CR3 Keyless Split Sample Cup Secobarbital- Methadone
Applicant
Guangzhou Wondfo Biotech Co., Ltd.
Product Code
DIS · Clinical Toxicology
Decision Date
Jan 13, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3150
Device Class
Class 2
Intended Use
CR3 Keyless Split Sample Cup Secobarbital-Methadone is a rapid test for the qualitative detection of Secobarbital and Methadone in human urine at a cutoff concentration of 300ng/mL for each of the drugs. The test may yield preliminary positive results when prescription drugs Secobarbital and Methadone are ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Secobarbital and Methadone in urine. The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only
Device Story
Lateral flow immunochromatographic assay for qualitative detection of Secobarbital and Methadone in human urine; competitive binding principle. Device consists of sample cup containing test membranes pre-coated with drug-protein conjugates and mouse monoclonal anti-drug antibody-dye conjugates. User immerses absorbent end in urine; capillary action draws sample across membrane. Analyte below cutoff allows antibody-dye to bind to test line (negative); analyte above cutoff binds antibody-dye, preventing test line formation (positive). Control line confirms proper flow. Used in point-of-care or home settings by lay users or clinicians. Preliminary positive results require confirmation via GC/MS. Benefits include rapid, on-site screening for drug presence to inform clinical or personal decision-making.
Clinical Evidence
Bench testing only. Precision studies (n=50 per concentration level) across 25 days confirmed performance at cutoffs. Interference testing evaluated 100+ compounds. Method comparison against GC/MS performed with 80 clinical samples. Lay-user study (n=260) demonstrated 80-100% agreement with GC/MS across various concentrations near cutoff. Flesch-Kincaid reading grade level <7 for instructions.
Technological Characteristics
Lateral flow immunochromatographic assay. Components: mouse monoclonal anti-drug antibody-dye conjugate, fixed drug-protein conjugates, anti-mouse IgG polyclonal antibodies. Form factor: split sample cup. No external energy source required. Standalone device. Stability: 18 months at 4-30°C.
Indications for Use
Indicated for qualitative detection of Secobarbital and Methadone in human urine at 300ng/mL cutoff. Intended for OTC and prescription use to provide preliminary results requiring confirmation by GC/MS.
Regulatory Classification
Identification
A barbiturate test system is a device intended to measure barbiturates, a class of hypnotic and sedative drugs, in serum, urine, and gastric contents. Measurements obtained by this device are used in the diagnosis and treatment of barbiturate use or overdose and in monitoring levels of barbiturate to ensure appropriate therapy.
Special Controls
*Classification.* Class II (special controls). A barbiturate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
K121987 — WONDFO AMPHETAMINE URINE TEST WONDFO BARBITURATES URINE TEST WONDFO BENZODIAZEPINES URINE TEST · Guangzhou Wondfo Biotech Co., Ltd. · Aug 1, 2012
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of a caduceus, which is a symbol often associated with healthcare and medicine.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
GUANGZHOU WONDFO BIOTECH CO., LTD. C/O JOE SHIA REGULATORY CONSULTANT 504 EAST DIAMOND AVE. SUITE F GAITHERSBURG MD 20878
January 13, 2015
Re: K143535
Trade/Device Name: CR3 Keyless Split Sample Cup Secobarbital-Methadone Regulation Number: 21 CFR 862.3150 Regulation Name: Barbiturate test system Regulatory Class: II Product Code: DIS. DJR Dated: December 9, 2014 Received: December 15, 2014
Dear Mr. Joe Shia:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Stayce Beck -S
For :
Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) k143535
Device Name
CR3 Keyless Split Sample Cup Secobarbital-Methadone
#### Indications for Use (Describe)
CR3 Keyless Split Sample Cup Secobarbital-Methadone is a rapid test for the qualitative detection of Secobarbital and Methadone in human urine at a cutoff concentration of 300ng/mL for each of the drugs.
The test may yield preliminary positive results when prescription drugs Secobarbital and Methadone are ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Secobarbital and Methadone in urine. The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.
For in vitro diagnostic use only
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement below.
{3}------------------------------------------------
## 510(k) SUMMARY
| 1. Date: | January 8, 2015 |
|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 2. Submitter: | Guangzhou Wondfo Biotech Co., Ltd.<br>No.8 Lizhishan Road, Science City, Luogang District, Guangzhou, P.R.<br>China 510663 |
| 3. Contact person: | Joe Shia<br>LSI International Inc.<br>504 East Diamond Ave., Suite F<br>Gaithersburg, MD 20878<br>Telephone: 240-505-7880<br>Fax: 301-916-6213<br>Email:shiajl@yahoo.com |
- CR3 Keyless Split Sample Cup Secobarbital-Methadone 4. Device Name:
| Classification: | Class II | |
|-----------------|------------------------------------------|------------|
| Product Code | CFR # | Panel |
| DIS | 21 CFR, 862.3150 Barbiturate Test System | Toxicology |
| DJR | 21 CFR, 862.3620 Methadone Test System | Toxicology |
- 5. Predicate Devices: K122904 Wondfo Multi-Drug Urine Test Cup
## 6. Intended Use:
CR3 Keyless Split Sample Cup Secobarbital-Methadone is a rapid test for the qualitative detection of Secobarbital and Methadone in human urine at a cutoff concentration of 300ng/mL for each of the drugs.
The test may yield preliminary positive results when prescription drugs Secobarbital and Methadone are ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Secobarbital and Methadone in urine. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.
For in vitro diagnostic use only.
{4}------------------------------------------------
# 7. Device Description:
The CR3 Keyless Split Sample Cup Secobarbital - Methadone test uses immunochromatographic assays for secobarbital and methadone. The test is a lateral flow system for the qualitative detection of secobarbital and methadone in human urine. The test is the first step in a two-step process. The second step is to send the sample for laboratory testing if preliminary positive results are obtained.
| Item | Device | Predicate - K122904 |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------|-------------------------------------|
| Indication(s) for use | For the qualitative determination of drugs of abuse in human urine | Same |
| Methodology | Competitive binding, lateral flow immunochromatographic assays based on the principle of antigen antibody immunochemistry. | Same |
| Results | Qualitative | Same |
| Specimen Type | Human urine | Same |
| Cut Off Values | Secobarbital: 300ng/ml<br>Methadone: 300ng/ml | Same for Secobarbital and Methadone |
| Configurations | Cup | Cup, Dipcard |
| Conditions for Use | Over-the-Counter & Prescription Use | Same |
- 8. Substantial Equivalence Information
## 9. Test Principle
The CR3 Keyless Split Sample Cup Secobarbital – Methadone test is a rapid test for the qualitative detection of Secobarbital and Methadone in urine samples and contains lateral flow chromatographic immunoassays for secobarbital and methadone. Each assay uses a mouse monoclonal anti-drug antibody-dye conjugate, fixed drug-protein conjugates, and anti-mouse IgG polyclonal antibodies coated on the test membranes. When the absorbent end of the test is immersed into a urine sample, the urine is absorbed into the device by capillary action and mixes with the antibody-dye conjugate, flowing across the pre-coated membrane. At analyte concentrations below the target cut-off, antibody-dye conjugates bind to the drug-protein conjugate immobilized in the Test Region (T) of the device. This produces a colored test line that indicates a negative result. When analyte concentration is above the cut-off, analyte molecules bind to the antibody-dye conjugate, preventing the antibody-dye conjugate from binding to the drug-protein conjugate immobilized in the Test Region (T) of the device. No colored band shows in the test region, indicating a potentially positive result. A band should form in the control region (C) of the device regardless of the presence of drug or metabolite in the sample.
{5}------------------------------------------------
## 10. Performance Characteristics
- 1. Analytical Performance
- a. Precision
Precision studies were carried out for samples with concentrations of -100% cut-off, -75% cut-off, -50% cut-off, -25% cut-off, at the cut-off, +25% cut-off, +50% cut-off and +100% cut-off. For each concentration, tests were performed two runs per day by three operators for 25 days. All sample aliquots were masked and randomized. The results obtained are summarized in the following tables:
| Result<br>BAR | -100%<br>cut-off | -75%<br>cut-off | -50%<br>cut-off | -25%<br>cut-off | cut-off | +25%<br>cut-off | +50%<br>cut-off | +75%<br>cut-off | +100%<br>cut-off |
|---------------|------------------|-----------------|-----------------|-----------------|---------|-----------------|-----------------|-----------------|------------------|
| W12010501CU5 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 43+/7- | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| W12010502CU5 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 42+/8- | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| W12010503CU5 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 42+/8- | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
### A. For Secobarbital (BAR) testing
| Result<br>MTD | -100%<br>cut-off | -75%<br>cut-off | -50%<br>cut- off | -25%<br>cut-off | cut-off | +25%<br>cut-off | +50%<br>cut-off | +75%<br>cut-off | +100%<br>cut-off |
|---------------|------------------|-----------------|------------------|-----------------|---------|-----------------|-----------------|-----------------|------------------|
| W12010501CU5 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 43+/7- | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| W12010502CU5 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 41+/9- | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| W12010503CU5 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 42+/8- | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
### B. For Methadone (MTD) testing
- b. Linearity
Not applicable.
- c. Stability
The CR3 Keyless Split Sample Cup Secobarbital – Methadone is stable at 4-30°C for 18 months as determined by conducting accelerated and real-time stability testing.
Control materials are not provided with the device. The labeling provides information on how to obtain control materials.
{6}------------------------------------------------
## d. Cut-off
A total of 125 secobarbital samples and 125 methadone samples equally distributed at concentrations of -50%, -25%, at the cut-off, +25%, +50% of their respective cut-offs. These samples were tested using three different lots by three different operators. Results were all positive at +25% and +50% cut-off and all negative at -25% and -50% cut-off for both secobarbital and methadone. The following cut-off values for the test devices have been verified.
| Test | Calibrator | Cut-off<br>(ng/ml) |
|---------------------|--------------|--------------------|
| Secobarbital (BAR ) | Secobarbital | 300 |
| Methadone (MTD) | Methadone | 300 |
## e. Interference
Potential interfering substances found in human urine of physiological conditions were added to urine containing target drugs (secobarbital or methadone) at 25% below and 25% above the cut-off. These urine samples were tested using three batches of the CR3Keyless Split Sample Cup Secobarbital - Methadone by three different operators. Compounds that showed no interference at a concentration of 100ug/mL are summarized below:
### Secobarbital
| Acetaminophen | Erythromycin | O-Hydroxyhippuric acid |
|------------------------|--------------------------|------------------------|
| Acetophenetidin | ß-Estradiol | D,L-Octopamine |
| Acetylsalicylic acid | Estrone-3-sulfate | Oxalic acid |
| Aminopyrine | Ethyl-p-aminobenzoate | Oxazepam |
| Amitryptyline | Fenoprofen | Oxolinic acid |
| Amoxicillin | Furosemide | Oxycodone |
| DL-Amphetamine sulfate | Gentisic acid | Oxymetazoline |
| Ampicillin | Hemoglobin | Papaverine |
| Apomorphine | Hydralazine | Penicillin-G |
| Ascorbic acid | Hydrochlorothiazide | Pentazocaine |
| Aspartame | Hydrocodone | Perphenazine |
| Atropine | Hydrocortisone | Phencyclidine |
| Benzilic acid | p-Hydroxyamphetamine | Phenelzine |
| Benzoic acid | p-Hydroxymethamphetamine | ß-Phenylethlamine |
| Benzoylecgonine | 3-Hydroxytyramine | Phenylpropanolamine |
| Bilirubin | Ibuprofen | Prednisolone |
| Brompheniramine | Imipramine | Prednisone |
| Caffeine | (-) Isoproterenol | Procaine |
{7}------------------------------------------------
Cannabidiol Cannabinol Chloralhydrate Chloramphenicol Chlorothiazide (±) Chlorpheniramine Chlorpromazine Chlorquine Cholesterol Clomipramine Clonidine Cocaine hydrochloride Codeine Cortisone
(-) Cotinine
- Creatinine Deoxycorticosterone Dextromethorphan Diazepam Diclofenac Diflunisal Digoxin Diphenhydramine Doxylamine Ecgonine hydrochloride Ecgonine methylester (IR,2S)(-)Ephedrine
#### Methadone
- Acetaminophen Acetophenetidin Acetylsalicylic acid Amobarbital Aminopyrine Amitryptyline Amoxicillin DL-Amphetamine sulfate Ampicillin Apomorphine
Isoxsuprine Ketamine Ketoprofen Labetalol Levorphanol Loperamide L-Phenylephrine Maprotiline Meperidine Meprobamate Morphine-3-B-D glucuronide Methadone Methamphetamine ±) - 3,4-Methylenedioxy- amphetamine hydrochloride (±)-3,4-Methylenedioxymethamphetamine hydrochloride Morphine Sulfate N-Acetylprocainamide Nalidixic acid Naloxone Naltrexone Naproxen Niacinamide Nifedipine Norcodein Norethindrone D-Norpropox yphene Noscapine
Ecgonine methylester (IR,2S)(-)Ephedrine Erythromycin ß-Estradiol Estrone-3-sulfate Ethyl-p-aminobenzoate Fenoprofen Furosemide Gentisic acid Hemoglobin
Promazine Promethazine D,L-Propanolol D-Propoxyphene Quinidine Quinine Ranitidine Salicylic acid Serotonin Sulfamethazine Sulindac Temazepam Tetracycline Tetrahydrozoline
Thebaine
Thiamine Thioridazine Triamterene Trifluoperazine Trimethoprim Trimipramine Tryptamine D, L-Tyrosine Uric acid Verapamil Zomepirac
Oxazepam Oxolinic acid Oxycodone Oxymetazoline Papaverine Penicillin-G Pentazocine Pentobarbital Perphenazine Phencyclidine
{8}------------------------------------------------
Ascorbic acid Aspartame Atropine Benzilic acid Benzoic acid Benzoylecgonine Bilirubin Brompheniramine Caffeine Cannabidiol Cannabinol Chloralhydrate Chloramphenicol Chlorothiazide (±) - Chlorpheniramine Chlorpromazine Chlorquine Cholesterol Clomipramine Clonidine Cocaine hydrochloride Codeine (-) Cotinine Cortisone Creatinine Deoxycorticosterone Dextromethorphan Diazepam Diclofenac Diflunisal Digoxin Diphenhydramine D-Norpropox yphene D-Propoxyphene D,L-Tyrosine
DL-Octopamine
Hydralazine Hydrochlorothiazide Hydrocodone Hydrocortisone p-Hydroxyamphetamine p-Hydroxymethamphetamine 3-Hydroxytyramine Ibuprofen Imipramine (-) Isoproterenol Isoxsuprine Ketamine Ketoprofen Labetalol Levorphanol Loperamide L-Phenylephrine Maprotiline Meperidine Meprobamate
Methamphetamine
Methoxyphenamine
(±) - 3,4-Methylenedioxyamphetamine hydrochloride (±)-3,4-Methylenedioxymethamphetamine hydrochloride Morphine Sulfate Morphine-3-B-D glucuronide N-Acetylprocainamide Nalidixic acid Naloxone Naltrexone Naproxen Niacinamide Nifedipine Norcodein Norethindrone Noscapine
Phenelzine Phenobarbital Phentermine β-Phenylethylamine Phenylpropanolamine Prednisolone Prednisone Procaine Promazine Promethazine Quinidine Quinine Ranitidine Salicylic acid Secobarbital Serotonin Sulfamethazine Sulindac Temazepam Tetracycline
Tetrahydrocortisone, 3-acetate
Tetrahydrocortisone 3-(ß-D-glucuronide) Tetrahydrozoline
Thebaine
Thiamine Thioridazine Triamterene Trifluoperazine Trimethoprim Trimipramine Tryptamine DL-Tryptophan Tyramine Uric acid Verapamil Zomepirac
{9}------------------------------------------------
- f. Specificity
To test the specificity, drug metabolites and other components that are likely to be present in urine samples were tested. The target drug (Secobarbital or Methadone), its drug metabolites and the related compounds were studied. These samples were tested using three batches of the CR3Keyless Split Sample Cup Secobarbital –Methadone by three different operators. The drug metabolites and other components were tested at different concentrations. The obtained lowest detectable concentration was used to calculate the cross-reactivity. Results are shown in the following tables.
| BAR<br>(Secobarbital,<br>Cut-off=300 ng/mL) | Result | %<br>Cross-Reactivity |
|---------------------------------------------|-------------------------|-----------------------|
| Secobarbital | Positive at 300 ng/mL | 100% |
| Amobarbital | Positive at 300 ng/mL | 100% |
| Alphenol | Positive at 150 ng/mL | 200% |
| Aprobarbital | Positive at 200 ng/mL | 150% |
| Butabarbital | Positive at 75 ng/mL | 400% |
| Butathal | Positive at 100 ng/mL | 300% |
| Butalbital | Positive at 2,500 ng/mL | 12% |
| Cyclopentobarbital | Positive at 600 ng/mL | 50% |
| Pentobarbital | Positive at 300 ng/mL | 100% |
| Phenobarbital | Positive at 100 ng/mL | 300% |
| MTD<br>(Methadone,<br>Cut-off=300 ng/mL) | Result | %<br>Cross-Reactivity |
|------------------------------------------|--------------------------|-----------------------|
| Methadone | Positive at 300 ng/mL | 100% |
| Doxylamine | Positive at 50,000 ng/mL | 0.6% |
# g. Effect of Specific Gravity and Urine pH
Twelve urine samples of normal, high, and low specific gravity ranges (1.000 to 1.035) were collected and spiked with either Secobarbital or Methadone at 25% below and 25% above the corresponding cut-off level. These samples were tested using three batches of the CR3 Keyless Split Sample Cup Secobarbital-Methadone by three different operators.
The pH of an aliquot negative urine pool was adjusted to pH ranges of 4.00 to 9.00 in 1 pH unit increments and spiked with Secobarbital or Methadone at 25% below and 25% above the
{10}------------------------------------------------
corresponding cut-off levels. These samples were tested using three batches of the CR3 Keyless Split Sample Cup Secobarbital-Methadone by three different operators.
The device performance was found not affected by varying specific gravity and pH.
- 2. Comparison Studies
The method comparison for the CR3 Keyless Split Sample Cup Secobarbital-Methadone was performed in-house with three laboratory assistants. Operators ran 80 (40 negative and 40 positive) unaltered clinical samples. The samples were masked and randomized. The obtained test results were compared to GC/MS results. The results are presented in the table below:
| Group<br>Operators | | Negative | Low<br>Negative by<br>GC/MS<br>(less than<br>-50%) | Near Cutoff<br>Negative by<br>GC/MS<br>(Between<br>-50% and<br>cutoff) | Near Cutoff<br>Positive by<br>GC/MS<br>(Between<br>the cutoff<br>and +50%) | High<br>Positive by<br>GC/MS<br>(greater<br>than +50%) |
|--------------------|----------|----------|----------------------------------------------------|------------------------------------------------------------------------|----------------------------------------------------------------------------|--------------------------------------------------------|
| | Positive | 0 | 0 | 4 | 15 | 20 |
| Viewer A | Negative | 10 | 13 | 13 | 5 | 0 |
| | Positive | 0 | 0 | 3 | 17 | 20 |
| Viewer B | Negative | 10 | 13 | 14 | 3 | 0 |
| Viewer C | Positive | 0 | 0 | 3 | 17 | 20 |
| | Negative | 10 | 13 | 14 | 3 | 0 |
Secobarbital
Discordant table:
| Viewer | Sample number | GC/MS result | Viewer result |
|----------|---------------|--------------|---------------|
| Viewer A | BARC3061 | 281 | positive |
| Viewer A | BARC3062 | 275 | positive |
| Viewer A | BARC3063 | 302 | negative |
| Viewer A | BARC3065 | 314 | negative |
| Viewer A | BAR 3212 | 278 | positive |
| Viewer A | BAR 3214 | 293 | positive |
| Viewer A | BAR 3218 | 305 | negative |
| Viewer A | BAR 3223 | 317 | negative |
| Viewer A | BAR 3229 | 309 | negative |
| Viewer B | BARC3061 | 281 | positive |
| Viewer B | BARC3063 | 302 | negative |
| Viewer B | BARC3064 | 297 | positive |
| Viewer B | BARC3065 | 314 | negative |
{11}------------------------------------------------
| Viewer B | BAR 3214 | 293 | positive |
|----------|----------|-----|----------|
| Viewer B | BAR 3227 | 314 | negative |
| Viewer C | BARC3062 | 275 | positive |
| Viewer C | BARC3063 | 302 | negative |
| Viewer C | BARC3064 | 297 | positive |
| Viewer C | BAR 3214 | 293 | positive |
| Viewer C | BAR 3218 | 305 | negative |
| Viewer C | BAR 3227 | 314 | negative |
# Methadone
| Group<br>Operators | | Negative | Low<br>Negative by<br>GC/MS<br>(less than<br>-50%) | Near Cutoff<br>Negative by<br>GC/MS<br>(Between<br>-50% and<br>cutoff) | Near Cutoff<br>Positive by<br>GC/MS<br>(Between<br>the cutoff<br>and +50%) | High<br>Positive by<br>GC/MS<br>(greater<br>than +50%) |
|--------------------|----------|----------|----------------------------------------------------|------------------------------------------------------------------------|----------------------------------------------------------------------------|--------------------------------------------------------|
| Viewer A | Positive | 0 | 0 | 3 | 17 | 20 |
| | Negative | 10 | 12 | 15 | 3 | 0 |
| Viewer B | Positive | 0 | 0 | 5 | 16 | 20 |
| | Negative | 10 | 12 | 13 | 4 | 0 |
| Viewer C | Positive | 0 | 0 | 4 | 17 | 20 |
| | Negative | 10 | 12 | 14 | 3 | 0 |
Discordant table:
| Viewer | Sample number | GC/MS result | viewer results |
|----------|---------------|--------------|----------------|
| Viewer A | MTDC3063 | 304 | negative |
| Viewer A | MTDC3065 | 291 | positive |
| Viewer A | MTD 3211 | 284 | positive |
| Viewer A | MTD 3214 | 299 | positive |
| Viewer A | MTD 3220 | 306 | negative |
| Viewer A | MTD 3226 | 301 | negative |
| Viewer B | MTDC3061 | 297 | positive |
| Viewer B | MTDC3062 | 313 | negative |
| Viewer B | MTDC3063 | 304 | negative |
| Viewer B | MTDC3064 | 286 | positive |
| Viewer B | MTDC3065 | 291 | positive |
| Viewer B | MTD 3211 | 284 | positive |
| Viewer B | MTD 3214 | 299 | positive |
| Viewer B | MTD 3224 | 311 | negative |
| Viewer B | MTD 3226 | 301 | negative |
| Viewer C | MTDC3061 | 297 | positive |
{12}------------------------------------------------
| Viewer C | MTDC3062 | 313 | negative |
|----------|----------|-----|----------|
| Viewer C | MTDC3065 | 291 | positive |
| Viewer C | MTD 3214 | 299 | positive |
| Viewer C | MTD 3216 | 278 | positive |
| Viewer C | MTD 3220 | 306 | negative |
| Viewer C | MTD 3224 | 311 | negative |
## Lay-user study
A lay user study was performed at three intended user sites with 260 lay persons, of which, 20 tested for drug-free samples, 120 for secobarbital samples, 120 for methadone samples. They had diverse educational and professional backgrounds and ranged in age from 21 to >50 years. Urine samples were prepared at the following concentrations; -100%, +/-75%, +/-50%, +/-25% of the cut-off by spiking drugs into drug free-pooled urine specimens. The concentrations of the samples were confirmed by GC/MS. Each sample was aliquoted into individual containers, blind-labeled and randomized. Each participant was provided with the package insert, 1 blind labeled sample and a device. The results are summarized below:
| Cup format | | Number<br>of<br>samples | OTC user | | % Agreement<br>With<br>GC/MS |
|--------------|---------------|-------------------------|----------|----------|------------------------------|
| Drug | Concentration | | Negative | Positive | |
| Drug -free | -100% | 20 | 20 | 0 | 100% |
| Secobarbital | -75% | 20 | 20 | 0 | 100% |
| | -50% | 20 | 20 | 0 | 100% |
| | -25% | 20 | 17 | 3 | 85% |
| | +25% | 20 | 4 | 16 | 80% |
| | +50% | 20 | 0 | 20 | 100% |
| | +75% | 20 | 0 | 20 | 100% |
| Methadone | -75% | 20 | 20 | 0 | 100% |
| | -50% | 20 | 20 | 0 | 100% |
| | -25% | 20 | 16 | 4 | 80% |
| | +25% | 20 | 4 | 16 | 80% |
| | +50% | 20 | 0 | 20 | 100% |
| | +75% | 20 | 0 | 20 | 100% |
Lay-users were also given surveys on the ease of understanding the package insert instructions. All lay users indicated that the device instructions can be easily followed A Flesch-Kincaid reading analysis was performed on the package insert and the score revealed a reading grade level of less than 7.
- 3. Clinical Studies
Not applicable.
{13}------------------------------------------------
# 11. Conclusion
Based on the test principle and performance characteristics of the device, it's concluded that CR³ Keyless Split Sample Cup Secobarbital – Methadone is substantially equivalent to the predicate.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.