ELECSYS LH ASSAY

K964694 · Boehringer Mannheim Corp. · CEP · Jan 24, 1997 · Clinical Chemistry

Device Facts

Record IDK964694
Device NameELECSYS LH ASSAY
ApplicantBoehringer Mannheim Corp.
Product CodeCEP · Clinical Chemistry
Decision DateJan 24, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1485
Device ClassClass 1

Intended Use

Immunoassay for the in vitro quantitative determination of human luteinizing hormone in human serum and plasma.

Device Story

Elecsys® LH Assay is an in vitro diagnostic immunoassay for quantitative determination of human luteinizing hormone (LH) in serum and plasma. Device utilizes sandwich principle: 20 µL sample incubated with biotinylated monoclonal LH-specific antibody and ruthenium-labeled monoclonal LH-specific antibody; complex binds to streptavidin-coated microparticles. Reaction mixture aspirated into measuring cell; microparticles magnetically captured on electrode surface. Voltage application induces chemiluminescent emission measured by photomultiplier. Results calculated via instrument-specific calibration curve generated by 2-point calibration and master curve from reagent bar code. Used on Elecsys® 2010 instrument. Provides clinicians with quantitative LH levels to assist in endocrine assessment.

Clinical Evidence

Bench testing only. Method comparison study (N=166) vs. Enzymun-Test® LH showed high correlation (r=0.993). Precision studies (N=60 per level) demonstrated %CVs ranging from 0.81% to 5.17%. Lower detection limit 0.10 mIU/mL. Linearity 0.1–200 mIU/mL. Interference testing performed for bilirubin, hemoglobin, lipemia, biotin, and rheumatoid factor. Specificity testing showed <0.1% cross-reactivity with FSH, HCG, TSH, HGH, and HPL.

Technological Characteristics

Sandwich immunoassay; electrochemiluminescence detection; streptavidin-coated microparticles; biotinylated monoclonal antibodies; ruthenium complex label. Requires Elecsys® 2010 instrument. Calibration via 2-point method and reagent bar code. Quantitative measurement range 0.1–200 mIU/mL.

Indications for Use

Indicated for the in vitro quantitative determination of human luteinizing hormone (LH) in human serum and plasma.

Regulatory Classification

Identification

A luteinizing hormone test system is a device intended to measure luteinizing hormone in serum and urine. Luteinizing hormone measurements are used in the diagnosis and treatment of gonadal dysfunction.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} JAN 24 1997 K964694 BOEHRINGER MANNHEIM CORPORATION # 510(k) Summary ## Introduction According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence. ## 1. Submitter name, address, contact Boehringer Mannheim Corporation 2400 Bisso Lane P.O. Box 4117 Concord, CA 94524-4117 (510) 674 - 0690 extension 8240 FAX 510 687 - 1850 Contact Person: Betsy Soares-Maddox Date Prepared: November 19, 1996 ## 2. Device name Proprietary name: Elecsys® LH Assay Common name: Electrochemiluminescence assay for the determination of human luteinizing hormone (LH). Classification name: System, Test, Human Luteinizing Hormone ## 3. Predicate device We claim substantial equivalence to the Enzymum® LH Assay (K900799). ## 4. Device Description Sandwich principle. Total duration of assay: 18 minutes. - 1st incubation (9min.): 20 µL of sample, a biotinylated monoclonal LH-specific antibody (75 µL) and a monoclonal LH-specific antibody labeled with a ruthenium complex (75 µL)** react to form a sandwich complex. - 2nd incubation (9min.): after addition of streptavidin-coated microparticles (30 µL), the complex becomes bound to the solid phase via interaction of biotin and streptavidin. **Tris(2,2'-bipyridyl)ruthenium(II) complex (Ru(bpy)₂⁺)** Continued on next page page 26 {1} page 27 # 510(k) Summary, Continued ## 4. Device Description - The reaction mixture is aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. Unbound substances are then removed with ProCell. Application of a voltage to the electrode then induces chemiluminescent emission which is measured by a photomultiplier (0.4 second read frame). - Results are determined via a calibration curve which is instrument-specifically generated by 2-point calibration and a master curve provided via the reagent bar code. ## 5. Intended use Immunoassay for the in vitro quantitative determination of human luteinizing hormone in human serum and plasma. ## 6. Comparison to predicate device The Boehringer Mannheim Elecsys® LH Assay is substantially equivalent to other products in commercial distribution intended for similar use. Most notably it is substantially equivalent to the currently marketed Enzymun® LH Assay (K900779). The following table compares the Elecsys® LH Assay with the predicate device, Enzymun® LH Assay. Specific data on the performance of the test have been incorporated into the draft labeling in attachment 5. Labeling for the predicate device is provided in attachment 6. ## Similarities: - Intended Use: Immunoassay for the in vitro quantitative determination of human luteinizing hormone (LH) - Sample type: Serum and plasma - Antibody: Same pair of monoclonal mouse anti-LH antibodies - Solid phase binding principle: Streptavidin/Biotin Continued on next page {2} 510(k) Summary, Continued 6. Comparison to predicate device cont. Differences: | Feature | Elecsys® LH | Enzymun-Test® LH | | --- | --- | --- | | Assay Standardization | Enzmun® LH | WHO # 80/552 | | Detection method | Electrochemiluminescence | ELISA/1-step sandwich assay using streptavidin technology | | Instrument required | Elecsys® 2010 | ES 300 | | Calibration Stability | A calibration is recommended every 7 days if kits is not consumed; 4 weeks with same reagent lot if reagent is consumed within 7 days. | Full calibration required every 2 weeks. One-point calibration required every run. | Performance Characteristics: | Feature | Elecsys® LH | | | Enzymun-Test® LH | | | | --- | --- | --- | --- | --- | --- | --- | | Precision | Modified NCCLS (mIU/mL): | | | Modified NCCLS (mIU/mL): | | | | Level | Low | Mid | High | Low | Mid | High | | N | 60 | 60 | 60 | 120 | 120 | 120 | | Within-Run Mean | 0.54 | 9.38 | 50.72 | 3.6 | 13.5 | 59.5 | | %CV | 1.82 | 1.13 | 0.81 | 2.9 | 3.8 | 1.5 | | Total Mean | 0.54 | 9.38 | 50.72 | 3.6 | 13.5 | 59.5 | | %CV | 5.17 | 1.97 | 1.99 | 4.4 | 4.7 | 3.9 | | Lower Detection Limit | 0.10 mIU/mL | | | 0.50 mIU/mL | | | Continued on next page {3} 510(k) Summary, Continued 6. Comparison to predicate device, (cont.) Performance Characteristics: | Feature | Elecsys® LH | Enzymun-Test® LH | | --- | --- | --- | | Linearity | 0.1 - 200 mIU/mL (with a deviation from a linear line of ±10%) | 0.5 - 150 mIU/mL (with a deviation from a linear line of ±10%) | | Method Comparison | Vs Enzymun-Test® LH Least Squares y = 1.00x - 0.199 r = 0.993 SEE = 1.141 N = 166 Passing/Bablok y = 0.964x + 0.040 r = 0.993 SEE = 0.456 N = 166 | Vs Enzymun-Test® LH Least Squares y = 0.93x + 0.42 r = 0.953 SEE = 5.079 N = 62 | | Interfering substances | No interference at: | No interference at: | | Bilirubin | 25.0 mg/dL | 64.5 mg/dL | | Hemoglobin | 1 g/dL | 1 g/dL | | Lipemia | 1500 mg/dL | 1250 mg/dL | | Biotin | 30 ng/mL | 40 ng/mL | | Rheumatoid Factor | 1500 U/mL | no interference | | Specificity | Level tested % Cross-reactivity | Level tested % Cross-reactivity | | FSH | 300 mIU/mL < 0.1 | 200 mU/mL 0.00 | | HCG | 600 IU/mL < 0.1 | 200 U/mL 0.00 | | TSH | 300 μIU/mL < 0.1 | 100 μU/mL 0.00 | | HGH | 600 μIU/mL < 0.1 | 200 ng/mL 0.00 | | HPL | 13.80 pmol/mL < 0.1 | --- |
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...