CHIRON DIAGNOSTICS AUTOMATED CHEMILUMINESCENCE SYSTEM (ASC) CENTAUR ANALYZER
Applicant
Chiron Diagnostics Corp.
Product Code
MOI · Immunology
Decision Date
Jul 17, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.6010
Device Class
Class 2
Indications for Use
The intended use of the ACS:Centaur system is an automated immunoassay analyzer designed to perform in vitro diagnostic immunochemical assays analysis on clinical specimens using the ACS ilmmunoassays. The system menu will include endocrine, anemia, reproductive, cardiac, therapeutic drug, and infectious disease assays. All assays are based on chemiluminescent technology.
Device Story
The ACS:Centaur is a stand-alone, automated immunoassay analyzer for clinical laboratory use. It processes clinical specimens to perform various immunochemical assays. The system automates sample aspiration, dispensing, dilution, reagent addition, incubation, and solid/liquid waste separation. It utilizes chemiluminescent technology, measuring photon emissions via a photomultiplier in photon counting mode to quantify antigens in samples. It supports both competitive binding and sandwich assays. The system features electronic fluid sensing, clog detection, and random access/batch processing. Data reduction, patient demographics, and results are managed via an integrated human interface with barcode reading and LIS connectivity. By automating complex immunoassay workflows, the device provides clinicians with rapid, standardized diagnostic results for endocrine, cardiac, and infectious disease management.
Clinical Evidence
Bench testing only. The device relies on established chemiluminescent immunoassay technology consistent with the predicate device. No clinical trial data was required for this 510(k) submission.
Technological Characteristics
Automated immunoassay analyzer; chemiluminescent detection using acridinium ester label and paramagnetic particles; photomultiplier tube for photon counting; electronic fluid sensing; automated pipetting and dilution; random access/batch processing; LIS connectivity; barcode reader; software-based data reduction.
Indications for Use
Indicated for in vitro diagnostic immunochemical analysis of clinical specimens for endocrine, anemia, reproductive, cardiac, therapeutic drug, and infectious disease conditions. For prescription use only.
Regulatory Classification
Identification
A tumor-associated antigen immunological test system is a device that consists of reagents used to qualitatively or quantitatively measure, by immunochemical techniques, tumor-associated antigens in serum, plasma, urine, or other body fluids. This device is intended as an aid in monitoring patients for disease progress or response to therapy or for the detection of recurrent or residual disease.
Special Controls
*Classification.* Class II (special controls). Tumor markers must comply with the following special controls: (1) A guidance document entitled “Guidance Document for the Submission of Tumor Associated Antigen Premarket Notifications (510(k)s) to FDA,” and (2) voluntary assay performance standards issued by the National Committee on Clinical Laboratory Standards.
{1}------------------------------------------------
K4 111418
7 1097
# ACS Centaur
# Summary of Safety and Effectiveness
As required by 21 CFR 807.92, the following 510(k) Summary is provided:
## 1. Submitters Information
| Contact person: | William J. Pignato<br>Director of Regulatory Affairs |
|-----------------|-------------------------------------------------------------------------|
| Address: | Chiron Diagnostics Corporation<br>63 North Street<br>Medfield, MA 02052 |
| Phone: | 508 359-3825 |
April 14, 1997 Date Summary Prepared:
#### 2. Device Information
| Proprietary Name: | ACS:Centaur Analyzer |
|------------------------|---------------------------------------|
| Common Name: | Automated Immunoassay Analyzer |
| Classification Name: | Photometric Analyzer for Clinical Use |
| Classification Number: | 21 CFR 862.2160, Class I |
## 3. Predicate Device Information
| Name: | ACS:180® Analyzer |
|----------------|--------------------------------------------------------------------------|
| Manufacturer: | Chiron Diagnostics Corporation (formally Ciba Corning Diagnostics Corp.) |
| 510(k) Number: | D.C. #K902336. |
## 4. Device Description
The ACS:Centaur system is a stand-alone, continuous operation, immunochemistry analyzer. The system performs the following functions.
- aspirates and dispenses samples .
- . performs dilutions
- adds reagents .
- incubates reaction vessels .
- separates solid and liquid waste .
- measures photon emissions .
- performs data reduction .
- collects and maintains patient demographics and results .
{2}------------------------------------------------
#### 5. Statement of Intended Use
The intended use of the ACS:Centaur system is an automated immunoassay analyzer designed to perform in vitro diagnostic immunochemical assays analysis on clinical specimens using the ACS ilmmunoassays. The system menu will include endocrine, anemia, reproductive, cardiac, therapeutic drug, and infectious disease assays. All assays are based on chemiluminescent technology.
## 6. Summary of Technological Characteristics
ACS assays are dedicated for use on the ACS:180® and the ACS:Centaur utilizes acridinium ester as label and paramagnetic particles as the solid phase. Like the ACS:180®, the ACS:Centaur measures the amount of light emitted during the chemiluminescent reaction. There is direct relationship between the amount of light emitted and amount of antigen in the patient sample. The system will measure both competitive binding assavs and sandwich assays.
The ACS:Centaur system uses a Master Curve and a two-point, user-initiated calibration to calibrate all the ACS assays. TThe Master Curve and the two-point calibration system eliminate the need to measure a full standard curve with each assay or to run calibrators each time the assay is run. The system stores the calibration for the interval specified in the assay product inserts.
Other Technological Features include:
- Photomultiplier used in the photon counting mode .
- Automated pipetting of sample and reagents ●
- Electronic Fluid Sensing on sample and reagents ●
- Clog detection for samples ●
- Random access and batch test processing ●
- . Varied test incubations
- . Human Interface features including barcode reading, LIS, automated data reduction.
- Automated dilution capabilities ●
{3}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the top half of the circle. The bottom half of the circle contains a stylized image of a human figure, with three overlapping profiles. The profiles are facing to the right and are connected by flowing lines.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
JUL 17 1997
Thomas F. Flynn
Manager, Requlatory Affairs & Compliance Chiron Diagnostics Corporation 63 North Street Medfield, Massachusetts 02052-1688
Re : K971418 ACS Centaur Analyzer Regulatory Class: I Product Code: JJE June 26, 1997 Dated: Received: June 27, 1997
Dear Mr. Flynn:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual reqistration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Requlations.
{4}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Steven Gutman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Page __ of ___________________________________________________________________________________________________________________________________________________________________
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: Chiron Diagnostics ACS:Centaur System ___________________________________________________________________________________________________________________________
Indications for Use:
The intended use of the ACS:Centaur system is an automated immunoassay analyzer The Intended use of the AOO.OChiaan of chamical assays analysis on clinical specimens using the ACS ilmmunoassays. The system menu will include endorine, specificals using the AOO minimanedation drug, and infectious disease assays. All anemia, reproductive, "Cardiac, therefound and is an indirect indicator of thyroid status.4
| (Division Sign-Off) |
|-----------------------------------------|
| Division of Clinical Laboratory Leviges |
| 510(k) Number |
(PLEASE DO NOT WRITE BELOW THIS LINE--CONTINUE ON ANOTHER PAGE, IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
**Prescription Use**
(Per 21 CFR 801.109)
OR
Over-The-Counter Use _ (Optional Format 1-2-96)
Page 7
\$\mathscr{A}\$
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.