HITACHI CLINICAL ANALYZER S TEST REAGENT CARTRIDGES FOR LD AND AMY
K130685 · Hitachi Chemical Diagnostics, Inc. · JFJ · Aug 9, 2013 · Clinical Chemistry
Device Facts
Record ID
K130685
Device Name
HITACHI CLINICAL ANALYZER S TEST REAGENT CARTRIDGES FOR LD AND AMY
Applicant
Hitachi Chemical Diagnostics, Inc.
Product Code
JFJ · Clinical Chemistry
Decision Date
Aug 9, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1070
Device Class
Class 2
Intended Use
The S TEST Reagent Cartridge Lactate Dehydrogenase (LD) is intended for the quantitative determination of LD in serum and plasma using the HITACHI Clinical Analyzer E40. The S TEST Reagent Cartridge Lactate Dehydrogenase (LD) is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only. Measurements of LD are used in the diagnosis and treatment of liver and cardiac diseases. The S TEST Reagent Cartridge Amylase (AMY) is intended for the quantitative determination of AMY in serum and plasma using the HITACHI Clinical Analyzer E40. The S TEST Reagent Cartridge Amylase (AMY) is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only. Measurements of AMY are mainly used in the diagnosis and treatment of pancreatic diseases.
Device Story
The S TEST Reagent Cartridges are plastic, single-use cartridges containing two reagent reservoirs and a reaction cell/photometric cuvette. Designed for use with the Hitachi Clinical Analyzer E40 in clinical or physician office laboratories. The cartridge includes a dot code label containing chemistry parameters and calibration factors. For LD testing, the device measures the conversion of lactic acid to pyruvic acid, reducing NAD to NADH, which increases absorbance. For AMY testing, the device measures the cleavage of Gal-G2-CNP substrate, where the production rate of CNP is proportional to amylase activity. The analyzer reads the barcode to apply lot-specific calibration factors to the change in absorbance, providing a quantitative result. This allows healthcare providers to assess liver, cardiac, or pancreatic function.
Clinical Evidence
Performance evaluated via nonclinical bench testing and clinical method comparison studies at three physician office laboratory (POL) sites. Nonclinical studies followed CLSI guidelines (EP17-A, EP6-A, EP5-A2, EP7-A2) for sensitivity, linearity, precision, and interference. Clinical accuracy was assessed by comparing the Hitachi E40 system against a reference laboratory method using 70-80 serum specimens per site. Regression analysis showed high correlation (r > 0.99) for both LD and AMY across all sites. Precision studies (n=30 replicates per sample) demonstrated total %CVs ranging from 2.7% to 9.3% for LD and 1.0% to 4.6% for AMY.
Technological Characteristics
Plastic reagent cartridges with two reservoirs and a reaction cell. Sensing principle: UV/visible spectrophotometry. Energy source: Hitachi Clinical Analyzer E40. Connectivity: Barcode-based parameter/calibration input. Reagents: LD (L-lactic acid, NAD, citric acid buffer); AMY (NaCl, CaCl2, Good's buffer, Gal-G2-CNP, KSCN).
Indications for Use
Indicated for quantitative determination of Lactate Dehydrogenase (LD) and Amylase (AMY) in human serum and plasma. Used in clinical or physician office laboratories for diagnosis and treatment of liver, cardiac, and pancreatic diseases.
Regulatory Classification
Identification
An amylase test system is a device intended to measure the activity of the enzyme amylase in serum and urine. Amylase measurements are used primarily for the diagnosis and treatment of pancreatitis (inflammation of the pancreas).
Predicate Devices
Roche cobas c systems (k100853)
Related Devices
K130141 — S TEST REAGENT CARTRIDGE ALP (ALKALINE PHOSPHATASE) · Hitachi Chemical Diagnostics, Inc. · Mar 22, 2013
K120945 — HITACHI CLINICAL ANALYZER S TEST REAGENT CARTRIDGE ALT, HITACHI CLINICAL ANALYZER S TEST REAGENT CARTRIDGE AST · Hitachi Chemical Diagnostics, Inc. · Jun 7, 2013
K131051 — HITACHI S TEST REAGENT CARTRIDGE TOTAL PROTEIN (TP) AND ALBUMIN (ALB) · Hitachi Chemical Diagnostics, Inc. · Jun 7, 2013
Submission Summary (Full Text)
{0}------------------------------------------------
130683
## SECTION 8 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. The assigned 510(k) number is K130685.
| 807.92 (a)(1): Name: | Hitachi Chemical Diagnostics |
|----------------------|----------------------------------------------------------------------------------------------------|
| Address: | 630 Clyde Court<br>Mountain View, CA 94043 |
| Contact: | Erika Ammirati, Ammirati Regulatory Consulting<br>Consultant to Hitachi Chemical Diagnostics, Inc. |
| Address: | 575 Shirlynn Court<br>Los Altos, CA 94022 |
| Phone: | (650) 949-2768 |
| FAX: | (650) 949-5347 |
#### 807.92 (a)(2): Device name- trade name and common name, and classification
#### Trade name:
S TEST Reagent Cartridge Lactate Dehydrogenase (LD) S TEST Reagent Cartridge Amylase (AMY)
Common Name: Routine chemistry analyzer for LD Routine chemistry analyzer for AMY
Classifications: 21 CFR § 862.1440 Lactate Dehydrogenase (LD) 21 CFR § 862.1070 Amylase (AMY)
## 807.92 (a)(3): Identification of the legally marketed predicate devices
Cobas c systems LDHI2 (Roche Diagnostics, Inc., Indianapolis, IN)- K100853 Cobas c systems AMYL2 (Roche Diagnostics, Inc., Indianapolis, IN)- K100853
## 807.92 (a)(4): Device Description
The Hitachi Clinical Analyzer is an automatic, bench-top, wet chemistry system intended for use in clinical laboratories or physician office laboratories. The instrument consists of a desktop analyzer unit, an operations screen that prompts the user for operation input and displays data, a printer, and a unit cover. The analyzer unit includes a single probe, an incubation rotor, carousels for sample cups and reagent cartridges, and a multi-wavelength photometer. The single-use reagent cartridges may be placed in any configuration on the carousel, allowing the user to develop any test panel where the reagent cartridges are available.
Page 1 of 9
O Hitachi Chemical Diagnostics, Inc. 630 Clyde Court. Mountain View. CA 84043-2239 Tel: 800 233 6278 Fax: 650 969 2745
| www.hcdiagnostics.com | |
|-----------------------|--|
|-----------------------|--|
{1}------------------------------------------------
The S TEST reagent cartridges are made of plastic and include two small reservoirs capable of holding two separate reagents (R1 and R2), separated by a reaction cell/photometric cuvette. The cartridges also include a dot code label that contains all chemistry parameters, calibration factors, and other production-related information, e.g., expiration dating. The dimensions of the reagent cartridges are: 13.5 mm (W) × 28 mm (D) × 20.2 mm (H).
System operation: After the sample cup is placed into the carousel, the analyzer pipettes the sample, pipettes the reagent, and mixes (stirs) the sample and reagent together. After the sample and reagent react in the incubator bath, the analyzer measures the absorbance of the sample, and based on the absorbance of the reactions, it calculates the concentration of analyte in the sample. The test system can measure analytes in serum or plasma and results are available in approximately 15 minutes per test. This submission is for Reagent Cartridge ALP.
Chemistry reactions: (LD) Lactate dehydrogenase in samples catalyzes the reaction of converting lactic acid to pyruvic acid. During this reaction, NAD is converted into NADH with an increase in absorbance at 340 nm. The LD activity can be determined by measuring the production rate of the resulting NADH.
(AMY) Amylase in blood samples reacts with the substrate alfa-2-chloro-4-nitropheny]galactopyranosylmaltoside (Gal-G2-CNP), and the substrate is cleaved into 4galactopyranosylmaltose (Gal-G2) and 2-chloro-4-nitrophenol (CNP). Amylase activity is determined by measuring the production rate of CNP (yellow).
## 807.92 (a)(5): Intended Use
The S TEST Reagent Cartridge Lactate Dehydrogenase (LD) is intended for the quantitative determination of LD in serum and plasma using the HITACHI Clinical Analyzer E40. The S TEST Reagent Cartridge Lactate Dehydrogenase (LD) is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.
Measurements of LD are used in the diagnosis and treatment of heart, liver, kidney, and blood diseases.
The S TEST Reagent Cartridge Amylase (AMY) is intended for the quantitative determination of AMY in serum and plasma using the HITACHI Clinical Analyzer E40. The S TEST Reagent Cartridge Amylase (AMY) is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.
Measurements of AMY are mainly used in the diagnosis and treatment of pancreatic diseases.
Page 2 of 9
Hitachi Chemical Diagnostics In Clyde Court. Mountain Viaw. CA 94043-2239 Tel. 800-233-6278
www.hcdlagnostics.com
{2}------------------------------------------------
## 807.92 (a)(6): Technological Similarities and Differences to the Predicate
The following chart describes similarities and differences between the LD test systems.
| Characteristic | Hitachi S TEST Systems | PREDICATE |
|-------------------------------|------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|
| Instrument Platform | Hitachi Clinical Analyzer<br>(originally cleared under K111753) | Roche cobas c systems - K100853 |
| Lactate Dehydrogenase (LD) | K number- K130685 | Roche K number- K100853 |
| Device Class, Regulation Code | Class II, Reserved, 21 CFR 862.1440 | Same, except Exempt (not POC) |
| Classification Product Code | CFJ | CFJ |
| Intended Use | Quantitative determination of LD | Same |
| Testing Environment | Physician office or clinical lab | Clinical lab |
| Test Principle | LD in the sample catalyzes the conversion of lactic acid to pyruvic acid. NAD is converted to NADH with an increase in absorbance. | UV assay- LD catalyzes the conversion of L-lactate to pyruvate (pyruvic acid); NAD is reduced to NADH in the process. |
| Specimen Type | Human serum or plasma | Same |
| Reportable Range | 10 to 1,000 U/L | Same |
| Detection Wavelength | 340/546 nm | 700/340 nm |
| Detection Limit | 10 U/L | Same |
| Linearity | 10 to 1,000 U/L | Same |
| Precision | %CVs range from 5.4% to 6.3% | %CVs range from 0.4% to 2.7%<br>(from product labeling) |
Page 3 of 9
## ത്ര Hitachi Chemical Diagnostics, Inc.
630 Clyde Court, Mountain View, CA 94043-2239 Tel: 800 233 6276 Fax: 650 969 2745
{3}------------------------------------------------
| Characteristic | Hitachi S TEST Systems | PREDICATE |
|-------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Instrument Platform | Hitachi Clinical Analyzer<br>(originally cleared under K111753) | Roche cobas c systems - K100853 |
| Amylase (AMY) | K number-K130685 | Roche K number- K100853 |
| Device Class, Regulation Code | Class II, 21 CFR 862.1070 | Same |
| Classification Product Code | JFJ | Same |
| Intended Use | Quantitative determination of AMY | Same |
| Testing Environment | Physician office or clinical lab | Clinical lab |
| Test Principle | Alpha amylases in blood samples react with the substrate alfa-2-chloro-4-nitrophenyl-galactopyranosylmaltoside (Gal-G2-CNP), and the substrate is cleaved into 4-galactopyranosylmaltose (Gal-G2) and 2-chloro-4-nitrophenol (CNP). Amylase activity is determined by measuring the production rate of CNP (yellow) | Defined oligosaccharides are cleaved under the catalytic action of alpha amylases. The fragments formed are completely hydrolyzed to p-nitrophenol (p-NP) and glucose by alpha-glucosidase. The color intensity of the p-NP formed is directly proportional to the amylase activity and is determined by measuring the increase in absorbance |
| Specimen Type | Human serum or plasma | Human serum, plasma, or urine |
| Reportable Range | 4 to 1,500 U/L | 3 to 1,500 U/L |
| Detection Wavelength | 405/546 nm | 700/415 nm |
| Detection Limit | 4 U/L | 3 U/L |
| Linearity | 3 to 1,700 U/L | 3 to 1,500 U/L |
| Precision | %CVs range from 2.7% to 3.7% | %CVs range from 0.7% to 2.4%<br>(from product labeling) |
The following chart describes similarities and differences between the AMY test systems.
## 807.92 (b)(1): Brief Description of Nonclinical Data
A series of studies were performed that evaluated the following nonclinical performance characteristics for each analytical sensitivity (limits of detection), linearity, 20-day in-house precision, interference testing, in-house method comparisons, and matrices comparison between serum and various plasma types.
## Analytical Sensitivity (Limits of Detection)- LD
The study followed CLSI EP17-A, and the limit of detection was found to be 7.9 U/L. The quantitation limit was found to be 10 U/L.
## Analytical Sensitivity (Limits of Detection)- AMY
The study followed CLSI EP17-A, and the limit of detection was found to be 2.2 U/L. The quantitation limit was found to be 4 U/L.
Page 4 of 9
itachi Chemical Diagnostic
Fax: 650 969 2745 View, CA 94043-2239 Tel: 800 233 6278
{4}------------------------------------------------
## Linearity- LD
The study followed CLSI EP-6A, and the range of linearity was 3 U/L to 1,196 U/L. The reportable range is 10 U/L to 1,000 U/L.
## Linearity- AMY
The study followed CLSI EP-6A, and the range of linearity was 3 U/L to 1,700 U/L. The reportable range is 4 U/L to 1,500 U/L.
## 20-day In-house Precision- LD
The studies followed CLSI EP5-A2, where three levels of samples were each tested in two runs, twice a day, for 20 days. The results were as follows:
## Precision Summary:
## LD- Low, Level 1, Summary
| LD | Within-Run | Total |
|------------|------------|-------|
| Mean (U/L) | 108.2 | 108.2 |
| SD (U/L) | 5.24 | 6.82 |
| %CV | 4.8% | 6.3% |
## LD- Middle, Level 2, Summary
| LD | Within-Run | Total |
|------------|------------|-------|
| Mean (U/L) | 159.3 | 159.3 |
| SD (U/L) | 9.15 | 8.85 |
| %CV | 5.7% | 5.6% |
## LD- High, Level 3, Summary
| LD | Within-Run | Total |
|------------|------------|-------|
| Mean (U/L) | 628.0 | 628.0 |
| SD (U/L) | 20.0 | 33.8 |
| %CV | 3.2% | 5.4% |
## 20-day In-house Precision- AMY
The studies followed CLSI EPS-A2, where three levels of samples were each tested in two runs, twice a day, for 20 days. The results were as follows:
## Precision Summary:
## AMY- Low, Level 1, Summary
| AMY | Within-Run | Total |
|------------|------------|-------|
| Mean (U/L) | 54.1 | 54.1 |
| SD (U/L) | 0.94 | 1.45 |
| %CV | 1.7% | 2.7% |
Page 5 of 9
#### chi Chemical Diagnostics, Inc
yde Court, Mountain View, CA 84043-2239 Tel: 800 233 6278 Fax: 650 969 2745
| man relations and consistencies and controllation and production promotives processional production promotive promotive processorial con un minimalize provincial pro 1971-194 | | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| | | |
| www.hcdlagnostics.com | | |
| | | |
| Репейского поставления и в сельно в присание пристемента в «Мении Как «Спостика» в «Макси Самиранские по собращенность по собращенности с мористика мести се подорение можени | | |
{5}------------------------------------------------
| AMY | Within-Run | Total |
|------------|------------|-------|
| Mean (U/L) | 188.5 | 188.5 |
| SD (U/L) | 1.50 | 6.99 |
| %CV | 0.8% | 3.7% |
## AMY- Middle, Level 2, Summary
#### AMY- High, Level 3, Summary
| AMY | Within-Run | Total |
|------------|------------|--------|
| Mean (U/L) | 1126.8 | 1126.8 |
| SD (U/L) | 8.85 | 39.5 |
| %CV | 0.8% | 3.5% |
## Interference Testing (per CLSI EP7-A2)
The data demonstrated that the LD test system was not affected by high levels of the following substances at the levels noted:
Unconjugated bilirubin: no interference up to 50 mg/dL
Lipemia: no interference up to 1,000 mg/dL
Ascorbic acid: no interference up to 50 mg/dL
Hemoglobin: Positive interference (increase in concentration) from hemolysis occurred at levels as low as 31mg/dL hemoglobin. Any level of hemolysis may cause interference. Do not use hemolyzed specimens.
Lack of interference was defined as recoveries between 90% and 110% of the neat value, and assay performance claims were established on the HITACHI Clinical Analyzer by testing two serum pools containing approximately 100 and 350 U/L LD.
The data demonstrated that the AMY test system was not affected by high levels of the following substances at the levels noted:
Hemoglobin: no interference up to 500 mg/dL Unconjugated bilirubin: no interference up to 50 mg/dL Lipemia: no interference up to 2,000 mg/dL Ascorbic acid: no interference up to 50 mg/dL
Lack of interference was defined as recoveries between 90% and 110% of the neat value, and assay performance claims were established on the HITACHI Clinical Analyzer by testing two serum pools containing approximately 150 and 300 U/L AMY.
## Method Comparison - LD
A total of 106 clinical specimens spanning the dynamic range (13 to 959 U/L), were assayed in singleton and in a blinded fashion by both the Hitachi E40 system and a standard
Page 6 of 9
◎ **Hitachi Chemical Diagnostics, Inc.**
630 Clyde Court, Mountain View, CA 94043-2239 Tel: 800 233 6278 Fax: 650 969 2745
.hcdiagnostics.com
{6}------------------------------------------------
laboratory system. The comparative data were analyzed by linear regression and are shown below. (CI = confidence interval)
## LD Regression Statistics:
| n | r | Slope<br>(95% CI) | y-intercept<br>(95% CI) | X mean | Y mean |
|-----|-------|------------------------|-------------------------|---------|---------|
| 106 | 0.991 | 1.01<br>(0.99 to 1.04) | 5.4<br>(-3.8 to 14.6) | 288 U/L | 297 U/L |
## Method Comparison - AMY
A total of 105 clinical specimens spanning the dynamic range (5 to 1,443 U/L), were assayed in singleton and in a blinded fashion by both the Hitachi E40 system and a standard laboratory system. The comparative data were analyzed by linear regression and are shown below. (CI = confidence interval)
## AMY Regression Statistics:
| n | r | Slope<br>(95% CI) | y-intercept<br>(95% CI) | X mean | Y mean |
|-----|-------|------------------------|-------------------------|---------|---------|
| 105 | 0.997 | 1.08<br>(1.06 to 1.10) | -3.3<br>(-8.7 to 2.1) | 218 U/L | 232 U/L |
## Matrices Comparisons- LD
A study was performed to validate the use of two plasma types as an alternative to serum for the Hitachi Clinical Analyzer with S TEST Reagent Cartridge LD. The plasma types were K3 EDTA and lithium heparin. Thirty-nine (39) matched serum/plasma samples that spanned the dynamic range (13 to 967) were assayed in singleton and the results were compared using linear regression (plasma = y-axis, each type). The performance characteristics were as follows.
## N = 39
Range (serum) = 32 to 804 U/L
| LD | Heparinized Plasma | K3 EDTA Plasma |
|-----------------------|---------------------|---------------------|
| Slope (95% CIs) | 0.99 (0.97 to 1.01) | 0.97 (0.94 to 1.00) |
| y-intercept (95% CIs) | -5.5(-10.7 to -0.3) | 0.1 (-8.9 to 9.0) |
| r | 0.998 | 0.994 |
## Matrices Comparisons- AMY
A study was performed to validate the use of two plasma types as an alternative to serum for the Hitachi Clinical Analyzer with S TEST Reagent Cartridge AMY. The plasma types were K3 EDTA and lithium heparin. Approximately 43matched serum/plasma samples that spanned the dynamic range (5 to 1,494) were assayed in singleton and the results were compared using linear regression (plasma = y-axis, each type). The performance characteristics were as follows.
N = 43 (serum) Range (serum) = 5 to 1,494 U/L
Page 7 of 9
#### chi Chemical Diagnostics Inc
t. Mountain View. CA 94043-2239 Tel: 800-233 6278 Fax: 650 969 2745
| 14 - 14 - 14 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - | | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| | | |
| www.hcdiaonostics.com | | |
| within the many and every and every of the count of the count of the first of the first of the first of the first of the first of the first of the first of the first of the f | | |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "HITACHI" in bold, black letters. Below the word "HITACHI" is the phrase "Inspire the Next" in a smaller font size. The text is all aligned to the left and the background is white.
| | AMY | Heparinized Plasma | K3 EDTA Plasma |
|-----------------------|-----|---------------------|---------------------|
| Slope (95% Cls) | | 1.02 (1.00 to 1.04) | 0.97 (0.95 to 0.99) |
| y-intercept (95% Cls) | | -8.4 (-17.2 to 0.3) | -6.6 (-14.3 to 1.0) |
| r | | 0.998 | 0.999 |
## 807.92 (b)(2): Brief Description of Clinical Data
Studies for precision and method comparison (accuracy) were performed at three external POL-type sites to evaluate the Hitachi E40 Clinical Analyzer with S TEST Reagent Cartridges for LD and AMY in one of its targeted intended use environments, the physician's office laboratory.
For the external site precision study, each site received three blinded serum samples (the Precision Panel, labeled A, B, and C) that were chosen to represent low, middle, and high concentrations of LD or AMY. Each sample was assayed six times per day for five days, reporting 30 results per level. Precision estimates for total precision were as follows:
| n = 30 replicates per sample per site | | | | | | |
|---------------------------------------|--------|-------|----------------------|-----|-----------------|-----|
| Site | Sample | Mean | Within-run Precision | | Total Precision | |
| | | | SD (U/L) | %CV | SD (U/L) | %CV |
| Site 1 | Low | 47.3 | 3.83 | 8.1 | 4.42 | 9.3 |
| Site 2 | Low | 49.8 | 3.00 | 6.0 | 3.20 | 6.4 |
| Site 3 | Low | 45.7 | 3.69 | 8.1 | 3.82 | 8.4 |
| Site 1 | Middle | 161.9 | 5.60 | 3.5 | 6.45 | 4.0 |
| Site 2 | Middle | 161.6 | 6.01 | 3.7 | 6.25 | 3.9 |
| Site 3 | Middle | 155.7 | 8.90 | 5.7 | 9.63 | 6.2 |
| Site 1 | High | 498.1 | 12.70 | 2.6 | 15.10 | 3.0 |
| Site 2 | High | 488.5 | 20.29 | 4.2 | 35.20 | 7.2 |
| Site 3 | High | 497.0 | 14.71 | 3.0 | 20.55 | 4.1 |
LD (U/L) n = 30 replicates per sample per site
AMY (U/L)
| n = 30 replicates per sample per site | |
|---------------------------------------|--|
|---------------------------------------|--|
| Site | Sample | Mean | Within-run Precision | | Total Precision | |
|--------|--------|-------|----------------------|-----|-----------------|-----|
| | | | SD (U/L) | %CV | SD (U/L) | %CV |
| Site 1 | Low | 53.2 | 2.43 | 4.6 | 2.44 | 4.6 |
| Site 2 | Low | 50.4 | 1.44 | 2.8 | 1.74 | 3.5 |
| Site 3 | Low | 51.2 | 2.08 | 4.1 | 1.93 | 3.8 |
| Site 1 | Middle | 116.9 | 1.58 | 1.4 | 1.65 | 1.4 |
| Site 2 | Middle | 111.5 | 1.73 | 1.6 | 2.11 | 1.9 |
| Site 3 | Middle | 113.3 | 2.29 | 2.0 | 3.51 | 3.1 |
Page 8 of 9
#### O Hitachi Chemical Diagnostics, Inc.
630 Clyde Court, Mountain View, CA 84043-2239 Tel: 800 233 6278 Fax: 650 969 2745
www.hcdiagnostics.com
{8}------------------------------------------------
| Site 1 | High | 1527.4 | 7.19 | 0.5 | 15.51 | 1.0 |
|--------|------|--------|-------|-----|-------|-----|
| Site 2 | High | 1428.6 | 16.06 | 1.1 | 29.24 | 2.0 |
| Site 3 | High | 1465.4 | 13.51 | 0.9 | 19.10 | 1.3 |
For the external method comparison studies, a series of approximately 70-80 serum specimens with LD values ranging from 16 to 938 U/L, and AMY values ranging from 27 to 1,146 U/L, were assayed on the Hitachi E40 Clinical Analyzer at three sites using S TEST Reagent Cartridges LD and AMY (y) and a comparative method as the reference method (x). Linear regression analyses (least squares) yielded the following results:
| Site # | n | Range<br>(U/L) | Regression<br>Equation | “r” | CI*<br>Slope | CI Intercept |
|--------|----|----------------|------------------------|-------|--------------|--------------|
| 1 | 87 | 16 to 938 | y = 0.96x + 2.5 | 0.997 | 0.94 to 0.97 | -2.3 to 7.4 |
| 2 | 78 | 23 to 877 | y = 0.96x + 4.7 | 0.998 | 0.94 to 0.97 | 0.6 to 8.9 |
| 3 | 86 | 17 to 914 | y = 0.91x +13.5 | 0.999 | 0.90 to 0.93 | 9.6 to 17.4 |
## POL ACCURACY DATA SUMMARY- LD (U/L)
*95% Confidence Interval
| Site # | n | Range (U/L) | Regression Equation | "r" | CI* Slope | CI* Intercept |
|--------|----|-------------|---------------------|-------|--------------|---------------|
| 1 | 76 | 29 to 1134 | $y = 1.05x - 1.2$ | 0.999 | 1.04 to 1.06 | -4.6 to 2.2 |
| 2 | 69 | 27 to 1146 | $y = 1.00x - 0.5$ | 0.995 | 0.98 to 1.03 | -6.6 to 5.7 |
| 3 | 71 | 29 to 1112 | $y = 0.98x + 3.3$ | 0.995 | 0.95 to 1.00 | -3.2 to 9.8 |
## POL ACCURACY DATA SUMMARY- AMY (UL)
*95% Confidence Interval
### 807.92 (b)(3): Conclusions from Nonclinical and Clinical Testing
Nonclinical and clinical testing was performed for the Hitachi E40 Clinical Analyzer with the S TEST Reagent Cartridge Lactate Dehydrogenase (LD) and the S TEST Reagent Cartridge Amylase (AMY). The test systems were shown to be safe and effective for their intended uses.
Page 9 of 9
C Hitachi Chemical Diagnostics, Inc. 630 Clyde Court, Mountain View, CA 94043-2239 Tel: 800 233 6278 Fax: 650 989 2745
www.hcdiagnostics.com
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for the Department of Health & Human Services (USA). The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three lines representing its wings. The eagle is facing to the right.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-(1609 Silver Spring, MD 20993-00012
August 9, 2013
August 9, 2013
Hitachi Chemical Diagnostics, Inc. C/O Erika Ammirati President, Ammirati Regulatory Consulting 575 Shirlynn Court LOS ALTOS CA 94022
Re: K130685
Trade/Device Name: S TEST Reagent Cartridge Amylase (AMY) S TEST Reagent Cartridge Lactate Dehydrogenase (LD)
Regulation Number: 21 CFR 862.1070 Regulation Name: Amylase test system Regulatory Class: II Product Codc: JFJ, CFJ Dated: July 29, 2013 Received: July 30, 2013
Dear Ms. Ammirati:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA is issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practive requirements as set forth in the quality systems (QS) regulation (2) CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{10}------------------------------------------------
Page 2-Ms. Ammirati
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
# Carol C. Benson -S for
Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{11}------------------------------------------------
## INDICATIONS FOR USE
510(k) Number (if Known): k130685
Device Name: S TEST Reagent Cartridge Lactate Dehydrogenase (LD) S TEST Reagent Cartridge Amylase (AMY)
Indications for Use:
The S TEST Reagent Cartridge Lactate Dehydrogenase (LD) is intended for the quantitative determination of LD in serum and plasma using the HITACHI Clinical Analyzer E40. The S TEST Reagent Cartridge Lactate Dehydrogenase (LD) is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only. Measurements of LD are used in the diagnosis and treatment of liver and cardiac diseases.
The S TEST Reagent Cartridge Amylase (AMY) is intended for the quantitative determination of AMY in serum and plasma using the HITACHI Clinical Analyzer E40. The S TEST Reagent Cartridge Amylase (AMY) is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only. Measurements of AMY are mainly used in the diagnosis and treatment of pancreatic diseases.
Prescription Use ________________ And/Or (21 CFR Part 801 Subpart D)
Over the Counter Use _ (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
YungWDchan-S
Division Sign-Off Office of In Vitro Diagnostics and Radiological Health (OIR)
510(k)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.