PhysCade System

K250749 · Physcade, Inc. · DQK · Jun 4, 2025 · Cardiovascular

Device Facts

Record IDK250749
Device NamePhysCade System
ApplicantPhyscade, Inc.
Product CodeDQK · Cardiovascular
Decision DateJun 4, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1425
Device ClassClass 2
AttributesAI/ML, Software as a Medical Device

AI Performance

OutputAcceptanceObservedDev DSTest DS
Earliest site of activationNot specifiedNot specifiedTraining: ~15M electrograms from 174 patients; Tuning: ~5M electrograms from 62 patients; Test: ~5M electrograms from 109 patients.Not specified
Activation timeNot specifiedNot specifiedTraining: ~15M electrograms from 174 patients; Tuning: ~5M electrograms from 62 patients; Test: ~5M electrograms from 109 patients.Not specified

Indications for Use

The PhysCade System is intended for the analysis, display, and storage of cardiac electrophysiological data and maps for analysis by a physician. The clinical significance of utilizing the PhysCade System to help identify areas with intra-cardiac atrial electrograms exhibiting local early activated potentials and other features of interest during atrial arrhythmias has not been established by clinical investigations.

Device Story

PhysCade System is an AI-enabled medical decision support system for cardiac electrophysiology labs. It processes intra-cardiac electrograms (EGMs) and surface ECG signals collected via compatible diagnostic EP catheters during mapping procedures. The system analyzes EGM signals to identify regions of interest, specifically the earliest site of activation (coPILOT output). Supporting outputs include activation times (coMAP), signal voltage, signal quality, and propagation sequences. Operated by trained nurses, technicians, or physicians, the system provides visual cues on a workstation display to assist clinicians in managing arrhythmias. By highlighting areas of interest, the device supports clinical decision-making during EP studies. It is a standalone workstation not connected to other medical equipment; data is imported via USB flash drive from cleared data acquisition systems (Boston Scientific LabSystem Pro, GE MacLab CardioLab, or Abbott Ensite NavX/Precision).

Clinical Evidence

Bench testing only. No clinical data provided. The device underwent rigorous software/firmware verification, cybersecurity validation, design verification, and usability testing. AI/ML accuracy was confirmed through design validation using development, tuning, and test cohorts totaling approximately 25 million electrograms from 345 patients.

Technological Characteristics

System consists of an off-the-shelf computer workstation, display, and proprietary AI software. Compatible with catheters having 0.5-3mm electrode size, 2-10mm inter-electrode spacing, and 8-64 electrodes in a spatial grid. Operates as a standalone system; data imported via USB. Software developed per IEC 62304. AI/ML algorithms used for signal analysis.

Indications for Use

Indicated for adult patients undergoing cardiac electrophysiology mapping procedures for the management of heart rhythm disorders (arrhythmias).

Regulatory Classification

Identification

A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION June 4, 2025 PhysCade, Inc. Grace Li Head of QA/RA 2100 Geng Road, Suite 210 Palo Alto, California 94303 Re: K250749 Trade/Device Name: PhysCade System Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II Product Code: DQK Dated: March 12, 2025 Received: March 12, 2025 Dear Grace Li: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K250749 - Grace Li Page 2 (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. {2} K250749 - Grace Li Page 3 For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Aneesh S. Deoras -S Aneesh Deoras Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. Submission Number (if known) K250749 Device Name PhysCade System Indications for Use (Describe) The PhysCade System is intended for the analysis, display, and storage of cardiac electrophysiological data and maps for analysis by a physician. The clinical significance of utilizing the PhysCade System to help identify areas with intra-cardiac atrial electrograms exhibiting local early activated potentials and other features of interest during atrial arrhythmias has not been established by clinical investigations. Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov “An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.” {4} K250749 PhysCade System 510(k) Summary Submitter's information PhysCade, Inc. 2100 Geng Road, Suite 210 Palo Alto, CA 94303 United States Contact: Raphael Michel President & CEO PhysCade, Inc. 2100 Geng Road, Suite 210 Palo Alto, CA 94303 Classification The classification for the new device is listed below. | 21 CFR Reference | Product Code | Class | Generic Device Name | Classification Description | | --- | --- | --- | --- | --- | | §870.1425 | DQK | II | Computer, Diagnostic, Programmable | Programmable diagnostic computer | New device The new device's indications for use statement is listed in the table below. | Device Name | Indications for Use | | --- | --- | | PhysCade System | The PhysCade System is intended for the analysis, display, and storage of cardiac electrophysiological data and maps for analysis by a physician. The clinical significance of utilizing the PhysCade System to help identify areas with intra-cardiac atrial electrograms exhibiting local early activated potentials and other features of interest during atrial arrhythmias has not been established by clinical investigations. | Predicate device The predicate device for the PhysCade System is shown in the table below. | K Number | Product Code | Predicate Device Name | Indications for Use | | --- | --- | --- | --- | | K201298 | DQK | Volta Medical VX1 | The VX1 assists operators in the real-time manual annotation of 3D anatomical and electrical maps of human atria for the presence of multipolar intra-cardiac atrial electrograms exhibiting spatiotemporal dispersion during atrial fibrillation or atrial tachycardia. The clinical significance of utilizing the VX1 software to help identify areas with intra-cardiac atrial electrograms exhibiting spatiotemporal dispersion for catheter ablation of atrial arrhythmias, such as atrial fibrillation, has not been established by clinical investigations. | {5} K250749 PhysCade System 510(k) Summary ## Device description The PhysCade™ System (PhysCade) is an artificial intelligence (AI) enabled device intended to assist clinicians in their management of patients with heart rhythm disorders (arrhythmias). PhysCade is a medical decision support system which post-processes electrograms (EGMs) collected inside the heart during electrophysiology (EP) mapping procedures using compatible diagnostic EP catheters. The PhysCade software has advanced algorithms that analyze the collected EGMs to provide information on regions of interest in the heart that may be useful to the clinician to support clinical decisions together with other available patient-related information. PhysCade provides specialized analyses of data from a compatible multipolar catheter. The primary output (coPILOT) indicates the predominant earliest site of activation relative to the catheter electrode array. Supporting outputs include determining activation times of successive beats in the EGM signal on each electrode (coMAP), voltage at each electrode, signal quality, and sequences of propagation over multiple beats of the arrhythmia on the catheter. The PhysCade System consists of a computer workstation, display, and custom software and is not connected to other devices or medical equipment. PhysCade is intended to be used in the cardiac electrophysiology lab, outside the sterile field. PhysCade is intended to be operated by nurses, technicians, physicians, or other personnel who are trained and approved by each treating facility to work in the electrophysiology lab. PhysCade uses AI algorithms that were developed and tested using datasets with the following characteristics: {6} K250749 PhysCade System 510(k) Summary | Features | Development Cohort | | Test Cohort | | --- | --- | --- | --- | | | Training | Tuning | | | Number of electrograms | ~15M | ~5M | ~5M | | Number of patients | 174 | 62 | 109 | | Age | 67.91 ± 8.80 | 68.97 ± 8.26 | 67.10 ± 7.66 | | AF Type | | | | | Non-Paroxysmal, n (%) | 124 (71.26) | 45 (72.58) | 82 (75.23) | | Paroxysmal, n (%) | 50 (28.74) | 17 (27.42) | 27 (24.77) | | Female, n (%) | 31 (17.82) | 6 (9.68) | 29 (26.61) | | Racial category (non-white), n (%)* | 36 (20.93) | 16 (25.81) | 20 (20.20) | | BMI, kg/m2 | 29.45 ± 5.72 | 29.45 ± 6.46 | 29.67 ± 6.67 | | CHADS2VASC | 2.57 ± 1.57 | 2.65 ± 1.36 | 2.70 ± 1.66 | | LA Size Enlarged, n (%)* | 90 (55.21) | 32 (55.17) | 54 (58.70) | | Coronary disease, n (%) | 35 (20.11) | 19 (30.65) | 27 (24.77) | | Myocardial Infarct, n (%) | 10 (5.75) | 6 (9.68) | 6 (5.50) | | Hypertension, n (%) | 128 (73.56) | 47 (75.81) | 70 (64.22) | | Diabetes Mellitus, n (%) | 32 (18.39) | 9 (14.52) | 21 (19.27) | | Prior Stroke/TIA, n (%) | 16 (9.20) | 3 (4.84) | 8 (7.34) | | All data complete except for *minor missingness in the categorical variables of racial category (disclosed in N=172/174 in training, N=62/62 in tuning and 99/109 in test cohorts), and left atrial enlargement (variable recorded in N=163/174 in training, N=58/62 in tuning and 92/109 in test cohorts). | | | | Characteristics - comparison with predicate device The table below lists the characteristics for both the new and predicate devices. {7} K250749 PhysCade System 510(k) Summary | ATTRIBUTE | SUBJECT DEVICE PhysCade System | PREDICATE DEVICE Volta VX1 K201298 | COMPARISON | | --- | --- | --- | --- | | Indications for use | The PhysCade System is intended for the analysis, display, and storage of cardiac electrophysiological data and maps for analysis by a physician. The clinical significance of utilizing the PhysCade System to help identify areas with intra-cardiac atrial electrograms exhibiting local early activated potentials and other features of interest during atrial arrhythmias has not been established by clinical investigations. | The VX1 assists operators in the real-time manual annotation of 3D anatomical and electrical maps of human atria for the presence of multipolar intra-cardiac atrial electrograms exhibiting spatiotemporal dispersion during atrial fibrillation or atrial tachycardia. The clinical significance of utilizing the VX1 software to help identify areas with intra-cardiac atrial electrograms exhibiting spatiotemporal dispersion for catheter ablation of atrial arrhythmias, such as atrial fibrillation, has not been established by clinical investigations. | The PhysCade System has the same intended use and similar indications for use as the predicate device. Both devices assist in review and analysis of electrical maps of the human atria. The predicate device indications for use, while broader, identifies electrical activity in areas of interest for catheter ablation of atrial arrhythmias. | | Regulation | 21 CFR 870.1425 | 21 CFR 870.1425 | Same | | FDA Product code | DQK | DQK | Same | | Classification | Programmable diagnostic computer, 21 CFR §870.1425 | Programmable diagnostic computer, 21 CFR §870.1425 | Same | | Population | Adults | Adults | Same | | System Type | Signal processing based atrial mapping system | Signal processing based atrial mapping system | Same | | Display(s) | Color monitor | Color monitor | Same | | Control | Standard keyboard / mouse | Standard keyboard / mouse | Same | Page 4 {8} K250749 PhysCade System 510(k) Summary | ATTRIBUTE | SUBJECT DEVICE PhysCade System | PREDICATE DEVICE Volta VX1 K201298 | COMPARISON | | --- | --- | --- | --- | | Inputs Required | Intra-cardiac atrial electrograms that are recorded with catheters, and surface ECG signals amplified and input to the PhysCade System on which they are displayed. | Intra-cardiac atrial electrograms that are recorded with catheters, amplified and redirected to an acquisition system on which they are displayed. | The subject and predicate devices analyze intra-cardiac atrial electrograms. The PhysCade System also inputs the ECG signal to assist in the analysis of intracardiac electrograms, which does not alter the intended use. | | Connections | A USB flash drive is used to export data from compatible analog-to- digital converter systems and import data into the PhysCade System. | A connection cable is used to connect the data acquisition system with an Advantech PCI-1713U analog-to-digital converter, which transmits the acquired information to a nearby computer that hosts the VX1 software. | Both devices digitize intra-cardiac data for input to their analysis systems. While the subject and predicate devices use different brands of analog-to- digital converters, and the subject device uses a USB drive for export while the predicate device uses a connector cable, the transfer of data is equivalent. | | Principal Mapping Output | Displays adjudications as visual cues after analyzing intra-cardiac atrial electrograms using signal processing techniques. | Displays adjudications as visual cues after analyzing intra-cardiac atrial electrograms in real-time using signal processing techniques. | The predicate device and subject device provide adjudications during a procedure. While the predicate device inputs data from a connector cable, the subject device uploads data from a USB flash drive. Both systems provide output during an EP study to support clinical decision-making. | {9} K250749 PhysCade System 510(k) Summary | ATTRIBUTE | SUBJECT DEVICE PhysCade System | PREDICATE DEVICE Volta VX1 K201298 | COMPARISON | | --- | --- | --- | --- | | Map Types Generated | The operator is provided with a display of the multipolar electrode map indicating locations where wave direction is earliest, and markings of electrogram activation time, by the PhysCade System during recordings of atrial fibrillation. | Real-time Dispersed Electrogram (DE) subtype of multipolar electrogram map- The operator is provided with display of the electrode locations where dispersed or non-dispersed electrograms have been recorded during atrial fibrillation or atrial tachycardia. | Both the subject and predicate devices generate multipolar maps of their analysis of intra-cardiac data. The predicate device focuses on dispersed electrograms, while the subject device focuses on direction to the earliest site. Both devices produce a multipolar map of electrical locations of interest during atrial fibrillation. | | Compatible Catheters | Any compatible catheter meeting the requirements listed below: **Compatibility requirements:** **Mapping catheter:** - Electrode size: 0.5 mm to 3 mm - Inter-electrode spacing: 2 to 10mm (edge-to-edge) - Number of selected electrodes: between 8 and 64 - Electrode configuration: a spatial grid of at least 2 electrodes along each side (e.g., 4X4, 8X8) | Any compatible catheter meeting the requirements listed below: **Compatibility Requirements:** **Mapping catheter:** - Electrode size: 1 mm - Inter-electrode Spacing: 2 to 3 mm - Number of selected dipoles: 10 **Coronary sinus catheter:** - Electrode size: 1 mm - Interelectrode Spacing: 2 to 3 mm - Number of selected dipoles: 5 | Both devices are compatible with catheters that are defined in their labeling which have been validated with their corresponding algorithms. The predicate device can also be used with Coronary Sinus catheters but this is not required for the subject device analysis. | Page 6 {10} K250749 PhysCade System 510(k) Summary | ATTRIBUTE | SUBJECT DEVICE PhysCade System | PREDICATE DEVICE Volta VX1 K201298 | COMPARISON | | --- | --- | --- | --- | | Data Acquisition System | The PhysCade System works with three data acquisition systems: - LabSystem Pro Acquisition (Boston Scientific), - MacLab CardioLab Acquisition System (General Electric), - Ensite NavX/Precision System (Abbott) | The VX1 works with two data acquisition systems: - LabSystem Pro Acquisition (Boston Scientific), - MacLab CardioLab Acquisition System (General Electric) | The subject device is compatible with the same two data acquisition systems as the predicate device and adds compatibility with a third data acquisition system - Ensite. These 510(k) cleared data acquisition systems all produce digitized and amplified signals that are numerically similar. | | Hardware Design and Materials | Off-the-shelf computer and monitor, connection cable, acquisition system, proprietary software algorithm. | Off-the-shelf analog / digital converter, computer and monitor, connection cable, acquisition system, proprietary software algorithm. | Both devices use off-the-shelf hardware with a proprietary software algorithm. | ## Performance testing The PhysCade System was subjected to extensive non-clinical testing including rigorous verification and validation testing. The testing consisted of: - Software/Firmware and cybersecurity verification and validation testing confirmed that the subject device meets the design input requirements and specifications - Design Verification testing confirmed that the subject device hardware meets performance specifications. - Design Validation confirmed that the AI/ML system is accurate for its intended use. - Usability testing demonstrated that the subject device meets the user needs specifications. The PhysCade System was developed and tested in accordance with IEC 62304 Edition 1.1 2015-06. ## Guidance documents referenced for testing The following guidance documents were referenced for testing: - Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions Page 7 {11} K250749 PhysCade System 510(k) Summary - Content of Premarket Submissions for Management of Cybersecurity in Medical Device - Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices - Off-The-Shelf Software Use in Medical Devices - Applying Human Factors and Usability Engineering to Medical Devices - Good Machine Learning Practice for Medical Device Development: Guiding Principles Conclusion A comparison has been conducted of the intended use/indications for use, performance characteristics, design features, technological/functional characteristics, and labeling for the PhysCade System. Intended to provide the physician with additional information that supports standard clinical practice, no new or different questions of safety or effectiveness are raised. The PhysCade System has the same intended use as the predicate device. Testing results demonstrate that minor differences in the technological characteristics do not raise questions of safety and effectiveness. Thus, the PhysCade System is substantially equivalent to the Volta VX1 predicate device (K201298). Page 8
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...