison<br>to Predicate | Technological characteristics of the subject Merit Hydrophilic Guide Wire<br>are substantially equivalent to those of the predicate, the Merit<br>Hydrophilic Guide Wire [K120644]. The difference between the devices<br>relates to the guide wire diameter size. The guide wire design remains<br>unchanged. Predicate device #2 [K863138] Terumo Radiofocus®<br>Glidewire® is the predicate for performance testing parameters.<br>Predicate device #3 [K130588] includes clarification of the Intended use<br>statement. |
{2}------------------------------------------------
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
| Safety &<br>Performance<br>Tests | No performance standards have been established under section 514 of<br>the Food, Drug and Cosmetic Act for these devices. A battery of testing<br>was conducted in accordance with protocols based on requirements<br>outlined in guidance's and industry standards and these were shown to<br>meet the acceptance criteria that were determined to demonstrate<br>substantial equivalence.<br><br>Where appropriate, the tests were based on the requirements of the<br>following documents:<br>FDA guidance Coronary and Cerebrovascular Guide Wire<br>Guidance January 1995. ISO 11070: 1998, Sterile Single-Use Intravascular Catheter<br>Introducers. ISO 11135-1: 2007 Sterilization of health care products-Ethylene<br>oxide- Part 1: Requirements for the development, validation and<br>routine control of a sterilization process for medical devices. ASTM F1980-07 Standard Guide for Accelerated Aging of Sterile<br>Barrier Systems for Medical Devices ISO 10993-1: 2009, Biological Evaluation of Medical Devices Part<br>1: Evaluation and Testing within a risk management process, and<br>the FDA Modified ISO 10993 Test Profile FDA Memo G95-1. The Merit Hydrophilic Guide Wire was compared to the predicate<br>device(s) for various performance attributes that support substantial<br>equivalence of the device. The difference in wire diameter between the<br>modified device and the cleared device [K120644] has raised no new<br>issues. Performance testing was based on Predicate device # 2<br>[K863138] Terumo Radiofocus® Glidewire®.<br><br>The following is a list of all significant testing that was successfully<br>completed:<br>Tensile Strength, Torque Strength, Torquability, Tip Flexibility, Coating<br>Adherence/Integrity (including Evaluation using Anatomical Model),<br>Catheter Compatibility (Durability), Surface, Fracture Test, Flex test, Size<br>Designation/ Dimensions and Radiopacity.<br>As all test results were comparable to the predicate devices and the<br>subject Merit Hydrophilic guide wire met the predeterminded acceptance<br>criteria applicable to the safety and efficacy of the device,this has<br>demonstrated the subject device is substantially equivalent to predicate<br>devices. |
|------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Summary of<br>Substantial<br>Equivalence | Based on the Indications for Use, design, safety and performance<br>testing, the subject Merit Medical Hydrophilic Guide Wire meets the<br>requirements that are considered essential for its intended use and is<br>substantively equivalent to the predicate device, the Merit Hydrophilic<br>Guide Wire manufactured by Merit Medical Systems Inc. |
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles a caduceus, a traditional symbol of medicine, but with a more abstract and modern design.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 29, 2013
Merit Medical Systems, Inc. Ms. Martha Folan Senior Regulatory Affairs Specialist Parkmore Business Park West Galway, Ireland
Re: K131710
Trade/Device Name: Merit Hydrophilic Guide Wire Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter guide wire Regulatory Class: II Product Code: DQX Dated: June 27, 2013 Received: July 1, 2013
Dear Ms. Martha Folan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or 10 devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Ms. Folan
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
## Bram D. Zuckerman -S
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
{5}------------------------------------------------
K131710
Merit Medical Systems, Inc. Merit Hydrophilic Guide Wire Special Premarket Notification 510(k)
#### Section 5
#### Indications for Use Statement
510 (k) Number:
Device Name: Merit Hydrophilic Guide Wire
Indications for Use:
The Merit Hydrophilic Guide Wire is intended to be used in the peripheral vascular system to facilitate the placement of devices during diagnostic and interventional procedures.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
# Bram D. Zuckerman -S 2013.07.29 18:10:49 -04'00'
Page 27