The Axis-Shield HoloTC RIA is an in-vitro diagnostic assay for quantitative measurement of the fraction of cobalamin (vitamin B12) bound to the carrier protein transcobalamin in human serum or plasma. Measurements obtained by this device are used in the diagnosis and treatment of vitamin B12 deficiency.
Device Story
In-vitro diagnostic competitive binding immunoassay; utilizes human serum or plasma samples. Procedure: transcobalamin captured via specific monoclonal antibody; cobalamin released from transcobalamin using dithiothreitol and sodium hydroxide; released cobalamin competes with 57Co-labeled vitamin B12 for limited intrinsic factor. Detection signal: radioactivity (57Co). Used in clinical laboratory settings by trained technicians. Output: quantitative concentration of holotranscobalamin (pmol/L). Healthcare providers use results to assess vitamin B12 status and guide clinical management of deficiency.
Clinical Evidence
Clinical performance evaluated across 3 study sites (n=425 total: 112 likely deficient, 313 not likely deficient). Deficiency defined as low B12 and elevated MMA. Concordance was 80% (95% CI: 76.3-83.7%), relative sensitivity 99.5% (95% CI: 97.3-100%), and relative specificity 76.3% (95% CI: 71.8-80.8%).
Technological Characteristics
Competitive binding radioimmunoassay. Components: magnetic particles coated with monoclonal anti-human TC antibodies, dithiothreitol (reductant), sodium hydroxide (extractant), 57Co-labeled vitamin B12 (tracer), and intrinsic factor (binder). Detection via radioactivity measurement. Standard: NCCLS EP5-T2 for precision. Calibrators: recombinant HoloTC, purity >97% (A280/A362 ratio 2.3 +/- 0.1).
Indications for Use
Indicated for quantitative measurement of cobalamin bound to transcobalamin in human serum or plasma to aid in the diagnosis and treatment of vitamin B12 deficiency.
Regulatory Classification
Identification
A vitamin B12 test system is a device intended to measure vitamin B12 in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of anemias of gastrointestinal malabsorption.
{0}------------------------------------------------
K030655
# 510(K) Summary Axis-Shield HoloTC RIA
## Summary of Safety and Effectiveness Information Supporting a Substantially Equivalent Determination
### Statement of Intended Use
The Axis-Shield HoloTC RIA is an in-vitro diagnostic assay for quantitative mcasurement of holotranscobalamin (vitamin B12 bound to transcobalamin) in human serum or plasma. Measurements obtained by this device are used in the diagnosis and treatment of vitamin B12 deficiency. HoloTC RIA is calibrated with HoloTC Calibrators. HoloTC controls are assayed for the verification of the accuracy and precision of the HoloTC RIA.
## Summary of Technological Characteristics
The Axis-Shield HoloTC RIA is a competitive binding immunoassay in which a specific monoclonal antibody is used to capture transcobalamin from the patient sample. Thereafter the procedure is as commonly used in vitamin B12 assays. The cobalamin (vitamin B12) is released from the transcobalamin using dithiothreitol and sodium hydroxide. The released cobalamin (vitamin B12) then competes for a limited amount of intrinsic factor with added 'Co labelled vitamin B12. The Axis-Shield HoloTC RIA differs from the predicate device in two main aspects:
- 1) The use of a transcobalamin specific antibody, this allows the quantitation of only the cobalamin bound to the protein transcobalamin as opposed to measurement of cobalamin bound to all proteins in the predicate device and
- 2) The detection signal is radioactivity (37Co) as opposed to chemiluminescence in the predicate device.
## Method Comparison
The Axis-Shield HoloTC RIA was compared to the Bayer Advia Centaur VB12 assay (K993571) using 392 patient samples with vitamin B12 concentrations ranging from 114-821 pmol/L.
Linear regression (least squares) vielded the following statistics:
\HoloTC pmol/L = 0.55\VB12-2 pmol/L r2 = 0.52
#### Reference Interval
Based on a Finnish population of normal individuals (n=303, age 22-88 years) the 95% central reference interval was found to be 37-171 pmol/L.
Analysis of covariance demonstrated that HoloTC levels depended on gender (but not age) in this reference population. The 90 % confidence intervals for the lower limit of the reference range are for the whole population, males and females 36-37 pmol/L, 37-39 pmol/L and 35-36 pmol/L, respectively.
{1}------------------------------------------------
#### External Evaluation
The Axis-Shield HoloTC RIA external laboratory performance was evaluated at two external study sites, UK and Denmark. Performance of the system was demonstrated by using three different levels of native and pooled serum samples in addition to kit controls Low and High. CV for the three serum samples were within the given acceptance criteria, <15 %.
#### Precision
Precision studies were done according to NCCLS standard. Precision of the system was demonstrated by using three levels of serum samples, low, medium and high (14, 67 and 139 pmol/L HoloTC, respectively). Within-run CV for duplicate measurements of serum low, medium and high were 11%, 5% and 8%, respectively and total precision was 11%, 6% and 8%, respectively.
#### Linearity
Dilution Linearity within the working range of 10-160 pmol/L was demonstrated to meet all criteria for linearity. Recovery test of HoloTC spiked serum samples was demonstrated to be within acceptance criteria. Dilution of high HoloTC sample into six dilutions showed a correlation coefficient r2 = 0.998, slope = 1.01 ± 0.02 and y-intercept = - 5 ± 2 pmol/L for dilutions ranging between 13 - 182 pmol/L HoloTC. Limit of Quantification (LOQ) and Limit of Detection (LOD) were demonstrated to be 8-176 pmol/L and 6 pmol/L, respectively. Cross-reactivity was tested against haptocorrin (HC) and cross-reactivity was well within the acceptance criteria of ±10% for up to 70,000pg/mL HC spiked serum samples.
#### Interference
Effect on quantification by endogenous interfering substances was tested. billirubin, haemoglobin, total protein and lipids showed less that 10% interference.
#### Serum/EDTA Plasma comparison
50 paired serum and EDTA plasma samples collected from the same donors were analysed and compared. No significant difference in measured HoloTC level was demonstrated between serum sample and EDTA sample. In-use stability of the reagent and calibrator kit showed stability up to three months after opening.
#### Sensitivity and Specificity
The estimated concordance and relative sensitivity and specificity of HoloTC RIA was calculated using samples from 3 study sites that were classified as being likely (n=112) and not likely (n=313) vitamin B12 deficient based on cobalamin and MMA levels. A general parametric approach was used. The estimated concordance, sensitivity, and specificity for each study was weighted and then all were combined to give the following estimates; concordance 80%, sensitivity 99.5%, specificity 76.3%.
#### Conclusion
{2}------------------------------------------------
In conclusion these data demonstrate that the Axis-Shield HoloTC RIA is as safe and effective as, and is substantially equivalent to, the Bayer Advia Centaur VB12 assay.
Contact information:
Axis-Shield C/O - R & R REGISTRATIONS Ronald G. Leonardi President San Diego, CA 92131 858-586-0751 858-586-1108 (fax)
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services (USA). The logo features a stylized abstract design with three wave-like shapes, possibly representing people or services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the abstract design.
Public Health Service
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
JAN 2 9 2004
Axis-Shield Biochemicals, ASA c/o Ronald G. Leonardi, Ph. D. President R & R Registrations P.O. Box 262069 San Diego, CA 92196-2069
k030655 Re:
.
Trade/Device Name: Axis-Shield Holo TC RIA Regulation Number: 21 CFR 862.1810 Regulation Name: Vitamin B12 test system Regulatory Class: Class II Product Code: CDD; JIS; JJX Dated: November 4, 2003 Received: November 5, 2003
Dear Dr. Leonardi:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your betermined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate rol use stated in the encreat date of the Medical Device Amendments, or to commerce proc to may 20, 2011 - 12:00 accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and Cosmetic For ( rol) lies the device, subject to the general controls provisions of the Act. The I ou may, meroloro, manes of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. 1
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is elabilition (toontrols. Existing major regulations affecting your device can may be subject to sail adaxt Federal Regulations (CFR), Parts 800 to 895. In addition, FDA be found in Ther 21, 0 wouncements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean I loast be advised that I Drivian that your device complies with other requirements of the Act that I Dr has Intact and regulations administered by other Federal agencies. You must of any I catal statutes and states and including, but not limited to: registration and listing (21 comply with an and and the Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{4}------------------------------------------------
#### Page 2
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Jean M. Cooper, US, DVM.
Yean M. Cooper, MS, D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
X030655
510(k) Number (if known)
Device Name:Axis-Shield HoloTC RIA
## INDICATIONS FOR USE:
The Axis-Shield HoloTC RIA is an in-vitro diagnostic assay for quantitative measurement of the fraction of cobalamin (vitamin B12) bound to the carrier protein transcobalamin in human serum or plasma. Measurements obtained by this device are used in the diagnosis and treatment of vitamin B12 deficiency.
(PLEASE DO NO WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
OR Over-The-Counter Use
(Optional Format 1-2-96)
Carol C. Benson for Jean Cooper, OhiM
(Division Sign-Off) Division of Clinical Laboratory De 510(k) Number -
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.