K120644 · Merit Medical Systems, Inc. · DQX · Sep 19, 2012 · Cardiovascular
Device Facts
Record ID
K120644
Device Name
MERIT HYDROPHILIC GUIDE WIRE
Applicant
Merit Medical Systems, Inc.
Product Code
DQX · Cardiovascular
Decision Date
Sep 19, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1330
Device Class
Class 2
Intended Use
The Merit Hydrophilic Guide Wire is intended to facilitate the placement of devices during diagnostic and interventional procedures.
Device Story
The Merit Hydrophilic Guide Wire is a catheter guide wire consisting of a jacketed core wire with a hydrophilic coating. It is available in various lengths with straight or angled tip configurations. The device is used by clinicians in diagnostic and interventional settings to facilitate the placement of other medical devices within the vasculature. The hydrophilic coating reduces friction, aiding navigation through vessels. The device is a sterile, single-use tool.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included coating adherence/integrity, catheter compatibility, surface testing, fracture testing, flex testing, size designation, biocompatibility (ISO 10993-1, -5, -7, -18, ASTM F756-08), and accelerated aging (ASTM F1980-07).
Technological Characteristics
Jacketed core wire with hydrophilic coating. Available in straight and angled tip configurations. Sterilized via ethylene oxide (ISO 11135-1). Biocompatibility per ISO 10993 series. Mechanical testing per FDA Coronary and Cerebrovascular Guide Wire Guidance and ISO 11070.
Indications for Use
Indicated for use in patients undergoing diagnostic and interventional procedures to facilitate the placement of medical devices.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
K092303 — MERIT LAUREATE HYDROPHILIC GUIDE WIRE · Merit Medical Systems, Inc. · Oct 27, 2009
K123609 — MERIT HYDROPHILIC GUIDE WIRE · Merit Medical Systems, Inc. · Jan 16, 2013
K131710 — MERIT HYDROPHILIC GUIDE WIRE · Merit Medical Systems, Inc. · Jul 29, 2013
K193653 — Hydrophilic Coating Guide Wire · Suzhou Hengrui Disheng Medical Co., Ltd. · Aug 26, 2020
K141707 — ULTRASKIN HYDROPHILIC GUIDE WIRE · Lepu Medical Technology (Beijing) Co., Ltd. · Jan 21, 2015
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Merit Medical. The logo consists of a stylized "M" symbol on the left, followed by the words "MERIT MEDICAL" in a bold, sans-serif font. The word "MEDICAL" is outlined in white, while the rest of the logo is in black. There is a registered trademark symbol to the right of the word "MEDICAL".
.
Merit Medical Ireland Ltd., KV20644
Parkmore Business Park West, Galway, Ireland.
SEP 1 9 2012
int Tel: +353 (0) 91 703 700 Int Fax: +353 {0} 91 771 888
Web: www.merit.com
## Section 5
### 510(k) Summary
| General<br>Provisions | Submitter Name:<br>Address:<br>Telephone Number:<br>Fax Number:<br>Contact Person:<br>Registration Number: | Merit Medical Systems, Inc.<br>1600 West Merit Parkway,<br>South Jordan, UT 84095<br>(801) 208-4349<br>(801) 253-6967<br>Stephanie Erskine<br>1721504 |
|-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Correspondent Name:<br>Address:<br>Telephone Number:<br>Fax Number:<br>Contact Person:<br>Date of Preparation:<br>Registration Number: | Merit Medical Ireland Ltd.<br>Parkmore Business Park,<br>Galway, Ireland<br>(353) 91 703 761<br>(353) 91 771 888<br>Mark Mullaney<br>March 1, 2012<br>9616662 |
| Subject<br>Device | Trade Name:<br>Common/Usual Name:<br>Classification Name: | Merit Hydrophilic Guide Wire<br>Hydrophilic Guide Wire<br>Catheter Guide Wire |
| Predicate<br>Devices | Trade Name:<br>Classification Name:<br>Premarket Notification:<br>Manufacturer: | Merit Hydrophilic Guide Wire<br>Catheter Guide Wire<br>K092303<br>Merit Medical Systems, Inc. |
| Classification | Class II<br>21 CFR § 870.1330, 74 DQX<br>Division of Cardiovascular Devices | |
| Intended Use | The Merit Hydrophilic Guide Wire is intended to facilitate the placement<br>of devices during diagnostic and interventional procedures. | |
| Device<br>Description | The Merit Hydrophilic Guide Wire consists of a jacketed core wire with a<br>hydrophilic coating applied to the jacket. The wire will be offered with<br>straight and angled tip configurations in various lengths. | |
Directors: F. Lampropoulos (American), R. Perry (American), K. Stanger (American), M. Butter (Irish) Registered in Dublin, Ireland No. 210890. Registered Office: 25/28 North Wall Quay, Dublin 2.
{1}------------------------------------------------
#### Comparison to Predicate ·
Technological characteristics of the subject Merit Hydrophilic Guide Wire are substantially equivalent to those of the predicate, the Merit Hydrophilic Guide Wire {K092303]. The differences between the devices relate to the hydrophilic coating process. The guide wire design remains unchanged.
{2}------------------------------------------------
No performance standards have been established under Section 514 of the Food, Drug and Cosmetic Act for these devices. However, a battery of tests was performed according to protocols based on the requirements of industry standards and guidances and were shown to meet the acceptance criteria that were determined to demonstrate the safety and efficacy of the device.
Where appropriate, the tests were based on the requirements of the following documents:
- . FDA Guidance Coronary and Cerebrovascular Guide Wire Guidance January 1995.
- . ISO 11070:1998, Sterile Single-Use Intravascular Catheter Introducers.
- . ISO 11135-1:2007, Sterilization of health care products - Ethylene oxide - Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices.
- ASTM F756-08 Standard Practice for Assessment of Hemolytic . Properties of Materials
- . ASTM F1980-07 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices.
- ISO 10993-1:2009, Biological Evaluation of Medical Devices Part 1: . Evaluation and Testing within a risk management process, and the FDA Modified ISO 10993 Test Profile FDA Memo G95-1.
- ISO 10993-5:2009 Biological Evaluation of Medical Devices -Part 5: Tests for In-Vitro Cytotoxicity
- . ISO 10993-7:2008 - Biological Evaluation of Medical Devices - Part 7: Ethylene Oxide Sterilization Residuals
- ISO 10993-18:2005 Biological Evaluation of Medical Devices -. Part 18: Chemical Characterization of Materials
The Merit Hydrophilic Guide Wire was compared to the predicate device for various performance attributes that demonstrate the safety or efficacy of the device. Differences in materials between the modified device and the cleared device [K092303] raised no issues of safety and effectiveness. The hydrophilic coating process was modified. The coating solution is made with the identical raw materials, equipment, and processes used for the preparation of the predicate coating solution.
The following is a list of all significant testing that was successfully completed:
- . Coating Adherence/Integrity
- Catheter Compatibility .
- Surface .
- Fracture Test .
- . Flex Test
- Size Designation .
- Biocompatibility .
- Chemical Characterization .
# Safety & ● Performance Tests
{3}------------------------------------------------
# Summary of Substantial Equivalence
Based on the indications for use, design, safety, and performance testing, the subject Merit Hydrophilic Guide Wire meets the requirements that are considered essential for its intended use and is substantially equivalent to the predicate device, the Merit Hydrophilic Guide Wire manufactured by Merit Medical Systems, Inc. Differences between the two devices do not raise any significant issues of safety or effectiveness.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or bird in flight. The symbol is composed of three curved lines that form the body and wings of the bird.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
SEP 19 2012
Merit Medical Systems, Inc. Mr. Mark Mullaney Regulatory Affairs Manager Parkmore Business Park West, Galway, Ireland
Re: K120644
Trade/Device Name: Merit Hydrophilic Guide Wire Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: DQX Dated: September 10, 2012 Received: September 12, 2012
Dear Mr. Mullaney:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Mark Mullaney
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act, or any Federal statutes and regulations administered by other Federal agencies. You must or ury 1 vith all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
CMA. Hellehn
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Merit Medical Systems, Inc. Merit Hydrophilic Guide Wire Premarket Notification 510(k)
#### Section 4
### Indications for Use Statement
K120644
510(k) Number (if known):
Device Name: Merit Hydrophilic Guide Wire
Indications for Use:
The Merit Hydrophilic Guide Wire is intended to facilitate the placement of devices during diagnostic and interventional procedures.
Prescription Use _ × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
. S. Allelu
diovascular Devices
510(k) Number K120644
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.