AUTOMATED GYLCOHEMOGLOBIN ANALYZER HLC-723G8

K131580 · Tosoh Bioscience, Inc. · PDJ · Jan 23, 2014 · Clinical Chemistry

Device Facts

Record IDK131580
Device NameAUTOMATED GYLCOHEMOGLOBIN ANALYZER HLC-723G8
ApplicantTosoh Bioscience, Inc.
Product CodePDJ · Clinical Chemistry
Decision DateJan 23, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1373
Device ClassClass 2

Intended Use

The Tosoh Automated Glycohemoglobin Analyzer HLC-723G8 is intended for in vitro diagnostic use for the quantitative measurement of % hemoglobin A1c (HbA1c) (DCCT/NGSP) and mmol/mol hemoglobin A1c (IFCC) in whole blood specimens. This test is to be used as an aid in diagnosis of diabetes and as an aid in identifying patients who may be at risk for developing diabetes.

Device Story

The Tosoh Automated Glycohemoglobin Analyzer HLC-723G8 is an automated HPLC system for measuring HbA1c in whole blood. It uses a cation-exchange column to separate hemoglobin fractions (A1a, A1b, F, LA1c, sA1c, A0) via stepwise elution with salt-based buffers. An LED photometer measures absorbance at 415 nm. The system integrates raw data to calculate relative percentages of hemoglobin fractions, producing a chromatogram and numerical results. Used in clinical laboratories, it aids physicians in diagnosing diabetes and identifying at-risk patients. The device flags samples containing Hemoglobin E (HbE) and suppresses results due to interference. It provides quantitative HbA1c values to support clinical decision-making, though it is not intended for use in specific patient populations with abnormal red cell turnover or hemoglobinopathies.

Clinical Evidence

No clinical trials were performed. Evidence consists of analytical bench testing, including precision (repeatability, between-run, between-day, between-lot, and between-instrument), linearity, and interference studies. Method comparison was conducted against an NGSP-certified reference laboratory method (Primus Ultra) using 120 patient samples. Results showed high correlation (Deming slope 0.996, intercept 0.073) and Total Error (TE) ranging from 2.8% to 5.8% across clinical decision points (5.0%–12.0% HbA1c).

Technological Characteristics

Non-porous ion-exchange HPLC; cation exchange column; visible wavelength detector; stepwise elution with three salt-concentration buffers. Connectivity includes printer and floppy disk drive. Software-controlled via microprocessors. Standards: EP5-A2 (precision), EP09-A2-IR (method comparison), EP7-A2 (interference).

Indications for Use

Indicated for use as an aid in the diagnosis of diabetes and identifying patients at risk for developing diabetes in human whole blood specimens. Contraindicated for patients with hemoglobinopathies (e.g., sickle cell trait), abnormal red cell turnover (e.g., hemolytic anemia, iron deficiency), hereditary spherocytosis, malignancies, severe chronic hepatic or renal disease, and pregnancy. Not for use in the diagnosis of gestational diabetes or as a replacement for glucose testing in type 1 diabetes, pediatric patients, or pregnant women.

Regulatory Classification

Identification

A hemoglobin A1c test system is a device used to measure the percentage concentration of hemoglobin A1c in blood. Measurement of hemoglobin A1c is used as an aid in the diagnosis of diabetes mellitus and as an aid in the identification of patients at risk for developing diabetes mellitus.

Special Controls

(b) Classification. Class II (special controls). Hemoglobin A1c test systems must comply with the following special controls: 1) The device must have initial and annual standardization verification by a certifying glycohemoglobin standardization organization deemed acceptable by FDA. 2) The premarket notification submission must include performance testing to evaluate precision, accuracy, linearity and interference, including the following: i) Performance testing of device precision must, at a minimum, use blood samples with concentrations near 5.0%, 6.5%, 8.0% and 12% hemoglobin A1c. This testing must evaluate precision over a minimum of 20 days using at least 3 lots of the device and 3 instruments, as applicable. ii) Performance testing of device accuracy must include a minimum of 120 blood samples that span the measuring interval of the new device and compare results of the new device to results of the standardized test method. Results must demonstrate little or no bias versus the standardized method. iii) Total error of the new device must be evaluated using single measurements by the new device compared to results of the standardized test method, and this evaluation must demonstrate a total error less than or equal to 6%. iv) Performance testing must demonstrate that there is little to no interference from common hemoglobin variants, including Hemoglobin C, Hemoglobin D, Hemoglobin E, Hemoglobin A2 and Hemoglobin S. 3) When assay interference from Hemoglobin F or interference with other hemoglobin variants with low frequency in the population is observed, a warning statement must be placed in a black box and must appear in all labeling material for these devices describing the interference and any affected populations.

*Classification.* Class II (special controls). The special controls for this device are:(1) The device must have initial and annual standardization verification by a certifying glycohemoglobin standardization organization deemed acceptable by FDA. (2) The premarket notification submission must include performance testing to evaluate precision, accuracy, linearity, and interference, including the following: (i) Performance testing of device precision must, at a minimum, use blood samples with concentrations near 5.0 percent, 6.5 percent, 8.0 percent, and 12 percent hemoglobin A1c. This testing must evaluate precision over a minimum of 20 days using at least three lots of the device and three instruments, as applicable. (ii) Performance testing of device accuracy must include a minimum of 120 blood samples that span the measuring interval of the device and compare results of the new device to results of a standardized test method. Results must demonstrate little or no bias versus the standardized method. (iii) Total error of the new device must be evaluated using single measurements by the new device compared to results of the standardized test method, and this evaluation must demonstrate a total error less than or equal to 6 percent. (iv) Performance testing must demonstrate that there is little to no interference from common hemoglobin variants, including Hemoglobin C, Hemoglobin D, Hemoglobin E, Hemoglobin A2, and Hemoglobin S. (3) When assay interference from Hemoglobin F or interference with other hemoglobin variants with low frequency in the population is observed, a warning statement must be placed in a black box and must appear in all labeling material for these devices describing the interference and any affected populations.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K13/570 ### 510(k) Summary JAN 2 3 2014 l of 10 # Tosoh Bioscience, Inc.'s Automated Glycohemoglobin Analyzer HLC-723G8 #### Submitter: #### Contact Person: : : : : ### Device Name: Classification Name: #### Predicate Device: Tosoh Bioscience, Inc 3600 Gantz Road Grove City, OH 43123 Robert L. Wick Regulatory Specialist 6000 Shoreline Ct., Ste. 101 South San Francisco, CA 94080 . Phone: 650-636-8117 Fax: 650-636-8121 : . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Email: Robert.Wick@Tosoh.com Automated Glycohemoglobin Analyzer HLC-723G8 Hemoglobin A1c Test System Class II PDJ 21 CFR 862.1373 : ・・・・・ K121291 Roche Diagnostics Corporation COBAS INTEGRA 800 Tina-quant HbA1cDx Gen.2 assay {1}------------------------------------------------ ### 510(k) Summary # Automated Glycohemoglobin Analyzer HLC-723G8 According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence. #### Device Description: The Tosoh Automated Glycohemoglobin Analyzer HLC-723G8 is an automated High Performance Liquid Chromatography (HPLC) system that separates and reports stable A 1c (sA1c) percentage in whole blood. The operational portion of the G8 is composed of a sampling unit, liquid pump, degasser, column, detector, microprocessors, sample loader, floppy disk drive unit, operation panel and a printer. The Tosoh Automated Glycohemoglobin Analyzer HLC-723G8 uses non-porous ion- exchange high performance liquid chromatography (HPLC) for rapid, accurate and precise separation of the stable form of HbA Ic from other hemoglobin fractions. The G8 uses a cation exchange column and separates the usual hemoglobin components in the blood into six fractions, A 1a, A 1b, F, LA 1c, sA Ic, and A0. The separation is done by eluting the hemoglobins from the column with a stepwise elution of three elution buffers containing different salt concentrations. The result report includes a sample ID, date, percentage and retention time of each fraction, sA1c percentage and total A1 percentage (A1a + A1b + sA1c), along with a chromatogram of the elution pattern of the hemoglobin fractions. If a sample contains a hemoglobin variant, the column elutes the material depending upon its charge. #### New Device Intended Use: The Tosoh Automated Glycohemoglobin Analyzer HLC-723G8 is intended for in vitro diagnostic use for the quantitative measurement of % hemoglobin A Ic (HbA 1c) (DCCT/NGSP) and mmol/mo hemoglobin A1c (IFCC) in whole blood specimens. This test is to be used as an aid in diagnosis of diabetes and as an aid in identifying patients who may be at risk for developing diabetes. {2}------------------------------------------------ ### Substantial Equivalence: ## Comparison between the Tosoh Automated Glycohemoglobin Analyzer HLC-723G8 and the COBAS INTEGRA 800 Tina-quant HbA1cDx Gen.2 assay ## Similarities | Parameter | Tosoh Automated<br>Glycohemoglobin Analyzer<br>HLC-723G8 (k)071132<br>(k)131580 | COBAS INTEGRA 800<br>Tina-quant<br>HbA1cDx Gen.2 assay<br>(k)121291 | |------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use | The Tosoh Automated<br>Glycohemoglobin Analyzer<br>HLC-723G8 is intended for in<br>vitro diagnostic use for the<br>measurement of hemoglobin<br>A1c (HbA1c) in whole blood<br>specimens. Hemoglobin A1c<br>measurements are used in the<br>clinical management of<br>diabetes to assess the long-<br>term efficacy of diabetic<br>control. This test is to be used<br>as an aid in diagnosis of<br>diabetes and identifying<br>patients who may be at risk for<br>developing diabetes. | This test is to be used as an aid<br>in diagnosis of diabetes and as<br>an aid in identifying patients<br>who may be at risk for<br>developing diabetes The<br>COBAS INTEGRA 800 Tina-<br>quant HbA1cDx Gen.2 assay<br>is an in vitro diagnostics<br>reagent system intended for<br>quantitative determination of<br>mmol/mol hemoglobin A1c<br>(IFCC) and % hemoglobin<br>A1c (DCCT/NGSP) in<br>hemolysate or whole blood on<br>the Roche COBAS INTEGRA<br>800 clinical chemistry<br>analyzer | | Specimen Type<br>Matrix | Human Whole Blood<br>EDTA Whole Blood | Human Whole Blood<br>EDTA, Li-Heparin, Na<br>Heparin, NaF/K- Oxalate | | Standardization | Certified via the National<br>Glycohemoglobin<br>Standardization<br>Program (NGSP) | Certified via the National<br>Glycohemoglobin<br>Standardization<br>Program (NGSP) | | Linearity | Linearity was previously<br>evaluated for this assay under<br>k071132. The reportable<br>range is 4.0% - 16.9%. | Linearity was previously<br>evaluated for this assay under<br>k072714. The reportable<br>range is 4.2-20.1% | | Interference | Substances were evaluated at<br>two HbA1c levels (6.5% and<br>≥8% HbA1c)<br><br>Lipemia<br>Conjugated Bilirubin<br>Unconjugated bilirubin<br>Rheumatoid Factor<br>Glucose<br>Total Protein<br>Ascorbic Acid | Substances were evaluated at<br>two HbA1c levels (~6.5 and<br>~8.9% HbA1c)<br><br>Lipemia<br>Conjugated Bilirubin<br>Unconjugated bilirubin<br>Rheumatoid Factor<br>Glucose<br>Total Protein<br>Ascorbic Acid | | | Acetaldehyde | | | Precision | HbA1c concentration values of<br>approximately 5%, 6.5%, 8%<br>and 12% | HbA1c concentration values of<br>approximately 5%, 6.5%, 8%<br>and 10% | | | Precision was determined to<br>be 0.7 - 1.8% (total precision)<br>and 0.3 - 0.9% (within run,<br>precision). | The between-analyzer and<br>between-lot precision was<br>equal to or less than 1.5% for<br>concentrations in the range of<br>5.3% to 12.1% HbA1c | | Hemoglobin Variant<br>Interference | HbA2, HbS, HbC, HbD, does<br>not interfere with the assay. | HbA2, HbS, HbC, HbD, does<br>not interfere with the assay. | | Bias | Concentration and Bias: | Concentration and Bias: | | | 6.0% -0.847% | 5.2% 1.21% | | | 6.5% -0.753% | 6.5% 1.29% | | | 7.0% -0.673% | 8.0% 1.34% | | Total Error - allowable bias<br>6% | Concentration and Total Error | Concentration and Total Error | | | 5.0 5.8% | 5.2 4.2% | | | 6.5 2.8% | 6.5 4.1% | | | 8.0 3.0% | 8.0 4.1% | | | 12.0 3.1% | | {3}------------------------------------------------ ### Differences : · | Parameter | Tosoh Automated<br>Glycohemoglobin Analyzer<br>HLC-723G8 (k)071132 | COBAS INTEGRA 800<br>Tina-quant<br>HbA1cDx Gen.2 assay<br>(k)121291 | |-------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Assay Principle | Ion-exchange HPLC | Quantitative turbidimetric<br>inhibition immunoassay | | Detection Method | Visible wavelength detector | Absorption spectrum and<br>measured biochromatically. | | Method Comparison | The method comparison study<br>was conducted utilizing the<br>Automated Glycohemoglobin<br>Analyzer HLC-723G8 and the<br>Primus model ultra2.<br><br>r = 0.991<br>(Y = 0.996x + 0.073) | The method comparison study<br>was conducted using utilizing<br>the Roche COBAS INTEGRA<br>800 Tina-quant HbA1cDx<br>Gen.2 assay and the Tosoh<br>Automated Glycohemoglobin<br>Analyzer HLC-723G8.<br><br>r = None Listed<br>(Y = 1.015x + 0.015) | | | Hemoglobin Variant<br>Interference | HbE does interfere with the<br>assay. | # Automated Glycohemoglobin Analyzer HLC-723G8 4 of 10 . {4}------------------------------------------------ ### PERFORMANCE CHARACTERISTICS #### Interference: This interference study was developed according to the CLSI guideline Interference Testing in Clinical Chemistry (EP7-A2). Interference studies were conducted on known concentrations of HbA l . They were spiked with increasing amounts of the substances below. Interference was determined as a variance greater than the assigned value x 1.00 ± 5%. | Potential<br>Interferent | Range tested | %A1c<br>Concentrations | Concentration in which<br>no significant<br>interference was<br>observed | |--------------------------|------------------|------------------------|--------------------------------------------------------------------------| | Acetylated Hb | 10 - 50 mg/dL | 6.5 and 9.5 | 50 mg/dL | | Albumin | 500 - 5000 mg/dL | 6.6 and 14.7 | 5000 mg/dL | | Aldehyde Hb | 5.0 - 25 mg/dL | 6.3 and 12.6 | 25 mg/dL | | Ascorbic Acid | 3.0 - 25 mg/dL | 6.4 and 10.8 | 25 mg/dL | | Carbamylated Hb | 5.0 - 25 mg/dL | 6.5 and 9.8 | 25 mg/dL | | Bilirubin C | 2.0 - 21 mg/dL | 6.5 and 14.3 | 21 mg/dL | | Labile Hb | 200 – 1000 mg/dL | 6.4 and 10.3 | 1000 mg/dL | | Lipemia | 100 - 1000 mg/dL | 6.4 and 14.1 | 1000 mg/dL | | Rheumatoid Factor | 110 - 550 IU/mL | 6.3 and 12.6 | 550 IU/mL | | Bilirubin F | 2.0 - 18 mg/dL | 6.5 and 14.3 | 18 mg/dL | {5}------------------------------------------------ | | Hemoglobin Variant Interference study | | |--|---------------------------------------|--| | | | | | Hemoglobin<br>Variant | Number of<br>Samples at<br>approximately<br>6% and ≥8% | %<br>HbA1c<br>reference<br>values | IFCC<br>mmol/ml | % Variant | Interference | |-----------------------|--------------------------------------------------------|-----------------------------------|-----------------|-----------|--------------| | HbS | 10 | 6.2 | 44 | 36.42 | No | | | 10 | 9.8 | 78.1 | 29.49 | No | | HbC | 10 | 6.7 | 47 | 36.54 | No | | | 10 | 8.0 | 59.4 | 35.06 | No | | HbE* | 10 | 6.9 | N/A* | N/A* | Yes | | | 10 | 9.8 | N/A* | N/A* | Yes | | HbD | 10 | 6.0 | 43.1 | 36.23 | No | | | 10 | 8.5 | 66.9 | 35.70 | No | | HbF | 10 | 7.0 | 56.5 | 24.38 | No | | | 10 | 8.7 | 74 | 7.43 | No | | HbA2 | 10 | 6.59 | 47.1 | 13.9 | No | | | 10 | 8.05 | 62.9 | 12.7 | No | *The G8 has known HbE interference. When a sample is suspected to contain HbE, a flag will be displayed. The %HbA Ic result will not be reported from the analyzer. The Tosoh Automated Glycohemoglobin Analyzer HLC-723G8 has known Hemoglobin E (HbE) interference. When a sample is suspected to contain HbE a flag will be displayed. The HbA1c result will not be reported from the analyzer. The percent relative bias from the reference method at low and high concentrations of %HbA I c in each sample is: | Hb Variant | Percent Relative Bias from Reference Method<br>at Low and High Concentrations of %HbA1c<br>Samples | | |------------|----------------------------------------------------------------------------------------------------|-----------| | | ~6.5 %HbA1c | ≥8 %HbA1c | | S | 0 | -5 | | C | 5 | -5 | | E | *Flag | *Flag | | D | 1 | -3 | | A2 | -2 | -2 | | F | 5 | 2 | #### Precision: The precision study was developed with reference to the CLSI protocol entitled: Evaluation of Precision Performance of Quantitative Measurement Methods (EP5-A2). {6}------------------------------------------------ Precision was assessed by assaying four concentrations of unaltered EDTA whole blood specimens with each using 3 lots and 3 instruments. Estimates of total and within-run precision were obtained from measurements of 2 replicates in a single run, 2 times a day for 20 nonconsecutive days for each specimen/reagent combination. This equaled a total of 40 runs and 80 determinants for each data set. There were three (3) data sets at HbA Ic concentrations of approximately 5%, 6.5%, 8% and 12% for each specimen. ### Within-run Precision Intra-assay (within run) Precision | | Mean | Standard<br>Deviation | Coefficient of<br>Variation | |--------------------------|-------|-----------------------|-----------------------------| | Sample | (%) | (%) | (%) | | EDTA Whole Blood 5% A1 | 4.98 | 0.04 | 0.86 | | EDTA Whole Blood 5% A2 | 5.00 | 0.05 | 1.00 | | EDTA Whole Blood 5% A3 | 4.93 | 0.03 | 0.68 | | EDTA Whole Blood 6.5% B1 | 6.50 | 0.02 | 0.29 | | EDTA Whole Blood 6.5% B2 | 6.51 | 0.02 | 0.35 | | EDTA Whole Blood 6.5% B3 | 6.44 | 0.03 | 0.44 | | EDTA Whole Blood 8% C1 | 7.89 | 0.02 | 0.26 | | EDTA Whole Blood 8% C2 | 7.90 | 0.03 | 0.32 | | EDTA Whole Blood 8% C3 | 7.81 | 0.03 | 0.34 | | EDTA Whole Blood 12% D1 | 11.90 | 0.02 | 0.18 | | EDTA Whole Blood 12% D2 | 11.89 | 0.03 | 0.23 | | EDTA Whole Blood 12% D3 | 11.79 | 0.03 | 0.27 | ### Total Precision . . | | Mean | Standard<br>Deviation | Coefficient of<br>Variation | |--------------------------|-------|-----------------------|-----------------------------| | Sample | (%) | (%) | (%) | | EDTA Whole Blood 5% A1 | 4.98 | 0.09 | 1.83 | | EDTA Whole Blood 5% A2 | 5.00 | 0.09 | 1.79 | | EDTA Whole Blood 5% A3 | 4.93 | 0.08 | 1.69 | | EDTA Whole Blood 6.5% B1 | 6.50 | 0.04 | 0.68 | | EDTA Whole Blood 6.5% B2 | 6.51 | 0.07 | 1.14 | | EDTA Whole Blood 6.5% B3 | 6.44 | 0.07 | 1.10 | | EDTA Whole Blood 8% C1 | 7.89 | 0.08 | 0.98 | | EDTA Whole Blood 8% C2 | 7.90 | 0.07 | 0.89 | | EDTA Whole Blood 8% C3 | 7.81 | 0.07 | 0.86 | | EDTA Whole Blood 12% D1 | 11.90 | 0.16 | 1.31 | | EDTA Whole Blood 12% D2 | 11.89 | 0.19 | 1.58 | | EDTA Whole Blood 12% D3 | 11.79 | 0.13 | 1.14 | {7}------------------------------------------------ #### Correlation: The methods comparison study was conducted with reference to the CLSI protocol entitled: Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline (EP9-A2-IR). Comparative testing was conducted between samples which were analyzed on the Automated Glycohemoglobin Analyzer HLC-723G8 (candidate method) and Primus ultra2 (comparative method) analyzed by a secondary NGSP reference laboratory. A total of 120 unaltered EDTA whole blood specimens were assayed in singleton. Frozen specimens were utilized for this study. The sample distribution is as follows: | Hemoglobin A1c level | Number of samples<br>tested at each A1c<br>Level | %samples tested | |----------------------|--------------------------------------------------|-----------------| | ≤ 5% | 5 | 4.2% | | 5 – 6% | 15 | 12.5% | | 6 – 6.5% | 30 | 25.0% | | 6.5 – 7% | 30 | 25.0% | | 7 – 8% | 20 | 16.7% | | 8 – 9% | 10 | 8.3 | | > 9% | 10 | 8.3% | | Total samples | 120 | 100% | | Regression Analysis | | | |----------------------------|-------------------------|-------------------------| | | Deming | Regular | | Slope: | 0.996 (0.971 to 1.021) | 0.987 (0.962 to 1.012) | | Intercept: | 0.073 (-0.110 to 0.255) | 0.139 (-0.043 to 0.321) | | Corr Coef (R): | 0.991 | | | Bias: | 0.047 (0.657 %) | | | Points<br>(Plotted/Total): | 120/120 | | | Result Ranges: | 4.50% to 16.5% | | #### Total Error Near the Cutoff Using the results of bias estimation (%Bias) in the method comparison study and precision estimates in the precision study, Total Error (TE) at four concentrations: (5.0%, 6.5%, 8.0% and 12.0) was calculated as follows: %TE =|%Bias| + 1.96 *%CV*(1+%Bias). The results are presented in the tables below. %TE of ≤6% is acceptable. | Decision Level % | % Bias | 1%CV | %TE | |-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|-------------|----------| | | 1.08 | 0 .<br>1.04 | W.V | | ﻟﻠﺘﻌﻠﻴﻘﺎﺕ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤ | 0 75<br>N.I. | 0.68 | data o O | {8}------------------------------------------------ | 8.0 | 0.54 | 0.98 | 3.0 | |------|------|------|-----| | 12.0 | 0.24 | 1.31 | 3.1 | Using the calculation: (TE)95 = Bias + 1.96*SD the results are shown in the following table: | Decision Level % | Bias | SD | (TE)95 | Goal at 6% | |------------------|-------|------|--------|------------| | 5.0 | 0.054 | 0.09 | 0.23 | 0.30 | | 6.5 | 0.049 | 0.04 | 0.13 | 0.39 | | 8.0 | 0.043 | 0.08 | 0.20 | 0.48 | | 12.0 | 0.029 | 0.16 | 0.34 | 0.72 | Image /page/8/Figure/4 description: This image is a scatter plot with the title "Scatter Plot". The x-axis is labeled "Ultra 2 (%)" and ranges from 5 to 15. The y-axis is labeled "G8 (%)" and ranges from 5 to 15. The plot shows a cluster of points that follow a linear trend, along with a legend indicating "Deming Regr", "1:1 Line", and "Med Dec Pt". ### Standards: | Number | FDA<br>Recognition<br>Number | Revision<br>Date | Title | |----------------|------------------------------|------------------|-------------------------------------------------------------------------------------------------------------------| | EP5-A2 | 7-110 | 10/31/2005 | Evaluation of Precision Performance of<br>Quantitative Measurement Methods: Approved<br>Guideline- Second Edition | | EP09-A2-<br>IR | 7-92 | 03/08/2004 | Method Comparison and Bias Estimation Using<br>Patient Samples; Approved Guideline-Second<br>Edition | | EP7-A2 | 7-127 | 05/21/2007 | Interference Testing in Clinical Chemistry;<br>Approved Guideline - Second Edition | {9}------------------------------------------------ ### Linearity and Detection Limit: Linearity was previously established for this assay under k071132. The reportable range for this device is 4.0-16.9% HbAlc. #### Calibrator and Controls: Calibrators (Tosoh A 1c Calibrator Set) and Controls (Canterbury Scientific Hemoglobin A 1c) are recommended for use with this device. The calibrators and controls were previously cleared under 510(k) numbers k071132 and k021484 respectively. #### Traceability The assigned HbA1c values of the Tosoh Automated Glycohemoglobin Analyzer are certified with The National Glycohemoglobin Standardization Program (NGSP). The NGSP certification expires in one year. See NGSP website for current certification at http://www.ngsp.org. · The final reportable result is traceable to both the International Federation if Clinical Chemistry (IFCC) and the Diabetes Control and Complications Trial (DCCT). The International Federation of Clinical Chemistry (IFCC) units of mmol/mol are calculated using the Master Equation NGSP (%) =0.09148 x IFCC (mmol/mol) + 2.52. HbA 1c results are provided to the customers using two different units: NGSP equivalent units (%) and IFCC equivalent units (mmol/mol). #### Conclusion: The Tosoh Bioscience, Inc. Automated Glycohemoglobin Analyzer HLC-723G8 is substantially equivalent to the Roche Diagnostics Corporation, COBAS INTEGRA 800 Tina-quant HbA 1cDx Gen.2 assay (k121291) for in vitro diagnostic use for the measurement of hemoglobin Alc (HbA1c) in whole blood specimens and to be used as an aid in diagnosis of diabetes and identifying patients who may be at risk for developing diabetes. {10}------------------------------------------------ Image /page/10/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight. The symbol is composed of three curved lines that form the body and wings of the bird. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 January 23, 2014 TOSOH BIOSCIENCE. INC. C/O ROBERT WICK 6000 SHORELINE COURT SUITE 101 SOUTH SAN FRANCISCO CA 94080 Re: K131580 Trade/Device Name: Automated Glycohemoglobin Analyzer HLC-723G8 Regulation Number: 21 CFR 862.1373 Regulation Name: Hemoglobin A1c Test System Regulatory Class: II Product Code: PDJ Dated: December 20, 2013 Received: December 20, 2013 Dear Mr. Wick: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {11}------------------------------------------------ Page 2-Mr. Wick if you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/defiult.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Courtney H.Lias-S Courtney H. Lias, Ph.D. Director Division of Chemistry. and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {12}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ### Indications for Use 510(k) Number (if known) k131580 #### Device Name Automated Glycohemoglobin Analyzer HLC-723G8 #### Indications for Use (Describe) The Tosoh Automated Glycohemoglobin Analyzer HLC-723G8 is intended for in vitro diagnostic use for the quantitative measurement of % hemoglobin Alc (HbA Ic) (DCCT/NGSP) and mmol/mo hemoglobin A Ic (IFCC) in whole blood specimens. This test is to be used as an aid in diagnosis of diabetes and as an aid in identifying patients who may be at risk for developing diabetes. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. FOR FDA USE ONLY .... Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) Stayce Beck FORM FDA 3881 (1/14) Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page. {13}------------------------------------------------ This section applies only to requirements of the Paperwork Reduction Act of 1995. # *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* *** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...