HITACHI CLINICAL ANALYZER S TEST REAGENT CARTRIDGE ALT, HITACHI CLINICAL ANALYZER S TEST REAGENT CARTRIDGE AST
K120945 · Hitachi Chemical Diagnostics, Inc. · CKA · Jun 7, 2013 · Clinical Chemistry
Device Facts
Record ID
K120945
Device Name
HITACHI CLINICAL ANALYZER S TEST REAGENT CARTRIDGE ALT, HITACHI CLINICAL ANALYZER S TEST REAGENT CARTRIDGE AST
Applicant
Hitachi Chemical Diagnostics, Inc.
Product Code
CKA · Clinical Chemistry
Decision Date
Jun 7, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1030
Device Class
Class 1
Intended Use
The S TEST Reagent Cartridge Alanine Amino Transferase (ALT) is intended for the quantitative measurement of the activity of the enzyme alanine amino transferase (ALT) in serum, lithium heparin plasma, K3 EDTA plasma, and sodium citrate plasma on the Hitachi Clinical Analyzer E40. The test system is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only. ALT measurements are used in the diagnosis and treatment of certain liver diseases (e.g., viral hepatitis and cirrhosis) and heart diseases. The S TEST Reagent Cartridge Aspartate Amino Transferase (AST) is intended for the quantitative measurement of the activity of the enzyme aspartate amino transferase (AST) in serum, lithium heparin plasma, K3 EDTA plasma, and sodium citrate plasma on the Hitachi Clinical Analyzer E40. The test system is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only. AST measurements are used in the diagnosis and treatment of certain liver diseases (e.g., viral hepatitis and cirrhosis) and heart diseases.
Device Story
The S TEST system consists of plastic reagent cartridges containing two reservoirs (R1, R2) and a reaction cell/photometric cuvette. Cartridges feature a dot code label containing chemistry parameters and calibration factors. The system is used with the Hitachi Clinical Analyzer E40 in clinical laboratories or physician office laboratories. The device takes serum or plasma samples as input. It performs quantitative photometric analysis based on NADH oxidation; ALT/AST activity is determined by measuring the rate of decrease in NADH absorbance at 340 nm. The analyzer processes the reaction and provides quantitative enzyme activity results. Healthcare providers use these results to aid in the diagnosis and treatment of liver and heart diseases. The system provides a standardized, automated method for enzyme quantification in point-of-care or clinical settings.
Clinical Evidence
Clinical performance evaluated at three external physician office laboratory (POL) sites. Study included precision testing (n=30 replicates per sample per site) and method comparison against traditional central laboratory methods. Method comparison showed high correlation (r=0.997-0.999) across dynamic ranges (ALT 8-244 U/L; AST 5-383 U/L). Precision studies demonstrated total %CVs ranging from 1.1% to 5.9%.
Technological Characteristics
Plastic reagent cartridges (13.5mm x 28mm x 20.2mm) with two reservoirs and integrated reaction cell. Sensing principle: quantitative photometric analysis (NADH oxidation). Energy source: Hitachi Clinical Analyzer E40. Connectivity: 2D dot code label for automated parameter/calibration input. Software: Embedded firmware for analyzer control and data processing. Sterilization: Not specified.
Indications for Use
Indicated for quantitative measurement of ALT and AST enzyme activity in human serum, lithium heparin plasma, K3 EDTA plasma, and sodium citrate plasma. Used in diagnosis and treatment of liver diseases (e.g., viral hepatitis, cirrhosis) and heart diseases. Intended for use in clinical laboratories and physician office laboratories (POLs). Prescription use only.
Regulatory Classification
Identification
An alanine amino transferase (ALT/SGPT) test system is a device intended to measure the activity of the enzyme alanine amino transferase (ALT) (also known as a serum glutamic pyruvic transaminase or SGPT) in serum and plasma. Alanine amino transferase measurements are used in the diagnosis and treatment of certain liver diseases (e.g., viral hepatitis and cirrhosis) and heart diseases.
Predicate Devices
Sekisui ALT-SL Assay for ALT (k974003)
Roche/Hitachi cobas 8000 (c502 module) for AST (k100853)
Related Devices
K132462 — S TEST REAGENT CARTRIDGE BUN AND S TEST REAGENT CARTRIDGE CRE · Hitachi Chemical Diagnostics, Inc. · Oct 28, 2013
K072140 — S40 CLINICAL ANALYZER, S TEST BIL, S TEST BUN, S TEST GLU · Alfa Wassermann Diagnostic Technologies, Inc. · Jun 24, 2008
K130685 — HITACHI CLINICAL ANALYZER S TEST REAGENT CARTRIDGES FOR LD AND AMY · Hitachi Chemical Diagnostics, Inc. · Aug 9, 2013
K131051 — HITACHI S TEST REAGENT CARTRIDGE TOTAL PROTEIN (TP) AND ALBUMIN (ALB) · Hitachi Chemical Diagnostics, Inc. · Jun 7, 2013
Submission Summary (Full Text)
{0}------------------------------------------------
# SECTION 8 510(k) SUMMARY
# JUN 0 7 2013
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. The assigned 510(k) number is K120945.
| 807.92 (a)(1): Name: | Hitachi Chemical Diagnostics | |
|----------------------|------------------------------|--|
| Address: | 630 Clyde Court | |
| | Mountain View, CA 94043 | |
| Phone: | (୧૨૦) ଜ୧୮୮ ୧૨૦) | |
| FAX: | (650) 969 2745 | |
| Contact: | Mr. Charles Tsou | |
Regulatory Correspondent: Ms. Erika Ammirati
#### 807.92 (a)(2): Device name- trade name and common name, and classification
#### Trade name:
S TEST Reagent Cartridge Alanine Amino Transferase (ALT) S TEST Reagent Cartridge Aspartate Amino Transferase (AST)
Common Name: Routine chemistry analyzer for ALT and AST
- Classification: ALT-21 CFR \$862.1030, Class I with exemption by 21 CFR §862.9, product code CKA
AST- 21 CRF \$862.1100, Class II, product code CIT
#### 807.92 (a)(3): Identification of the legally marketed predicate devices
ALT: K974003- ALT SL Assay (Sekisui Diagnostics, Ltd, PEI, Canada) AST: K100853- ASTL assay (cobas, Roche Diagnostics, Indianapolis, IN)
#### 807.92 (a)(4): Device Description
The Hitachi Clinical Analyzer is an automatic, bench-top, wet chemistry system intended for use in clinical laboratories or physician office laboratories. The instrument consists of a desktop analyzer unit, an operations screen that prompts the user for operation input and displays data, a printer, and a unit cover. The analyzer unit includes a single probe, an incubation rotor, carousels for sample cups and reagent cartridges, and a multi-wavelength photometer. The single-use reagent cartridges may be placed in any configuration on the carousel, allowing the user to develop any test panel where the reagent cartridges are available.
. Page 1 of 8
{1}------------------------------------------------
The S TEST reagent cartridges are made of plastic and include two small reservoirs capable of holding two separate reagents (R1 and R2), separated by a reaction cell/photometric cuvette. The cartridges also include a dot code label that contains all chemistry parameters, calibration factors, and other production-related information, e.g., expiration dating. The dimensions of the reagent cartridges are: 13.5 mm (W) × 28 mm (D) × 20.2 mm (H).
System operation: After the sample cup is placed into the carousel, the analyzer pipettes the sample, pipettes the reagent, and mixes (stirs) the sample and reagent together. After the sample and reagent react in the incubator bath, the analyzer measures the absorbance of the sample, and based on the absorbance of the reactions, it calculates the concentration of analyte in the sample. The test system can measure analytes in serum or plasma and results are available in approximately 15 minutes per test. This submission is for reagent cartridge test systems for glucose.
#### ALT Chemistry Reactions:
Alanine aminotransferase (ALT) catalyzes the reaction from L-alanine and alfa-ketoglutaric acid to pyruvic acid and glutamic acid. When the produced pyruvic acid is converted into lactic acid by lactate dehydrogenase (LD), NADH is converted into NAD with a decrease in absorbance at 340 nm. The ALT activity can be determined by measuring the decreased rate of NADH.
Image /page/1/Figure/5 description: This image shows two chemical reaction equations and a timeline. The first reaction shows L-Alanine and Alfa-Ketoglutaric acid converting into Pyruvic acid and Glutamic acid with the enzyme ALT. The second reaction shows Pyruvic acid and NADH converting into Lactic acid and NAD with the enzyme LD. The timeline shows the addition of Reagent 1/Sample (200 µL / 15 μι) at 0 minutes, Reagent 2 (100 µL) at 7.5 minutes, and measurement at 8.5 minutes, with enzyme activity calculation at 12.5 minutes.
#### AST Chemistry Reactions:
Aspartate aminotransferase (AST) catalyzes a reaction from L-aspartic acid and aketogulutaric acid to oxaloacetic acid and glutamic acid. When the produced oxaloacetic acid is converted into malic acid by malate dehydrogenase (MD), NADH is converted into NAD with a decrease in absorbance at 340 nm. The AST activity can be determined by measuring the decreased rate of NADH.
Image /page/1/Figure/10 description: This image shows two chemical reactions. The first reaction shows L-Aspartic acid and Alfa-Ketoglutaric acid reacting with AST to produce Oxaloacetic acid and Glutamic acid. The second reaction shows Oxaloacetic acid and NADH reacting with MD to produce Malic acid and NAD. The image also contains the text 'Page 2 of 8'.
{2}------------------------------------------------
### AST (Continued)
Image /page/2/Figure/1 description: This image shows a timeline of an experiment. At time zero, Reagent 1 and a sample are combined in volumes of 200 μL and 15 μL, respectively. At time 7.5, 100 μL of Reagent 2 is added, and measurements are taken at 8.5. The enzyme activity is calculated at 12.5 minutes, and the measurement is 340/546 nm.
#### 807.92 (a)(5): Intended Use
S TEST Reagent Cartridge Alanine Amino Transferase ALT
#### Indications for Use:
The S TEST Reagent Cartridge Alanine Amino Transferase (ALT) is intended for the quantitative measurement of the activity of the enzyme alanine amino transferase (ALT) in serum, lithium heparin plasma, K3 EDTA plasma, and sodium citrate plasma on the Hitachi Clinical Analyzer E40. The test system is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only. ALT measurements are used in the diagnosis and treatment of certain liver diseases (e.g., viral hepatitis and circhosis) and heart diseases.
#### S TEST Reagent Cartridge Aspartate Amino Transferase AST
#### Indications for Use:
The S TEST Reagent Cartridge Aspartate Amino Transferase (AST) is intended for the quantitative measurement of the activity of the enzyme aspartate amino transferase (AST) in serum, lithium heparin plasma, K3 EDTA plasma, and sodium citrate plasma on the Hitachi Clinical Analyzer E40. The test system is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only. AST measurements are used in the diagnosis and treatment of certain liver diseases (e.g., viral hepatitis and cirrhosis) and heart diseases.
{3}------------------------------------------------
# 807.92 (a)(6): Technological Similarities and Differences to the Predicate
The following chart describes similarities and differences between the two test systems.
| Characteristic | Hitachi S TEST Systems | PREDICATE(S) |
|-------------------------------|-----------------------------------------------------------------------------------------------------|------------------------------|
| ALT Test System | K number- K120945 | Sekisui ALT-SL- K974003 |
| Device Class, Regulation Code | Class I (reserved), 21 CFR 862.1030 | Class I, 21 CFR 862.1030 |
| Classification Product Code | CKA | Same |
| Intended Use | Quantitative determination of ALT | Same |
| Testing Environment | Physician office or clinical lab | Clinical lab |
| Test Principle | NADH oxidation of pyruvate formed<br>by L-alanine and alpha-ketoglutarate<br>in the presence of ALT | Same |
| Specimen Type | Human serum or plasma | Human serum |
| Reportable Range | 6 to 400 U/L | 10 to 600 U/L |
| Detection Wavelength | 340/546 nm | 340/415 nm |
| Detection Limit | 2.2 U/L | 10 U/L |
| Linearity | 6 to 400 U/L | 10 to 600 U/L |
| Precision | %CVs range from 2.3% to 5.6% | %CVs range from 2.4% to 3.6% |
| AST Test System | K number- K120945 | Roche cobas K number- K100853 |
|-------------------------------|--------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|
| Device Class, Regulation Code | Class II, 21 CFR 862.1100 | Same |
| Classification Product Code | CIT | Same |
| Intended Use | Quantitative determination of AST | Same |
| Testing Environment | Physician office or clinical lab | Clinical lab |
| Test Principle | NADH oxidation of pyruvate formed<br>by L-aspartate and alpha-<br>ketoglutarate in the presence of AST | NADH oxidation of pyruvate formed<br>by L-aspartate and 2-oxyglutarate in<br>the presence of AST |
| Specimen Type | Human serum or plasma | Human serum or plasma |
| Reportable Range | 4 to 400 U/L | 5 to 700 U/L |
| Detection Wavelength | 340/546 nm | 700/340 nm |
| Detection Limit | 1.4 U/L | 5 U/L |
| Linearity | 4 to 400 U/L | 5 to 700 U/L |
| Precision | %CVs range from 1.4% to 3.2% | %CVs range from 0.4% to 3.1% |
### 807.92 (b)(1): Brief Description of Nonclinical Data
A series of studies were performed that evaluated the following nonclinical performance characteristics for ALT and AST: analytical sensitivity (limits of detection), linearity, 20-day in-house precision, interference testing, in-house method comparisons, and matrices comparison between serum and various plasma options.
{4}------------------------------------------------
### Analytical Sensitivity (Limits of Detection)
The studies followed CLSI EP17. The LoD for ALT was calculated to be 2.2 U/L and the LoD for AST was calculated to be 1.4 U/L
#### Linearitv
The studies followed CLSI EP-6A. The ALT S TEST is linear between 6 and 400 U/L and the AST S Test is linear between 4 and 400 U/L.
#### 20-day In-house Precision
The studies followed CLSI EP5-A2, where three levels of samples were each tested fourtimes a day for 20 days. The results were as follows:
#### Precision Summary:
| | | /TT/T<br>Mean (U/L | Within-Run<br>0/al<br>( | .<br>1%CV<br>Total . |
|------------|---------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| 1<br>1<br> | Level | 00 0<br>and and the been and the best of the state of the state of the state of the state of the state of the state of the state of the state of the state of the state of the state o | | |
| per level | evel | 70 | Acres of the cases of a career a<br>- - | l (<br>i |
| <br>. | t<br>l'evel . | 706 7<br>LOU. | <br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br> | to to<br>- |
| | | Mean (U/L | Within-Run %C | Total %CV |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|-----------|---------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| AST | Level | -------- | C<br>i | c<br>And Chand |
| n= 80 per level | Level 2 | 109.6 | | 1 |
| of the control of children and the different of the different of the different of the different of the different of the different of the different of the different of the dif | Level 3 | 304.2 | ﺮ | , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , |
### Interference Testing
The studies followed CLSI EP7-A2. The data demonstrated that the S TEST for ALT and the S Test for AST were not affected by high levels of the following substances at the levels noted:
### ALT
- Hemoglobin: no interference up to 250 mg/dL .
- Unconjugated bilirubin no interference up to 25 mg/dL .
- Triglyceride: no interference up to 500 mg/dL .
- Ascorbic acid: no interference up to 50 mg/dL .
### AST
- Hemoglobin: no interference up to 31 mg/dL (slight visual hemolysis) for samples at . approximately 40 U/L, and up to 125 mg/dL (moderate visual hemolysis) for samples at approximately 100 U/L
- Unconjugated bilirubin no interference up to 50 mg/dL .
- Triglyceride: no interference up to 500 mg/dL .
- Ascorbic acid: no interference up to 50 mg/dL .
### Method Comparisons
Method comparison studies evaluated at least serum samples; matched aliquots were assayed with both the Hitachi Clinical Analyzer with S TEST ALT and AST reagent cartridges and routine laboratory methods. The data were analyzed by linear regression (Hitachi = y-axis), and the results were as follows:
Page 5 of 8
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "HITACHI" in large, bold, black letters. Below the word "HITACHI" is the phrase "Inspire the Next" in a smaller, black font. The text is simple and clear, with a focus on the brand name and a tagline.
#### ALT (U/L) Regression Statistics (range = 12 to 392 U/L) :
| n | r | Slope (95% CI) | y-intercept (95% CI) |
|-----|-------|---------------------|----------------------|
| 103 | 0.999 | 1.09 (1.08 to 1.09) | 2.3 (1.4 to 3.3) |
#### AST (U/L) Regression Statistics (range = 5 to 369 U/L):
| n | r | Slope (95% CI) | y-intercept (95% CI) |
|-----|-------|---------------------|----------------------|
| 169 | 0.997 | 1.09 (1.08 to 1.10) | -3.7 (-4.8 to -2.7) |
#### Matrices Comparisons
A study was performed to validate the use of sodium citrate, lithium heparinized, and K3 EDTA plasma as alternatives to serum for the Hitachi Clinical Analyzer with S TEST ALT and AST reagent cartridges Approximately 35 matched serum/plasma samples that spanned the ALT and AST dynamic ranges were assayed in singleton and the results were compared using least squares liner regression (plasma = y-axis). The performance characteristics were as follows.
#### ALT Range (serum) = 6-392 U/L
| | Na Citrate Plasma n = 28 | Heparinized Plasma n = 31 | EDTA Plasma = 30 |
|-----------------------|--------------------------|---------------------------|---------------------|
| Slope (95% CIs) | 0.99x (0.97 to 1.01) | 1.02 (1.00 to 1.04) | 1.01 (0.98 to 1.04) |
| y-intercept (95% CIs) | 0.2 (-1.9 to 2.3) | 0.2 (-1.9 to 2.3) | 0.4 (-2.3 to 3.1) |
| r | 0.998 | 0.998 | 0.997 |
#### AST Range (serum) = 5 to 369 U/L
| | Na Citrate Plasma n = 38 | Heparinized Plasma n = 38 | EDTA Plasma n = 39 |
|-----------------------|--------------------------|---------------------------|---------------------|
| Slope (95% CIs) | 1.02 (1.00 to 1.04) | 1.04 (1.02 to 1.06) | 0.98 (0.94 to 1.02) |
| y-intercept (95% CIs) | -2.6 (-4.4 to -0.9) | 0.6 (-1.1 to 2.2) | 3.0 (-1.0 to 7.0) |
| r | 0.999 | 0.999 | 0.992 |
#### 807.92 (b)(2): Brief Description of Clinical Data
Studies for precision and method comparison (accuracy) were performed at three external POL-type sites to evaluate the Hitachi Clinical Analyzer with S TEST ALT and AST reagent cartridges in one of its targeted intended use environments, the physician's office laboratory.
For the external site precision study, each site received three blinded serum samples that were chosen to represent low, intermediate, and high concentrations of each analyte. Each sample was assaved six times per day for five days, reporting 30 results per level per analyte. Precision estimates for within-run precision and total precision were as follows (NOTE: precision samples at Site 3 were different than the precision samples at Sites 1 and 2, as materials had been consumed):
Precision Samples at Sites 1 and 2: A, B, and C Precision samples at Site 3: D, E, and F
{6}------------------------------------------------
16
| Site | Sample | Mean | Within-run Precision | | Total Precision | |
|--------|--------|-------|----------------------|------|-----------------|------|
| | | | SD (U/L) | %CV | SD (mg/dL) | %CV |
| Site 1 | A | 51.8 | 2.0 | 3.9% | 2.0 | 3.9% |
| Site 2 | A | 51.7 | 2.4 | 4.7% | 2.4 | 4.7% |
| Site 3 | D | 24.3 | 0.9 | 3.8% | 1.4 | 5.7% |
| Site 1 | B | 143.8 | 2.2 | 1.5% | 3.4 | 2.4% |
| Site 2 | B | 139.5 | 2.5 | 1.8% | 2.8 | 2.0% |
| Site 3 | E | 77.4 | 1.1 | 1.4% | 2.0' | 2.6% |
| Site 1 | C | 319.7 | 3.8 | 1.2% | 5.0 | 1.6% |
| Site 2 | C | 305.9 | 4.1 | 1.3% | 7.7 | 2.5% |
| Site 3 | F | 194.4 | 2.2 | 1.1% | 4.3 | 2.2% |
#### ALT (U/L) = 30 renlicates ner sample ner site
## AST (U/L)
| n = 30 replicates per sample per site | | | | | | |
|---------------------------------------|--------|-------|----------------------|------|-----------------|------|
| Site | Sample | Mean | Within-run Precision | | Total Precision | |
| | | | SD (U/L) | %CV | SD (mg/dL) | %CV |
| Site 1 | A | 76.3 | 1.9 | 2.5% | 2.0 | 2.7% |
| Site 2 | A | 75.4 | 1.7 | 2.2% | 1.6 | 2.1% |
| Site 3 | D | 41.0 | 0.9 | 2.2% | 0.9 | 2.1% |
| Site 1 | B | 156.8 | 2.1 | 1.3% | 3.8 | 2.4% |
| Site 2 | B | 154.4 | 2.5 | 1.6% | 4.1 | 2.7% |
| Site 3 | E | 106.1 | 0.9 | 0.8% | 2.0 | 1.9% |
| Site 1 | C | 356.0 | 5.8 | 1.6% | 6.9 | 1.9% |
| Site 2 | C | 348.9 | 7.3 | 2.1% | 20.6 | 5.9% |
| Site 3 | F | 256.9 | 2.6 | 1.0% | 4.9 | 1.9% |
For the external site method comparisons studies, each POL site received 50 (ALT) to 60 (AST) blinded serum samples that were chosen to represent as full a range of analyte concentrations as possible, and a central laboratory received a matched aliquot for each serum sample. Each sample was assayed by the Hitachi system at the POL sites, and by traditional methods at the central laboratory. The results were analyzed by least squares linear regression (Hitachi = y-axis), and the performance characteristics were as follows:
| Site # | n | Range | Regression<br>Equation | "r" | CI*<br>Slope | CI Intercept |
|--------|----|----------|------------------------|-------|--------------|--------------|
| 1 | 49 | 8 to 238 | y = 1.09x -0.8 | 0.998 | 1.07 to 1.11 | -2.3 to 0.8 |
| 2 | 50 | 9 to 227 | y = 1.05 x +0.1 | 0.997 | 1.03 to 1.07 | -1.7 to 1.9 |
| 3 | 50 | 8 to 244 | y = 1.11x +0.2 | 0.997 | 1.08 to 1.14 | -2.0 to 2.3 |
CV DATA SUMMARY- ALT (I/L)
*95% Confidence Interval
{7}------------------------------------------------
| Site # | | Range | Regression | 46 - 99 | Cl* | CI Intercept |
|--------|----|----------|----------------|---------|--------------|--------------|
| | | | Equation | | Slope | |
| | 64 | 5 to 359 | y = 1.00x -0.1 | 0.999 | 0.99 to 1.01 | -1.5 to 1.4 |
| | 62 | 6 to 383 | Y = 1.04x -0.3 | 0.998 | 1.02 to 1.05 | -2.1 to 1.5 |
| | 64 | 5 to 375 | y = 1.05x +0.7 | 0.999 | 1.04 to 1.06 | -1.1 to 2.5 |
| | | | | | | |
# POL ACCURACY DATA SUMMARY- AST (U/L)
*95% Confidence Interval
### 807.92 (b)(3): Conclusions from Nonclinical and Clinical Testing
Nonclinical and clinical testing was performed for the Hitachi Clinical Analyzer with the S TEST ALT and AST reagent cartridges. The test systems were shown to be safe and effective for their intended uses.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES". The text is in all caps and is in a bold, sans-serif font. The text is centered on the image and is the only element present. The image is simple and straightforward.
Image /page/8/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The text is in all caps and appears to be in a sans-serif font. The logo is black and white.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 7, 2013
Hitachi Chemical Diagnostics, Inc. C/O Charles Tsou 630 Clyde Court MOUNTAIN VIEW CA 94043
Re: K120945
Trade/Device Name: S TEST Reagent Cartridge Aspartate Amino Transferase (AST) S TEST Reagent Cartridge Alanine Amino Transferase (ALT) Regulation Number: 21 CFR 862.1100 Regulation Name: Aspartate amino transferase (AST/SGOT) test system Regulatory Class: II Product Code: CIT, CKA Dated: June 04, 2013 Received: June 05, 2013
Dear Mr. Tsou:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. - Please-note: - CDRH does.not evaluate.information.related.to.contract.liability. warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{9}------------------------------------------------
Page 2-Mr. Tsou
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm 1 15809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
# Carol C. Benson -S for
Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{10}------------------------------------------------
# Indications for Use
#### 510(k) Number (if Known): k120945
Device Name:
S TEST Reagent Cartridge Alanine Amino Transferase (ALT) S TEST Reagent Cartridge Aspartate Amino Transferase (AST)
Indications for Use:
The S TEST Reagent Cartridge Alanine Amino Transferase (ALT) is intended for the quantitative measurement of the activity of the enzyme alanine amino transferase (ALT) in serum, lithium heparin plasma, K3 EDTA plasma, and sodium citrate plasma on the Hitachi Clinical Analyzer E40. The test system is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only. ALT measurements are used in the diagnosis and treatment of certain liver diseases (e.g., viral hepatitis and cirrhosis) and heart diseases.
The S TEST Reagent Cartridge Aspartate Amino Transferase (AST) is intended for the quantitative measurement of the activity of the enzyme aspartate amino transferase (AST) in serum, lithium heparin plasma, K3 EDTA plasma, and sodium citrate plasma on the Hitachi Clinical Analyzer E40. The test system is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only. AST measurements are used in the diagnosis and treatment of certain liver diseases (e.g., viral hepatitis and cirrhosis) and heart diseases.
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use _(21_CFR 807_Subpart_C)_
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH. Office of In Vitro Diagnostics and Radiological Health (OIR)
AND/OR
YungW.Chan-S
Division Sign-Off Office of In Vitro Diagnostics and Radiological Health
510(k) k120945
Page 1 of 1
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.