ADVIA CENTAUR SYSTEM WITH STREAMLAB ANALYTICAL WORKCELL

K082638 · Siemens Healthcare Diagnostics, Inc. · KLI · Feb 26, 2009 · Clinical Chemistry

Device Facts

Record IDK082638
Device NameADVIA CENTAUR SYSTEM WITH STREAMLAB ANALYTICAL WORKCELL
ApplicantSiemens Healthcare Diagnostics, Inc.
Product CodeKLI · Clinical Chemistry
Decision DateFeb 26, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1700
Device ClassClass 2

Intended Use

The ADVIA Centaur with StreamLAB® Analytical Workcell is an automated immunoassay analyzer designed to perform in vitro diagnostic immunochemical assay analysis on clinical specimens. The system menu will include assays based on chemiluminescent technology, such as Thyroxine, along with other various chemiluminescent assays that may be adaptable to the analyzer depending on the reagent used to induce the chemiluminescent reaction. The ADVIA Centaur T4 assay is for in vitro diagnostic use in the quantitative determination of thyroxine (T4) in serum on the ADVIA Centaur and ADVIA Centaur XP systems. Measurements obtained by this device are used in the diagnosis and treatment of thyroid diseases.

Device Story

The ADVIA Centaur with StreamLAB Analytical Workcell is an integrated laboratory automation system (LAS) and immunochemistry analyzer. It automates pre- and post-analytical sample handling, including bar code identification, transport, tracking, centrifugation, de-capping, and tube sealing. The system receives test requests from a Laboratory Information System (LIS). Samples are either loaded directly onto the Centaur analyzer or onto the StreamLAB, which routes them to the Centaur. The Centaur performs chemiluminescent immunochemical assays using magnetic-particle solid phase and chemiluminescent labels, with detection via a photomultiplier tube in photon counting mode. The system is intended for use in clinical laboratories by trained personnel. Output consists of quantitative assay results reported to the LIS, which healthcare providers use to diagnose and treat thyroid and other conditions.

Clinical Evidence

Bench testing only. Split-sample method comparison studies were conducted using the ADVIA Centaur T4 Assay to compare results processed directly on the predicate device versus the proposed integrated system. Data were analyzed by linear regression.

Technological Characteristics

Chemiluminescent immunoassay using magnetic-particle solid phase and chemiluminescent label. Photomultiplier tube optical system. System comprises an automated analyzer (ADVIA Centaur) and a laboratory automation track (StreamLAB). Connectivity via LAS interface for sample tracking and LIS interface for test requests/results. Software-controlled sample routing and identification. No specific materials or ASTM standards cited.

Indications for Use

Indicated for the quantitative determination of thyroxine (T4) in serum for the diagnosis and treatment of thyroid diseases in patients requiring clinical laboratory testing.

Regulatory Classification

Identification

A total thyroxine test system is a device intended to measure total (free and protein bound) thyroxine (thyroid hormone) in serum and plasma. Measurements obtained by this device are used in the diagnosis and treatment of thyroid diseases.

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE A. 510(k) Number: k082638 B. Purpose for Submission: This is a new 510(k) to support the integration of a StreamLAB Laboratory Automation System (LAS) (k043546) to the ADVIA Centaur Analyzer (k041133) using the ADVIA Centaur T4 Assay (k905532) to show acceptable performance. C. Measurand: Thyroxine D. Type of Test: Quantitative Chemiluminescent Immunoassay E. Applicant: Siemens Healthcare Diagnostics Inc. F. Proprietary and Established Names: Advia Centaur System with StreamLAB Analytical Workcell G. Regulatory Information: 1. Regulation section: 21CFR Sec.862.1700 Total thyroxine test system. 21CFR Sec.- 862.2160-Discrete photometric chemistry analyzer for clinical use. 2. Classification: Class II and I, respectively 3. Product code: KLI - Enzyme Immunoassay, Non-Radiolabeled, Total Thyroxine JJE - Analyzer, Chemistry (Photometric, Discrete), For Clinical Use 4. Panel: Chemistry (75) H. Intended Use: 1. Intended use(s): See indication(s) for use below. 2. Indication(s) for use: The ADVIA Centaur with StreamLAB Analytical Workcell is an automated immunoassay analyzer designed to perform in vitro diagnostic immunochemical assay analysis on clinical specimens. The system menu will include assays based on chemiluminescent technology, such as Thyroxine, along with other various {1} chemiluminescent assays that may be adaptable to the analyzer depending on the reagent used to induce the chemiluminescent reaction. The ADVIA Centaur T4 assay is for in vitro diagnostic use in the quantitative determination of thyroxine (T4) in serum on the ADVIA Centaur and ADVIA Centaur XP systems. Measurements obtained by this device are used in the diagnosis and treatment of thyroid diseases. 3. Special conditions for use statement(s): Prescription use 4. Special instrument requirements: ADVIA Centaur with StreamLAB® Analytical Workcell I. Device Description: The ADVIA Centaur is a continuous operation, immunochemistry analyzer designed to perform in vitro diagnostic testing on clinical specimens. The StreamLAB® Analytical Workcell is a laboratory automation system (LAS) designed to automate sample handling and processing in the clinical laboratory. The ADVIA Centaur with StreamLAB® Analytical Workcell combines the features of both the analyzer and the laboratory automation system. The StreamLAB routes samples to the Centaur analyzer based on test request information it (StreamLAB) receives from the Laboratory Information System (LIS) and the test map established for the Centaur analyzer. StreamLAB and Centaur communicate sample and analyzer status via Centaur's Laboratory Automation System (LAS) interface. Via its LIS interface, the Centaur analyzer interfaces separately with the hospital's LIS to receive its test instructions (test requests) and to report results for each sample. Centaur's test instructions and test results for each sample are not processed through the StreamLAB. The StreamLAB performs the following pre and post-analytical functions. - Sample bar code identification (previously performed by the Centaur) - Sample transport and tracking - Sample centrifugation (optional functionality) - Sample de-capping (optional functionality) - Sample transport and tracking - Tube sealing (optional functionality) The Centaur continues to perform the following functions, when connected to the StreamLAB. All functions except reading the sample tube bar code. When Centaur is connected to 2 of 7 {2} StreamLAB, samples can be loaded directly onto Centaur and/or loaded onto StreamLAB and routed to Centaur. For samples loaded onto StreamLAB, StreamLAB reads the sample tube bar code (sample identification) and passes it electronically to Centaur via the LAS interface to Centaur. # J. Substantial Equivalence Information: 1. Predicate device name(s): ADVIA Centaur T4 Assay ADVIA Centaur Analyzer StreamLAB Laboratory Automation System (LAS) 2. Predicate 510(k) number(s): k905532, k041133, and k043546 respectively 3. Comparison with predicate: | Feature | Predicate device: ADVIA Centaur (K041133) | Proposed device: ADVIA Centaur with StreamLAB Analytical Workcell | | --- | --- | --- | | Principles of Operation | Chemiluminescence using magnetic-particle solid phase and chemiluminescent label | | | Optical System | Photomultiplier tube used in photon counting mode | | | Sample Containers | Sample cups or primary tubes | Sample cups* or primary tubes | | Sample Loading | Load directly onto the Centaur | Load directly onto the Centaur and/or load onto the StreamLAB | | Serum & Plasma Sample Preparation | Manually centrifuged samples Manually decapped sample tubes | Manually centrifuged samples or automatically centrifuged samples by the StreamLAB. Manually decapped sample tubes or automatically decapped tubes by StreamLAB. | | Sample Identification of Bar-coded Tubes | Tube bar code (identification) is read by the Centaur. | Tube bar code (identification) is read by the Centaur (when tubes are placed directly on the Centaur); or tube bar code read by StreamLAB and communicated electronically to the Centaur (when tubes are loaded onto StreamLAB). | | Test Orders | Unidirectional communication with external LIS | | | Test Results | Unidirectional communication with external LIS | | | Laboratory Automation | Centaur's software communicates with Lab Automation System via LAS interface. Centaur performs direct sampling from tube on the track. | | | * Sample cups cannot be used on the StreamLAB. Sample cups must be loaded directly onto the Centaur when connected to the StreamLAB. The StreamLAB does not impact Centaur's capability to accept sample cups. | | | {3} 4 of 7 K. Standard/Guidance Document Referenced (if applicable): None were referenced L. Test Principle: Chemiluminescence using magnetic-particle solid phase and chemiluminescent label M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: The precision of ADVIA Centaur T4 Assay was evaluated by analyzing three levels of a quality control product in triplicate once a day for five (5) days on the Advia Centaur System with StreamLAB Analytical Workcell. The data were analyzed using Analysis of Variance (ANOVA). The results are similar to the precision of the predicate device. | | | | S.D. (%CV) | | | --- | --- | --- | --- | --- | | LEVEL | n | Mean | Repeatability | Within Laboratory | | 1 | 15 | 5.45 μg/dL | 0.13 (2.4) | 0.25 (4.7) | | 2 | 15 | 8.26 μg/dL | 0.18 (2.1) | 0.28 (3.4) | | 3 | 15 | 14.49 μg/dL | 0.39 (2.7) | 0.44 (3.0) | b. Linearity/assay reportable range: Provided in k905532 c. Traceability, Stability, Expected values (controls, calibrators, or methods): Provided in k905532 d. Detection limit: Provided in k905532 e. Analytical specificity: Provided in k905532 f. Assay cut-off: Provided in k905532 2. Comparison studies: a. Method comparison with predicate device: Split-sample method comparison studies were conducted using the ADVIA Centaur T4 Assay. Samples were processed on the predicate device and on the proposed device. The data were analyzed by linear regression and the results are summarized in the table below. {4} | Slope | Intercept | r | Sy,x | n | Range | | --- | --- | --- | --- | --- | --- | | 1.050 | -0.299 | 0.993 | 0.36 | 66 | 1.60 – 17.00 μg/dL | ![img-0.jpeg](img-0.jpeg) T4 METHOD COMPARISON ADVIA Centaur with StreamLAB Analytical Workcell (SLAB) VS ADVIA Centaur (CNTR) b. Matrix comparison: Not applicable 3. Clinical studies: a. Clinical Sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): 4. Clinical cut-off: Not applicable 5. Expected values/Reference range: Provided in k905532 N. Instrument Name: Advia Centaur System with StreamLAB Analytical Workcell {5} O. System Descriptions: 1. Modes of Operation: The ADVIA Centaur is a continuous operation, immunochemistry analyzer designed to perform in vitro diagnostic testing on clinical specimens. Pre-analytical and Post-analytical activity is performed by StreamLAB. Centaur’s software communicates with Lab Automation System via LAS interface. Centaur performs direct sampling from tube on the track. The ADVIA Centaur with StreamLAB Analytical Workcell combines the features of both the analyzer and the laboratory automation system. 2. Software: FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types: Yes ☐ X or No ☐ The applicant provided software documentation that supports the device was designed and developed under good software LifeCycle processes. 3. Specimen Identification: Sample tube bar code (identification) is read by the Centaur (when tubes are placed directly on the Centaur); or sample tube bar code read by StreamLAB and communicated electronically to the Centaur (when tubes are loaded onto StreamLAB). 4. Specimen Sampling and Handling: Load directly onto the Centaur and/or load onto the StreamLAB. Manually centrifuged samples or automatically centrifuged samples by the StreamLAB. Manually decapped sample tubes or automatically decapped tubes by StreamLAB. Sample cups cannot be used on the StreamLAB. Sample cups must be loaded directly onto the Centaur when connected to the StreamLAB. The StreamLAB does not impact Centaur’s capability to accept sample cups. 5. Calibration: Provided in k041133 6 of 7 {6} 6. Quality Control: Provided in k041133 P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above: Q. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. R. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 7 of 7
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...