Vivio System

K183710 · Avicena, LLC · DQD · Oct 4, 2019 · Cardiovascular

Device Facts

Record IDK183710
Device NameVivio System
ApplicantAvicena, LLC
Product CodeDQD · Cardiovascular
Decision DateOct 4, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1875
Device ClassClass 2

Indications for Use

The Vivio System is intended to be used as part of a physical assessment of a patient by healthcare professionals for diagnostic decision support. It may be used for the detection and amplification of sounds from the heart and arteries with the use of selective frequencies. It can be used on any person undergoing a physical assessment.

Device Story

The Vivio System is a digital electronic stethoscope used by healthcare professionals in clinical settings (hospitals, offices) to assist in physical assessments. The device features a membrane placed on the skin over the carotid artery; vibrations from the arterial pulse and heart sounds displace the membrane. An infrared light photo reflector detects membrane movement and generates a current; internal electronics condition this signal and transmit it to a tablet running the Avicena App. The system displays heart sounds (audio/visual) and heart rate, and provides an arterial pulse waveform visualization to confirm proper device placement. The device is powered by an internal rechargeable battery and does not function while connected to mains power. Clinicians use the output to support diagnostic decision-making. The device is self-contained with no user-replaceable parts.

Clinical Evidence

No clinical or animal performance testing was conducted. Evidence consists of bench testing, including performance, user, and usability testing. Compliance with standards including AAMI ES 60601-1, IEC 60601-1-2, ISO 10993-1, and IEC 62304 was demonstrated.

Technological Characteristics

Digital electronic stethoscope; infrared light photo reflector sensing principle; internal rechargeable battery; no user-replaceable diaphragm. Complies with AAMI ES 60601-1 (safety), IEC 60601-1-2 (EMC), and ISO 10993-1 (biocompatibility). Software developed per IEC 62304.

Indications for Use

Indicated for low frequency (<200 Hz) or Bell Mode operation to acquire and transmit heart sound data to a tablet/UI for heart rate calculation and audio/visual playback of heart sounds. Contraindicated for patients with history of carotid sinus hypersensitivity, carotid artery stenosis, or open skin lesions at the examination site.

Regulatory Classification

Identification

Manual stethoscope: A mechanical device used to project the sounds associated with the heart, arteries, and veins and other internal organs. Electronic stethoscope: An electrically amplified device used to project the sounds associated with the heart, arteries, and veins and other internal organs.

Special Controls

*Classification.* Class II (special controls). The device, when it is a lung sound monitor, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. October 4, 2019 Avicena, LLC % James West Principal Consultant West Device Innovations, LLC 23412 West 46th Street Shawnee, Kansas 66226 Re: K183710 Trade/Device Name: Vivio System Regulation Number: 21 CFR 870.1875 Regulation Name: Stethoscope Regulatory Class: Class II Product Code: DQD Dated: August 22, 2019 Received: August 27, 2019 Dear James West: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for Stephen Browning Assistant Director DHT2A: Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K183710 Device Name Vivio System Indications for Use (Describe) The Vivio System is indicated for low frequency (<200 Hz) or Bell Mode operation to acquire and transmit heart sound data to the tablet/UI (with Avicena App) for calculation of heart rate and display and playback of heart sounds (audio and visual). | Type of Use (Select one or both, as applicable) | | |-----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------| | <div> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <input type="checkbox"/> Over The Counter Use (21 CFR 801 Subpart </div> | |X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the word "AVICENA" in all caps, with a logo to the left of the word. The logo is a stylized drawing of a heart with an oval inside. There are dots around the outside of the heart and some squiggly lines inside the oval. # Traditional 510(k) Summary for K183710 | Summary Date: | 02 Oct. 2019 | |----------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Applicant Name: | Avicena, LLC, 117 East Colorado Blvd. Suite 510 Pasadena CA<br>91105, Ph: 626.344.9665, Email: info@avicenaheart.com<br>Establishment Registration Number: FDA Registration is pending. | | Submission<br>Correspondent: | On behalf of Avicena LLC., the following consultant is assigned the<br>responsibility of submission correspondence:<br>Jim West, Principal Consultant at West Device Innovations, LLC<br>23412 West 46'th Street, Shawnee, KS 66226, Ph: 913.991.6288,<br>Email: jwest@westdeviceinnovations.com | | Trade Name: | Vivio System | | Common Name: | Digital Electronic Stethoscope | | Description: | The Vivio Unit is an electronic stethoscope containing a membrane,<br>placed on the skin over the carotid artery, that is displaced due to<br>vibrations produced by the arterial pulse waveform and heart<br>sounds associated with the aortic valve opening and closing. An<br>infrared light photo reflector transmits and detects light to/from the<br>membrane and outputs a current relative to the detected light (or<br>membrane movement). Sensor electronics inside the Vivio Unit<br>generate a current to operate the sensor, read the current produced<br>by the sensor and output a voltage relative to the current. Filter<br>electronics condition the sensor signal, process the data and<br>communicate the data to the tablet with the Avicena App. The Vivio<br>Unit is powered by an internal, rechargeable battery. | | Classification<br>Regulation, Class &<br>Product Code & Panel: | 21 CFR 870.1875(b)<br>Class II<br>Product Code: DQD<br>Panel: Cardiovascular | | Avicena | | K183710 | |----------------------------------|--|----------| | Vivio Device - 510(k) Submission | | Page 1/5 | {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the word "AVICENA" in all capital letters next to a logo. The logo is a drawing of a cell with some dots and lines inside and outside of it. The word "AVICENA" is in a simple, sans-serif font. ## Intended Use of Device | Product Requirements Document - Avicena Vivio Device | | |------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indications for Use: | The Vivio System is indicated for low frequency (<200 Hz) or Bell<br>Mode operation to acquire and transmit heart sound data to the<br>tablet/UI (with Avicena App) for calculation of heart rate and display<br>and playback of heart sounds (audio and visual). | | Intended Use: | The Vivio System is intended to be used as part of a physical<br>assessment of a patient by healthcare professionals for diagnostic<br>decision support. It may be used for the detection and amplification<br>of sounds from the heart and arteries with the use of selective<br>frequencies. It can be used on any person undergoing a physical<br>assessment. | | Intended User: | The Vivio System is intended to be used by Healthcare Providers. | | Intended Use<br>Environment: | The Vivio System is intended to be used in a clinical setting such as<br>a hospital, physician's office, or other patient care facility. | | Targeted Patient<br>Population: | It can be used on any person undergoing a physical assessment.. | | Contraindications: | Subjects with a hypersensitive response to even slight pressure<br>over the carotid pulse may faint when the carotid is palpated<br>(carotid sinus hypersensitivity). For this reason, the following are<br>contraindications for the use of the Vivio System:<br>• Patients with a history of carotid sinus hypersensitivity<br>(fainting in response to touching or positioning the neck)<br>• Patients with a known history of carotid artery stenosis<br>• Open skin lesions at the site of examination | | Predicate Device(s): | 510(k) Number: K083903<br>Manufacturer: 3M Company<br>Trade Name: 3M Littmann Electronic Stethoscope, Model 3200<br>Product Code: DQD<br>Classification: 870.1875 | |----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Reference Device(s): | 510(K) Number: K122129<br>Manufacturer: AtCor Medical | | Avicena | | | |----------------------------------|--|---------------------| | Vivio Device - 510(k) Submission | | K183710<br>Page 2/5 | {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows a logo with the word "AVICENA" in bold, sans-serif font. To the left of the word is a stylized graphic that resembles a cell or seed with an outer shell and inner components. The graphic is outlined in black and has a hand-drawn quality, with small dots and lines adding texture and detail. | Trade Name: SphygmoCor XCEL<br>Product Code: DSK: DXN<br>Classification: Computer, Blood-Pressure; System, Measurement,<br>Blood-Pressure, Non-Invasive | |---------------------------------------------------------------------------------------------------------------------------------------------------------| | | ## Comparison to Predicate The Vivio System combines the following features into one product that consists of a physical stethoscope for collection of heart sounds, and software in the form of an app on a tablet to record, playback and visualize the sounds recorded. The Vivio System, Model 1.0 under review and the predicate have similar indications for use. The devices are substantially equivalent. Table 1 below is a summary of the differences between the devices and a commentary regarding the reasons the differences do not introduce additional safety or efficacy concerns. The proposed device and the predicate device have the same/equivalent intended use, intended user, intended use environment and targeted patient population. The device under consideration uses for a predicate device, the 3M Littmann Model 3200 Electronic Stethoscope (cleared under 510(k) K083903) used in conjunction with the Zargis StethAssist phonocardiogram software. Additionally a reference device is included, the AtCor Medical SphygmoCor XCEL (cleared under 510(k) K122129). It should be noted that phonocardiograms are 510(k) exempt devices per 870.2390. Essentially, the use of the Littmann Model 3200 with StethAssist software is related to heart sounds, heart sound waveforms, and heart rate visualization, and SphygmoCor for arterial pulse waveform visualization which can be used by the clinician as an assessment to indicate adequate system application/placement and is not presented for clinical interpretation. The Vivio product differs from the predicate in the areas of: - signal export capability ● - replaceable diaphragm ● - latex indication label - category AP Equipment classification labeling - sharp edges caution statement - electric warning labeling ● - visualization of the arterial pulse waveform for assessing adequate device placement ● While the differences listed affect the design and form of the product they do not introduce any safety, performance or efficacy concerns. | Avicena | | | |----------------------------------|--|---------------------| | Vivio Device - 510(k) Submission | | K183710<br>Page 3/5 | {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the word "AVICENA" in a simple, sans-serif font. To the left of the word is a stylized graphic that appears to be a heart shape with some internal markings. The graphic is drawn with a rough, hand-drawn quality, giving it a slightly imperfect and organic feel. # Avicena Vivio Device Traditional 510(k) Summary SECTION 001 | | | | | Table 1 Differences between Vivio System and Predicate Efficacy/Safety Concerns | |----------------------------|---------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------| | | | | | | | | | | | | | | Vivio System<br>(Subject Device)<br>(K183710) | 3M Model 3200 with<br>Zargis StethAssist<br>(Predicate Device)<br>(K083903) | Comments on Efficacy/Safety | | | Information | | | | | | Signal Export | No export functionality. | Can export .wav and .zsa<br>files and print a report. | The Vivio System does not allow<br>for long term storage, and as<br>such has no mechanism to export<br>data. This has no impact on<br>safety or efficacy. | | | Arterial Pulse<br>Waveform | Displays on tablet. | No. | Used as an assessment to<br>indicate adequate system<br>application/placement and is<br>not presented for clinical<br>interpretation. | | | Design | | | | | | Replaceable<br>diaphragm | No, the system is<br>self-contained and not<br>modifiable by the<br>users. | Yes, the diaphragm can<br>be replaced | It is not possible for a user to<br>modify the Vivio Device. While<br>this is a difference between the<br>predicate, it offers no risk to<br>safety or efficacy. | | | Labeling | | | | | | No latex<br>indication | No | Yes | Our product has no latex in its<br>packaging or construction, but we<br>have not added this declaration<br>to our labels. | | | Arterial Pulse<br>Waveform | Yes | No | The following is included in the<br>Instructions for Use: The arterial<br>pulse waveform is used as an<br>assessment to indicate adequate<br>system application/placement<br>and is not presented for clinical<br>interpretation. | | | Category AP<br>Equipment | No | Yes | The Vivio System has not been<br>tested for flammable anesthetic<br>mixture with air. It is not intended<br>for use in a surgical environment. | | | Electric shock<br>warning | Usage is not allowed<br>when device is<br>plugged into wall<br>power. The Vivio | To reduce the risk<br>associated with an<br>electrical shock do not<br>use the stethoscope on | Since the Vivio Unit does not<br>have a removable diaphragm,<br>there is not an equivalent<br>warning. The device additionally | | | Avicena | | K183710 | |----------------------------------|--|----------| | Vivio Device - 510(k) Submission | | Page 4/5 | {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the logo for Avicena. The logo consists of a stylized heart shape with an oval inside, containing some text. To the right of the heart shape is the word "AVICENA" in all capital letters. ## Avicena Vivio Device Traditional 510(k) Summary | device will not pair with<br>the tablet when<br>plugged into mains<br>power. | patients without the<br>stethoscope's diaphragm<br>cover in place. | will not function while connected<br>to mains power as an additional<br>safety measure. | |------------------------------------------------------------------------------|--------------------------------------------------------------------|-----------------------------------------------------------------------------------------| |------------------------------------------------------------------------------|--------------------------------------------------------------------|-----------------------------------------------------------------------------------------| #### Summary of Clinical and Non-Clinical Testing This submission does not include animal or clinical performance testing. This submission includes data from bench testing to evaluate the performance of the Avicena Vivio System - Electronic Stethoscope. Testing involved performance and bench testing as well as user and usability testing. Although there are no specific performance standards for this device, the device was tested for compliance with voluntary standards as outlined in the following table. Testing to the following standards was conducted per the standards requirements: - CFR 21CFR820: Part 820 QUALITY SYSTEM REGULATION, ● - IEC 62304:2006 + A1:2015 Medical device software Software life cycle processes, ● - ISO 14971:2012 Medical devices Application of risk management to medical devices, ● - ASTM D4169-16 Performance Testing of Shipping Containers and Systems, ● - . AAMI ES 60601-1:2005 + A1:2012 + A2:2010 +C1:2009 - Medical electrical equipment -Part 1: General requirements for basic safety and essential performance - . IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance – Collateral standard: Electromagnetic compatibility - Requirements and tests - ISO 10993-1:2018 Biological evaluation of medical devices Part 1: Evaluation and ● testing within a risk management process and U.S. Food and Drug Administration (FDA) Guidance Document for the Use of ISO 10993-1 (issued 06/16/2016) - . IEC 60601-2-27: 2011/(R)2016 Medical electrical equipment - Part 2-27 Particular requirements for the basic safety and essential performance of electrocardiographic monitoring equipment ## Conclusions Avicena believes the proposed Vivio System and its predicate the 3M Littmann Model 3200 Electronic Stethoscope are substantially equivalent in their intended users, intended use environment and indications for use. They share similar design and technology output characteristics as well as recording the same biological functions. The differences that exist between the devices do not affect the relative safety and/or effectiveness. | Avicena | | |----------------------------------|---------------------| | Vivio Device - 510(k) Submission | K183710<br>Page 5/5 |
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...