The ProTron Technologies, Electronic Stethoscope, Model Stethotron is intended for medical diagnostic purpose only. Stethotron has two modes of operation, amplification with filtering (Mode 1) and amplification only (Mode 2). Mode 1 filters out sounds below 350 Hz and above 1,000 Hz and used to listen specifically for vascular pathology. Mode 2 is used to listen for heart, lung, arteries, and other body sounds from 20 Hz to 20,000 Hz during routine physical examinations. Stethotron can be used on adult or paediatric patients undergoing physical examination.
Device Story
Battery-powered electronic stethoscope; captures body sounds (heart, lung, arteries) via chestpiece-integrated microphone. Two user-selectable modes: Mode 1 (350-1,000 Hz) for vascular pathology; Mode 2 (20-20,000 Hz) for general auscultation. Device includes volume control, mode selection, and ON/OFF switch. Operated by clinicians during routine physical exams. Output delivered via eartips to clinician; aids in diagnostic decision-making by amplifying and filtering acoustic signals. Benefits include enhanced sound detection for diagnostic assessment.
Clinical Evidence
Bench testing only. Performance verified using a computerized audio analysis system with a frequency oscillator (20-10,000 Hz) and calibrated microphone to measure response and gain in both filtered and non-filtered modes. Biocompatibility evaluated per ISO 10993-Part 1 for limited skin contact.
Technological Characteristics
Battery-powered (2x AAA alkaline) electronic stethoscope. Plastic casing with integrated microphone, volume control, and mode selection switch. Frequency response: Mode 1 (350-1,000 Hz), Mode 2 (20-20,000 Hz). Biocompatible materials per ISO 10993-Part 1. Analog signal processing.
Indications for Use
Indicated for adult and pediatric patients undergoing physical examination for the auscultation of heart, lung, arterial, and other body sounds, including vascular pathology assessment.
Regulatory Classification
Identification
Manual stethoscope: A mechanical device used to project the sounds associated with the heart, arteries, and veins and other internal organs. Electronic stethoscope: An electrically amplified device used to project the sounds associated with the heart, arteries, and veins and other internal organs.
Special Controls
*Classification.* Class II (special controls). The device, when it is a lung sound monitor, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
Predicate Devices
3M Littman Electronic Stethoscope, Model 2000 (K961848)
Submission Summary (Full Text)
{0}------------------------------------------------
K082771
p1/3
Page | 13
# 5. 510(k) Summary
[As required by 21 CFR 807.92]
FEB 2 7 2009
| Submitted by | ProTron Technologies, LLC<br>141 Oakdene Ave<br>Leonia, NJ 07605<br>Phone 201.297.7377 |
|---------------------|---------------------------------------------------------------------------------------------|
| Contact person | Leon M. Dondysh |
| Date Prepared | December 20, 2008 |
| Device trade name | ProTron Technologies, LLC, Electronic Stethoscope Model:<br>Stethotron |
| Common name | Electronic Stethoscope |
| Classification name | Stethoscope;<br>21 CFR Sec. 870.1875(b) Electronic stethoscope, Product code DQ<br>Class II |
#### Intended Use
The ProTron Technologies, Electronic Stethoscope, Model Stethotron is intended for medical diagnostic purpose only. Stethotron has two modes of operation, amplification with filtering (Mode 1) and amplification only (Mode 2). Mode 1 filters out sounds below 350 Hz and above 1,000 Hz and used to listen specifically for vascular pathology. Mode 2 is used to listen for heart, lung, arteries, and other body sounds from 20 Hz to 20,000 Hz during routine physical examinations. Stethotron can be used on adult or paediatric patients undergoing physical examination.
# Predicate Device to which Substantial Equivalence is Claimed
The ProTron Technologies, LLC Electronic Stethoscope, Model Stethotron is substantial equivalent to the 3M Littman Electronic Stethoscope, Model 2000 cleared for market under Premarket Notification k961848.
{1}------------------------------------------------
K082771
p²/3
Summary of how the differences in technological characteristic compare to the predicate device [807.92(a){6)]:
| ProTron Technologies,<br>LLC Electronic<br>Stethoscope, Model<br>Stethotron | 3M Littman Electronic<br>Stethoscope, Model 2000 | Discussion |
|---------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Two AAA alkaline batteries<br>(maximum 1.5 volts) | One AAA alkaline battery<br>(Maximum 3 volts) | A 1.5 volt increase in power<br>does not materially effect<br>the use or safety of the<br>device |
| Two modes of operation,<br>Mode 1 amplifies the range<br>from 350 Hz to 1,000 Hz;<br>Mode 2 amplifies the full<br>range, 20 Hz to 20,000 Hz | Three frequency response<br>modes to choose from: Bell<br>(20-200 Hz), Diaphragm<br>(100-500 Hz) and Extended<br>Range (20-1,000 Hz). | Sounds from 20-1,000 Hz<br>can be amplified with both<br>devices. The collection of<br>sounds in Mode 2 from 1-50<br>Hz does not significantly<br>alter the effectiveness of<br>the device. |
| No automatic shutoff | Automatic shutoff | Automatic shutoff only<br>serves to prolong battery<br>life and does not<br>significantly alter the<br>effectiveness or safety of<br>the device. |
#### Description of Device
The ProTron Technologies, LLC Electronic Stethoscope, Model Stethotron is a battery The From To Fron Tronstic device with two user selectable modes of operation for powerod analog alagrious aring sounds of the heart, lungs, arteries, and internal detecting, antiplitying and filters out sounds below 350 Hz and above organs: Mode 2 provides amplification only, 20 Hz to 20,000 Hz. Stethotron has 1,000 Hz. Mode 2 provided amplitudes and non-filtrated sounds enabling the user adjust the volume controls on both mirated anastpiece of the device is sized for use with adult or sound to the level desired. The endeaphragm design incorporating a microphone. All pediatic patients and to a bell and trend of electronics casing positioned below the bifurcation election are contained in a plactic sto the headset that terminates at the eartips. The of the stethoscope yoke that oblinests to witch, volume control, and mode of operation electionic casing includes the ON/Offication and filtering electronics, a selection control on the lace. It somethis and is powered by two AAA alkaline batteries.
December 20, 2008 ProTron Technologies, LLC Electronic Stethoscope, Model Stethotron Submission to FDA
{2}------------------------------------------------
K082771
P3/3
# Test Data
Test data includes electrical testing demonstrating the The ProTron Technologies, LLC Electronic Stethoscope, Model Stethotron to be safe and effective under label conditions.
Non-clinical Testing Brief Description:
Testing to determine response and gain are equivalent to the predicate electronic stethoscope was preformed with a computerized audio analysis system sweeps a frequency oscillator from 20 to 10,000 Hz. This frequency sweep is presented to a transdicer on which the stethoscope is placed. The computer measures the response from the stethoscope with a small calibrated microphone placed in the eartip. The signal from the microphone is fed back to the computer where it is recorded, stored, and can be printed. The test was done in each of the two amplification modes, filtered and nonfiltered.
### Biocompatability
All components of the Electronic Stethoscope, Model "Stethotron" have been reviewed for biocompatibility. With respect to ISO 10993-Part 1 Biological Evaluation of Medical Devices for limited (<24 hour) skin contact for both patient and/or health care professional exposure. Each component with potential skin contact with either the user profononal oxpeed or for possible health concerns. ProTron Technologies Company concludes that all of the components of the Electronic Stethoscope, Model "Stethotron" that make skin contact would have no potential for adverse health concern, all materials that make only to has this specified role that is equivalent to this device.
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FFR 2 7 2009
ProTron Technologies, LLC c/o Mr. Leon M. Dondysh 141 Oakdene Avenue Leonia, NJ 07605
Re: K082771
Trade Name: Electronic Stethoscope Model Stethotron Regulation Number: 21 CFR 870.1875 Regulation Name: Stethoscope Regulatory Class: Class II (two) Product Code: DQD . Dated: January 9, 2009 Received: January 13, 2009
Dear Mr. Dondysh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
### Page 2 - Mr. Leon M. Dondysh
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark Teller for
D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Page | 12
# 4. Indications for Use Statement
Indications for Use
510(k) Number (if known): K082771
Device Name: ProTron Technologies, LLC Electronic Stethoscope, Model Stethotron
Indications for Use:
The ProTron Technologies, Electronic Stethoscope, Model Stethotron is intended for medical diagnostic purpose only. Stethotron has two modes of operation, amplification with filtering (Mode 1) and amplification only (Mode 2). Mode 1 filters out sounds below 350 Hz and above 1,000 Hz and used to listen specifically for vascular pathology. Mode 2 is used to listen for heart, lung, arteries, and other body sounds from 20 Hz to 20,000 Hz during routine physical examinations. Stethotron can be used on adulf or paediatric patients undergoing physical examination.
| 1 "<br>th<br>4 11 3 8.2" | |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 182591 .<br>તું<br>- 1854 | .<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>-14 |
ANDIOR
Over-The-Counter Use (21 CFR 801 Subpert C)
(Please Do NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | |
|------------------------------------|---------|
| Division of Cardiovascular Devices | |
| 510(k) Number | K082771 |
Page 1 ofPage 4-1 December 20, 2008 Pro Tron Technologies, LLC Electronic Stethoscope, Model Stethotron Submission to FDA
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.