Inessa System

K230545 · Triple Jump Israel , Ltd. · QFG · Dec 20, 2023 · Clinical Chemistry

Device Facts

Record IDK230545
Device NameInessa System
ApplicantTriple Jump Israel , Ltd.
Product CodeQFG · Clinical Chemistry
Decision DateDec 20, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5730
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The Inessa System is intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin. The Inessa System is able to reliably and securely communicate with compatible, digitally connected devices, including automated insulin dosing software, to receive, execute, and confirm commands from these devices. The Inessa System is intended for single patient, home use and requires a prescription. The Inessa System is indicated for use in individuals 6 years of age and greater.

Device Story

Inessa System is an Alternate Controller Enabled (ACE) infusion pump for subcutaneous insulin delivery. System comprises a handheld controller and a skin-adhered patch pump (reusable motor/electronics unit + sterile disposable cartridge with reservoir, cannula, and needle). Controller communicates with pump via Bluetooth Low Energy (BLE). Integrated Bolus Calculator uses user-inputted blood glucose, carbohydrate intake, and insulin parameters (duration, correction factor, ratio) to calculate correction boluses, meal boluses, and 'Insulin on Board'. Used in home environment by patients; requires prescription. Device receives commands from compatible automated insulin dosing software. Provides basal and bolus delivery; includes occlusion detection and data logging. Benefits patients by automating insulin delivery and facilitating interoperability with other diabetes management devices.

Clinical Evidence

No clinical efficacy trials were conducted. Evidence consists of bench verification and validation, including software testing, delivery volume accuracy, occlusion detection, and insulin stability. Human factors and usability validation testing confirmed that use-related risks are mitigated for both pediatric (6+) and adult populations. On-body adhesive testing confirmed performance throughout the intended wear life.

Technological Characteristics

Wearable patch pump; reusable pump unit with disposable cartridge. Materials biocompatible per ISO 10993-1. Connectivity via BLE. Power: rechargeable Li-ion battery. Software-controlled basal/bolus delivery. Standards: ES 60601-1 (electrical safety), IEC 60601-1-2 (EMC), ISO 11137 (sterilization), ASTM 1980-21 (shelf life).

Indications for Use

Indicated for subcutaneous insulin delivery in patients 6 years and older with diabetes mellitus requiring insulin. Contraindicated for patients unable to perform 4+ blood glucose tests daily, maintain HCP contact, follow instructions, or recognize pump alarms/signals. Not for use at low atmospheric pressures (<700hPA) or extreme temperatures.

Regulatory Classification

Identification

An alternate controller enabled infusion pump (ACE pump) is a device intended for the infusion of drugs into a patient. The ACE pump may include basal and bolus drug delivery at set or variable rates. ACE pumps are designed to reliably and securely communicate with external devices, such as automated drug dosing systems, to allow drug delivery commands to be received, executed, and confirmed. ACE pumps are intended to be used both alone and in conjunction with digitally connected medical devices for the purpose of drug delivery.

Special Controls

Alternate controller enabled infusion pumps must comply with the following special controls:

*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include the following: (i) Evidence demonstrating that device infusion delivery accuracy conforms to defined user needs and intended uses and is validated to support safe use under actual use conditions. (A) Design input requirements must include delivery accuracy specifications under reasonably foreseeable use conditions, including ambient temperature changes, pressure changes ( *e.g.,* head-height, backpressure, atmospheric), and, as appropriate, different drug fluidic properties.(B) Test results must demonstrate that the device meets the design input requirements for delivery accuracy under use conditions for the programmable range of delivery rates and volumes. Testing shall be conducted with a statistically valid number of devices to account for variation between devices. (ii) Validation testing results demonstrating the ability of the pump to detect relevant hazards associated with drug delivery and the route of administration ( *e.g.,* occlusions, air in line, etc.) within a clinically relevant timeframe across the range of programmable drug delivery rates and volumes. Hazard detection must be appropriate for the intended use of the device and testing must validate appropriate performance under the conditions of use for the device.(iii) Validation testing results demonstrating compatibility with drugs that may be used with the pump based on its labeling. Testing must include assessment of drug stability under reasonably foreseeable use conditions that may affect drug stability ( *e.g.,* temperature, light exposure, or other factors as needed).(iv) The device parts that directly or indirectly contact the patient must be demonstrated to be biocompatible. This shall include chemical and particulate characterization on the final, finished, fluid contacting device components demonstrating that risk of harm from device-related residues is reasonably low. (v) Evidence verifying and validating that the device is reliable over the ACE pump use life, as specified in the design file, in terms of all device functions and in terms of pump performance. (vi) The device must be designed and tested for electrical safety, electromagnetic compatibility, and radio frequency wireless safety and availability consistent with patient safety requirements in the intended use environment. (vii) For any device that is capable of delivering more than one drug, the risk of cross-channeling drugs must be adequately mitigated. (viii) For any devices intended for multiple patient use, testing must demonstrate validation of reprocessing procedures and include verification that the device meets all functional and performance requirements after reprocessing. (2) Design verification and validation activities must include appropriate design inputs and design outputs that are essential for the proper functioning of the device that have been documented and include the following: (i) Risk control measures shall be implemented to address device system hazards and the design decisions related to how the risk control measures impact essential performance shall be documented. (ii) A traceability analysis demonstrating that all hazards are adequately controlled and that all controls have been validated in the final device design. (3) The device shall include validated interface specifications for digitally connected devices. These interface specifications shall, at a minimum, provide for the following: (i) Secure authentication (pairing) to external devices. (ii) Secure, accurate, and reliable means of data transmission between the pump and connected devices. (iii) Sharing of necessary state information between the pump and any digitally connected alternate controllers ( *e.g.,* battery level, reservoir level, pump status, error conditions).(iv) Ensuring that the pump continues to operate safely when data is received in a manner outside the bounds of the parameters specified. (v) A detailed process and procedure for sharing the pump interface specification with digitally connected devices and for validating the correct implementation of that protocol. (4) The device must include appropriate measures to ensure that safe therapy is maintained when communications with digitally connected alternate controller devices is interrupted, lost, or re-established after an interruption ( *e.g.,* reverting to a pre-programmed, safe drug delivery rate). Validation testing results must demonstrate that critical events that occur during a loss of communications (*e.g.,* commands, device malfunctions, occlusions, etc.) are handled appropriately during and after the interruption.(5) The device design must ensure that a record of critical events is stored and accessible for an adequate period to allow for auditing of communications between digitally connected devices and to facilitate the sharing of pertinent information with the responsible parties for those connected devices. Critical events to be stored by the system must, at a minimum, include: (i) A record of all drug delivery (ii) Commands issued to the pump and pump confirmations (iii) Device malfunctions (iv) Alarms and alerts and associated acknowledgements (v) Connectivity events ( *e.g.,* establishment or loss of communications)(6) Design verification and validation must include results obtained through a human factors study that demonstrates that an intended user can safely use the device for its intended use. (7) Device labeling must include the following: (i) A prominent statement identifying the drugs that are compatible with the device, including the identity and concentration of those drugs as appropriate. (ii) A description of the minimum and maximum basal rates, minimum and maximum bolus volumes, and the increment size for basal and bolus delivery, or other similarly applicable information about drug delivery parameters. (iii) A description of the pump accuracy at minimum, intermediate, and maximum bolus delivery volumes and the method(s) used to establish bolus delivery accuracy. For each bolus volume, pump accuracy shall be described in terms of the number of bolus doses measured to be within a given range as compared to the commanded volume. An acceptable accuracy description (depending on the drug delivered and bolus volume) may be provided as follows for each bolus volume tested, as applicable: Number of bolus doses with volume that is <25 percent, 25 percent to <75 percent, 75 percent to <95 percent, 95 percent to <105 percent, 105 percent to <125 percent, 125 percent to <175 percent, 175 to 250 percent, and >250 percent of the commanded amount. (iv) A description of the pump accuracy at minimum, intermediate, and maximum basal delivery rates and the method(s) used to establish basal delivery accuracy. For each basal rate, pump accuracy shall be described in terms of the amount of drug delivered after the basal delivery was first commanded, without a warmup period, up to various time points. The information provided must include typical pump performance, as well as worst-case pump performance observed during testing in terms of both over-delivery and under-delivery. An acceptable accuracy description (depending on the drug delivered) may be provided as follows, as applicable: The total volume delivered 1 hour, 6 hours, and 12 hours after starting delivery for a typical pump tested, as well as for the pump that delivered the least and the pump that delivered the most at each time point. (v) A description of delivery hazard alarm performance, as applicable. For occlusion alarms, performance shall be reported at minimum, intermediate, and maximum delivery rates and volumes. This description must include the specification for the longest time period that may elapse before an occlusion alarm is triggered under each delivery condition, as well as the typical results observed during performance testing of the pumps. (vi) For wireless connection enabled devices, a description of the wireless quality of service required for proper use of the device. (vii) For any infusion pumps intended for multiple patient reuse, instructions for safely reprocessing the device between uses.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD &amp; DRUG ADMINISTRATION # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ## I Background Information: A 510(k) Number K230545 B Applicant Triple Jump Israel Ltd. C Proprietary and Established Names Inessa System D Regulatory Information | Product Code(s) | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | QFG | Class II | 21 CFR 880.5730 - Alternate Controller Enabled Infusion Pump | CH - Clinical Chemistry | | NDC | Class II | 21 CFR 868.1890 - Predictive pulmonary-function value calculator | CH - Clinical Chemistry | ## E Purpose for Submission New device ## II Intended Use/Indications for Use: A Intended Use(s): See Indications for Use below. B Indication(s) for Use: The Inessa System is intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin. The Inessa System is able to reliably and securely communicate with compatible, digitally connected devices, including automated insulin dosing software, to receive, execute, and confirm commands from these Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993-0002 www.fda.gov {1} devices. The Inessa System is intended for single patient, home use and requires a prescription. The Inessa System is indicated for use in individuals 6 years of age and greater. ## C Special Conditions for Use Statement(s): Rx - For Prescription Use Only The Inessa System is NOT recommended for patients who are: - Unable to perform at least four (4) Blood Glucose tests per day. - Unable to maintain contact with their Healthcare Provider (HCP). - Unable to use the Inessa System according to the instructions. The Inessa System is not recommended for people whose vision or hearing does not allow recognition of insulin pump signals and alarms. The Inessa System is intended for one-person use ONLY. Do not share the system with anyone - including other family members. The Inessa System should NOT be used: - At low atmospheric pressures below 700hPA (10,000’). - In extreme temperatures. The Inessa System may be affected by strong radiation or magnetic fields. In the event you are required to undergo imaging examinations (X-ray, MRI scan, CT scan, or similar tests), always remove the Patch Pump from your body, leaving it outside the treatment area (together with your Controller). ## III Device/System Characteristics: The Inessa System is an Alternate Controller Enabled (ACE) infusion pump system intended for subcutaneous delivery of insulin at set and variable rates. The Inessa System includes a handheld controller and a skin-adhered patch pump: - Patch pump: a skin adhered syringe pump designed for insulin delivery at set basal and/or bolus doses. The patch pump includes two parts: - Pump: a reusable part that includes motor, electronics, drive mechanism, and rechargeable battery. - Cartridge: a sterile disposable part that includes insulin reservoir, dosing mechanism, delivery channels, adhesive base, soft cannula, and insertion needle. The cartridge is preassembled with an inserter and is provided in a sterile disposable kit. - Controller: a handheld user interface providing instructions to the pump and receiving information from the pump using wireless Bluetooth Low Energy (BLE) communication. The Inessa System also includes a Bolus Calculator, accessible through the controller. Based on user inputs of blood glucose (current and targeted), carbohydrate intake (meals), patient's insulin characteristics (i.e., insulin duration of action, insulin correction factor, insulin-to-carbs ratio), this feature calculates values for: - Correction bolus needed to correct elevated blood glucose level; - Meal bolus needed to cover carbohydrates in an upcoming meal; and K230545 - Page 2 of 10 {2} - "Insulin on Board", an estimation of how much active insulin remains in the body from previous boluses. ## IV Substantial Equivalence Information: ### A Predicate Device Name(s): Omnipod DASH Insulin Management System with interoperable technology ### B Predicate 510(k) Number(s): K191679 ### C Comparison with Predicate(s): | Device & Predicate Device(s): | K230545 | K191679 | | --- | --- | --- | | Device Trade Name | Inessa System | Omnipod DASH Insulin Management System with interoperable technology | | General Device Characteristic Similarities | | | | Intended Use/Indications For Use | Intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin. Intended to be interoperable with connected devices including CGMs and automated insulin dosing algorithms. | Same | | Operating Environment | Home | Same | | Insulin Delivery Modes | Both basal and bolus | Same | | Insulin Basal Rate Delivery Range | 0.05 – 30U/h in 0.05U/h increments | Same | | Insulin Bolus Delivery Range | 0.05 – 30U in 0.05U increments | Same | | General Device Characteristic Differences | | | | Specific Drug/Biologic Use | U-100 Insulin. System has been tested with Novolog and Humalog. | U-100 Insulin. System has been tested with Novolog, Humalog, | K230545 - Page 3 of 10 {3} | | | Admelog and Apidra. | | --- | --- | --- | | Intended Population | Persons with Diabetes Mellitus ages 6 and up | Persons with Diabetes Mellitus ages 2 and up | V Standards/Guidance Documents Referenced: - ISO 2859-1:1999 - Sampling Procedures for Inspection by Attributes - Part 1: Sampling Schemes Indexed by Acceptable Quality Limit (AQL) for Lot - by - Lot Inspection - ISO 14971:2019 - Medical Device – Application of Risk Management to Medical Device - AAMI TIR38:2019 - Medical Device Safety Assurance Case guidance - AAMI TIR69: 2017 - Technical Information Report Risk Management of Radio Frequency - IEC 62304:2015 - Medical Device Software - Software Life-Cycle Processes - AAMI TIR 57: 2016 - Principles for Medical Device Security – Risk Management - AAMI TIR 36:2007 - Validation of Software for Regulated Processes - ES60601-1:2005/(R)2012 &amp; A1:2012, C1:2009/(R)2012 &amp; A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021] - IEC 60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSION - Medical Electrical Equipment part 1-2: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests - IEC 60601-1-8 Edition 2.2 2020-07 CONSOLIDATED VERSION - Medical Electrical Equipment - Part 1-8: General Requirements for Basic Safety and Essential Performance - Collateral Standard: General Requirements, Tests and Guidance for Alarm Systems in Medical Electrical Equipment and Medical Electrical Systems - 60601-1-11 Edition 2.1 2020-07 CONSOLIDATED VERSION – Medical Electrical Equipment - Part 1-11: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment - IEC 62133-2:2017 Edition1.0 2017-02 - Secondary Cells and Batteries Containing Alkaline or Other Non-Acid Electrolytes - Safety Requirements for Portable Sealed Secondary Cells, and for Batteries Made from Them, for Use in Portable Applications - Part 2: Lithium Systems - ANSI IEEE C63.27-2017 - American National Standard for Evaluation of Wireless Coexistence - ANSI AAMI ISO 11137-1: 2006/(R)2015 - Sterilization of Health Care Products - Radiation - Part 1: Requirements for Development, Validation and Routine Control of a Sterilization Process for Medical Devices - ISO 11137-2:2013 Third edition - Sterilization of Health Care Products - Radiation - Part 2: Establishing the Sterilization Dose - ISO 11607-1 Second Edition 2019-02 - Packaging for Terminally Sterilized Medical Devices - Part 1: Requirements for Materials, Sterile Barrier Systems and Packaging Systems - ISO 11607-2 second edition 2019-02 - Packaging for Terminally Sterilized Medical Devices - Part 2: Validation Requirements for Forming, Sealing and Assembly Processes - ISO 11737-1:2018 - Sterilization of Medical Devices – Microbiological Methods - Part 1: Determination of a Population of Microorganisms on Products - ISO 11737-2 third edition 2019-12 - Sterilization of Medical Devices - Microbiological Methods - Part 2: Tests of Sterility Performed in the Definition, Validation and Maintenance of a Sterilization Process K230545 - Page 4 of 10 {4} - ASTM D4169-16:2016 - Standard Practice for Performance Testing of Shipping Containers and Systems - ASTM F1980-21 - Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices - ASTM F1929-15:2015 - Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration - ASTM F88/F88M-15:2021 - Standard Test Method for Seal Strength of Flexible Barrier Materials - ISO 10993-1:2018 - Biological Evaluation of Medical Devices -- Part 1: Evaluation and Testing Within a Risk Management Process - ISO 10993-3:2014 - Biological Evaluation of Medical Devices - Part 3: Tests for Genotoxicity, Carcinogenicity and Reproductive Toxicity - ISO 10993-5:2009 - Biological Evaluation of Medical Devices - Part 5: Tests for In Vitro Cytotoxicity - ISO 10993-6:2016 - Tests for Local Effects After Implantation - ISO 10993-10:2021 - Test for Irritation and Delayed – Type Hypersensitivity - ISO 10993-11:2017 - Test for Systemic Toxicity - ISO 10993-12 fifth edition:2021 - Sample Preparation and Reference Materials - ISO 10993-17:2012 - Biological Evaluation of Medical Devices - Part 17: Establishment of Allowable Limits for Leachable Substances - ISO 10993-18:2020 Biological Evaluation of Medical Devices - Part 18: Chemical Characterization of Materials - ISO 10993-33:2015 Biological Evaluation of Medical Devices - Part 33: Guidance on Tests to Evaluate Genotoxicity - Supplement to ISO 10993-3 - ASTM F756-17 Standard Practice for Assessment of Hemolytic Properties of Materials - ASTM F2888-19: Standard Practice for Platelet Leukocyte Count - An In-Vitro Measure for Hemocompatibility Assessment of Cardiovascular Materials - ASTM F2382-18 Standard Test Method for Assessment of Circulating Blood-Contacting Medical Device Materials on Partial Thromboplastin Time (PTT) - ISO /TS 21726 First edition 2019-02 Biological evaluation of medical devices - Application of the threshold of toxicological concern (TTC) for assessing biocompatibility of medical device constituents - ISO 15223-1:2021 Medical Devices - Symbols to be Used with Medical Devices Labels, Labeling, and Information to be Supplied - Part 1: General Requirements - ISO 20417:2021 Medical Devices - Information to Be Supplied by the Manufacturer - ISO 14155:2011 Clinical Investigation of Medical Devices for Human Subjects - Good Clinical Practice - IEC 60601-1-6:2020 -edition 3.2 General Requirements for Basic Safety and Essential Performance –Collateral Standard: Usability - IEC 62366-1:2020 edition 1.1 consolidated version Medical Devices - Application of Usability Engineering to Medical Devices - AAMI / ANSI HE75-2009 (R) 2018 Human Factors Engineering - Design of Medical Devices - AAMI TIR 42:2021- Evaluation of Particulates Associated with Vascular Medical Devices K230545 - Page 5 of 10 {5} VI Performance Characteristics (if/when applicable): A Analytical Performance: 1. Basal delivery accuracy: To assess basal delivery accuracy, 58 patch pumps were tested for each basal rate by delivering insulin at low, medium, and high basal rates (0.05, 1.00, and 30.0 U/hr). All 58 patch pumps were pre-conditioned for simulated shipping and handling, and 29 of which were treated by accelerated aging for simulated 6 month of shelf life. Two different insulin products were used for testing (NovoLog and Humalog). The insulin was pumped into a container on a scale and the weight of the liquid at various time points was used to assess basal delivery accuracy. The following tables report the typical basal performance (median) observed, along with the lowest and highest results observed for the low, medium, and high basal rate settings for all pumps tested with no warmup period. Table 1: Amount of fluid delivered after 1, 6, and 12 hours with 0.05 U/hr (low) basal rate setting | 0.05 U/hr Basal Duration | 1 hour | 6 hours | 12 hours | | --- | --- | --- | --- | | Total expected delivery volume | 0.05 U | 0.3 U | 0.6 U | | Median amount delivered | 0.05 U | 0.28 U | 0.56 U | | [min, max] | [0.04, 0.05] | [0.26, 0.30] | [0.52, 0.62] | Table 2: Amount of fluid delivered after 1, 6, and 12 hours with 1 U/hr (medium) basal rate setting | 1 U/hr Basal Duration | 1 hour | 6 hours | 12 hours | | --- | --- | --- | --- | | Total expected delivery volume | 1 U | 6 U | 12 U | | Median amount delivered | 1.03 U | 6.13 U | 12.17 U | | [min, max] | [0.88, 1.39] | [5.80, 6.53] | [11.65, 12.73] | Table 3: Amount of fluid delivered after 1 and 6 hours with 30 U/hr (high) basal rate setting | 30 U/hr Basal Duration | 1 hour | 6 hours | | --- | --- | --- | | Total expected delivery volume | 30 U | 180 U | | Median amount delivered | 30.14 U | 177.88 U | | [min, max] | [29.20, 30.78] | [171.44, 181.28] | Note: A measurement at the 12-hour period with 30.0 U/hr basal rate is not applicable to the Inessa System as the reservoir will empty prior to this time point. 2. Bolus delivery accuracy: To assess bolus delivery accuracy, 58 patch pumps were tested for each bolus size by delivering a minimum, intermediate, and maximum bolus amounts (0.05, 5.00, and 30.0 Units). All 58 pump were pre-conditioned for simulated shipping and handling, and 29 of which were treated by accelerated aging for simulated 6 month of shelf life. K230545 - Page 6 of 10 {6} Two different insulin products were used for testing (NovoLog and Humalog). The insulin was pumped into a container on a scale and the weight of the liquid at various time points was used to assess bolus delivery accuracy. The number of total and consecutive boluses delivered in this testing for each delivery volume is described in Table 4 below: Table 4: Summary of bolus testing protocol | Bolus size (units) | Number of pumps tested | Consecutive boluses per pump | Total boluses | | --- | --- | --- | --- | | 0.05 units | 58 | 500 | 29,000 | | 5.0 units | 58 | 25 | 1450 | | 30 units | 58 | 6 | 348 | The actual bolus volume delivered was compared to the expected bolus volume for minimum, intermediate, and maximum boluses. Tables 5-7 below show the number (and %) of boluses within the specified range of each target bolus volume. Table 5: Amount of fluid delivered after a 0.05 U bolus request | Units delivered after a 0.05 U bolus request (% of commanded units) | | | | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | <25% | 25-75% | 75-90% | 90-95% | 95-105% | 105-110% | 110-125% | 125-175% | 175-250% | >250% | | Number and percent of boluses | 21/29000 (0.07%) | 213/29000 (0.73%) | 1209/29000 (4.2%) | 2718/29000 (9.4%) | 23666/29000 (82%) | 804/29000 (2.8%) | 94/29000 (0.32%) | 103/29000 (0.36%) | 118/29000 (0.41%) | 54/29000 (0.19%) | Table 6: Amount of fluid delivered after a 5.0 U bolus request | Units delivered after a 5.0 U bolus request (% of commanded units) | | | | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | <25% | 25-75% | 75-90% | 90-95% | 95-105% | 105-110% | 110-125% | 125-175% | 175-250% | >250% | | Number and percent of boluses | 0/1450 - | 0/1450 - | 9/1450 (0.62%) | 9/1450 (0.62%) | 1432/145 (99%) | 0/1450 - | 0/1450 - | 0/1450 - | 0/1450 - | 0/1450 - | Table 7: Amount of fluid delivered after a 30 U bolus request | Units delivered after a 25 U bolus request (% of commanded units) | | | | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | <25% | 25-75% | 75-90% | 90-95% | 95-105% | 105-110% | 110-125% | 125-175% | 175-250% | >250% | | Number and percent of boluses | 0/348 - | 0/348 - | 0/348 - | 1/348 (0.29%) | 347/348 (99.71%) | 0/348 - | 0/348 - | 0/348 - | 0/348 - | 0/348 - | K230545 - Page 7 of 10 {7} K230545 - Page 8 of 10 3. Occlusion detection: Occlusion detection testing was conducted using 29 pumps and 3 delivery profiles: 5U bolus, 1.0 U/hr basal rate, and 0.05 U/hr basal rate. Each pump was tested for the time between occlusion and pump alarm sequentially and for the 3 delivery profiles. To test the time between occlusion and pump alarm, each pump was physically occluded by closing the soft cannula after priming, and a 5U bolus, a 1.0 U/hr basal rate, or a 0.05 U/hr basal rate were initiated. A timer was used to measure the time between initiation of delivery and the occlusion alarm being sounded. The typical time to occlusion detection in the table below is the average for the samples measured and the maximum time is the absolute maximum. Results are presented in the table below. Table 8: Occlusion detection testing | | Typical time to occlusion detection | Maximum time to occlusion detection | | --- | --- | --- | | 5.0 U Bolus | 0.05 minutes | 0.05 minutes | | 1.0 U/hr Basal | 17 minutes | 36 minutes | | 0.05 U/hr Basal | 3.95 hours | 13.9 hours | After pumps alarmed, the occlusions were cleared and the total amount of fluid released was measured. Typical volumes were 0.1 U, and maximum volumes were 0.55 U for 1 U/hr basal rate. B Other Supportive Instrument Performance Characteristics Data: 1. Hazard analysis A comprehensive hazard analysis was reviewed, in which design inputs and outputs, risks, and risk mitigations for hardware and software associated with proper functioning of the insulin pump were reviewed. The sponsor performed a hazard analysis to account for the unique intended use, design elements, and risks of their ACE pump. This analysis identified hazards which could reasonably be anticipated to impact the proper use of the device, traced all identified risks to adequate design controls, and demonstrated that design features were appropriately implemented and validated. 2. Human Factors Human factors validation tests were conducted with the Inessa System. All study participants received training that was consistent with the training that patients would receive with the commercial product. Usability evaluations assessed comprehension and usability of the device for critical device tasks. Results of the study demonstrated that the Inessa System is validated for its intended use. {8} K230545 - Page 9 of 10 3. Biocompatibility Biocompatibility testing was performed per ISO 10993-1:2018, FDA Guidance Document: Use of International Standard ISO 10993-1 “Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process,” and FDA special controls for alternate controller enabled infusion pumps. All endpoints were tested adequately, and results were acceptable. 4. Sterility The sterility information was reviewed and found to be acceptable. 5. Insulin Compatibility and Stability The Inessa System was found to be compatible with U-100 insulin Novolog (insulin aspart) and Humalog (insulin lispro) for up to 3 days. 6. Additional Bench Testing In addition to the performance testing described above, other device verification testing was conducted to demonstrate that the system meets its intended use and is safe, reliable, and all safety and reliability critical requirements have been adequately verified. Summaries for reliability and safety testing follow: | Testing to Support System Reliability and Safety | | --- | | Delivery accuracy performance under worst-case environmental conditions | | Performance after rapid environmental changes | | Unintended Insulin delivery | | Reusable lifetime testing | | Controller lifetime testing | | Disposable components lifetime testing | | Battery Alarm | | Reliability | 7. Electromagnetic Compatibility and Wireless Coexistence Electromagnetic compatibility, electromagnetic immunity and wireless coexistence testing was performed for the pump. All tests demonstrated that the device would perform as expected in the home healthcare environment. 8. Basic Safety and Essential Performance (Electrical Safety) The sponsor provided verification evidence for compliance with the IEC 60601-1 and applicable collateral standards. Verification results support the finding of substantial equivalent for this device. {9} K230545 - Page 10 of 10 ## 9. Data Logging The sponsor provided a summary of pump and controller logging capability which enable the device to record critical events including insulin delivery, pump commands and confirmations, connectivity states, malfunctions, and alarms. These were reviewed and found to be adequate. ## 10. Interoperability A plan and approach for interoperability were provided according to the FDA Guidance “Design Considerations and Pre-market Submission Recommendations for Interoperable Medical Devices” and determined to be adequate to support and clearly specify expectations, requirements, and interface specifications to potential interoperable devices. In addition, their plan covered their approach to working with connected device companies regarding contractual approaches, interfaces for data communication and exchange, and post-market reporting procedures and responsibilities (e.g., who is responsible for investigating and reporting complaints, malfunctions, and adverse events). The sponsor additionally provided validated software protocols intended to ensure secure, accurate, and reliable communication with digital interfacing devices, as well as failsafe design features to mitigate the risks associated with interruption of communication with digitally connected devices. These protocols were reviewed and found to be adequate. ## 11. Software and Cybersecurity Detailed Information on software and cybersecurity of the device was reviewed and found acceptable. ## VII Proposed Labeling: The labeling supports the finding of substantial equivalence for this device. ## VIII Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...