QFG · Alternate Controller Enabled Insulin Infusion Pump
Clinical Chemistry · 21 CFR 880.5730 · Class 2
Overview
| Product Code | QFG |
|---|---|
| Device Name | Alternate Controller Enabled Insulin Infusion Pump |
| Regulation | 21 CFR 880.5730 |
| Device Class | Class 2 |
| Review Panel | Clinical Chemistry |
Identification
An alternate controller enabled infusion pump (ACE pump) is a device intended for the infusion of drugs into a patient. The ACE pump may include basal and bolus drug delivery at set or variable rates. ACE pumps are designed to reliably and securely communicate with external devices, such as automated drug dosing systems, to allow drug delivery commands to be received, executed, and confirmed. ACE pumps are intended to be used both alone and in conjunction with digitally connected medical devices for the purpose of drug delivery.
Classification Rationale
Class II (special controls). The special controls for this device are:
Special Controls
Alternate controller enabled infusion pumps must comply with the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include the following: (i) Evidence demonstrating that device infusion delivery accuracy conforms to defined user needs and intended uses and is validated to support safe use under actual use conditions. (A) Design input requirements must include delivery accuracy specifications under reasonably foreseeable use conditions, including ambient temperature changes, pressure changes ( *e.g.,* head-height, backpressure, atmospheric), and, as appropriate, different drug fluidic properties.(B) Test results must demonstrate that the device meets the design input requirements for delivery accuracy under use conditions for the programmable range of delivery rates and volumes. Testing shall be conducted with a statistically valid number of devices to account for variation between devices. (ii) Validation testing results demonstrating the ability of the pump to detect relevant hazards associated with drug delivery and the route of administration ( *e.g.,* occlusions, air in line, etc.) within a clinically relevant timeframe across the range of programmable drug delivery rates and volumes. Hazard detection must be appropriate for the intended use of the device and testing must validate appropriate performance under the conditions of use for the device.(iii) Validation testing results demonstrating compatibility with drugs that may be used with the pump based on its labeling. Testing must include assessment of drug stability under reasonably foreseeable use conditions that may affect drug stability ( *e.g.,* temperature, light exposure, or other factors as needed).(iv) The device parts that directly or indirectly contact the patient must be demonstrated to be biocompatible. This shall include chemical and particulate characterization on the final, finished, fluid contacting device components demonstrating that risk of harm from device-related residues is reasonably low. (v) Evidence verifying and validating that the device is reliable over the ACE pump use life, as specified in the design file, in terms of all device functions and in terms of pump performance. (vi) The device must be designed and tested for electrical safety, electromagnetic compatibility, and radio frequency wireless safety and availability consistent with patient safety requirements in the intended use environment. (vii) For any device that is capable of delivering more than one drug, the risk of cross-channeling drugs must be adequately mitigated. (viii) For any devices intended for multiple patient use, testing must demonstrate validation of reprocessing procedures and include verification that the device meets all functional and performance requirements after reprocessing. (2) Design verification and validation activities must include appropriate design inputs and design outputs that are essential for the proper functioning of the device that have been documented and include the following: (i) Risk control measures shall be implemented to address device system hazards and the design decisions related to how the risk control measures impact essential performance shall be documented. (ii) A traceability analysis demonstrating that all hazards are adequately controlled and that all controls have been validated in the final device design. (3) The device shall include validated interface specifications for digitally connected devices. These interface specifications shall, at a minimum, provide for the following: (i) Secure authentication (pairing) to external devices. (ii) Secure, accurate, and reliable means of data transmission between the pump and connected devices. (iii) Sharing of necessary state information between the pump and any digitally connected alternate controllers ( *e.g.,* battery level, reservoir level, pump status, error conditions).(iv) Ensuring that the pump continues to operate safely when data is received in a manner outside the bounds of the parameters specified. (v) A detailed process and procedure for sharing the pump interface specification with digitally connected devices and for validating the correct implementation of that protocol. (4) The device must include appropriate measures to ensure that safe therapy is maintained when communications with digitally connected alternate controller devices is interrupted, lost, or re-established after an interruption ( *e.g.,* reverting to a pre-programmed, safe drug delivery rate). Validation testing results must demonstrate that critical events that occur during a loss of communications (*e.g.,* commands, device malfunctions, occlusions, etc.) are handled appropriately during and after the interruption.(5) The device design must ensure that a record of critical events is stored and accessible for an adequate period to allow for auditing of communications between digitally connected devices and to facilitate the sharing of pertinent information with the responsible parties for those connected devices. Critical events to be stored by the system must, at a minimum, include: (i) A record of all drug delivery (ii) Commands issued to the pump and pump confirmations (iii) Device malfunctions (iv) Alarms and alerts and associated acknowledgements (v) Connectivity events ( *e.g.,* establishment or loss of communications)(6) Design verification and validation must include results obtained through a human factors study that demonstrates that an intended user can safely use the device for its intended use. (7) Device labeling must include the following: (i) A prominent statement identifying the drugs that are compatible with the device, including the identity and concentration of those drugs as appropriate. (ii) A description of the minimum and maximum basal rates, minimum and maximum bolus volumes, and the increment size for basal and bolus delivery, or other similarly applicable information about drug delivery parameters. (iii) A description of the pump accuracy at minimum, intermediate, and maximum bolus delivery volumes and the method(s) used to establish bolus delivery accuracy. For each bolus volume, pump accuracy shall be described in terms of the number of bolus doses measured to be within a given range as compared to the commanded volume. An acceptable accuracy description (depending on the drug delivered and bolus volume) may be provided as follows for each bolus volume tested, as applicable: Number of bolus doses with volume that is <25 percent, 25 percent to <75 percent, 75 percent to <95 percent, 95 percent to <105 percent, 105 percent to <125 percent, 125 percent to <175 percent, 175 to 250 percent, and >250 percent of the commanded amount. (iv) A description of the pump accuracy at minimum, intermediate, and maximum basal delivery rates and the method(s) used to establish basal delivery accuracy. For each basal rate, pump accuracy shall be described in terms of the amount of drug delivered after the basal delivery was first commanded, without a warmup period, up to various time points. The information provided must include typical pump performance, as well as worst-case pump performance observed during testing in terms of both over-delivery and under-delivery. An acceptable accuracy description (depending on the drug delivered) may be provided as follows, as applicable: The total volume delivered 1 hour, 6 hours, and 12 hours after starting delivery for a typical pump tested, as well as for the pump that delivered the least and the pump that delivered the most at each time point. (v) A description of delivery hazard alarm performance, as applicable. For occlusion alarms, performance shall be reported at minimum, intermediate, and maximum delivery rates and volumes. This description must include the specification for the longest time period that may elapse before an occlusion alarm is triggered under each delivery condition, as well as the typical results observed during performance testing of the pumps. (vi) For wireless connection enabled devices, a description of the wireless quality of service required for proper use of the device. (vii) For any infusion pumps intended for multiple patient reuse, instructions for safely reprocessing the device between uses.
Recent Cleared Devices (20 of 26)
Showing 20 most recent of 26 cleared devices.
| Record | Device Name | Applicant | Decision Date | Decision |
|---|---|---|---|---|
| K253743 | MiniMed Flex pump | Medtronic Minimed, Inc. | Mar 12, 2026 | SESE |
| K253470 | MiniMed 780G Insulin Pump (MMT-1884); MiniMed 780G Insulin Pump Kit (MMT-1894) | Medtronic Minimed, Inc. | Jan 15, 2026 | SESE |
| K253074 | Tandem Mobi insulin pump with interoperable technology | Tandem Diabetes Care, Inc. | Nov 6, 2025 | SESE |
| K252770 | iLet ACE Pump | Beta Bionics, Inc. | Sep 29, 2025 | SESE |
| K250792 | t:slim X2 insulin pump with interoperable technology | Tandem Diabetes Care, Inc. | Sep 24, 2025 | SESE |
| K251032 | MiniMed 780G insulin pump | Medtronic Minimed, Inc. | Jul 1, 2025 | SESE |
| K250930 | twiist system | Deka Research & Development Corp. | Apr 2, 2025 | SESE |
| K234027 | embecta Insulin Delivery System | Embecta Medical I, LLC | Aug 30, 2024 | SESE |
| K241078 | Tandem Mobi Insulin Pump with interoperable technology | Tandem Diabetes Care, Inc. | Jul 26, 2024 | SESE |
| K241178 | DEKA ACE Pump System | Deka Research and Development | May 24, 2024 | SESE |
| K240309 | Tandem Mobi insulin pump with interoperable technology | Tandem Diabetes Care, Inc. | Mar 21, 2024 | SESE |
| K233952 | DEKA ACE Pump System | Deka Research and Development | Mar 13, 2024 | SESE |
| K230545 | Inessa System | Triple Jump Israel , Ltd. | Dec 20, 2023 | SESE |
| K232380 | t:slim X2 Insulin Pump with Interoperable Technology | Tandem Diabetes Care, Inc. | Nov 3, 2023 | SESE |
| K231826 | Omnipod 5 ACE Pump | Insulet Corporation | Oct 18, 2023 | SESE |
| K233044 | Tandem Mobi insulin pump with interoperable technology | Tandem Diabetes Care, Inc. | Oct 5, 2023 | SESE |
| K213134 | Accu-Chek Solo micropump system with interoperable technology | Roche Diabetes Care GmbH | Aug 10, 2023 | SESE |
| K213536 | DEKA ACE Pump System | Deka Research and Development | Jul 25, 2023 | SESE |
| K223213 | Tandem Mobi Insulin Pump with Interoperable Technology | Tandem Diabetes Care, Inc. | Jul 10, 2023 | SESE |
| K231485 | iLet® ACE Pump | Beta Bionics, Inc. | Jun 21, 2023 | SESE |
Top Applicants
- Tandem Diabetes Care, Inc. — 10 clearances
- Beta Bionics, Inc. — 3 clearances
- Deka Research and Development — 3 clearances
- Insulet Corporation — 3 clearances
- Medtronic Minimed, Inc. — 3 clearances