Wondfo Methadone Urine Test is an immunochromatographic assay for the qualitative determination of methadone in human urine at a cutoff concentration of 300 ng/mL. The test is available in a dip card format and a cup format. It is intended for prescription use and over the counter use. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. Wondfo Morphine (2000) Urine Test is an immunochromatographic assay for the qualitative determination of Morphine in human urine at a cutoff concentration of 2000 ng/mL. The test is available in a dip card format and a cup format. It is intended for prescription use and over the counter use. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.
Device Story
Lateral flow chromatographic immunoassay for qualitative detection of methadone and morphine in human urine. Available in dip card or integrated cup formats. Principle: competitive binding; urine sample migrates via capillary action; drug in sample competes with drug-protein conjugate for antibody-dye conjugate binding sites. Negative result: colored test line appears (drug concentration below cutoff). Positive result: no test line (drug concentration at or above cutoff). Control line confirms proper flow. Used in home or clinical settings; operated by lay users or healthcare professionals. Provides preliminary screening results; requires GC/MS confirmation. Assists in identifying potential drug use; supports clinical decision-making.
Clinical Evidence
No clinical trials performed. Performance established via analytical bench testing and a lay-user study. Lay-user study (n=140 per format/drug) compared results against GC/MS, showing high agreement (97.1%–98.6% total agreement). Analytical precision, specificity, and cross-reactivity studies confirmed performance at cutoff concentrations.
Technological Characteristics
Lateral flow immunochromatographic assay. Components: mouse monoclonal antibody-dye conjugate (gold chloride), immobilized drug-protein conjugate, anti-mouse IgG polyclonal antibody on membrane. Formats: dip card or integrated cup. Qualitative readout. No external energy source. Standalone device.
Indications for Use
Indicated for qualitative detection of methadone (300 ng/mL cutoff) or morphine (2000 ng/mL cutoff) in human urine. Intended for prescription and over-the-counter use in individuals suspected of drug use. Provides preliminary results; requires confirmatory testing (GC/MS).
Regulatory Classification
Identification
An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.
Special Controls
*Classification.* Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Predicate Devices
ACON One Step Drug Screen Test Card (k020771)
Related Devices
K131232 — WONDFO METHADONE URINE TEST (MTD 200), WONDOFO MORPHINE URINE TEST (MOP 100) · Guangzhou Wondfo Biotech Co., Ltd. · May 31, 2013
K253567 — Healgen Accurate Urine Drug Screen Dip Card; Healgen Accurate Home Urine Drug Test Dip Card · Healgen Scientific,, LLC · Dec 23, 2025
K143535 — CR3 Keyless Split Sample Cup Secobarbital- Methadone · Guangzhou Wondfo Biotech Co., Ltd. · Jan 13, 2015
Submission Summary (Full Text)
{0}
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
k112310
B. Purpose for Submission:
New device
C. Measurand:
Methadone, Morphine
D. Type of Test:
Qualitative lateral flow chromatographic immunoassay
E. Applicant:
Guangzhou Wondfo Biotech Co., Ltd.
F. Proprietary and Established Names:
Wondfo Methadone Urine Test
Wondfo Morphine (2000) Urine Test
G. Regulatory Information:
1. Regulation section:
21 CFR §862.3620, Methadone Test System
21 CFR §862.3650, Morphine Test System
2. Classification:
All are Class II
3. Product code:
DJR, DJG
4. Panel:
{1}
Page 2 of 20
Toxicology (91)
## H. Intended Use:
1. Intended use(s):
See indications for use below.
2. Indications(s) for use:
### Wondfo Methadone Urine Test:
Wondfo Methadone Urine Test is an immunochromatographic assay for the qualitative determination of methadone in human urine at a cutoff concentration of 300 ng/mL. The test is available in a dip card format and a cup format. It is intended for prescription use and over the counter use.
The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.
### Wondfo Morphine (2000) Urine Test:
Wondfo Morphine (2000) Urine Test is an immunochromatographic assay for the qualitative determination of Morphine in human urine at a cutoff concentration of 2000 ng/mL. The test is available in a dip card format and a cup format. It is intended for prescription use and over the counter use.
The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.
3. Special conditions for use statement(s):
For prescription and over-the counter use.
4. Special instrument requirements:
Not Applicable
## I. Device Description:
The tests are in different formats: dip card and cup
{2}
Page 3 of 20
The Dip Card format is a test with an integrated cup. It is a rapid test for the qualitative detection of Methadone and Morphine. It is a lateral flow chromatographic immunoassay. When the absorbent end of the test device is immersed into the urine sample, the urine is absorbed into the device by capillary action, mixes with the antibody-dye conjugate, and flows across the pre-coated membrane. When sample drug levels are zero or below the target cut off (the detection sensitivity of the test), antibody-dye conjugate binds to the drug-protein conjugate immobilized in the Test Region (T) of the device. This produces a colored Test line, which regardless of its intensity, indicates a negative result.
When sample drug levels are at or above the target cutoff, the free drug in the binding sample binds to the antibody-dye conjugate preventing the antibody-dye conjugate from binding to the drug-protein conjugate immobilized in the Test region (T) of the device. This prevents the development of a distinct colored band in the test region, indicating a potentially positive result.
To serve as a procedure control, a colored line will appear at the Control Region (C), if the test has been performed properly because of the antibody-dye conjugate binding to antimouse IgG immobilized in the Control Region of the device.
The Cup format is the same function or performance specification of dip card. It is composed of an integrated cup and a functional dip card. When the key is inserted to the cup and rotated to activate the test, the urine is absorbed into the device by capillary action, mixes with the antibody-dye conjugate, and flows across the pre-coated membrane. When sample drug levels are zero or below the target cut off (the detection sensitivity of the test), antibody-dye conjugate binds to the drug-protein conjugate immobilized in the Test Region (T) of the device. This produces a colored Test line, which regardless of its intensity, indicates a negative result.
When sample drug levels are at or above the target cutoff, the free drug in the binding sample binds to the antibody-dye conjugate preventing the antibody-dye conjugate from binding to the drug-protein conjugate immobilized in the Test region (T) of the device. This prevents the development of a distinct colored band in the test region, indicating a potentially positive result.
To serve as a procedure control, a colored line will appear at the Control Region (C), if the test has been performed properly because of the antibody-dye conjugate binding to antimouse IgG immobilized in the Control Region of the device.
J. Substantial Equivalence Information:
1. Predicate device name(s):
ACON One Step Drug Screen Test Card
{3}
Page 4 of 20
2. Predicate 510(k) number(s):
k020771
3. Comparison with predicate:
| Reagent Similarities and Differences | | |
| --- | --- | --- |
| Feature | Candidate Device: Wondfo Methadone Urine Test and Wondfo Morphine (2000) Urine Test (k112310) | Predicate Device: ACON One Step Drug Screen Card (k020771) |
| Intended Use /Indications for Use | For the qualitative determination of Methadone, Morphine in individual human urine. | Same (different number of drugs detected) |
| Methodology | Competitive binding, lateral flow immunochromatographic assays based on the principle of antigen-antibody immunochemistry | Same |
| Type of Test | Immunoassay principles that rely on antigen-antibody interactions to indicate positive or negative result | Same |
| Results | Qualitative | Same |
| Specimen Type | Human Urine | Same |
| Cut Off Value | Methadone: 300 ng/ml
Morphine: 2000 ng/ml | Same |
| Configurations | Cup, Dip Card | Card, dip card with integrated cup (same) |
| Intended Use | OTC Use & Prescription Use | Prescription Use |
K. Standard/ Guidance Document Referenced (if applicable):
- In Vitro Diagnostic Devices; Guidance for the Preparation of 510(k) Submission
- Premarket Submission and Labeling Recommendations for Drugs of Abuse Screening Tests, Draft Guidance
L. Test Principle:
Immunochromatograph assay using a lateral flow, one step system for the qualitative detection of Methadone and Morphine in human urine. Each assay uses a mouse monoclonal antibody-dye conjugate against drug with gold chloride and fixed drug-protein conjugate and anti-mouse IgG polyclonal antibody in membrane.
4
{4}
Page 5 of 20
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
The precision study was performed by taking negative urine samples and spiking these with: 600, 525, 450, 375, 300, 225, 150, 75 and 0 ng/ml of methadone, corresponding to ± 25, 50, 75, and 100% of the 300 ng/ml cut-off value; and 4000, 3500, 300, 2500, 2000, 1500, 1000, 500 and 0 ng/ml of morphine, corresponding to ± 25, 50, 75, and 100% of the 2000 ng/ml cut-off value. Two sets of operators performed the testing (group A ran the cup format and group B ran the dip card format) and the samples were divided into 12 sets of 25 (one set per lot per run for each format). Three lots of the cup and three lots of the dip card formats of the Wondfo Methadone Urine Test and Wondfo Morphine (2000) Urine Test were used. Each of three operators tested 2 aliquots at each concentration for each lot per day (2 runs per day for 25 days), resulting in a total of 50 determinations by each operator at each concentration. The operators were blinded. A summary of the results is presented in the tables below.
#### a. Cup Format
##### Methadone
| Lot | -100% Cut-off | -75% Cut-off | -50% Cut-off | -25% Cut-off | Cut-off | +25% Cut-off | +50% Cut-off | +75% Cut-off | + 100% Cut-off |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Neg/Pos | Neg/Pos | Neg/Pos | Neg/Pos | Neg/Pos | Ne/Pos | Neg/Pos | Neg/Pos | Neg/Pos |
| Lot 1 | 50/0 | 50/0 | 50/0 | 50/0 | 6/44 | 0/50 | 0/50 | 0/50 | 0/50 |
| Lot 2 | 50/0 | 50/0 | 50/0 | 50/0 | 4/46 | 0/50 | 0/50 | 0/50 | 0/50 |
| Lot 3 | 50/0 | 50/0 | 50/0 | 50/0 | 5/45 | 0/50 | 0/50 | 0/50 | 0/50 |
##### Morphine
| Lot | -100% Cut-off | -75% Cut-off | -50% Cut-off | -25% Cut-off | Cut-off | +25% Cut-off | +50% Cut-off | +75% Cut-off | + 100% Cut-off |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Neg/Pos | Neg/Pos | Neg/Pos | Neg/Pos | Neg/Pos | Ne/Pos | Neg/Pos | Neg/Pos | Neg/Pos |
| Lot 1 | 50/0 | 50/0 | 50/0 | 50/0 | 6/44 | 0/50 | 0/50 | 0/50 | 0/50 |
| Lot 2 | 50/0 | 50/0 | 50/0 | 50/0 | 4/46 | 0/50 | 0/50 | 0/50 | 0/50 |
| Lot 3 | 50/0 | 50/0 | 50/0 | 50/0 | 4/46 | 0/50 | 0/50 | 0/50 | 0/50 |
#### b. Dip Card Format
{5}
Page 6 of 20
## Methadone
| Lot | -100% Cut-off | -75% Cut-off | -50% Cut-off | -25% Cut-off | Cut-off | +25% Cut-off | +50% Cut-off | +75% Cut-off | + 100% Cut-off |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Neg/Pos | Neg/Pos | Neg/Pos | Neg/Pos | Neg/Pos | Ne/Pos | Neg/Pos | Neg/Pos | Neg/Pos |
| Lot 1 | 50/0 | 50/0 | 50/0 | 50/0 | 5/45 | 0/50 | 0/50 | 0/50 | 0/50 |
| Lot 2 | 50/0 | 50/0 | 50/0 | 50/0 | 7/43 | 0/50 | 0/50 | 0/50 | 0/50 |
| Lot 3 | 50/0 | 50/0 | 50/0 | 50/0 | 5/45 | 0/50 | 0/50 | 0/50 | 0/50 |
## Morphine
| Lot | -100% Cut-off | -75% Cut-off | -50% Cut-off | -25% Cut-off | Cut-off | +25% Cut-off | +50% Cut-off | +75% Cut-off | + 100% Cut-off |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Neg/Pos | Neg/Pos | Neg/Pos | Neg/Pos | Neg/Pos | Ne/Pos | Neg/Pos | Neg/Pos | Neg/Pos |
| Lot 1 | 50/0 | 50/0 | 50/0 | 50/0 | 5/45 | 0/50 | 0/50 | 0/50 | 0/50 |
| Lot 2 | 50/0 | 50/0 | 50/0 | 50/0 | 5/45 | 0/50 | 0/50 | 0/50 | 0/50 |
| Lot 3 | 50/0 | 50/0 | 50/0 | 50/0 | 6/44 | 0/50 | 0/50 | 0/50 | 0/50 |
b. Linearity/assay reportable range:
Not Applicable, the assay is intended for qualitative use
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Control standards are not supplied with this device; however, this device has internal process controls. A colored line appearing in the control region confirms sufficient sample volume and adequate membrane wicking. Users are informed that the test is invalid if a line fails to appear in the control region.
Stability
Accelerated and real time studies have been conducted. Protocols and acceptance criteria were described and found to be acceptable. The manufacturer claims an 18 month shelf life.
d. Detection limit:
Analytical performance of the device around the cutoff is described in Section f. (Assay cut-off) below.
e. Analytical specificity:
{6}
Page 7 of 20
Specificity and cross-reactivity was performed by using three batches of both the cup and dip card formats of the Wondfo Methadone Urine Test and Wondfo Morphine (2000) Urine Tests. The samples were split into two aliquots and one group of operators ran the cup format on one set of aliquots and a separate group of operators ran the dip card format on the other set of aliquots. Thirty drug-free urine samples were tested and all were found to be negative, five were confirmed by GC/MS to have a concentration of zero. Cross-reactivity was determined by spiking the target drug, drug metabolites and the same class compounds that may cross-react with the target drugs into drug-free urine at three different concentrations and measuring the samples in triplicate using three batches of cups and three batches of dip cards.
Methadone
| MTD (Methadone) (Methadone, Cutoff = 300 ng/ml) | Result (Positive at 300 ng/ml) | % Cross Reactivity 100% |
| --- | --- | --- |
| Doxylamine | Positive at 50000 ng/ml | 0.6% |
Morphine
| MOR(Morphine) (Morphine, Cutoff= 2000 ng/ml) | Result (Positive at 2000 ng/ml) | % Cross Reactivity 100% |
| --- | --- | --- |
| Codeine | Positive at 2000 ng/mL | 100% |
| Ethylmorphine | Positive at 5000 ng/mL | 40% |
| Hydrocodone | Positive at 12,500 ng/mL | 16% |
| Hydromorphine | Positive at 5000 ng/mL | 40% |
| Levorphanol | Positive at 75000 ng/mL | 2.7% |
| s-Monoacetylmorphine | Positive at 5000 ng/mL | 40% |
| Morphine 3-b-glucuronide | Positive at 2000 ng/mL | 100% |
| Norcodeine | Positive at 12500 ng/mL | 16% |
| Normorphone | Positive at 50000 ng/mL | 4% |
| Oxycodone | Positive at | 8% |
7
{7}
Page 8 of 20
| | 25000 ng/mL | |
| --- | --- | --- |
| Oxymorphine | Positive at 25000 ng/mL | 8% |
| Procaine | Positive at 150000 ng/mL | 1.3% |
| Thebaine | Positive at 100000 ng/mL | 2% |
The following unrelated compounds were found not to cross-react when tested spiked (100 µg/mL) into drug-free urine, as well as into urine spiked with ± 25% of the cut-off concentration of methadone or Morphine individually:
## Methadone
Acetaminophen
Acetophenetidin
N-Acetylprocainamide
Acetylsalicylic Acid
Aminopyrine
Amitryptyline
Amobarbital
Amoxicillin
Ampicillin
L-Ascorbic Acid
DL-Amphetamine Sulfate
Apomorphine
Aspartame
Atropine
Benzilic Acid
Benzoic Acid
Benzoylecgonine
Benzphetamine
Bilirubin
(+/-) Brompheniramine
Caffeine
Cannabidiol
Cannabinol
Chloralhydrate
Chloramphenicol
Chlorothiazide
(+/-) Chlorpheniramine
Chlorpromazine
Chlorquine
Cholesterol
Clomipramine
Deoxycorticosterone
Dextrmethorphan
Diazepam
Diclofenac
Diflunisal
Digoxin
Diphenydramine
Ecgonine Hydrochloride
Ecgonine Methyl Ester
(-)-Ephedrine
[1R,2S]-Ephedrine
(L)-Epinephrine
Erythromycin
- Estradiol
Estrone-3-Sulfate
Ethyl-p-aminobenzoate
Fenoprofen
Furosemide
Gentisic acid
Hemoglobin
Hydralazine
Hydrochlorothiazide
Hydrocodone
Hydrocortisone
O-Hydroxylhippuric Acid
p-Hydroxylamphetamine
p-Hydroxymethamphetamine
3-Hydroxytyramine
Ibuprofen
Imipramine
Iproniazid
Meprobamate
Methamphetamine
Methoxyphenamine
(+/-)-3,4-Methylenedioxy-amphetamine Hydrochloride
(+/-)-3,4-Methylenedioxy-methamphetamine Hydrochloride
Morphine-3-D-glucoronide
Morphine Sulfate
Nalidixic Acid
Naloxone
Naltrexone
Naproxen
Niacinamide
Nifedipine
Norcodein
Norethindrone
D-Norpropoxyphene
Noscapine
D,L-Octopamine
Oxalic Acid
Oxazepam
Oxolinic Acid
Oxycodone
Oxymetazoline
Papaverine
Penicillin-G
Petazocine Hydrochloride
Pentobarbital
Perphenazine
Phencyclidine
DL-Propranolol
D-Propoxyphene
D-Pseudoephedrine
Quinacrine
Quinidine
Quinine
Ranitidine
Salicylic Acid
Secobarbital
Serotonin
Sulfamethazine
Sulindac
Tetracycine
Tetrahydrocortisone, 3-acetate
Tetrahydrocortisone 3-(D-glucoronide)
Tetrahydrozoline
Thebaine
Thiamine
Thioridazine
DL-Tyrosine
Tolbutamide
Triamterene
Trifluoperazine
Trimethoprim
Trimipramine
Tryptamine
DL-Tryptophan
Tyramine
Uric acid
{8}
Page 9 of 20
| Clonidine | (+/-)-Isoproterenol | Phenelzine | Verapamil |
| --- | --- | --- | --- |
| Cocathylene | Isoxsuprine | Phenobarbital | Zomepirac |
| Temazepam | Ketamine | Phentermine | |
| Cocaine Hydrochloride | Ketoprofen | Trans-2-phenylcyclo-propylamine hydrochloride | |
| Codeine | Labetalol | L-Phenylephrine | |
| Cortisone | Levorphanol | □-Phenylethylamine | |
| (-) Cotinine | Loperamide | Phenylpropanolamine | |
| Creatinine | Mephentermine | Prednisolone | |
| | Maprotiline | Prednisone | |
| | Meperidine | Procaine | |
| | | Promazine | |
| | | Promethazine | |
## Morphine
| 4-Acetamidophenol | Deoxycorticosterone | Meprobamate | |
| --- | --- | --- | --- |
| Acetophenetidin | Dextrmethorphan | Methadone | |
| N-Acetylprocainamide | Diazepam | Methoxyphenamine | DL-Propranolol |
| Acetylsalicylic Acid | Diclofenac | (+)-3,4-Methylenedioxy-amphetamine Hydrochloride | D-Propoxyphene |
| Aminopyrine | Diflunisal | | D-Pseudoephedrine |
| Amitryptyline | Digoxin | (+)-3,4-Methylenedioxy-methamphetamine Hydrochloride | Quinidine |
| Amobarbital | Diphenydramine | | Quinine |
| Amoxicillin | Doxylamine | Nalidixic Acid | Ranitidine |
| Ampicillin | Ecgonine Hydrochloride | Nalorphine | Salicylic Acid |
| L-Ascorbic Acid | Ecgonine Methyl Ester | Naloxone | Secobarbital |
| DL-Amphetamine | (-)-Ephedrine | Naltrexone | Serotonin |
| Apomorphine | Erythromycin | Naproxen | Sulfamethazine |
| Aspartame | □-Estradiol | Niacinamide | Sulindac |
| Atropine | Estrone-3-Sulfate | Nifedipine | Temazepam |
| Benzilic Acid | Ethyl-p-aminobenzoate | Norethindrone | Tetracycine |
| Benzoic Acid | Fenoprofen | D-Norpropoxyphene | Tetrahydrocortisone, 3-acetate |
| Benzoylecgonine | Furosemide | Noscapine | |
| Benzphetamine | Gentisic acid | D,L-Octopamine | Tetreahydrocortisone 3-(□-D-glucoronide |
| Bilirubin | Hemoglobin | Oxalic Acid | Tetrahydrozoline |
| (+/-) Brompheniramine | Hydralazine | Oxazepam | Thiamine |
| Caffeine | Hydrochlorothiazide | Oxolinic Acid | Thioridazine |
| Cannabidiol | Hydrocortisone | Oxymetazoline | DL-Tyrosine |
| Chloralhydrate | O-Hydroxylhippuric Acid | Papaverine | Tolbutamide |
| Chloramphenicol | p-Hydroxymethamphetamine | Penicillin-G | Triamterene |
| Chlordiazepoxide | 3-Hydroxytyramine | Petazocine | Trifluoperazine |
| Chlorothiazide | Ibuprofen | Pentobarbital | Trimethoprim |
| (+/-) Chlorpheniramine | Imipramine | Perphenazine | Trimipramine |
| Chlorpromazine | Iproniazid | Phencyclidine | Tryptamine |
| Chlorquine | (+/-)-Isoproterenol | Phenelzine | DL-Tryptophan |
| Cholesterol | Isoxsuprine | Phenobarbital | Tyramine |
| Clomipramine | Ketamine | Phentermine | |
9
{9}
Page 10 of 20
Clonidine
Ketoprofen
L-Phenylephrine
Uric acid
Cocaine Hydrochloride
Labetalol
L-Phenylethylamine
Verapamil
Cortisone
Loperamide
Phenylpropanolamine
Zomepirac
(-) Cotinine
Maprotiline
Prednisone
Creatinine
Meperidine
The pH of a drug-free urine pool was adjusted to pH 4, pH 5, pH 6, pH 7, pH 8 or pH 9; the urine was spiked with ± 25% of the cut-off concentration of methadone or morphine, individually and tested using three batches of cup and three batches of dip card of the Wondfo Methadone Urine Test and Wondfo Morphine (2000) Urine Tests, respectively. The spiked, pH-adjusted urine was tested in triplicate. Altering the pH of the urine sample did not affect the accuracy of any of the test results.
The specific gravity of a drug-free urine pool was measured by obtaining urine samples with specific gravities of 1.000-1.035 and spiking with ± 25% of the cut-off concentration of methadone and morphine, individually using three batches of cup and three batches of dip card of the Wondfo Methadone Urine Test and Wondfo Morphine (2000) Urine Test, respectively. The spiked urines of different specific gravities were tested in triplicate. The results indicate that there is stable performance of the Wondfo Methadone Urine Test and Wondfo Morphine Urine Tests when urine has a specific gravity between 1.00-1.035.
f. Assay cut-off:
The assay cut off was investigated by using three batches of both the cup and dip card of the Wondfo Methadone Urine Test and Wondfo Morphine (2000) Urine Test to test 150 urine samples each. These samples came from the Shezhen Drug Addiction Recovery Center (25 Drug urine samples each with Methadone or morphine) and from drug-free urine samples (125 samples) spiked with methadone or morphine (125 samples for each drug) diluted from the International Drug Standard (Sigma) to concentrations that are plus and minus 25% and 50% of the cutoff concentrations. Drug concentrations were confirmed by GC/MS. Results were read by three laboratory assistants with relevant experience. The cutoff studies were performed by two separate groups of operators (one for the strip format and one for the cassette format). Three operators in each group performed the readings and they were blinded to the sample. Each result was confirmed by at least two assistants.
a. Cup
Methadone
| [Drug] (ng/ml) | Cut-off range | Batch 1 | | Batch 2 | | Batch 3 | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | - | + | - | + | - | + | - | + |
| ~150 | -50% Cutoff | 90 | 0 | 90 | 0 | 90 | 0 | 270 | 0 |
| ~225 | -25% Cutoff | 90 | 0 | 90 | 0 | 90 | 0 | 270 | 0 |
{10}
Page 11 of 20
| ~300 | Cut-off | 13 | 77 | 12 | 78 | 10 | 80 | 35 | 235 |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| ~375 | +25% Cut-off | 0 | 90 | 0 | 90 | 0 | 90 | 0 | 270 |
| ~450 | +50% Cut-off | 0 | 90 | 0 | 90 | 0 | 90 | 0 | 270 |
## Morphine
| [Drug] (ng/ml) | Cut-off range | Batch 1 | | Batch 2 | | Batch 3 | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | - | + | - | + | - | + | - | + |
| ~1000 | -50% Cutoff | 90 | 0 | 90 | 0 | 90 | 0 | 270 | 0 |
| ~1500 | -25% Cutoff | 90 | 0 | 90 | 0 | 90 | 0 | 270 | 0 |
| ~2000 | Cut-off | 10 | 80 | 10 | 80 | 12 | 78 | 32 | 238 |
| ~2500 | +25% Cut-off | 0 | 90 | 0 | 90 | 0 | 90 | 0 | 270 |
| ~3000 | +50% Cut-off | 0 | 90 | 0 | 90 | 0 | 90 | 0 | 270 |
b. Dip Strip:
## Methadone
| [Drug] (ng/ml) | Cut-off range | Batch 1 | | Batch 2 | | Batch 3 | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | - | + | - | + | - | + | - | + |
| ~150 | -50% Cutoff | 90 | 0 | 90 | 0 | 90 | 0 | 270 | 0 |
| ~225 | -25% Cutoff | 90 | 0 | 90 | 0 | 90 | 0 | 270 | 0 |
| ~300 | Cut-off | 11 | 79 | 12 | 78 | 8 | 82 | 31 | 239 |
| ~375 | +25% Cut-off | 0 | 90 | 0 | 90 | 0 | 90 | 0 | 270 |
| ~450 | +50% Cut-off | 0 | 90 | 0 | 90 | 0 | 90 | 0 | 270 |
## Morphine
| [Drug] (ng/ml) | Cut-off range | Batch 1 | | Batch 2 | | Batch 3 | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | - | + | - | + | - | + | - | + |
| ~1000 | -50% Cutoff | 90 | 0 | 90 | 0 | 90 | 0 | 270 | 0 |
| ~1500 | -25% Cutoff | 90 | 0 | 90 | 0 | 90 | 0 | 270 | 0 |
| ~2000 | Cut-off | 13 | 77 | 11 | 79 | 10 | 80 | 34 | 236 |
| ~2500 | +25% Cut-off | 0 | 90 | 0 | 90 | 0 | 90 | 0 | 270 |
11
{11}
Page 12 of 20
The cut-off values were determined to be:
| Test | Calibrator | Cut-off (ng/ml) |
| --- | --- | --- |
| Methadone | Methadone | 300 |
| Morphine | Morphine | 2000 |
# 2. Comparison studies:
a. Method comparison with predicate device:
1. Performance of the Wondfo Methadone Urine Test and Wondfo Morphine (2000) Urine Test (cup and dip stick formats) was established by comparing 80 samples of each analyte against GC/MS. These samples came from drug-free urine (10 samples each), the Shezhen Drug Addiction Recovery Center (70 samples for each analyte) and from clinical samples with concentrations ranges as follows: drug free urine, $< -50\%$ of the cut-off, $-50\%$ of the cut-off to the cut-off, cut-off to $+50\%$ of the cut-off, and $> +50\%$ of the Cut-off. There were two groups of operators (one for the cup format and one for the dip stick format) and they were blinded to the samples. Each result was read by three laboratory assistants with relevant experience and a lay person with no experience other than reading the instructions for use.
a. Cup
Methadone:
| Wondfo Device Result | | Drug-free urine | < -50% of the cut-off | -50% of the cut-off to the cut-off | cut-off to +50% of the cut-off | > +50% of the Cut-off |
| --- | --- | --- | --- | --- | --- | --- |
| Viewer A | + | 0 | 0 | 2 | 19 | 21 |
| | - | 10 | 12 | 16 | 0 | 0 |
| Viewer B | + | 0 | 0 | 1 | 19 | 21 |
| | - | 10 | 12 | 17 | 0 | 0 |
| Viewer C | + | 0 | 0 | 2 | 19 | 21 |
| | - | 10 | 12 | 16 | 0 | 0 |
| Lay Person | + | 0 | 0 | 3 | 19 | 21 |
| | - | 10 | 12 | 15 | 0 | 0 |
{12}
Page 13 of 20
## Morphine
| Wondfo Device Result | | Drug-free urine | < -50% of the cut-off | -50% of the cut-off to the cut-off | cut-off to +50% of the cut-off | > +50% of the Cut-off |
| --- | --- | --- | --- | --- | --- | --- |
| Viewer A | + | 0 | 0 | 1 | 18 | 22 |
| | - | 10 | 20 | 9 | 0 | 0 |
| Viewer B | + | 0 | 0 | 1 | 18 | 22 |
| | - | 10 | 20 | 9 | 0 | 0 |
| Viewer C | + | 0 | 0 | 1 | 18 | 22 |
| | - | 10 | 20 | 9 | 0 | 0 |
| Lay Person | + | 0 | 0 | 2 | 18 | 22 |
| | - | 10 | 20 | 8 | 0 | 0 |
b. Dip Card
## Methadone:
| Wondfo Device Result | | Drug-free urine | < -50% of the cut-off | -50% of the cut-off to the cut-off | cut-off to +50% of the cut-off | > +50% of the Cut-off |
| --- | --- | --- | --- | --- | --- | --- |
| Viewer A | + | 0 | 0 | 1 | 19 | 21 |
| | - | 10 | 12 | 17 | 0 | 0 |
| Viewer B | + | 0 | 0 | 2 | 19 | 21 |
| | - | 10 | 12 | 16 | 0 | 0 |
| Viewer C | + | 0 | 0 | 1 | 19 | 21 |
| | - | 10 | 12 | 17 | 0 | 0 |
| Lay Person | + | 0 | 0 | 2 | 19 | 21 |
| | - | 10 | 12 | 16 | 0 | 0 |
## Morphine
| Wondfo Device Result | | Drug-free urine | < -50% of the cut-off | -50% of the cut-off to the cut-off | cut-off to +50% of the cut-off | > +50% of the Cut-off |
| --- | --- | --- | --- | --- | --- | --- |
| Viewer A | + | 0 | 0 | 1 | 18 | 22 |
| | - | 10 | 20 | 9 | 0 | 0 |
| Viewer B | + | 0 | 0 | 1 | 18 | 22 |
13
{13}
Page 14 of 20
| B | - | 10 | 20 | 9 | 0 | 0 |
| --- | --- | --- | --- | --- | --- | --- |
| Viewer C | + | 0 | 0 | 1 | 18 | 22 |
| | - | 10 | 20 | 9 | 0 | 0 |
| Lay Person | + | 0 | 0 | 1 | 18 | 22 |
| | - | 10 | 20 | 9 | 0 | 0 |
The discordant results are listed in the table below.
a. Cup:
| Viewer | Sample Number | GC/MS result | Viewer Result |
| --- | --- | --- | --- |
| Viewer A | MTDC63 | 278 | positive |
| Viewer A | MTD 212 | 280 | positive |
| Viewer B | MTD 212 | 280 | positive |
| Viewer C | MTD 212 | 280 | positive |
| Viewer C | MTD 216 | 280 | positive |
| Lay person | MTDC63 | 278 | positive |
| Lay person | MTD 212 | 280 | positive |
| Lay person | MTD 216 | 280 | positive |
| Viewer A | OPIC61 | 1906 | positive |
| Viewer B | OPIC65 | 1923 | positive |
| Viewer C | OPIC61 | 1906 | positive |
| Lay person | OPIC61 | 1906 | positive |
| Lay person | OPIC65 | 1923 | positive |
b. Dip Card:
{14}
Page 15 of 20
| Viewer | Sample Number | GC/MS result | Viewer Result |
| --- | --- | --- | --- |
| Viewer A | MTD 212 | 280 | positive |
| Viewer B | MTDC63 | 278 | positive |
| Viewer B | MTD 216 | 280 | positive |
| Viewer C | MTD 212 | 280 | positive |
| Lay person | MTDC63 | 278 | positive |
| Lay person | MTD 212 | 280 | positive |
| Viewer A | OPIC65 | 1923 | positive |
| Viewer B | OPIC61 | 1906 | positive |
| Viewer C | OPIC61 | 1906 | positive |
| Lay person | OPIC65 | 1923 | positive |
The overall agreement between the Wondfo devices and GC/MS is presented in the table below:
| % Agreement | MTD (Cup) | MTD (Dip Card) | MOR (Cup) | MOR (Dip Card) |
| --- | --- | --- | --- | --- |
| Positive | 100% | 100% | 100% | 100% |
| Negative | 95.8% | 96.7% | 97.5% | 97.5% |
| Overall | 97.9% | 98.4% | 98.8% | 98.8% |
The overall agreement between the lay viewer and experienced viewer is presented in the table below:
| % Agreement | MTD (Cup) | MTD (Dip Card) | MOR (Cup) | MOR (Dip Card) |
| --- | --- | --- | --- | --- |
| Positive | 100% | 100% | 100% | 100% |
| Negative | 96.6% | 98.2% | 97.4% | 100% |
| Overall | 98.3% | 99.1% | 98.7% | 100% |
B. A lay-user study was performed to assess the suitability of the device for home use. Six drug-free urine sample pools were spiked to $\pm 25\%$, $\pm 50\%$, $\pm 75\%$ of the cutoff of target drug, additionally a negative urine pool with no drug was tested. These concentrations were confirmed by GC/MS and aliquoted into 40 individual containers per concentration (20 aliquots per concentration for each strip format). The testing was performed at three sites by 140 blinded consumers divided between three sites (140 users for cup, 140 for dip card for each of the Wondfo
{15}
Page 16 of 20
Methadone Urine Test and Wondfo Morphine (2000) Urine Test). The lay users were chosen from likely intended users at the Guangzhou No. 8 People's Hospital, Guangzhou Mental Hospital, and Guangdong Provincial No. 2 People's Hospital. Each participant received the package inset, 1 blinded sample, and either a test cup or dip card. The lay persons test result was compared to the GC/MS result to demonstrate accuracy by lay-users. The following are the results of the lay-user study pooled together from all three sites for the cup format:
Methadone
| % of Cutoff | Number of Samples | Methadone GC/MS Concentration (ng/mL) | Lay person results: Pos/Neg | Percentage of Correct Results |
| --- | --- | --- | --- | --- |
| -100% of the Cut-off | 20 | 0 | 0/20 | 100 |
| -75 % Cutoff | 20 | 69.5 | 0/20 | 100 |
| -50 % Cutoff | 20 | 186.6 | 0/20 | 100 |
| -25 % Cutoff | 20 | 248.2 | 3/17 | 85.0 |
| +25% Cutoff | 20 | 386.7 | 19/1 | 95.0 |
| +50 % Cutoff | 20 | 469.8 | 20/0 | 100 |
| +75 % Cutoff | 20 | 550.6 | 20/0 | 100 |
16
{16}
Page 17 of 20
Morphine:
| % of Cutoff | Number of Samples | Morphine GC/MS Concentration (ng/mL) | Lay person results: Pos/Neg | Percentage of Correct Results |
| --- | --- | --- | --- | --- |
| -100% of the Cut-off | 20 | 0 | 0/20 | 100 |
| -75 % Cutoff | 20 | 478.7 | 0/20 | 100 |
| -50 % Cutoff | 20 | 1080.8 | 0/20 | 100 |
| -25 % Cutoff | 20 | 1409.5 | 2/18 | 90.0 |
| +25% Cutoff | 20 | 2625.4 | 19/1 | 95.0 |
| +50 % Cutoff | 20 | 2900.6 | 20/0 | 100 |
| +75 % Cutoff | 20 | 3605.2 | 20/0 | 100 |
The following are the results of the lay-user study pooled together from all three sites for the dip card format:
Methadone:
| % of Cutoff | Number of Samples | Methadone GC/MS Concentration (ng/mL) | Lay person results: Pos/Neg | Percentage of Correct Results |
| --- | --- | --- | --- | --- |
| -100% of the Cut-off | 20 | 0 | 0/20 | 100 |
| -75 % Cutoff | 20 | 86.2 | 0/20 | 100 |
| -50 % Cutoff | 20 | 173.5 | 0/20 | 100 |
| -25 % Cutoff | 20 | 203.4 | 2/18 | 90.0 |
| +25% Cutoff | 20 | 401.6 | 19/1 | 95.0 |
| +50 % Cutoff | 20 | 466.5 | 20/0 | 100 |
| +75 % Cutoff | 20 | 560.8 | 20/0 | 100 |
17
{17}
Page 18 of 20
Morphine:
| % of Cutoff | Number of Samples | Morphine GC/MS Concentration (ng/mL) | Lay person results: Pos/Neg | Percentage of Correct Results |
| --- | --- | --- | --- | --- |
| -100% of the Cut-off | 20 | 0 | 0/20 | 100 |
| -75 % Cutoff | 20 | 478.7 | 0/20 | 100 |
| -50 % Cutoff | 20 | 1080.8 | 0/20 | 100 |
| -25 % Cutoff | 20 | 1409.5 | 2/18 | 90.0 |
| +25% Cutoff | 20 | 2625.4 | 20/0 | 100 |
| +50 % Cutoff | 20 | 2900.6 | 20/0 | 100 |
| +75 % Cutoff | 20 | 3605.2 | 20/0 | 100 |
The overall percent agreement between the Lay person and the GC/MS method was:
| % Agreement | Methadone (Cup) | Methadone (Dip Card) | Morphine (Cup) | Morphine (Dip Card) |
| --- | --- | --- | --- | --- |
| Total | 97.1% | 97.8% | 97.8% | 98.6% |
C. The subject’s were to fill out a questionnaire to evaluate labeling access.
i. Cup format
**Methadone:**
The participants’s ages ranged from 23-59 and there were slightly more males than females. They come from a variety of career and education backgrounds and 2 had used a home drug kit before. They all understood the storage and expiration of the device and that the test could not be reused. They all understood to read the result within 5 minutes and how to interpret the results. Two of the participants responded that the cup test was difficult, 69/140 said the test was very easy and 69/140 said it was easy. 72/140 said the instructions were very clear and 68/140 said they were clear. 88/140 said the interpretation of the results was very clear and 52/140 said they were clear. When asked what they would do if the device failed to show a result, 109/140 said they would
18
{18}
Page 19 of 20
repeat the test with a new test device 2/140 said they would turn to another method, and 29/140 said they would call the manufacturer if the device failed to show a result.
## Morphine:
The participant’s ages ranged from 22-59 and they were evenly distributed between males and females. They come from a variety of career and education backgrounds and three had used a home drug kit before. They all understood the storage and expiration of the device and that the test could not be reused. They all understood to read the result within 5 minutes and how to interpret the results. One of the participants responded that the strip test was difficult, 70/140 said the test was very easy and 69/140 said it was easy. 66/140 said the instructions were very clear and 74/140 said they were clear. 79/140 said the interpretation of the results was very clear and 61/140 said they were clear. When asked what they would do if the device failed to show a result, 105/140 said they would repeat the test with a new test device, 2/140 said they would turn to another method, and 33/140 said they would call the manufacturer if the device failed to show a result.
## ii. Dip Card format
## Methadone:
The participant’s ages ranged from 23-64 and there were slightly more males than females. They come from a variety of career and education backgrounds and only 1 subject had used a home drug kit before. They all understood the storage and expiration of the device and that the test could not be reused. They all understood to read the result within 5 minutes and how to interpret the results. Two of the participants responded that the strip test was difficult, 70/140 said the test was very easy and 58/140 said it was easy. 55/140 said the instructions were very clear and 85/140 said they were clear. 90/140 said the interpretation of the results was very clear and 50/140 said they were clear. When asked what they would do if the device failed to show a result, 114/140 said they would repeat the test with a new test device, 2/140 said they would turn to another method, and 24/140 said they would call the manufacturer if the device failed to show a result.
## Morphine:
The participant’s ages ranged from 21-62 and there were slightly more males than females. They come from a variety of career and education backgrounds and none of the participants had used a home drug kit before. They all understood the storage and expiration of the device and that the test could not be reused. They all understood to read the result within 5 minutes and how to interpret the results. One of the participants responded that the strip test was difficult, 60/140 said the test was very easy and 79/140 said it was easy. 79/140 said the instructions were very clear and 61/140 said they were clear. 92/140
{19}
Page 20 of 20
said the interpretation of the results was very clear and 48/140 said they were clear. When asked what they would do if the device failed to show a result, 107/140 said they would repeat the test with a new test device, 3/140 said they would turn to another method, and 30/140 said they would call the manufacturer if the device failed to show a result.
Additionally, a Flesh-Kincaid reading analysis revealed that both package inserts had a reading grade level of 7.
b. Matrix comparison:
Not applicable.
3. Clinical studies:
a. Clinical Sensitivity:
Not Applicable
b. Clinical specificity:
Not Applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not Applicable
4. Clinical cut-off:
Not Applicable
5. Expected values/Reference range
Specific ranges for each analyte/methodology are listed in the package insert.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.