ADVIA Chemistry Enzymatic Hemoglobin Ale (Alc_E) Assay

K200256 · Siemens Healthcare Diagnostics, Inc. · PDJ · Jul 2, 2021 · Clinical Chemistry

Device Facts

Record IDK200256
Device NameADVIA Chemistry Enzymatic Hemoglobin Ale (Alc_E) Assay
ApplicantSiemens Healthcare Diagnostics, Inc.
Product CodePDJ · Clinical Chemistry
Decision DateJul 2, 2021
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 862.1373
Device ClassClass 2

Intended Use

The ADVIA® Chemistry Enzymatic Hemoglobin A1c (A1c E) assay is an in vitro diagnostic assay for the quantitative determination of mmol/mol HbA1c (IFCC) and % HbA1c (DCCT/NGSP) in human anticoagulated venous whole blood and hemolysate for use on the ADVIA® Chemistry Systems. Measurement of Hemoglobin A 1 c is used as an aid in the diagnosis and monitoring of long-term blood glucose control in patients with diabetes mellitus, and as an aid in the identification of patients at risk for developing diabetes mellitus.

Device Story

ADVIA® Chemistry Enzymatic Hemoglobin A1c (A1c E) assay is an in vitro diagnostic test for quantitative determination of HbA1c in human anticoagulated venous whole blood and hemolysate. Device operates on ADVIA® Chemistry Systems; utilizes enzymatic methodology to measure HbA1c levels. Assay consists of three liquid, ready-to-use reagents (R1, R2, and Pretreatment Solution). Automated application lyses whole blood specimens on-system; manual application uses Pretreatment Solution to create hemolysate. Healthcare professionals use results to aid in diagnosis and monitoring of long-term blood glucose control in patients with diabetes mellitus and identification of at-risk patients. Modification extends low end of analytical measuring range for total hemoglobin (tHb).

Clinical Evidence

Bench testing only. Verification and validation activities performed based on risk analysis to confirm performance following the modification of the total hemoglobin measurement range.

Technological Characteristics

Quantitative enzymatic assay. Reagents: R1, R2, and Pretreatment Solution. Measuring range: 3.8-14.0% HbA1c (DCCT/NGSP); 18.03-129.50 mmol/mol HbA1c (IFCC). Total hemoglobin analytical measuring range: 74.74-320.15 μmol/L. Instrument platform: ADVIA® 1800 Clinical Chemistry System. Standardization: Traceable to IFCC reference calibrators; NGSP certified.

Indications for Use

Indicated for patients with diabetes mellitus or those at risk for developing diabetes mellitus. Used for diagnosis and monitoring of long-term blood glucose control.

Regulatory Classification

Identification

A hemoglobin A1c test system is a device used to measure the percentage concentration of hemoglobin A1c in blood. Measurement of hemoglobin A1c is used as an aid in the diagnosis of diabetes mellitus and as an aid in the identification of patients at risk for developing diabetes mellitus.

Special Controls

(b) Classification. Class II (special controls). Hemoglobin A1c test systems must comply with the following special controls: 1) The device must have initial and annual standardization verification by a certifying glycohemoglobin standardization organization deemed acceptable by FDA. 2) The premarket notification submission must include performance testing to evaluate precision, accuracy, linearity and interference, including the following: i) Performance testing of device precision must, at a minimum, use blood samples with concentrations near 5.0%, 6.5%, 8.0% and 12% hemoglobin A1c. This testing must evaluate precision over a minimum of 20 days using at least 3 lots of the device and 3 instruments, as applicable. ii) Performance testing of device accuracy must include a minimum of 120 blood samples that span the measuring interval of the new device and compare results of the new device to results of the standardized test method. Results must demonstrate little or no bias versus the standardized method. iii) Total error of the new device must be evaluated using single measurements by the new device compared to results of the standardized test method, and this evaluation must demonstrate a total error less than or equal to 6%. iv) Performance testing must demonstrate that there is little to no interference from common hemoglobin variants, including Hemoglobin C, Hemoglobin D, Hemoglobin E, Hemoglobin A2 and Hemoglobin S. 3) When assay interference from Hemoglobin F or interference with other hemoglobin variants with low frequency in the population is observed, a warning statement must be placed in a black box and must appear in all labeling material for these devices describing the interference and any affected populations.

*Classification.* Class II (special controls). The special controls for this device are:(1) The device must have initial and annual standardization verification by a certifying glycohemoglobin standardization organization deemed acceptable by FDA. (2) The premarket notification submission must include performance testing to evaluate precision, accuracy, linearity, and interference, including the following: (i) Performance testing of device precision must, at a minimum, use blood samples with concentrations near 5.0 percent, 6.5 percent, 8.0 percent, and 12 percent hemoglobin A1c. This testing must evaluate precision over a minimum of 20 days using at least three lots of the device and three instruments, as applicable. (ii) Performance testing of device accuracy must include a minimum of 120 blood samples that span the measuring interval of the device and compare results of the new device to results of a standardized test method. Results must demonstrate little or no bias versus the standardized method. (iii) Total error of the new device must be evaluated using single measurements by the new device compared to results of the standardized test method, and this evaluation must demonstrate a total error less than or equal to 6 percent. (iv) Performance testing must demonstrate that there is little to no interference from common hemoglobin variants, including Hemoglobin C, Hemoglobin D, Hemoglobin E, Hemoglobin A2, and Hemoglobin S. (3) When assay interference from Hemoglobin F or interference with other hemoglobin variants with low frequency in the population is observed, a warning statement must be placed in a black box and must appear in all labeling material for these devices describing the interference and any affected populations.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" stacked on top of each other. July 2, 2021 Siemens Healthcare Diagnostics, Inc. Julie Warren Regulatory Affairs Professional 511 Benedict Avenue Tarrytown, New York 10591 Re: K200256 Trade/Device Name: ADVIA® Chemistry Enzymatic Hemoglobin Ale (Alc E) Assay Regulation Number: 21 CFR 864.7470 Regulation Name: Glycosylated Hemoglobin Assay Regulatory Class: Class II Product Code: LCP, PDJ Dated: October 12, 2020 Received: October 14, 2020 Dear Julie Warren: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Marianela Perez-Torres, Ph.D. Deputy Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K200256 Device Name ADVIA® Chemistry Enzymatic Hemoglobin Ale (Alc E) Assay Indications for Use (Describe) The ADVIA® Chemistry Enzymatic Hemoglobin A1c (A1c E) assay is an in vitro diagnostic assay for the quantitative determination of mmol/mol HbA1c (IFCC) and % HbA1c (DCCT/NGSP) in human anticoagulated venous whole blood and hemolysate for use on the ADVIA® Chemistry Systems. Measurement of Hemoglobin A 1 c is used as an aid in the diagnosis and monitoring of long-term blood glucose control in patients with diabetes mellitus, and as an aid in the identification of patients at risk for developing diabetes mellitus. Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 20px;">☑</span> | |----------------------------------------------|-----------------------------------------| | Over-The-Counter Use (21 CFR 801 Subpart C) | <span style="font-size: 20px;">☐</span> | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots arranged in a circular pattern. #### 510(k) Summary – K200256 This 510(k) Summary is being submitted in accordance with the requirements of the Safe Medical Device Act of 1990 and 21 CFR 807.92. ### 1. Submitter Siemens Healthcare Diagnostics Inc. 511 Benedict Avenue Tarrytown, NY 10591 USA Establishment Registration Number: 2432235 | Submission Contact Person: | Julie Warren | |----------------------------|---------------------------------------| | Phone: | (302) 631-8722 | | Email: | julie.warren@siemens-healthineers.com | | Facsimile: | (302) 631-6299 | | Date of Preparation | June 17, 2021 | #### 2. Device Information Federal (USA) law restricts this device to sale by or on the order of a licensed healthcare professional. For in vitro diagnostic use. | Proprietary Name | ADVIA® Chemistry Enzymatic Hemoglobin A1c<br>(A1c_E) Assay | | |----------------------|------------------------------------------------------------|--------------------------------| | Common or Usual Name | Hemoglobin A1c Test System | Assay, Glycosylated Hemoglobin | | Classification Name | Hemoglobin A1c Test System | Glycosylated hemoglobin assay | | Regulation | 21 CFR 862.1373 | 21 CFR 864.7470 | | Classification | Class II | Class II | | Product Code | PDJ | LCP | | Review Panel | Clinical Chemistry | Hematology | {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" is a graphic of orange dots in a circular pattern. ### 3. Device Description The ADVIA® Chemistry Enzymatic Hemoglobin A1c (A1c E) assay measures hemoglobin A1c in human anticoagulated whole blood and hemolysate. The assay consists of three reagents (R1, R2, and Pretreatment Solution), which are liquid and ready to use. The assay offers both an automated and a manual application. The automated application (A1c_E) lyses the anticoagulated whole blood specimen on the system for the automated application (A1c E). Samples may also be lysed manually using the ADVIA® Chemistry A1c_E Pretreatment Solution to obtain hemolysate for the manual application (A1c_EM). The two applications yield the same results. ## 4. Intended Use/Indications for Use The ADVIA® Chemistry Enzymatic Hemoglobin A1c (A1c E) assay is an in vitro diagnostic assay for the quantitative determination of mmol/mol HbA1c (IFCC) and % HbA1c (DCCT/NGSP) in human anticoagulated venous whole blood and hemolysate for use on the ADVIA® Chemistry Systems. Measurement of Hemoglobin A1c is used as an aid in the diagnosis and monitoring of long-term blood glucose control in patients with diabetes mellitus, and as an aid in the identification of patients at risk for developing diabetes mellitus. ### 5. Reason for Submission This Special 510(k) is being filed to seek FDA clearance of a modification to the ADVIA Chemistry Enzymatic Hemoglobin A1c (A1c_E) assay to extend the low end of the analytical measuring range of total hemoglobin (tHb). A Special 510(k) Premarket Notification is the requested pathway because the following: - . There is no change to the intended use or indications for use. - There is no change in the fundamental scientific technology. - There is no change to the principle of operation. - . There is no change to the formulation. - There is no change to the instrument parameters related to sample volume, reagent volume, mix speed, wavelengths, or read times. {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the words is a cluster of orange dots of varying sizes. ## 6. Medical Device to Which Equivalence is Claimed The predicate device for this Special 510(k) submission is the ADVIA® Chemistry Enzymatic Hemoglobin A1c (A1c_E) assay cleared under K171771 on December 4, 2017. The classification information is as follows. | Trade Name | Common/Usual Name | Classification | Product<br>Code | Panel | |---------------------------------------------------------------|---------------------------------------------------------------------|----------------|-----------------|--------------------------------------| | ADVIA® Chemistry<br>Enzymatic Hemoglobin<br>A1c (A1c_E) Assay | Hemoglobin A1c Test<br>System;<br>Assay, Glycosylated<br>Hemoglobin | Class II | PDJ; LCP | Clinical<br>Chemistry;<br>Hematology | ### 7. Comparison of Technological Characteristics with the Predicate Device The similarities and differences to the predicate device were compared side by side in the following table. | Attribute | Predicate Device<br>ADVIA® Chemistry<br>Enzymatic Hemoglobin A1c (A1c_E) Assay,<br>K171771 | Modified Device<br>ADVIA® Chemistry<br>Enzymatic Hemoglobin<br>A1c (A1c_E) Assay | |----------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------| | Intended Use/<br>Indications for<br>Use | The ADVIA® Chemistry Enzymatic Hemoglobin A1c (A1c_E) assay is an <i>in vitro</i> diagnostic assay for the quantitative determination of mmol/mol HbA1c (IFCC) and % HbA1c (DCCT/NGSP) in human anticoagulated whole blood and hemolysate for use on the ADVIA® Chemistry systems.<br>Measurement of Hemoglobin A1c is used as an aid in the diagnosis and monitoring of patients with diabetes mellitus, and as an aid in the identification of patients at risk for developing diabetes mellitus. | Same | | Attribute | Predicate Device<br>ADVIA® Chemistry<br>Enzymatic Hemoglobin A1c (A1c_E) Assay,<br>K171771 | Modified Device<br>ADVIA® Chemistry<br>Enzymatic Hemoglobin<br>A1c (A1c_E) Assay | | Technology; Type of Test | Quantitative, enzymatic | Same | | Measuring Range | 3.8 to 14.0% HbA1c (DCCT/NGSP)<br>18.03 -129.50 mmol/mol HbA1c (IFCC) | Same | | Analytical<br>Measuring Range<br>for total<br>hemoglobin | 88.98-320.15 μmol/L | 74.74-320.15 μmol/L | | Specimen Types | Whole blood and Hemolysate | Same | | Anticoagulant<br>Types | Dipotassium EDTA<br>Lithium Heparin<br>Sodium Fluoride/Disodium EDTA<br>Tripotassium EDTA | Same | | Standardization<br>and Certification | Assay standardization is traceable to International<br>Federation of Clinical Chemistry (IFCC) reference<br>calibrators. Assay is certified with the National<br>Glycohemoglobin Standardization Program<br>(NGSP). The NGSP certification must be renewed<br>annually. | Same | | Instrument<br>Platform | ADVIA® 1800 Clinical Chemistry System | Same | | Reporting Units | % HbA1c NGSP/DCCT and mmol/mol IFCC | Same | {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots arranged in a pattern. {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots. ### 8. Summary of Design Control Activities A risk analysis and verification activities were conducted as part of design control activities. ### 8.1 Risk Analysis A risk analysis was conducted in accordance with ISO 14971:2012 standard, Medical Devices – Application of Risk Management to Medical Devices to identify and assess the risks associated with the device modification described in this submission and the risks associated with compliance to the FDA Special Controls set forth in 21 CFR 862.1373(b). The normal condition and fault condition/failure mode were identified for each risk. Risk mitigations were identified, and effectiveness confirmed through verification activities. ### 8.2 Verification Activities Based on the risk analysis, risk mitigations identified were verified for effectiveness in accordance with relevant standards and established protocols with predetermined acceptance criteria. Testing verified all acceptance criteria were met. #### 9. Conclusion The ADVIA® Chemistry Enzymatic Hemoglobin A1c (A1c E) Assay modified device is substantially equivalent to the predicate in intended use/indications for use, fundamental technology, principle of operation, formulation, and method parameters related to sample and reagent volume, mix speed, wavelengths, and read times. The change to the device was to extend the low end of the measuring range for total hemoglobin (tHb). The change is to Siemens Healthcare Diagnostics Inc. own device. Performance data were needed to evaluate the change. Risk analysis and verification activities were performed as part of design control activities. The risks identified to the assay Special Controls set forth in 21 CFR 862.1373(b) and the device modification were acceptable and mitigations are effective. The verification study of linearity was done in accordance with the CLSI standard recognized by the FDA. This study along with other verification activities demonstrate equivalent performance to the predicate and effective risk mitigations. The studies met pre-determined acceptance criteria. ADVIA® Chemistry Enzymatic Hemoglobin A1c (A1c E) Assay Siemens Healthcare Diagnostics Inc. 510(k) Summary Page 5 of 6 {8}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots arranged in a circular pattern. Results from design control activities including the risk analysis and the verification activities support substantial equivalence between the change device and the predicate. The changes to not impact the safety or effectiveness of the device. #### END OF SUMMARY
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...