Tosoh Automated Glycohemoglobin Analyzer HLC-723GR01

K250073 · Tosoh Bioscience, Inc. · PDJ · Oct 3, 2025 · Clinical Chemistry

Device Facts

Record IDK250073
Device NameTosoh Automated Glycohemoglobin Analyzer HLC-723GR01
ApplicantTosoh Bioscience, Inc.
Product CodePDJ · Clinical Chemistry
Decision DateOct 3, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1373
Device ClassClass 2

Intended Use

The Tosoh Automated Glycohemoglobin Analyzer HLC-723GR01 is intended for in vitro diagnostic use for the quantitative measurement of % hemoglobin A1c (HbA1c) (DCCT/NGSP) and mmol/mol hemoglobin A1c (IFCC) in human venous whole blood and hemolysate specimens using ion-exchange high performance liquid chromatography (HPLC). This test is to be used as an aid in diagnosis of diabetes and identifying patients who may be at risk for developing diabetes, and for monitoring of long-term blood glucose control in individuals with diabetes mellitus.

Device Story

The Tosoh Automated Glycohemoglobin Analyzer HLC-723GR01 is an automated clinical laboratory instrument for HbA1c quantification. It processes human venous whole blood or hemolysate samples via ion-exchange high-performance liquid chromatography (HPLC). The system hemolyzes and dilutes samples, injecting them onto a TSKgel GR01 HbA1c Column. Separation of hemoglobin fractions (A1a, A1b, HbF, LA1c+, SA1c, A0) occurs via step-gradient elution with three buffers of varying ionic strength. A flow cell monitors absorbance changes at 415nm (with 500nm background correction). The device identifies common hemoglobin variants (D, S, C, E) based on retention time windows. Output includes sample ID, date, percentage/retention time of fractions, SA1c percentage, and chromatograms. Results are stored in internal memory (up to 150,000 results) and exportable via USB. Healthcare providers use these results to diagnose diabetes and monitor long-term glycemic control, facilitating clinical management of diabetes mellitus.

Clinical Evidence

Bench testing only. Precision study (n=720) across three sites/analyzers showed total CVs 0.95-1.12% (NGSP) and 1.14-1.87% (IFCC). Linearity confirmed 3.9-16.9% HbA1c range (R²=1.000). Method comparison (n=135) against NGSP secondary reference laboratory showed high correlation (R=1.000) for both whole blood and hemolysate. Interference testing confirmed no significant bias from endogenous/exogenous substances or hemoglobin variants (HbS, HbC, HbD, HbE, HbA2, HbF).

Technological Characteristics

Ion-exchange HPLC; TSKgel GR01 cation exchange column; three-buffer step gradient elution; 415 nm absorbance detector. Automated dilution/hemolysis. Barcode reader for sample ID. Calibrated every 30 days or upon column change/QC failure. Standards: CLSI EP05-A3, EP06, EP07, EP09c-A3; IEC 60601-1-2.

Indications for Use

Indicated for use in clinical laboratories for quantitative measurement of HbA1c in human venous whole blood and hemolysate specimens to aid in the diagnosis of diabetes, identification of patients at risk for diabetes, and monitoring of long-term blood glucose control in individuals with diabetes mellitus.

Regulatory Classification

Identification

A hemoglobin A1c test system is a device used to measure the percentage concentration of hemoglobin A1c in blood. Measurement of hemoglobin A1c is used as an aid in the diagnosis of diabetes mellitus and as an aid in the identification of patients at risk for developing diabetes mellitus.

Special Controls

(b) Classification. Class II (special controls). Hemoglobin A1c test systems must comply with the following special controls: 1) The device must have initial and annual standardization verification by a certifying glycohemoglobin standardization organization deemed acceptable by FDA. 2) The premarket notification submission must include performance testing to evaluate precision, accuracy, linearity and interference, including the following: i) Performance testing of device precision must, at a minimum, use blood samples with concentrations near 5.0%, 6.5%, 8.0% and 12% hemoglobin A1c. This testing must evaluate precision over a minimum of 20 days using at least 3 lots of the device and 3 instruments, as applicable. ii) Performance testing of device accuracy must include a minimum of 120 blood samples that span the measuring interval of the new device and compare results of the new device to results of the standardized test method. Results must demonstrate little or no bias versus the standardized method. iii) Total error of the new device must be evaluated using single measurements by the new device compared to results of the standardized test method, and this evaluation must demonstrate a total error less than or equal to 6%. iv) Performance testing must demonstrate that there is little to no interference from common hemoglobin variants, including Hemoglobin C, Hemoglobin D, Hemoglobin E, Hemoglobin A2 and Hemoglobin S. 3) When assay interference from Hemoglobin F or interference with other hemoglobin variants with low frequency in the population is observed, a warning statement must be placed in a black box and must appear in all labeling material for these devices describing the interference and any affected populations.

*Classification.* Class II (special controls). The special controls for this device are:(1) The device must have initial and annual standardization verification by a certifying glycohemoglobin standardization organization deemed acceptable by FDA. (2) The premarket notification submission must include performance testing to evaluate precision, accuracy, linearity, and interference, including the following: (i) Performance testing of device precision must, at a minimum, use blood samples with concentrations near 5.0 percent, 6.5 percent, 8.0 percent, and 12 percent hemoglobin A1c. This testing must evaluate precision over a minimum of 20 days using at least three lots of the device and three instruments, as applicable. (ii) Performance testing of device accuracy must include a minimum of 120 blood samples that span the measuring interval of the device and compare results of the new device to results of a standardized test method. Results must demonstrate little or no bias versus the standardized method. (iii) Total error of the new device must be evaluated using single measurements by the new device compared to results of the standardized test method, and this evaluation must demonstrate a total error less than or equal to 6 percent. (iv) Performance testing must demonstrate that there is little to no interference from common hemoglobin variants, including Hemoglobin C, Hemoglobin D, Hemoglobin E, Hemoglobin A2, and Hemoglobin S. (3) When assay interference from Hemoglobin F or interference with other hemoglobin variants with low frequency in the population is observed, a warning statement must be placed in a black box and must appear in all labeling material for these devices describing the interference and any affected populations.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION October 3, 2025 Tosoh Bioscience, Inc. Corike Nuibe Sr. Regulatory Affairs Specialist 3600 Gantz Road Grove City, Ohio 43123 Re: K250073 Trade/Device Name: Tosoh Automated Glycohemoglobin Analyzer HLC-723GR01 Regulation Number: 21 CFR 862.1373 Regulation Name: Hemoglobin A1c Test System Regulatory Class: Class II Product Code: PDJ, LCP Dated: December 30, 2024 Received: January 10, 2025 Dear Corike Nuibe: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K250073 - Corike Nuibe Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K250073 - Corike Nuibe Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Leslie Landree-S 2025.10.03 18:14:57 -04'00' for Joshua M. Balsam, Ph.D. Branch Chief Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K250073 | | | Device Name Tosoh Automated Glycohemoglobin Analyzer HLC-723GR01 | | | Indications for Use (Describe) | | | The Tosoh Automated Glycohemoglobin Analyzer HLC-723GR01 is intended for in vitro diagnostic use for the quantitative measurement of % hemoglobin A1c (HbA1c) (DCCT/NGSP) and mmol/mol hemoglobin A1c (IFCC) in human venous whole blood and hemolysate specimens using ion-exchange high performance liquid chromatography (HPLC). | | | This test is to be used as an aid in diagnosis of diabetes and identifying patients who may be at risk for developing diabetes, and for monitoring of long-term blood glucose control in individuals with diabetes mellitus. | | | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {4} TOSOH TOSOH BIOSCIENCE Subject Device: Tosoh Automated Glycohemoglobin Analyzer HLC-723GR01 # 510k Summary (Summary of Safety and Effectiveness) This 510(k) Summary is being submitted in accordance with requirements of Title 21 CFR 862.1373. The assigned 510(k) Number is: K250073 Date of Preparation: October 1st, 2025 ## 1. Applicant Information Name: Tosoh Bioscience, Inc. Address: 3600 Gantz Road Grove City OH 43123 USA Contact: Corike Nuibe (Sr. Regulatory Affairs Specialist) Email: corike.nuibe@tosoh.com Phone: (385) 279-8114 ## 2. Device Name/Trade Name Device Trade Name: Tosoh Automated Glycohemoglobin Analyzer HLC-723GR01 Classification Name: Hemoglobin A1c test system Product Code: PDJ, LCP Regulation Number: 21 CFR Part 862.1373 and 21 CFR Part 864.7470 Regulation Name: Hemoglobin A1c test system, Glycosylated hemoglobin assay Device Class: Class II 510(k) Review Panel: Clinical Chemistry Regulation Medical Specialty: Clinical Chemistry ## 3. Predicate Device Trade name: Tosoh Automated Glycohemoglobin Analyzer HLC-723G8 510(k) submitter/holder: Tosoh Bioscience, Inc. 510(k) Numbers: K200904 ## 4. Description of the Device The Tosoh Automated Glycohemoglobin Analyzer HLC-723GR01 is intended for in vitro diagnostic use for the quantitative measurement of % hemoglobin A1c (HbA1c) Page 1 of 13 {5} TOSOH TOSOH BIOSCIENCE Subject Device: Tosoh Automated Glycohemoglobin Analyzer HLC-723GR01 (DCCT/NGSP) and mmol/mol hemoglobin A1c (IFCC) in human venous whole blood and hemolysate specimens using ion-exchange high performance liquid chromatography (HPLC). The Tosoh Automated Glycohemoglobin Analyzer HLC-723GR01 analyzer is to be used in the clinical laboratory setting. The analyzer hemolyzes and dilutes whole blood specimens and subsequently injects the diluted and hemolyzed sample into the injection valve and onto the TSKgel GR01 HbA1c Column. Manually hemolyzed and prediluted samples can be loaded in sample cups directly for analysis on the analyzer. Separation of hemoglobin fractions is achieved by using differences in ionic interactions between the cation exchange group on the column resin surface and the hemoglobin components by a step gradient elution with three elution buffers of different ionic strengths. Changes in light absorbance at 415nm caused by hemoglobin fractions eluting from the column are monitored. The GR01 software has chromatographic peak retention time windows for the six expected hemoglobin fractions, A1a, A1b, HbF, LA1c+, SA1c, and A0. In addition, the software has retention time windows for the presumptive identification of the common hemoglobin variants, D, S, C, and E. An analysis requires 50 seconds per sample to complete. A printout of the results includes the sample ID, date, percentage, and retention time of each hemoglobin fraction, the SA1c percentage along with a chromatogram of the elution pattern of the hemoglobin fractions. The analyzer is equipped with external USB ports. These can be used to store assay results, to update and backup program versions, to attach an external device, for example an external printer or an optional handheld barcode scanner, and for software installation or backup of data. The last 150,000 assay results are automatically saved to the analyzer's internal memory and can be retrieved as list data in .CSV format and chromatograms in .PDF format. The Tosoh Automated Glycohemoglobin Analyzer HLC-723GR01 system consists of the following components. - GR01 HbA1c Elution Buffer No. 1, No. 2, No. 3 - TSKgel® GR01 HbA1c Column - Hemoglobin A1c Controls Levels 1 and 2 - 21 CFR 862.1660, Product Code JJX - Hemoglobin A1c Calibrator Set - 21 CFR 862.1150, Product Code JIT - Hemolysis and Wash Solution (L) and (LL) - 21 CFR 864.8540, Product Code GGK Page 2 of 13 {6} TOSOH TOSOH BIOSCIENCE Subject Device: Tosoh Automated Glycohemoglobin Analyzer HLC-723GR01 # 5. Indications for Use The Tosoh Automated Glycohemoglobin Analyzer HLC-723GR01 is intended for in vitro diagnostic use for the quantitative measurement of % hemoglobin A1c (HbA1c) (DCCT/NGSP) and mmol/mol hemoglobin A1c (IFCC) in human venous whole blood and hemolysate specimens using ion-exchange high performance liquid chromatography (HPLC). This test is to be used as an aid in diagnosis of diabetes and identifying patients who may be at risk for developing diabetes, and for monitoring of long-term blood glucose control in individuals with diabetes mellitus. # 6. Substantial Equivalence Information The Tosoh Automated Glycohemoglobin Analyzer HLC-723GR01 is substantially equivalent to the claimed predicate device; the Tosoh Automated Glycohemoglobin Analyzer HLC-723G8 (k200904), based on comparisons of the intended use and technological characteristics. Table 1. Comparison Table of Subject Device to Predicate Device | Feature | Candidate Device: Tosoh Automated Glycohemoglobin Analyzer HLC-723GR01 | Predicate Device: Tosoh Automated Glycohemoglobin Analyzer HLC-723G8, v5.24F | | --- | --- | --- | | Intended Use | The Tosoh Automated Glycohemoglobin Analyzer HLC-723GR01 is intended for in vitro diagnostic use for the quantitative measurement of % hemoglobin A1c (HbA1c) (DCCT/NGSP) and mmol/mol hemoglobin A1c (IFCC) in human venous whole blood and hemolysate specimens using ion-exchange high performance liquid chromatography (HPLC). This test is to be used as an aid in diagnosis of diabetes and identifying patients who may be at risk for developing diabetes, and for monitoring of long-term blood glucose control in individuals with diabetes mellitus. | Tosoh Automated Glycohemoglobin Analyzer HLC-723G8 is intended for in vitro diagnostic use for the measurement of % hemoglobin A1c (HbA1c) (DCCT/NGSP) and mmol/mol hemoglobin A1c (IFCC) in venous whole blood specimens using ion-exchange high-performance liquid chromatography (HPLC). This test is to be used as an aid in diagnosis of diabetes and identifying patients who may be at risk for developing diabetes, and for monitoring of long- term blood glucose control in individuals with diabetes mellitus. | Page 3 of 13 {7} TOSOH TOSOH BIOSCIENCE Subject Device: Tosoh Automated Glycohemoglobin Analyzer HLC-723GR01 | Feature | Candidate Device: Tosoh Automated Glycohemoglobin Analyzer HLC-723GR01 | Predicate Device: Tosoh Automated Glycohemoglobin Analyzer HLC-723G8, v5.24F | | --- | --- | --- | | Assay Method | Ion-Exchange HPLC | Same | | Detection Method | Analyte passes through a flow cell where changes in absorbance are measured at 415nm and recorded as a digital chromatogram. An additional filter at 500nm corrects for background absorbance. | Same | | Applications/Test Time | 50 Seconds | 1.6 Minutes | | Instrument Platform | Automated Glycohemoglobin Analyzer HLC-723GR01 | Automated Glycohemoglobin Analyzer HLC-723G8 | | Sample Matrix | • Human Venous Whole Blood • Hemolysate | Human Venous Whole Blood | | Sample Anticoagulants | K2 EDTA, K3 EDTA | Same | | Traceability/Standardization | Traceable to the Diabetes Control and Complications Trial (DCCT) reference method and IFCC. Certified via the National Glycohemoglobin Standardization Program (NGSP) | Same | | Measuring Range | 3.9 – 16.9 %HbA1c 19 – 161 mmol/mol HbA1c | 4.0 – 16.9 %HbA1c | | Hemoglobin Variant Interference | Accurate and reportable HbA1c% results in the presence of HbAC (38.8%), HbAD (42.7%), HbAE (31.4%), HbAS (42.7%). Non-clinically significant interference defined as ≤ ± 7% relative difference in the results from a comparative method at 6.5% or 8.0% HbA1c. | Accurate and reportable HbA1c% results in the presence of HbAC (39%), HbAD (39.5%), HbAE (26.0%), HbAS (39%). Non-clinically significant interference defined as ≤ ± 6% relative difference in the results from a comparative method at 6.5% or 8.0% HbA1c. | | | Accurate HbA1c% in the presence of HbA2 up to 6.7%. Non-clinically significant interference is defined as ≤ ± 7% relative difference in the results from a comparative method at 6.5% or 8.0% HbA1c. | Accurate HbA1c% in the presence of HbA2 up to 5.5%. Non-clinically significant interference is defined as ≤ ± 6% relative difference in the results from a comparative method at 6.5% or 8.0% HbA1c. | | | Accurate HbA1c% in the presence of HbF up to 32.2%. | Accurate HbA1c% in the presence of HbF up to 25%. Non-clinically significant interference is defined as ≤ ± 6% relative difference in the results from a comparative method at 6.5% or 8.0% HbA1c. | Both the subject device (Tosoh Automated Glycohemoglobin Analyzer HLC-723GR01 and the Predicate device (Tosoh Automated Glycohemoglobin Analyzer HLC-723G8, v5.24F) have the same intended use, principle of operation, intended user, use environment and similar Page 4 of 13 {8} TOSOH TOSOH BIOSCIENCE Subject Device: Tosoh Automated Glycohemoglobin Analyzer HLC-723GR01 technological characteristics. The analytical performance data does not raise any new questions of safety or effectiveness for the subject device. Therefore, the subject device is substantially equivalent to the predicate device. ## 7. Summary of Non-Clinical Performance Testing ### a. Precision/Reproducibility Precision of the Tosoh Automated Glycohemoglobin Analyzer HLC-723GR01 (GR01 analyzer) was evaluated based on CLSI EP05-A3: Evaluation of Precision of Quantitative Measurement Procedures, 3rd Edition. Repeatability (within-run precision) and reproducibility (inter-run precision) of the GR01 analyzer was determined measuring %HbA1c in both K2 EDTA venous whole blood samples and corresponding hemolysates. HbA1c concentrations of approximately 5.0% (31 mmol/mol), 6.5% (48 mmol/mol), 8.0% (64 mmol/mol), and 12.0% (108 mmol/mol) were tested over two (2) runs per day, with two (2) replicates per run, at three (3) sites/instruments, with three (3) reagent lots, for 20 non-consecutive days per reagent lot, for a total of 60 days on three (3) GR01 analyzers. Following are the results shown for each sample type (whole blood and hemolysate), per analyzer and for the three GR01 analyzers combined, for NGSP units (%). A summary for the Precision data is shown for IFCC units (mmol/mol) for both Whole Blood and Hemolysate samples. ## Results of Precision Study for Whole Blood Samples (NGSP %) Table 2. Summary of Precision Analysis for all Analyzers Combined (Whole Blood/NGSP (%) | %HbA1c Target [Actual Mean] | Repeatability | Between Run | Between Day | Between Lot | Between Site/Instrument | Total Variation / Reproducibility | | --- | --- | --- | --- | --- | --- | --- | | SD | CV | SD | CV | SD | CV | SD | CV | SD | CV | SD | CV | | 5.0% [4.91%] | 0.024 | 0.48% | 0.000 | 0.00% | 0.024 | 0.48% | 0.028 | 0.57% | 0.025 | 0.52% | 0.051 | 1.03% | | 6.5% [6.88%] | 0.026 | 0.38% | 0.015 | 0.22% | 0.043 | 0.62% | 0.050 | 0.73% | 0.011 | 0.17% | 0.074 | 1.07% | | 8.0% [8.24%] | 0.025 | 0.31% | 0.016 | 0.20% | 0.049 | 0.59% | 0.058 | 0.71% | 0.017 | 0.21% | 0.084 | 1.01% | | 12.0% [11.79%] | 0.031 | 0.26% | 0.023 | 0.19% | 0.063 | 0.53% | 0.078 | 0.66% | 0.032 | 0.27% | 0.112 | 0.95% | | Abbreviations: 'SD': Standard Deviation | 'CV': Coefficient of Variation | {9} TOSOH TOSOH BIOSCIENCE Subject Device: Tosoh Automated Glycohemoglobin Analyzer HLC-723GR01 Results of Precision Study for Hemolysate Samples (NGSP %) Table 3. Summary of Precision Analysis for all Analyzers Combined (Hemolysate/NGSP (%)) | %HbA1c Target [Actual Mean] | Repeatability | Between Run | Between Day | Between Lot | Between Site/Instrument | Total Variation / Reproducibility | | --- | --- | --- | --- | --- | --- | --- | | SD | CV | SD | CV | SD | CV | SD | CV | SD | CV | SD | CV | | 5.0% [4.90%] | 0.017 | 0.35% | 0.012 | 0.25% | 0.024 | 0.50% | 0.029 | 0.60% | 0.027 | 0.56% | 0.052 | 1.05% | | 6.5% [6.87%] | 0.028 | 0.40% | 0.011 | 0.15% | 0.042 | 0.61% | 0.057 | 0.83% | 0.005 | 0.07% | 0.077 | 1.12% | | 8.0% [8.25%] | 0.025 | 0.30% | 0.012 | 0.14% | 0.050 | 0.61% | 0.057 | 0.69% | 0.016 | 0.20% | 0.082 | 1.00% | | 12.0% [11.82%] | 0.032 | 0.27% | 0.021 | 0.18% | 0.065 | 0.55% | 0.073 | 0.62% | 0.043 | 0.37% | 0.114 | 0.96% | | Abbreviations: 'SD': Standard Deviation | 'CV': Coefficient of Variation | Results of Precision Study for Whole Blood Samples (IFCC mmol/mol) Table 4. Summary of Precision Analysis for all Analyzers Combined (Whole Blood/IFCC (mmol/mol)) | HbA1c Target (mmol/mol) [Actual Mean] | Repeatability | Between Run | Between Day | Between Lot | Between Site/Instrument | Total Variation / Reproducibility | | --- | --- | --- | --- | --- | --- | --- | | SD | CV | SD | CV | SD | CV | SD | CV | SD | CV | SD | CV | | 31 [30] | 0.188 | 0.62% | 0.000 | 0.00% | 0.280 | 0.93% | 0.365 | 1.21% | 0.252 | 0.84% | 0.557 | 1.85% | | 48 [52] | 0.155 | 0.30% | 0.070 | 0.14% | 0.402 | 0.78% | 0.551 | 1.07% | 0.169 | 0.33% | 0.723 | 1.40% | | 64 [67] | 0.153 | 0.23% | 0.127 | 0.19% | 0.514 | 0.77% | 0.542 | 0.81% | 0.138 | 0.21% | 0.785 | 1.18% | | 108 [105] | 0.166 | 0.16% | 0.186 | 0.18% | 0.696 | 0.66% | 0.862 | 0.82% | 0.387 | 0.37% | 1.200 | 1.14% | | Abbreviations: 'SD': Standard Deviation | 'CV': Coefficient of Variation | Results of Precision Study for Hemolysate Samples (IFCC mmol/mol) Table 5. Summary of Precision Analysis for all Analyzers Combined (Hemolysate/IFCC (mmol/mol)) | HbA1c Target (mmol/mol) [Actual Mean] | Repeatability | Between Run | Between Day | Between Lot | Between Site/Instrument | Total Variation / Reproducibility | | --- | --- | --- | --- | --- | --- | --- | | SD | CV (%) | SD | CV (%) | SD | CV (%) | SD | CV (%) | SD | CV (%) | SD | CV (%) | | 31 [30] | 0.125 | 0.42% | 0.140 | 0.47% | 0.279 | 0.93% | 0.378 | 1.26% | 0.246 | 0.82% | 0.562 | 1.87% | | 48 [52] | 0.141 | 0.27% | 0.076 | 0.15% | 0.401 | 0.78% | 0.598 | 1.16% | 0.145 | 0.28% | 0.752 | 1.46% | | 64 [67] | 0.137 | 0.20% | 0.131 | 0.20% | 0.520 | 0.78% | 0.559 | 0.84% | 0.154 | 0.23% | 0.801 | 1.20% | | 108 [106] | 0.191 | 0.18% | 0.194 | 0.18% | 0.693 | 0.66% | 0.860 | 0.81% | 0.504 | 0.48% | 1.244 | 1.18% | | Abbreviations: 'SD': Standard Deviation | 'CV': Coefficient of Variation | {10} TOSOH TOSOH BIOSCIENCE Subject Device: Tosoh Automated Glycohemoglobin Analyzer HLC-723GR01 b. Method Comparison The method comparison study was designed in accordance with CLSI EP 21: Evaluation of Total Analytical Error for Quantitative Medical Laboratory Measurement Procedures - Second Edition, and parts of CLSI EP09c: Measurement Procedure Comparison and Bias Estimation Using Patient Samples - Third Edition. Hemoglobin A1c (HbA1c) test results of the Tosoh Automated Glycohemoglobin Analyzer HLC 723GR01 (GR01 analyzer) in an Intended Use setting were compared with results obtained with comparator testing run on the Tosoh Automated Glycohemoglobin Analyzer HLC-723G8 (G8 analyzer); comparator testing was performed in a National Glycohemoglobin Standardization Program (NGSP) Secondary Reference Laboratory (SRL). One Hundred and thirty-five (135) K2EDTA venous whole blood with HbA1c concentrations that spanned the GR01 measuring range (3.9% HbA1c to 16.9% NGSP units) were evaluated. Each sample was measured in triplicate with the G8 comparator device, and the mean value was defined as the "true" HbA1c value. Each sample was also measured in a single or one of three investigational GR01 analyzers placed at three (3) separate Moderate to High Complexity Clinical Laboratories. Sample distribution was as follows: Table 6. Method Comparison Study - Sample Distribution | %HbA1c Level | Whole Blood | | Hemolysate | | | --- | --- | --- | --- | --- | | | N | % of Samples | N | % of Samples | | ≤ 5.0% | 5 | 3.7 | 5 | 3.7 | | >5.0 – 6.0% | 16 | 11.9 | 15 | 11.1 | | >6.0 – 6.5% | 31 | 23.0 | 32 | 23.7 | | > 6.5 – 7.0% | 31 | 23.0 | 32 | 23.7 | | >7.0 – 8.0% | 21 | 15.6 | 20 | 14.8 | | >8.0 – 9.0% | 9 | 6.7 | 10 | 7.4 | | >9.0 | 22 | 16.3 | 21 | 15.6 | | Total | 135 | 100 | 135 | 100 | Bias between the Candidate and NGSP method was calculated using Passing-Bablok regression analysis for both NGSP. The regression parameters, slope and intercept, were the following: Table 7. Summary of Method Comparison - Passing-Bablok Parameter Estimates | Specimen Type | N | Slope (95% CI) | Intercept (95% CI) | | --- | --- | --- | --- | | Whole Blood | 135 | 1.0045 (1.0000 to 1.0135) | -0.0222 (-0.0743 to 0.0000) | | Hemolysate | 135 | 1.0125 (1.0039 to 1.0189) | -0.0788 (-0.1226 to -0.0241) | Bias was estimated based on the Passing-Bablok regression parameters and was as follows: {11} TOSOH TOSOH BIOSCIENCE Subject Device: Tosoh Automated Glycohemoglobin Analyzer HLC-723GR01 Table 8. Bias Estimation based on Passing-Bablok | Specimen Type | Decision Level | Bias | % Bias | | --- | --- | --- | --- | | Whole Blood | 5.0% | 0.000 | 0.006 | | | 6.5% | 0.007 | 0.108 | | | 8.0% | 0.014 | 0.172 | | | 12.0% | 0.032 | 0.265 | | Hemolysate | 5.0% | -0.016 | -0.326 | | | 6.5% | 0.002 | 0.038 | | | 8.0% | 0.021 | 0.265 | | | 12.0% | 0.071 | 0.593 | - Total Error near Medical Decision Points Using the results of bias estimation (%Bias) in the method comparison study and precision estimates in the precision study (%CV), Total Error (TE) at four HbA1c concentrations (5.0%, 6.5%, 8.0% and 12.0%) was calculated as follows: %TE = %Bias + 1.96*%CV*(1+%Bias/100). The results are presented below. Table 9. Total Error near Medical Decision Points | | HbA1c Level | %Bias | %CV | %Total Error | | --- | --- | --- | --- | --- | | Whole Blood | Sample 1 [5.0%] | 0.006 | 1.03 | 2.02 | | | Sample 2 [6.5%] | 0.108 | 1.07 | 2.21 | | | Sample 3 [8.0%] | 0.172 | 1.01 | 2.16 | | | Sample 4 [12.0%] | 0.265 | 0.95 | 2.13 | | Hemolysate | Sample 1 [5.0%] | -0.326 | 1.05 | 2.38 | | | Sample 2 [6.5%] | 0.038 | 1.12 | 2.23 | | | Sample 3 [8.0%] | 0.265 | 1.00 | 2.23 | | | Sample 4 [12.0%] | 0.593 | 0.96 | 2.49 | c. Matrix Comparison The data of the Matrix Comparison Evaluation support the use of K2-EDTA and K3-EDTA blood collection tubes. There was no clinical or statistical difference as the total bias between the two collection tubes met the acceptance criteria of ≤±6%, thus they may be used interchangeably for testing HbA1c on the GR01 HPLC Analyzer. Page 8 of 13 {12} TOSOH TOSOH BIOSCIENCE Subject Device: Tosoh Automated Glycohemoglobin Analyzer HLC-723GR01 d. Traceability, Stability, Expected values (Calibrators) The assigned HbA1c values of the Tosoh Automated Glycohemoglobin Analyzer are certified with The National Glycohemoglobin Standardization Program (NGSP). See NGSP website for current certification at http://www.ngsp.org. The final reportable result is traceable to both the International Federation of Clinical Chemistry (IFCC) and the Diabetes Control and Complications Trial (DCCT). The International Federation of Clinical Chemistry (IFCC) units of mmol/mol are calculated using the Master Equation NGSP (%) = [0.09148 x IFCC (mmol/mol)] + 2.152]. HbA1c results are provided to the customers using two different units: NGSP equivalent units (%) and IFCC equivalent units (mmol/mol). Calibrators (Tosoh A1c Calibrator Set) and Controls (Tosoh A1c Control Set) were used in the performance studies for use with this device. The calibrators and controls are both 510(k) exempt. e. Linearity The results of the linearity study support the claimed measuring range of 3.9% to 16.9% HbA1c (NGSP units), and 19 to 161 mmol/mol (IFCC units). f. Analytical Specificity i. Endogenous Interference The endogenous interference study was performed in compliance with CLSI EP07-A3, Interference Testing in Clinical Chemistry; Third Edition. Interference studies were conducted at known concentrations of HbA1c samples from both diabetic and non-diabetic donors. Specimens were spiked with increasing amounts of the interfering substance and then assayed ten (10) times. The mean result of each sample was then compared to the mean result of the untreated, reference sample. Significant interference was defined as an observed difference in HbA1c concentration greater than ± 5% from the untreated reference sample. Table 10. Endogenous Interfering Substances Tested | Potential Interfering Substance | Concentration with No Interference | | --- | --- | | Albumin | 20 g/dL | | Ascorbic Acid | 300 mg/dL | | Bilirubin - Conjugated | 100 mg/dL | | Bilirubin - Unconjugated | 100 mg/dL | | Rheumatoid Factor | 750 IU/mL | | Triglycerides | 6,000 mg/dL | Page 9 of 13 {13} TOSOH TOSOH BIOSCIENCE Subject Device: Tosoh Automated Glycohemoglobin Analyzer HLC-723GR01 Albumin, ascorbic acid, conjugated and unconjugated bilirubin, triglyceride and rheumatoid factor do not interfere with the Tosoh Automated Glycohemoglobin Analyzer HLC-723GR01 assay up to the stated concentrations. ii. Drug Interference The exogenous interference study was performed in compliance with CLSI EP07-A3, Interference Testing in Clinical Chemistry -Third Edition. Interference studies were conducted at known concentrations of HbA1c samples from both diabetic and non-diabetic donors. Specimens were spiked with increasing amounts of the interfering substance and then assayed ten (10) times. The mean result of each sample was then compared to the mean result of the untreated, reference sample. Significant interference was defined as an observed difference in HbA1c concentration greater than $\pm 5\%$ from the untreated reference sample. Table 11. Exogenous Interfering Substances Tested | Potential Interfering Substance | Concentration with No Interference | | | --- | --- | --- | | Acetaminophen | 20 | mg/dL | | Acetylcysteine | 330 | mg/dL | | Ampicillin | 1,000 | mg/dL | | Cefoxitin | 2,500 | mg/dL | | Cyclosporine | 0.7 | mg/dL | | Doxycycline | 50 | mg/dL | | Heparin | 5.000 | U/L | | Ibuprofen | 50 | mg/dL | | Levodopa | 20 | mg/dL | | Metformin | 5 | mg/dL | | Methyldopa | 30 | mg/dL | | Metronidazole | 200 | mg/dL | | Phenylbutazone | 400 | mg/dL | | Rifampicin | 6.4 | mg/dL | | Salicylic Acid | 60 | mg/dL | | Theophylline | 10 | mg/dL | The exogenous substances evaluated do not interfere with the Tosoh Automated Glycohemoglobin Analyzer HLC-723GR01 assay at therapeutic doses. iii. Cross Reactivity with Hemoglobin Derivatives A cross-reactivity study was performed to assess potential interferences from Acetylated hemoglobin (Hb), Carbamylated Hb, Aldehyde Hb, and Labile HbA1c with samples from both diabetic and non-diabetic donors. Specimens were spiked with increasing amounts of the interfering substance and then assayed ten (10) times. The mean result of each sample was then compared to the mean result of the untreated, reference sample. Significant interference was defined as an observed difference in HbA1c concentration Page 10 of 13 {14} greater than ± 5% from the untreated reference sample. The following results were concluded as not interfering with the assay.Acetylated Hb up to 50 mg/dLCarbamylated Hb up to 25 mg/dLAldehyde Hb up to 25 mg/dL Labile HbA1c up to 2000 mg/dL In conclusion, Acetylated hemoglobin (Hb), Carbamylated Hb, Aldehyde Hb, and Labile HbA1c with HbA1c do not cross-react with the assay up to the concentrations shown above. ## Hemoglobin Variant Interference The hemoglobin variant interference study was performed by comparing HbA1c results of samples with the presence of common Hb variants measured with both the GR01 analyzer and the reference method, Tosoh Automated Glycohemoglobin Analyzer HLC-723G8, Ver. 5.24F (“G8 v5.24F”), which is which has been demonstrated to be free from interference by these hemoglobin variants in compliance to CLSI EP07: Interference Testing in Clinical Chemistry; Approved Guideline -- Third Edition. Interference due to the presence of certain levels of common Hb variants (HbAC, HbAD, HbAE, and HbAS) and hemoglobinopathy, elevated HbA2 and elevated Fetal Hemoglobin (HbF) is well-known when measuring %HbA1c in clinical specimens. Interference studies were performed to identify the level of (1) Hb variants and HbA2, and (2) HbF likely to cause interference with HbA1c measurements. ### (1) Common Hemoglobin Variants and Elevated HbA2 The hemoglobin variant study was performed using K2 EDTA venous whole blood samples known to contain hemoglobin variants HbS, HbC, HbE, HbD and elevated HbA2. The samples containing the hemoglobin variants were tested using the GR01 and using the comparator method (G8 v5.24F), which has been demonstrated to be free from interference by these hemoglobin variants. Non-clinically significant interference was defined as ≤ ±7% relative difference in the results from the comparative method at 6.5% or 8.0% HbA1c. The following table summarizes the test results. ### (2) Fetal Hemoglobin (HbF) No clinically significant interference with HbA1c results was observed up to 32.2% HbF. Test results for the Hemoglobin Variance Interference evaluations are as follows: {15} TOSOH TOSOH BIOSCIENCE Subject Device: Tosoh Automated Glycohemoglobin Analyzer HLC-723GR01 Table 12. Hemoglobin Variant Interference Study – Relative Difference at Medical Decision Points | Variant | n | HbA1c Range | Variant Range | Mean (Range) of Relative Bias from Reference Method at Approximately 6.5% and 8.0% HbA1c | | | --- | --- | --- | --- | --- | --- | | | | | | ~6.5 % HbA1c | ~8.0 % HbA1c | | HbC | 25 | 4.6% to 14.3% | 30.0% to 38.8% | 2.79 (0.74 to 4.65) | 1.80 (-0.61 to 3.85) | | HbD | 22 | 5.7% to 10.9% | 24.3% to 42.7% | 0.50 (-3.08 to 5.26) | -1.16 (-4.71 to 1.27) | | HbE | 34 | 5.0% to 12.8% | 19.5% to 31.4% | 1.54 (-3.57 to 4.92) | 2.08 (0.00 to 7.19) | | HbS | 29 | 4.7% to 14.4% | 26.6% to 42.7% | 2.49 (1.57 to 3.91) | 2.71 (1.99 to 4.43) | | HbA2 | 33 | 4.7% to 14.3% | 3.5% to 6.7% | 2.51 (0.77 to 4.65) | 1.26 (-0.61 to 2.99) | | HbF | 33 | 5.5% to 11.7% | 3.8% to 32.2% | -1.2 (-3.0 to 1.5) | -2.6 (-4.2 to 0.8) | In conclusion, no clinically significant interference with HbA1c results was observed for the common hemoglobin variants HbAC, HbAD, HbAE or HbAS at the levels stated in the table above. No clinically significant interference with HbA1c results was observed up to 6.7% HbA2. No clinically significant interference with HbA1c results was observed up to 32.2% HbF. ## Expected Values/Reference Range The sponsor provided the following information in labeling regarding expected values: The "Classification and Diagnosis of Diabetes: Standards of Medical Care in Diabetes - 2021", American Diabetes Association gives the following criteria for the diagnosis of diabetes and categories of increased risk for diabetes. Table 13. Reference Ranges | | In NGSP units | In IFCC units | Comment | | --- | --- | --- | --- | | HbA1c | ≥ 6.5 % | (≥ 48 mmol/mol) | Cutoff point to diagnose diabetes | | HbA1c | 5.7 - 6.4 % | (39 – 47 mmol/mol) | Increased risk for diabetes (prediabetes) | The HbA1c reference range (non-diabetic) is reported as below HbA1c: 4.0 - 6.0 % (mean 5.0 %, SD 0.5 %) This range can be converted into IFCC units as: HbA1c: 20 - 42 mmol/mol, using the following Master Equation: $$ \text{IFCC (mmol/mol)} = 10.93 \times \text{NGSP}(\%) - 23.50 $$ Page 12 of 13 {16} TOSOH TOSOH BIOSCIENCE Subject Device: Tosoh Automated Glycohemoglobin Analyzer HLC-723GR01 ## 8. Proposed Labeling The labeling is sufficient and satisfies the requirements of 21 CFR Part 809.10. ## 9. Conclusion The information and data contained in this 510(k) document demonstrate that the Tosoh Automated Glycohemoglobin Analyzer HLC-723GR01 is an accurate, reliable, precise test that correlates well with current cleared methods and NGSP standardized testing for the quantitation of HbA1c. Furthermore, the performance criteria as stipulated by the Special Control requirements for HbA1c systems that diagnose diabetes have been met. Overall, the contents of this submission demonstrate that the Tosoh Automated Glycohemoglobin Analyzer HLC-723GR01 is substantially equivalent in intended use and technological characteristics to the Tosoh Automated Glycohemoglobin Analyzer HLC-723G8 v5.24F (k200904). Therefore, the Tosoh Automated Glycohemoglobin Analyzer HLC-723GR01 is safe and effective for its intended use, and meets the Federal Food, Drug and Cosmetic Act criteria for 510(k) clearance of this device. Page 13 of 13
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...