SEALITE SCIENCES, INC. AQUALITE LH

K964925 · Sealite Sciences, Inc. · CEP · Jan 22, 1997 · Clinical Chemistry

Device Facts

Record IDK964925
Device NameSEALITE SCIENCES, INC. AQUALITE LH
ApplicantSealite Sciences, Inc.
Product CodeCEP · Clinical Chemistry
Decision DateJan 22, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1485
Device ClassClass 1

Intended Use

The AquaLite® LH Bioluminescent Immunoassay (BIA) Kit (or the AquaLite® LH assay) is intended to be used for the quantitative determination of human LH in sera and plasma. The AquaLite® LH assay is for in vitro diagnostic use.

Device Story

Bioluminescent immunoassay (BIA) kit for quantitative LH measurement in serum/plasma. Uses polystyrene tubes pre-coated with polyclonal anti-LH antibody. Sample and recombinant aequorin-conjugated monoclonal anti-LH antibody added to form sandwich immune complex. After incubation and washing, tubes placed in luminometer. Calcium trigger solution injected, causing aequorin to oxidize luciferin, producing 469 nm light flash. Light intensity (RLU) measured by luminometer; proportional to LH concentration. Calibration curve relates RLU to LH concentration (IU/L). Used in clinical laboratory settings by trained personnel. Results assist clinicians in diagnosing gonadal dysfunction.

Clinical Evidence

Bench testing only. Sensitivity: 0.01 IU/L. Specificity: low cross-reactivity with TSH (4.9%), FSH (0.68%), and hCG (1.2%). Precision: intra-assay CV 5.9-7.55%, inter-assay CV 8.17-8.9%. Method comparison (N=62) vs. commercial kit: slope 0.89, y-intercept 1.53, correlation coefficient 0.938. Linearity and spike/recovery studies performed. No significant interference from hemoglobin, bilirubin, albumin, or triglycerides. No high-dose hook effect up to 5,000 IU/L.

Technological Characteristics

Solid-phase bioluminescent immunoassay. Materials: polystyrene tubes, recombinant aequorin (photoprotein), polyclonal/monoclonal antibodies. Energy source: luminometer (external). Connectivity: standalone. Software: none (manual calculation of calibration curve). Sterilization: not specified.

Indications for Use

Indicated for the quantitative determination of human luteinizing hormone (LH) in serum and plasma samples to aid in the diagnosis of gonadal dysfunction.

Regulatory Classification

Identification

A luteinizing hormone test system is a device intended to measure luteinizing hormone in serum and urine. Luteinizing hormone measurements are used in the diagnosis and treatment of gonadal dysfunction.

Related Devices

Submission Summary (Full Text)

{0} JAN 22 1997 K964925 EXHIBIT D SUMMARY OF SAFETY AND EFFECTIVENESS 383851.1 {1} 383851.1 # I. GENERAL INFORMATION Trade or proprietary name - SeaLite Sciences, Inc. AquaLite® LH Common or usual name - Bioluminescent immunoassay (BIA) Classification name - FDA has classified LH test systems intended for the measurement of LH in the diagnosis of gonadal dysfunction. # II. INDICATIONS FOR USE The AquaLite® LH Bioluminescent Immunoassay (BIA) Kit (or the AquaLite® LH assay) is intended to be used for the quantitative determination of human LH in sera and plasma. The AquaLite® LH assay is for in vitro diagnostic use. # III. DEVICE DESCRIPTION The AquaLite® LH Bioluminescent Immunoassay Kit uses a polyclonal anti-LH antibody that is pre-coated onto polystyrene tubes (solid phase). Samples (serum or plasma and appropriate calibrators or controls, are pipetted (50 μL) into the pre-coated tubes. Anti-LH Conjugate (150μL) is then added to the tubes. The conjugate uses the photoprotein, AquaLite® (recombinant aequorin; Patent Nos. 5, 422, 266 and 5, 486, 455 which is covalently linked to an anti-LH monoclonal antibody. LH in the sample simultaneously combines with polyclonal antibody on the solid phase and conjugate antibody to form an immune complex or "sandwich" bound to the solid phase. Complex formation is complete after a 60-minute incubation period at room temperature (18°C to 25°C) on a standard orbital shaker. The tubes are then washed to remove unbound conjugate. The washed tubes are placed in a luminometer that is capable of reading a triggered, flash-type reaction in 12 × 75 mm tubes. An injected calcium trigger solution causes AquaLite® to oxide its self-contained luciferin molecule. This reaction produces a flash of light at 469 nm, which is measured by the luminometer. The intensity of the light is directly proportional to the concentration of the LH in the sample. To calculate results, the light intensity (in relative light units, RLU) of the LH calibrators is plotted against LH concentration (in International Units per liter, IU/L) to yield a calibration curve. This curve is used to relate the light intensity generated from the samples and controls to LH concentration in IU/L. Note that the numerical value for LH in mIU/mL is the same as for IU/L (International System). For example, 15.6 mIU/mL equals 15.6 IU/L. Note: Samples that generate signals greater than the signal from the highest calibrator are off-scale. These samples must be diluted with Calibrator A and re-assayed. Remember to multiply the results from diluted samples by the dilution factor used. D-1 {2} 383851.1 D-2 # IV. ALTERNATIVE PRACTICES AND PROCEDURES There are several assay technologies commonly employed to measure the presence of human LH in serum or plasma. They include: radioimmunoassay (RIA) and enzyme-linked immunosorbent immunoassay (ELISA). # V. MARKETING HISTORY The modified SeaLite Sciences, Inc. AquaLite® LH that is the subject of this submission is not currently marketed. # VI. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH Caution: Use Universal Precautions. No known test method can offer complete assurance that products derived from human serum are pathogen-free; therefore, handle all materials of human origin as though they were potentially infectious. Sodium azide is used as a preservative. This preservative may react with metallic plumbing to from explosive metal azides. Flush with large volumes of water when disposing of materials containing sodium azide. As an *in vitro* diagnostic test, there are not direct adverse effects on the health of a patient from the use of this product. However, failure of the device to perform as indicated, the contamination of reagents, the use of reagents past the labeled expiration dates, the use of improper specimens, or human error during the performance of the test may lead to erroneous results and possible improper patient management. # VII. SUMMARY OF STUDIES Studies on SeaLite Sciences, Inc. AquaLite® LH were conducted at SeaLite Sciences. The results are summarized below: ## Performance Characteristics ### 1. Sensitivity The sensitivity or detection limit of the AquaLite® LH assay is 0.01 IU/L. Sensitivity is determined by adding the mean signal of twenty (20) replicates of the zero level calibrator plus two (2) standard deviations above this mean. The LH concentration (IU/L) corresponding to this calculated signal is defined as the analytical sensitivity of the assay. {3} # 2. Specificity The AquaLite® LH assay measures intact LH. The following human sialoglycoprotein hormones were supplied by the World Health Organization’s National Institute for Biological Standards and Control (London, England). Aliquots of these preparations were diluted to the following levels in zero calibrator and assayed. Percent cross-reactivity (%) is reported below: | Substance | WHO/NIBSC Lot Number | Tested at (IU/L) | % Cross-reactivity | | --- | --- | --- | --- | | TSH | 2nd IRP 80-558 | 1,000 | 4.9 | | FSH | I.S. 83-575 | 1,000 | 0.68 | | hCG | IRP 75-551 | 2,500 | 1.2 | # 3. High Dose Hook Effect No high dose hook effect occurs prior to 5,000 IU/L LH. # 4. Precision (a) Intra-assay precision. Tri-level commercial controls containing LH at the following concentrations were assayed to determine intra-assay precision. (Total N = 20 per solution.) | Mean LH Level (IU/L) | SD | % CV (calibration values) | | --- | --- | --- | | 1.069 | 0.0807 | 7.55% | | 9.821 | 0.578 | 5.9% | | 40.129 | 2.728 | 6.8% | 383851.1 {4} (b) *Inter-assay precision.* Tri-level commercial controls containing LH at the following concentrations were assayed in duplicate repetitively. Ten assays were performed using ten sets of calibration values. Inter-assay precision observed for the solutions (Total N = 2 x 10 = 20) is shown below. | Mean LH Level (IU/L) | SD | % CV | | --- | --- | --- | | 1.585 | 0.131 | 8.2% | | 11.303 | 0.924 | 8.17% | | 40.09 | 3.60 | 8.9% | ## 5. Method Comparison The AquaLite® LH assay was used to test patient samples (N=62) that were previously assayed by a commercially available kit. The samples ranged from 0.5 to 62.0 IU/L. A slope of 0.89 with a y-intercept of 1.53 was obtained. The correlation coefficient was 0.938. ## 6. Linearity and Nonparallelism Three human serum samples containing the levels of endogenous LH shown below were diluted in parallel using Calibrator A with 0.9% sodium chloride added. | Sample ID | Dilution Factor | Found (IU/L) | Expected (IU/L) | Percent (%) | | --- | --- | --- | --- | --- | | SLS#4 | Undiluted | 80.83 | -- | -- | | | 1:2 | 39.55 | 40.42 | 97.9 | | | 1:4 | 19.575 | 20.21 | 96.9 | | | 1:8 | 13.4 | 10.1 | 130 | | SLS#18 | Undiluted | 63.82 | -- | -- | | | 1:2 | 30.436 | 31.91 | 95.4 | | | 1:4 | 15.84 | 15.95 | 99.4 | | | 1:8 | 9.09 | 8.0 | 113 | | SLS#26 | Undiluted | 75.46 | -- | -- | | | 1:2 | 44.4 | 37.73 | 115 | | | 1:4 | 19.65 | 18.86 | 104 | | | 1:8 | 10.4 | 9.43 | 111 | 383851.1 D-4 {5} # 7. Spike and Recovery Normal male human serum samples were spiked with LH in the amounts noted below. The spiked samples were assayed using the AquaLite® LH assay. All values are in IU/L. | Endogenous LH | Added LH | Measured LH | Expected LH | Percent Recovered | | --- | --- | --- | --- | --- | | 6.83 | 50 | 60.35 | 56.83 | 106.1 | | 11.82 | 50 | 56.59 | 61.82 | 91.6 | | 3.14 | 50 | 57.69 | 53.14 | 108.6 | | 0.922 | 50 | 48.88 | 51.0 | 96 | | 2.17 | 50 | 53.29 | 52.17 | 101.5 | | 0.16 | 50 | 52.7 | 50.2 | 104.6 | | 2.06 | 50 | 56.89 | 52.06 | 109.3 | | 2.53 | 50 | 56.02 | 52.53 | 106.6 | # 8. Recovery in Serum and Plasma Blood Samples from 2 normal subjects were prepared as sera (standard technique and SST tubes) as well as heparin, EDTA, oxalate, and citrate plasmas. LH was quantified using the AquaLite® LH assay. Recovered LH was compared with LH recovered in serum (standard technique). The data demonstrate that there are no significant differences among serum and SST serum nor among serum and heparin, EDTA, and citrate plasmas when using the AquaLite® LH assay. Oxalate plasmas is not recommended. # 9. Effect of Common Interferents Pooled normal male human serum was spiked with preparations of hemoglobin, bilirubin, human serum albumin and triglycerides to the levels shown below. Equal amounts of LH were spiked into normal male serum as well as the normal male serum aliquots containing potential interferents. LH was quantified using the AquaLite® LH assay. Recovered LH was compared with the LH recovered in normal male serum. The data (in IU/L) demonstrate that the AquaLite® LH assay is not 383851.1 {6} significantly affected by hemoglobin, bilirubin, human serum albumin or triglycerides at the levels tested. | LH Spike | Neat Serum | Hemoglobin (at 500 mg/dL) | | Bilirubin (at 20 mg/dL) | | Albumin (at 12 mg/dL) | | Triglycerides (at 3000 mg/dL) | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | IU/L | % | IU/L | % | IU/L | % | IU/L | % | | 0 | 4.8 | - | - | - | - | - | - | - | - | | 65 | 70.0 | 68.7 | 98.2 | 72.4 | 103 | 71.0 | 101 | 69.1 | 98.7 | ## VIII. CONCLUSIONS DRAWN FROM STUDIES The data from the studies conducted demonstrated that the performance of SeaLite Sciences, Inc. AquaLite® LH is similar and substantially equivalent to that of other commercially available assays for LH. 383851.1 D-6
Innolitics
510(k) Summary
Decision Summary
Classification Order