ADVIA CHEMISTRY HEMOGLOBIN A1C, ADVIA CHEMISTRY A1C CALIBRATORS

K081895 · Siemens Medical Solutions Diagnostics · LCP · Jan 8, 2009 · Hematology

Device Facts

Record IDK081895
Device NameADVIA CHEMISTRY HEMOGLOBIN A1C, ADVIA CHEMISTRY A1C CALIBRATORS
ApplicantSiemens Medical Solutions Diagnostics
Product CodeLCP · Hematology
Decision DateJan 8, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.7470
Device ClassClass 2

Intended Use

The ADVIA® Chemistry Hemoglobin A1c method is for in vitro diagnostic use in the quantitative determination of Hemoglobin A1c, a diabetes marker, in whole blood on the ADVIA Chemistry systems. Such measurements are used for monitoring the long-term glycemic control of persons with diabetes. The A1c and total hemoglobin (tHb) values generated as part of the HbA1cN and HbA1cI results are intended for use in the calculation of the A1c / total hemoglobin ratio, and must not be used individually for diagnostic purposes. *Note: HbA1cN reports HbA1c in % and HbA1cI reports HbA1c in mmol/mol The ADVIA® Chemistry A1c Calibrators are for in vitro diagnostic use in the calibration of the A1c and total hemoglobin methods (Automated and Manual Pretreatment) on the ADVIA Chemistry Systems.

Device Story

ADVIA Chemistry Hemoglobin A1c Assay measures A1c and total hemoglobin in whole blood samples; utilizes automated or manual pretreatment to lyse red blood cells and hydrolyze hemoglobin chains. Total hemoglobin measured via conversion to alkaline hematin; A1c measured via latex agglutination inhibition where A1c competes with synthetic latex agglutinator for anti-A1c antibody. Change in scattered light/absorbance inversely proportional to A1c concentration. System calculates A1c/total hemoglobin ratio. Used in clinical laboratories on ADVIA Chemistry systems; operated by trained laboratory personnel. Results assist clinicians in monitoring long-term glycemic control in diabetic patients. Calibrators are lyophilized whole blood at multiple levels used to calibrate the assay.

Clinical Evidence

Bench testing only. Precision studies (N=80 per analyzer/method) showed total imprecision CVs ranging from 1.1% to 4.4%. Linearity assessed across 2.9-15.4% A1c range with R values ≥ 0.998. Method comparison against Tosoh G7 (N=80) yielded R values ≥ 0.991. Interference studies confirmed no significant bias from bilirubin, triglycerides, rheumatoid factor, labile A1c, or carbamylated A1c. HbF >10% causes negative bias.

Technological Characteristics

Latex agglutination inhibition assay for HbA1c; alkaline hematin method for total hemoglobin. Reagents include anti-HbA1c coupled latex (mouse), A1c hapten-polymer, porcine pepsin, and sodium hydroxide. Lyophilized calibrators. Analyzers: Siemens ADVIA 1200, 1650, 1800, 2400. Traceable to NGSP and IFCC. Automated or manual pretreatment.

Indications for Use

Indicated for the quantitative determination of Hemoglobin A1c in whole blood to monitor long-term glycemic control in patients with diabetes.

Regulatory Classification

Identification

A glycosylated hemoglobin assay is a device used to measure the glycosylated hemoglobins (A1a , A1b , and A1c ) in a patient's blood by a column chromatographic procedure. Measurement of glycosylated hemoglobin is used to assess the level of control of a patient's diabetes and to determine the proper insulin dosage for a patient. Elevated levels of glycosylated hemoglobin indicate uncontrolled diabetes in a patient.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K081895 # 510(k) Summary JAN - 8 2009 | Submitter information<br>Contact person: | Philip Liu<br>Manager, Regulatory Affairs & Compliance | | | |-----------------------------------------------------|------------------------------------------------------------------------------|---------------------------------------------------------------------------|--| | Address: | Siemens Healthcare Diagnostics<br>511 Benedict Avenue<br>Tarrytown, NY 10591 | | | | Phone: | 914-524-2443<br>914-524-2500 (fax) | | | | Date summary prepared: | November 12, 2008 | | | | Device Trade or Proprietary Names: | | ADVIA® Chemistry Hemoglobin A1c Assay<br>ADVIA® Chemistry A1c Calibrators | | | Device Common/Usual Name or<br>Classification Name: | | Glycosylated Hemoglobin Assay<br>Calibrators | | | Classification Number/Class: | | LCP / Class II | | JIX / Class II This 510(k) summary of safety and effectiveness is being submitted in accordance with the requirements of 21 CFR 807.92. The assigned 510(k) number is: _k081895 ### Assay Predicate Device: | | Predicate Device | | |---------------|----------------------------------------------------------|--| | Device Name | G7 Automated HPLC Analyzer - HbA1c Variant Analysis Mode | | | Common name | Glycosylated Hemoglobin Test | | | 510(k) Number | k011434 | | | Manufacturer | Tosoh Medics, Inc. | | ## Calibrator Predicate Device: | | Predicate Device | | |---------------|---------------------------------------------|--| | Device Name | VITROS Chemistry Products Calibrator Kit 18 | | | Common name | Calibrator | | | 510(k) Number | k041764 | | | Manufacturer | Ortho-Clinical Diagnostics | | Siemens Healthcare Diagnostics (formerly Siemens Medical Solutions Diagnostics) ADVIA Chemistry Hemoglobin A1c Assay and Calibrators Premarket Notification - 510(k) 510(k) Summary {1}------------------------------------------------ ### Device Description: The concentration of A1c and the concentration of total hemoglobin are measured and the ratio is reported (either as % or mmol/mol). There are two different sample pretreatment methods available on the ADVIA Chemistry System. The first method is an Automated Pretreatment that uses 4 reagents: A1c Denaturant Reagent, Total Hemoglobin Reagent (tHb 2), A1c Agglutinator Reagent (R1) and A1c Antibody Reagent (R2). In this Automated Pretreatment step, the whole blood sample is mixed with the A1c Denaturant Readent. The red blood cells are lysed and the hemoglobin chains are hydrolyzed by the protease present in the reagent. The second method is a Manual Pretreatment that uses the same reagents as the first method except that the A1c Denaturant Reagent is replaced with the Hemoglobin Denaturant Reagent. For this method, there is an off-line pretreatment that is followed bv a 10 minute incubation. For the measurement of total hemoglobin, the Total Hemoglobin Reagent is used. The method is based on the conversion of all hemoglobin derivatives into alkaline hematin in an alkaline solution of a nonionic detergent A latex agglutination inhibition method is used for the measurement of specific A1c. The A1c present in the sample competes with the agglutinator (synthetic latex containing multiple copies of the immunoreactive portion of A1c) for the anti-A1c antibody; thereby reducing the rate of agglutination. A concentration curve is obtained by monitoring the change in scattered light as a change of absorbance. The actual change in absorbance is inversely proportional to the concentration of A1c in the sample. The HbA1cN NGSP result (%) or the HbA1cl IFCC result (mmol/mol) is calculated using the A1c and total hemoglobin values. The ADVIA® Chemistry A1c Calibrators are used to calibrate the methods. The calibrators consist of four (4) levels of lyophilized whole blood containing varying concentrations of HbA1c and total hemoglobin. There is a single level calibration for total hemoglobin (Cal 1) and a multi-level calibration (six levels) for A1c. Four calibrator levels (designated Cal 1 - 4) are provided in a single kit and each level is 0.5 g/vial. The other two levels consist of Saline (Cal 0) and Cal 5 (prepared by the system from Cal 4 using 1.4 times the volume used for Cal 4). The target value of each calibrator (calibration) level is: - Calibration Level 1 (Cal 0): 0.00 umol/L A1c, 0.0 g/dL Total Hemoglobin ● - Calibration Level 2 (Cal 1): 2.30 µmol/L A1c, 11.0 g/dL Total Hemoglobin . - Calibration Level 3 (Cal 2): 3.65 umol/L A1c . - Calibration Level 4 (Cal 3): 5.15 umol/L A1c . - Calibration Level 5 (Cal 4): 6.80 umol/L A1c . - Calibration Level 6 (Cal 5): 8.20 umol/L A1c Siemens Healthcare Diagnostics (formerly Siemens Medical Solutions Diagnostics) ADVIA Chemistry Hemoglobin A1c Assay and Calibrators Premarket Notification - 510(k) 510(k) Summary {2}------------------------------------------------ ### Storage is at 2 - 8°C. CAUTION! POTENTIAL BIOHAZARD: Contains human source material. While each human serum or plasma donor unit used in the manufacture of this product was tested by FDA-approved methods and found nonreactive for hepatitis B surface antigen (HBsAg), antibody to hepatitis C (HCV), and antibody to HIV-1/2, all products manufactured using human source material should be handled as potentially infectious. Because no test method can offer complete assurance that hepatitis B or C viruses, HIV, or other infectious agents are absent, these products should be handled according to established good laboratory practices. ### Statement of Intended Use: The ADVIA® Chemistry Hemoglobin A1c method is for in vitro diagnostic use in the quantitative determination of Hemoglobin A1c, a diabetes marker, in whole blood on the ADVIA Chemistry systems. Such measurements are used for monitoring the long-term alvcemic control of persons with diabetes. The A1c and total hemoglobin (tHb) values generated as part of the HbA1cN and HbA1cl methods are intended for use in the calculation of the A1c / total hemoglobin ratio, and must not be used individually for diagnostic purposes. *Note: HbA1cN reports HbA1c in % and HbA1cl reports HbA1c in mmol/mol The ADVIA® Chemistry A1c Calibrators are for in vitro diagnostic use in the calibration of the A1c and total hemoglobin methods (Automated and Manual Pretreatment) on the ADVIA Chemistry Systems. #### Comparisons to the Predicate Device: #### Method Similarities | | ADVIA Chemistry<br>Hemoglobin A1c<br>( <i>new device</i> ) | Tosoh G7 Automated HPLC Analyzer<br>- HbA1c Variant Analysis Mode<br>( <i>predicate device</i> ) | |-----------------|-----------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------| | Intended Use | For <i>in vitro</i> diagnostic use for the<br>measurement of hemoglobin A1c | For <i>in vitro</i> diagnostic use for the<br>measurement of hemoglobin A1c | | Specimen Type | Human Whole Blood<br>(EDTA and Lithium Heparin) | Human Whole Blood | | Standardization | Traceable to NGSP***<br>Traceable to IFCC by Master Equation | Traceable to NGSP*** | {3}------------------------------------------------ ## Method Differences | | ADVIA Chemistry<br>Hemoglobin A1c<br>(new device) | Tosoh G7 Automated HPLC Analyzer<br>- HbA1c Variant Analysis Mode<br>(predicate device) | |-------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------| | Assay Principle | For Hemoglobin:<br>Conversion of all hemoglobin<br>derivatives into alkaline hematin<br>For HbA1c:<br>Latex agglutination inhibition | High Performance Liquid<br>Chromatography<br>(non-porous ion exchange) | | Determination /<br>Calculation of results | Ratio of HbA1c/ Total Hemoglobin | Measurement of HbA1c as a<br>percentage of total hemoglobin | | Calibration | Multipoint calibration (6 levels) for<br>Hemoglobin A1c<br>Single point calibration for Total<br>Hemoglobin | 2 point Calibration (high and low) | | Analytical Range | 2.9 - 15.4% | 4.2 - 20.8% | ## Calibrator Similarities | | ADVIA Chemistry<br>A1c Calibrators<br>(new device) | VITROS Chemistry Products Calibrator<br>Kit 18<br>(predicate device) | |---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use | For in vitro diagnostic use in the<br>calibration of the A1c and total<br>hemoglobin methods (Automated and<br>Manual Pretreatment) on the ADVIA<br>Chemistry Systems | For in vitro diagnostic use<br>only. VITROS Chemistry Products<br>Calibrator Kit 18 is used in conjunction<br>with VITROS Chemistry Products FS<br>Calibrator 1 to calibrate VITROS 5, 1<br>FS Chemistry Systems for the<br>calculation of percent glycated<br>hemoglobin (%A1c). | | Calibrator Levels | 4 levels provided | 4 levels provided | | Matrix / Ingredients | Processed human whole blood | Hemolysate derived from human and<br>ovine blood, surfactants, stabilizer, and<br>preservatives | | Form | Lyophilized | Lyophilized | | Storage | 2 - 8°C | 2 - 8°C | | Shelf Life Stability | stable until the expiration date | stable until the expiration date | | Reconstituted Vial /<br>Use Stability | 7 days stored at 2 - 8°C | 2 days stored at 2 - 8ºC | ## Calibrator Differences | | ADVIA Chemistry<br>A1c Calibrators<br>(new device) | VITROS Chemistry Products Calibrator<br>Kit 18<br>(predicate device) | |---------------------------------------|----------------------------------------------------|----------------------------------------------------------------------| | Instructions for Use<br>(Preparation) | Reconstitute with DI water before use | Reconstitute with diluent before use | Siemens Healthcare Diagnostics (formerly Siemens Medical Solutions Diagnostics) ADVIA Chemistry Hemoglobin A1c Assay and Calibrators Premarket Notification - 510(k) 510(k) Summary . {4}------------------------------------------------ #### Performance: Substantial equivalence was demonstrated by testing several performance characteristics including imprecision, method comparison, interfering substances, matrix equivalency, and analytical range. The following tables summarize the precision (total), interfering substances, analytical range, and method comparison results. All of the evaluation studies gave acceptable results compared to the Predicate Device. These studies support that the ADVIA Chemistry Hemoglobin A1c assay (with the ADVIA Chemistry A1c Calibrators) is substantially equivalent to the Tosoh G7 HPLC Hemoglobin A1c method that is currently marketed. #### Imprecision | ADVIA Chemistry<br>Hemoglobin A1c<br>ADVIA 1800 - Automated | | ADVIA Chemistry<br>Hemoglobin A1c<br>ADVIA 1800 - Manual | | Tosoh G7 HPLC<br>HbA1c | | |-------------------------------------------------------------|-----------------|----------------------------------------------------------|-----------------|------------------------|-----------------| | Level<br>HbA1c (%) | Total CV<br>(%) | Level<br>HbA1c (%) | Total CV<br>(%) | Level<br>HbA1c (%) | Total CV<br>(%) | | 5.5 | 1.9 | 5.0 | 1.9 | 5.8 | 1.12 | | 9.6 | 1.3 | 8.8 | 1.4 | 10.9 | 0.71 | #### Correlation #### (y = ADVIA Chemistry Hemoglobin A1c, x = comparison method/system) | System (y) | N | Regression<br>Equation | r | Sy.x | Sample Range<br>(%) | |------------------------------------------------------|----|------------------------|-------|------|---------------------| | ADVIA 1800 Automated –<br>Least Squares | 80 | $y = 0.98x + 0.17$ | 0.997 | 0.17 | 4.40 - 12.60 | | ADVIA 1800 Automated –<br>Passing Bablok / Pearson's | 80 | $y = 1.00x + 0.06$ | 1.00 | -- | 4.40 - 12.60 | | ADVIA 1800 Manual –<br>Least Squares | 80 | $y = 0.95x + 0.26$ | 0.995 | 0.19 | 4.40 - 12.60 | | ADVIA 1800 Manual –<br>Passing Bablok / Pearson's | 80 | $y = 0.97x + 0.10$ | 1.00 | -- | 4.40 - 12.60 | x = NGSP Reference Method (Tosoh G7), y = ADVIA Chemistry System #### Interfering Substances (ADVIA Chemistry Hemoglobin A1c on ADVIA 1800) #### Automated Sample Pretreatment | Interfering Substance | Interferent Conc.<br>(mg/dL) | HbA1c conc.<br>(%) | Effect<br>(% change) | |----------------------------|------------------------------|--------------------|----------------------| | Triglycerides (Intralipid) | 500 mg/dL | 5.2 | - 7.8 | | Bilirubin, free | 60 mg/dL | 5.3 | - 1.9 | | Bilirubin, conjugated | 60 mg/dL | 5.2 | - 3.8 | Siemens Healthcare Diagnostics (formerly Siemens Medical Solutions Diagnostics) ADVIA Chemistry Hemoglobin A1c Assay and Calibrators Premarket Notification - 510(k) 510(k) Summary Page 5 of 6 {5}------------------------------------------------ #### Manual Sample Pretreatment | Interfering Substance | Interferent Conc.<br>(mg/dL) | HbA1c conc.<br>(%) | Effect<br>(% change) | |----------------------------|------------------------------|--------------------|----------------------| | Triglycerides (Intralipid) | 500 mg/dL | 5.2 | - 7.7 | | Bilirubin, free | 60 mg/dL | 5.1 | - 2.0 | | Bilirubin, conjugated | 60 mg/dL | 5.1 | - 5.9 | | Rheumatoid Factor | 1100 IU/mL | 4.8 | 0 | #### Analytical Range | Platform | Analytical Range | | | |-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------|----| | All ADVIA Chemistry ' ' | - Galibrator Leval 6-(12-3-18.6%)<br>1 -1 -1 -1 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- | 2.4 /0 - | 10 | #### Conclusions: The ADVIA Chemistry Hemoglobin A1c method and Calibrators are substantially equivalent to other products in commercial distribution intended for similar use. Most notably, it is substantially equivalent to the currently marketed Tosoh G7 Automated HPLC Analyzer- HbA1c Variant Analysis Mode. The Tosoh G7 Automated HPLC Analyzer is certified by the National Glycohemoglobin Standardization Program (NGSP) and it is used as a reference method for establishing traceable results of other methods to the Diabetes Control and Complications Trial (DCCT). The ADVIA Chemistry A1c Calibrators are substantially equivalent to the Predicate Device VITROS Chemistry Products Calibrator Kit 18. {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an emblem featuring a stylized eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans. #### DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Siemens Healthcare Diagnostic c/o Dr. Philip Liu Manager, Regulatory Affairs and Compliance 511 Benedict Avenue Tarrytown, NY 10591 JAN - 8 2009 k081895 Re: Trade/Device Name: ADVIA Chemistry Hemoglobin A1c Assay, ADVIA Chemistry Alc Calibrators Regulation Number: 21 CFR 864.7470 Regulation Name: Glycosylate hemoglobin assay. Regulatory Class: Class II Product Code: LCP, JIX Dated: November 18, 2008 Received: November 19, 2008 Dear Dr. Liu: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820). {7}------------------------------------------------ Page 2 - This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Gregory C. Hammett, Ph.D. Courthey C. Harper, Ph.D. Acting Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {8}------------------------------------------------ # Indication for Use 510(k) Number (if known): k081895 Device Name(s): ADVIA® Chemistry Hemoglobin A1c ADVIA® Chemistry A1c Calibrators Indications For Use: The ADVIA® Chemistry Hemoglobin A1c method is for in vitro diagnostic use in the quantitative determination of Hemoglobin A1c, a diabetes marker, in whole blood on the ADVIA Chemistry systems. Such measurements are used for monitoring the long-term glycemic control of persons with diabetes. The A1c and total hemoglobin (tHb) values generated as part of the HbA1cN and HbA1cI results are intended for use in the calculation of the A1c / total hemoglobin ratio, and must not be used individually for diagnostic purposes. *Note: HbA1cN reports HbA1c in % and HbA1cI reports HbA1c in mmol/mol The ADVIA® Chemistry A1c Calibrators are for in vitro diagnostic use in the calibration of the A1c and total hemoglobin methods (Automated and Manual Pretreatment) on the ADVIA Chemistry Systems. Prescription Use X (21 CFR Part 801 Subpart D) And/Or Over the Counter Use (21 CFR Part 801 Subpart C), #### (PLEASE DO NOT WRITE BELOW THIS LINE: CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD) Carol C. Benson Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) K081895
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...